INTUITIVE SURGICAL DA VINCI S SURGICAL SYSTEM, MODEL IS2000, WITH DA VINCI CONNECT AND ONSITE

K081207 · Intuitive Surgical, Inc. · NAY · Dec 19, 2008 · Gastroenterology, Urology

Device Facts

Record IDK081207
Device NameINTUITIVE SURGICAL DA VINCI S SURGICAL SYSTEM, MODEL IS2000, WITH DA VINCI CONNECT AND ONSITE
ApplicantIntuitive Surgical, Inc.
Product CodeNAY · Gastroenterology, Urology
Decision DateDec 19, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The da Vinci Connect is an accessory to the da Vinci S Surgical System and is intended for use by trained Surgical proctors to communicate and to provide surgical advice to the operating surgeon when using the da Vinci S Surgical System, Model IS2000. In addition, the Remote Proctor Interface can be used to view surgical procedures related to the use of the da Vinci S Surgical System, Model IS2000. The da Vinci OnSite is an accessory to the da Vinci S Surgical System and is intended for use by trained Intuitive Surgical Customer Service personnel to obtain da Vinci S Surgical System information for the purpose of diagnosing faults with the da Vinci S Surgical System, Model IS2000.

Device Story

The da Vinci S Surgical System (IS2000) is a computer-assisted, minimally invasive surgical platform. This modification adds 'da Vinci Connect' and 'da Vinci OnSite' connectivity features. Connect allows remote proctors to view surgical video, communicate via audio, and provide instructional telestration to the operating surgeon. OnSite allows Intuitive Surgical customer support to remotely access system information for fault diagnosis. Hardware additions include a router, workgroup bridge, antenna, and a Proctor PC mounted on the Vision System Cart. The system uses a Surgeon Console for control, a Patient Side Cart for instrument manipulation, and an InSite Vision System for 3D/2D image processing. The surgeon operates the system to perform complex surgeries; remote proctors provide guidance; customer service personnel perform diagnostics. These features facilitate remote collaboration and maintenance, potentially improving surgical outcomes through expert guidance and reducing system downtime.

Clinical Evidence

Bench testing only. No clinical data provided. Substantial equivalence is supported by design analysis, risk assessment, and bench testing of the modified hardware and software components.

Technological Characteristics

Computer-assisted surgical system; Surgeon Console (SS2000), Patient Side Cart (PS2000), and InSite Vision System (VS2000). Modifications include added networking hardware (router, workgroup bridge, antenna, Proctor PC) and Ethernet connectivity. System utilizes stereo endoscopes for 3D/2D visualization. Software-based remote communication and diagnostic tools integrated via customized PC.

Indications for Use

Indicated for use by trained surgical proctors to provide remote surgical advice and viewing of procedures, and by trained customer service personnel to perform remote diagnostic fault-finding on the da Vinci S Surgical System (Model IS2000).

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 081207 ## 510(k) SUMMARY # DEC 1 9 2008 This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. #### 5 December 2008 Date Submitter Intuitive Surgical, Inc. 1266 Kifer Road Sunnyvale, CA 94086 ER Number 2955842 9028901 Owner/Operator James Farnworth Contact Director, Regulatory Affairs Direct: 408-523-8687 Fax: 408-523-1390 james.farnworth@intusurg.com Name: Intuitive Surgical® da Vinci® S" Surgical System, Model IS2000, Subject with da Vinci Connect™, da Vinci OnSite™ Device Classification Name: System, Surgical, Computer Controlled Instrument (2 CFR 876.1500) Common Name: Endoscopic Instrument Control System, Endoscopic Instruments, and accessories Predicate Devices Intuitive Surgical da Vinci S Surgical System (Model IS2000) with Endoscopic and EndoWrist Instruments (legally marketed under K050369 and K063220). The da Vinci S Surgical System, Model IS2000 with da Vinci Connect and Device Description da Vinci OnSite, is a minor modification to the existing computer assisted device designed to facilitate complex surgery using minimally invasive approach. An overview of the IS2000 consists of the following three main integrated sub-systems: The Surgeon Console, Model SS2000: The control center of the surgical system, and allows the surgeon to control critical aspects of the surgical procedure (including movement of the endoscopic instruments and endoscope). Instrument and camera movements are controlled by the Master Tool Manipulators (MTM), two hand operated controllers, and foot pedals at the Surgeon Console's base. These allow the surgeon to be as dexterous as in "open" surgery. while operating in a minimally invasive #### 510(k) Number: {1}------------------------------------------------ | Device<br>Description<br>(continued) | environment. The surgeon has the option to change the surgical view from<br>full screen mode to a multi-image mode, which displays an optional 3-<br>dimensional High Definition (3-D HD), 3-D, or 2-D standard definition<br>image of the operative field. | |--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | The Patient Side Cart (PSC, Model PS2000): The operative component of<br>the surgical system within the sterile field. Its primary function is to<br>support the endoscopic instrument arms and camera arm, during surgical<br>procedures. Endoscopic instruments are held in a fixed position (with<br>respect to the patient) by unique Patient Side Manipulators (PSM), and the<br>endoscope is held by the Endoscope Camera Manipulator (ECM). The<br>PSM's and ECM are attached to surgical arms on the PSC known as Set-up<br>Joint (SUJ) arms. Commands from the Surgeon Console are relayed to the<br>PSC, via a cable. | | | <i>InSite</i> ® Vision System (Model VS2000): Houses the system's image<br>processing equipment and is operated by a person outside the sterile field<br>during surgery. The cart provides space for an optional touch screen<br>monitor, as well as ancillary surgical equipment. The VS2000 consists of a<br>stereo endoscope, endoscopic camera, and various accessories, including a<br>light source and light guides. The <i>InSite</i> Vision System provides two<br>independent images that are relayed to the viewer located in the Surgeon<br>Console, where they are fused to form a 3-D image (or alternatively a 2-D)<br>of the surgical field for the surgeon's reference. | | Indications for<br>Use | da Vinci Connect: | | | The da Vinci Connect is an accessory to the da Vinci S Surgical System<br>and is intended for use by trained Surgical proctors to communicate and to<br>provide surgical advice to the operating surgeon when using the da Vinci<br>S Surgical System, Model IS2000. In addition, the Remote Proctor<br>Interface can be used to view surgical procedures related to the use of the<br>da Vinci S Surgical System, Model IS2000. | | | da Vinci OnSite: | | | The da Vinci OnSite is an accessory to the da Vinci S Surgical System<br>and is intended for use by trained Intuitive Surgical Customer Service<br>personnel to obtain da Vinci S Surgical System information for the<br>purpose of diagnosing faults with the da Vinci S Surgical System, Model<br>IS2000. | | Comparison to<br>Predicate Device | This 510(k) notification is being submitted for minor design<br>modifications to the Surgeon Console and Vision System Cart of the<br>Endoscopic Instrument Control System, Model IS2000. | | | • da Vinci Connect allows a skilled remote proctoring surgeon to | {2}------------------------------------------------ provide advice to the console surgeon via audio, video or using instructional telestration, during a surgical procedure. - da Vinci OnSite allows trained Intuitive Surgical, Inc. Customer . Support personnel to obtain system information for the purpose of debugging and diagnosing problems with the system. Minor hardware modifications to the Surgeon Console (SS2000) to support da Vinci Connect and da Vinci OnSite consists of: #### Comparison to Predicate Device (continued) - . Router, - Workgroup Bridge, . - . Antenna, - Rear Service Panel Modification to allow connection to hospital ethernet . Minor hardware modifications to the Vision System Cart (VS2000) to support da Vinci Connect and da Vinci OnSite consists of: - . a da Vinci Connect PC (also known as "Proctor PC") mounted on the Vision System Cart, with customized software. - a Uninterruptable Power Supply (UPS) with a battery backup, connected . to an existing Medical Grade Isolation Transformer, for the Proctor PC only. - . Rear Panel modification, two (2) additional User Video Input connections, and an Ethernet connector. Device operation, functionality, and methods of use for the subject device are identical to the predicate device. The technology, materials, manufacturing methods, and performance are essentially the same for the predicate device as the subject device. Technological The technological characteristics of the subject device are essentially the Characteristics same as for the predicate device. - Performance Design analysis and comparison, as well as bench testing and risk Data analysis activities, have been conducted to confirm that the characteristics of the modified device are substantially equivalent to the predicate devices cited. - Conclusion The da Vinci S Surgical System (Model IS2000) is a sophisticated computer assisted device designed to facilitate complex surgery using a minimally invasive approach. The indications for use, remains unchanged from the predicate device. The Intuitive Surgical da Vinci S Surgical System with da Vinci Connect and da Vinci OnSite does not raise any new safety or effectiveness issues and is substantially equivalent to, and have been determined to be substantially equivalent to devices in commercial distribution, prior to May 28, 1976. {3}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" around the perimeter. Inside the circle is an abstract image of an eagle. #### Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 ### DEC 1 9 2008 Director, Regulatory Affairs 1266 Kifer Road Sunnyvale, California 94086 Intuitive Surgical, Inc. % Mr. James Farnworth Re: K081207 Trade/Device Name: Intuitive Surgical® da Vinciço Sm Surgical System, Model IS200, with da Vinci Connect™, da Vinci Onsite™ Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: NAY Dated: December 5, 2008 Received: December 8, 2008 Dear Mr. Farnworth: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ Page 2 - Mr. James Famworth This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Mark N Millman Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use 510(k) Number (if known): K081207 Device Name: Intuitive Surgical da Vinci S Surgical System, Model IS2000 with da Vinci OnSite and da Vinci Connect Indications For Use: Indications For Use - da Vinci Connect: The da Vinci Connect is an accessory to the da Vinci S Surgical System and is intended for use by trained Surgical proctors to communicate and to provide surgical advice to the operating surgeon when using the da Vinci S Surgical System, Model IS2000. In addition, the Remote Proctor Interface can be used to view surgical procedures related to the use of the da Vinci S Surgical System, Model IS2000. Indications For Use - da Vinci OnSite: The da Vinci OnSite is an accessory to the da Vinci S Surgical System and is intended for use by trained Intuitive Surgical Customer Service personnel to obtain da Vinci S Surgical System information for the purpose of diagnosing faults with the da Vinci S Surgical System, Model IS2000. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Neill R. Oglin Cor mem Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of General, Restorative, and Neurological Devices | Page 1 of 1 | |-------------| |-------------| | 510(k) Number | K081207 | |---------------|---------| |---------------|---------|
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...