← Product Code [NAY](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/NAY) · K171294

# da Vinci X Surgical System (K171294)

_Intuitive Surgical, Inc. · NAY · May 26, 2017 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/NAY/K171294

## Device Facts

- **Applicant:** Intuitive Surgical, Inc.
- **Product Code:** [NAY](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/NAY.md)
- **Decision Date:** May 26, 2017
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The Intuitive Surgical Endoscopic Instrument Control System (da Vinci X Surgical System, Model IS4200) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

## Device Story

Software-controlled, electro-mechanical surgical system; assists surgeons in minimally invasive procedures. Input: surgeon manipulation of master controls and foot pedals at surgeon console. Transformation: system translates surgeon hand/foot movements into precise movement of EndoWrist instruments and endoscope mounted on patient-side cart arms. Output: real-time 3D endoscopic visualization of anatomy and instrumentation; precise tissue manipulation (grasping, cutting, dissection, suturing, electrocautery). Used in operating rooms by trained physicians. Surgeon views 3D image on console; system provides enhanced control and visualization compared to manual laparoscopy. Benefits: improved surgical precision, minimally invasive access, reduced trauma.

## Clinical Evidence

No clinical data. Evidence consists of hardware verification, human factors testing, preclinical validation, and software verification/validation testing.

## Technological Characteristics

Electro-mechanical, software-controlled system. Comprises surgeon console, patient-side cart, and vision cart. Patient cart features four arms for endoscope/instrument manipulation. Compatible with existing da Vinci Xi instruments/accessories. Connectivity: integrated system components. Sterilization: compatible with standard surgical instrument sterilization protocols.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- da Vinci Xi Surgical System ([K131861](/device/K131861.md))

## Reference Devices

- da Vinci Si Surgical System ([K081137](/device/K081137.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the address of the Food and Drug Administration. The address is 10903 New Hampshire Avenue, Document Control Center - WO66-G609, Silver Spring, MD 20993-0002. The text is black and the background is white.

May 26, 2017

Intuitve Surgical, Inc. Mr. Brandon Hansen Project Manager, Regulatory Affairs 1266 Kifer Road Sunnyvale, California 94086

Re: K171294

Trade/Device Name: da Vinci X Surgical System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: NAY Dated: April 28, 2017 Received: May 2, 2017

Dear Mr. Hansen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Actinclude requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-

Image /page/0/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles a stylized caduceus or a family of three people.

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Image /page/1/Picture/9 description: The image contains the text "Jennifer R. Stevenson -S" in a large, sans-serif font. The text is arranged in two lines, with "Jennifer R." on the first line and "Stevenson -S" on the second line. The background of the image is a light blue color, and the letters are black.

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# da Vinci® X Surgical System

## Special 510(k): Device Modification

DEPARTMENT OF HEALTH AND HUMAN SERVICES

Food and Drug Administration

**Indications for Use**

| Form Approved: OMB No.   | 0910-0120        |
|--------------------------|------------------|
| Expiration Date:         | January 31, 2017 |
| See PRA Statement below. |                  |

| 510(k) Number (if known) | K171294                    |
|--------------------------|----------------------------|
| Device Name              | da Vinci X Surgical System |

**Indications for Use (*Describe*)**

The Intuitive Surgical Endoscopic Instrument Control System (da Vinci X Surgical System, Model IS4200) is intended to
 assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp
 endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and
 accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation,
 ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and
 accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic
 surgical procedures, general thoracoscopic surgical procedures and thoracoscopically assisted cardiotomy procedures. The
 system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac
 revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an
 operating room environment in accordance with the representative, specific procedures set forth in the Professional
 Instructions for Use.

Type of Use (*Select one or both, as applicable*)

| <span style="padding-right: 20px;"></span> <svg height="12" width="12"> <rect height="12" style="fill:none;stroke:#000000;stroke-width:2" width="12"></rect> <line style="stroke:#000000;stroke-width:2" x1="0" x2="12" y1="0" y2="12"></line> <line style="stroke:#000000;stroke-width:2" x1="12" x2="0" y1="0" y2="12"></line> </svg> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="padding-right: 20px;"></span> <svg height="12" width="12"> <rect height="12" style="fill:none;stroke:#000000;stroke-width:2" width="12"></rect> </svg> Over-The-Counter Use (21 CFR 801 Subpart C) |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the
 time to review instructions, search existing data sources, gather and maintain the data needed and complete
 and review the collection of information. Send comments regarding this burden estimate or any other aspect
 of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

*'An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
 information unless it displays a currently valid OMB number.'*

| FORM FDA 3881 (8/14) | Page 1 of 1 | PSC Publishing Services (301) 443-6740 EF |
|----------------------|-------------|-------------------------------------------|
|----------------------|-------------|-------------------------------------------|

.

:

1 N T U I T I V E
Surgica.

Page | 5-2

{3}------------------------------------------------

da Vinci® X Surgical System

Special 510(k): Device Modification

P 1/3

## 510(k) Summary

| 510(k) Owner:                         | Intuitive Surgical, Inc.<br>1266 Kifer Road<br>Sunnyvale, CA 94086                                                                   |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| Contact:                              | Brandon Hansen<br>Regulatory Affairs<br>Phone Number: 408-523-7485<br>Fax Number: 408-523-8907<br>Email: brandon.hansen@intusurg.com |
| Date Summary Prepared:                | May 1, 2017                                                                                                                          |
| Trade Name:                           | da Vinci® X Surgical System                                                                                                          |
| Common Name:                          | Endoscopic instrument control system                                                                                                 |
| Classification:                       | Class II<br>21 CFR 876.1500, Endoscope and Accessories                                                                               |
| Product Codes:                        | NAY (System, Surgical, Computer Controlled Instrument)                                                                               |
| Classification Advisory<br>Committee: | General and Plastic Surgery                                                                                                          |
| Predicate Device:                     | da Vinci® Xi™ Surgical System device, K131861                                                                                        |

#### Device Description

The da Vinci X Surgical System, Model IS4200 is a modification to the Patient Side Cart of the da Vinci Xi Surgical System, Model IS4000, cleared under K131861. The arms of the IS4000 have been grafted onto the setup structure of the da Vinci Si Surgical System Patient Cart (K081137). The da Vinci X Patient Cart utilizes the same electronics as the da Vinci Xi Patient Cart with some minor modifications. The da Vinci X Patient Cart is run by the same software and is to be used with the Surgeon Console and Vision Cart of the da Vinci Xi Surgical System. All of the da Vinci Xi instruments and accessories, including advanced instruments (e.g., Single Site, Vessel Sealer, Stapler, etc.), are compatible with the da Vinci X Surgical System.

The da Vinci X Surgical System is a software-controlled, electro-mechanical system designed for surgeons to perform minimally invasive surgery. The da Vinci X Surgical System consists of a Surgeon Console, a Patient Cart, and a Vision Cart and is used with an Endoscope, EndoWrist Instruments, and Accessories.

The surgeon seated at the Surgeon Console controls all movement of the EndoWrist Instruments and Endoscope by using two Master Controls and a set of foot pedals. The surgeon views the three-dimensional endoscopic image on a High Resolution Stereo Viewer (3D Viewer), which

{4}------------------------------------------------

#### K171294

#### da Vinci® X Surgical System

provides him/her a view of patient anatomy and instrumentation, along with icons and other user interface features.

The Vision Cart includes the supporting electronic and video processing equipment for the system.

The Patient Side Cart is positioned at the operating room table and has four endoscope/instrument arms that are positioned over the target patient anatomy. An endoscope attaches onto one arm and provides the surgeon a high resolution, three-dimensional view of the patient anatomy. A suite of EndoWrist Instruments are attached/detached from the arms, enabling the surgeon to perform various surgical tasks. Accessories such as cannulas, obturators, seals, and drapes are also needed to perform procedures with the system. The da Vinci X Surgical System Patient Cart has the same arms as the da Vinci Xi Surgical System Patient Cart and is compatible with all of the base, Single-Site, and advanced instruments and accessories (with the exception of the column drape) cleared for use with the da Vinci Xi Surgical System (K131861).

## Intended Use/Indications for Use:

The Intuitive Surgical Endoscopic Instrument Control System (da Vinci X Surgical System, Model IS4200) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

#### Technological Characteristics:

In terms of technological characteristics, the da Vinci X Surgical System is substantially equivalent to the currently marketed da Vinci Xi Surgical System device, cleared under K131861. The da Vinci X Surgical System uses the same Surgeon Console, Vision Cart, Software, Instruments, and Accessories as the predicate da Vinci Xi Surgical System.

#### Performance Data:

Performance test data demonstrate that the subject device is substantially equivalent to the predicate device and that the design output meets the design input requirements. Hardware verification, human factors testing, and preclinical validation testing were performed. Software 

{5}------------------------------------------------

da Vinci@ X Surgical System

Special 510(k): Device Modification

testing was also conducted to support this modification. The software testing included user interface, algorithm, and software verification/validation testing.

### Summary:

Based on the intended use, indications for use, technological characteristics, and performance data, the da Vinci X Surgical System is substantially equivalent to the currently marketed da Vinci Xi Surgical System device, cleared under K131861.

Image /page/5/Picture/7 description: The image shows the logo for Intuitive Surgical. The word "INTUITIVE" is in all caps and is the main focus of the logo. Below the word "INTUITIVE" is the word "SURGICAL" in smaller letters.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/NAY/K171294](https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/NAY/K171294)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
