← Product Code [KNW](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/KNW) · K193268

# APrioCore Plus (K193268)

_Apriomed AB · KNW · Feb 27, 2020 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/KNW/K193268

## Device Facts

- **Applicant:** Apriomed AB
- **Product Code:** [KNW](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/KNW.md)
- **Decision Date:** Feb 27, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1075
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

AprioCore plus is intended for use in obtaining core biopsy specimens from soft tissue such as breast, kidney, liver, lung, thyroid, lymph nodes and various soft tissue masses performed under image guidance techniques for soft tissue biopsy.

## Device Story

AprioCore plus is a manually operated, sterile, single-use, semi-automatic, spring-loaded biopsy needle. It features a beveled needle with centimeter markings and an echogenic distal tip for ultrasound visualization; it is also visible under CT and fluoroscopy. The device includes an inner stylet with a specimen notch for tissue collection. Used by physicians trained in percutaneous image-guided biopsy, the device is operated manually to penetrate soft tissue and retrieve core samples. The design includes a shorter tip compared to the predicate, allowing for biopsies closer to sensitive anatomical structures. The housing facilitates various hand positioning options. The device is compatible with Gangi-SoftGuard and Gangi-HydroGuard coaxial needles. By providing a reliable mechanism for tissue extraction under imaging, it assists clinicians in diagnostic procedures, potentially benefiting patients through accurate tissue sampling for pathology.

## Clinical Evidence

Bench testing only. No clinical data provided. Performance verified through mechanical testing including depth projection, penetration force, loading force, activation force, mechanical durability, tissue sample extraction, tip configuration, imaging visibility, and coaxial compatibility. Biocompatibility testing confirmed compliance with ISO 10993 standards (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity).

## Technological Characteristics

Spring-loaded semi-automatic biopsy needle. Materials: Stainless steel AISI 304 and AISI 302 (ISO 9626 compliant). Features: Beveled needle, centimeter markings, echogenic distal tip. Energy source: Manual spring-loaded mechanism. Sterilization: Ethylene Oxide. Compatibility: Coaxial needles. Form factor: Various gauges (16G, 18G) and lengths (8.5-19.5 cm).

## Regulatory Identification

A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Predicate Devices

- Quick-Core Biopsy Needle ([K973565](/device/K973565.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on top and "ADMINISTRATION" on the bottom.

February 27, 2020

AprioMed AB Katrin Svensson Director, Quality and Regulatory Affairs Virdings Alle 28 Uppsala, Sweden 754 50

Re: K193268

Trade/Device Name: APrioCore plus Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-Urology Biopsy Instrument Regulatory Class: Class II Product Code: KNW Dated: January 20, 2020 Received: January 22, 2020

Dear Katrin Svensson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K193268

Device Name AprioCore plus

Indications for Use (Describe)

AprioCore plus is intended for use in obtaining core biopsy specimens from soft tissue such as breast, kidney, liver, lung, thyroid, lymph nodes and various soft tissue masses performed under image guidance techniques for soft tissue biopsy.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### 5 510(K) SUMMARY

As Required by 21 CFR 807.92(c) 510(k) Summary

### 5.1 Submitter/510(k) Holder

AprioMed AB Virdings Allé 28 Uppsala, SWEDEN 754 50

Contact Person: Katrin Svensson, Director Quality & Regulatory Affairs Telephone: +46 (0) 73- 345 14 40

Date Prepared: November 22, 2019

### 5.2 Device

| Proprietary Name:     | AprioCore plus                                       |
|-----------------------|------------------------------------------------------|
| Common/Usual Names:   | Instrument, Biopsy                                   |
| Classification Names: | Instrument, Biopsy                                   |
| Regulation number:    | 876.1075, Gastroenterology-Urology Biopsy Instrument |
| Product Code:         | KNW                                                  |
| 510(k) number:        | TBD                                                  |

# 5.3 Predicate Devices identification

| Proprietary Name:     | Quick-Core Biopsy Needle                             |
|-----------------------|------------------------------------------------------|
| Common/Usual Names:   | Instrument, Biopsy                                   |
| Classification Names: | Instrument, Biopsy                                   |
| Regulation number:    | 876.1075, Gastroenterology-Urology Biopsy Instrument |
| Product Code:         | KNW                                                  |
| 510(k) number:        | K973565                                              |

{4}------------------------------------------------

# 5.4 Device Description

AprioCore plus is a manually operated, sterile, single use, semi-automatic, spring-loaded biopsy needle. AprioCore plus is provided in a variety of sizes (gauge and length) and is compatible with AprioMed devices Gangi-SoftGuard Coaxial Needle and Gangi-HydroGuard Coaxial Needle.

#### ર્સ્ડ Technological Characteristics

AprioCore plus is a spring-loaded semi-automatic biopsy needle that is visible in CT (computed tomography) and fluoroscopy. The device is equipped with a beveled needle with centimeter markings and an echogenic distal tip for visualization during ultrasound imaging.

During use, the position of the device is monitored using technique. The inner stylet is equipped with a slot (specimen notch) to collect a biopsy specimen. Compared to predicate device, AprioCore plus has a shorter tip, thereby making it possible to take a biopsy specimen closer to nearby sensitive anatomy/structure. The housing of the device has been designed to facilitate different hand positioning options during use.

The materials used for construction of AprioCore plus needles are typical for this type of medical device. The only material in direct patient contact is the stainless steel AISI 304 and stainless steel AISI 302.

#### 5.6 Indications for Use

AprioCore plus is intended for use in obtaining core biopsy specimens from soft tissue such as breast, kidney, liver, lung, thyroid, lymph nodes and various soft tissue masses performed under image guidance techniques for soft tissue biopsy.

### Substantial Equivalence 5.7

AprioMed AB has determined that AprioCore plus are substantially equivalent to the predicate device based on the indications for use, compliance with internationally recognized design and performance standards, material specifications, use environments, and performance. The differences between subject devices and the predicate device used for the same indications for use do not raise new issues of safety and effectiveness.

A substantial equivalence comparison table of subject devices. AprioCore plus and the predicate device, Quick-Core Biopsy Needle (K973565) is provided in Table 5-1.

{5}------------------------------------------------

| Parameter                           | Subject Devices                                                                                                                                                                                                                                                  | Predicate Device                                                                                                                                                                                                                                                  |
|-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Proprietary Name                    | AprioCore plus                                                                                                                                                                                                                                                   | Quick-Core Biopsy Needle                                                                                                                                                                                                                                          |
| 510(k)                              | This 510(k) application                                                                                                                                                                                                                                          | K973565                                                                                                                                                                                                                                                           |
| Device Classification<br>Name       | Biopsy needle                                                                                                                                                                                                                                                    | Biopsy needle                                                                                                                                                                                                                                                     |
| Product Code                        | KNW                                                                                                                                                                                                                                                              | KNW                                                                                                                                                                                                                                                               |
| Regulation<br>description           | Gastroenterology-Urology Biopsy<br>Instrument                                                                                                                                                                                                                    | Gastroenterology-Urology Biopsy<br>Instrument                                                                                                                                                                                                                     |
| Regulation Number                   | 21 CFR 876.1075                                                                                                                                                                                                                                                  | 21 CFR 876.1075                                                                                                                                                                                                                                                   |
| Intended/use<br>Indications for Use | AprioCore plus is intended for use in<br>obtaining core biopsy specimens from soft<br>tissue such as breast, kidney, liver, lung,<br>thyroid, lymph nodes and various soft<br>tissue masses performed under image<br>guidance techniques for soft tissue biopsy. | Quick-Core Biopsy Needles are intended<br>for soft tissue biopsy. The product is<br>intended for use by physicians trained and<br>experienced in diagnostic and<br>interventional techniques. Standard<br>technique for soft tissue biopsy should be<br>employed. |
| Anatomical sites                    | Specimens from soft tissue such as<br>breast, kidney, liver, lung, thyroid, lymph<br>nodes and various soft tissue masses                                                                                                                                        | Soft tissue biopsy                                                                                                                                                                                                                                                |
| Device type                         | Semi-automatic biopsy gun, spring-<br>operated                                                                                                                                                                                                                   | Semi-automatic biopsy gun, spring-<br>operated                                                                                                                                                                                                                    |
| Visualization<br>technique          | Conventional imaging guidance equipment<br>excluding MRI                                                                                                                                                                                                         | Conventional imaging guidance equipment<br>excluding MRI                                                                                                                                                                                                          |
| Single Use                          | Yes                                                                                                                                                                                                                                                              | Yes                                                                                                                                                                                                                                                               |
| Sterile                             | Yes, Ethylene Oxide                                                                                                                                                                                                                                              | Yes, Ethylene Oxide                                                                                                                                                                                                                                               |
| Needle material                     | Stainless Steel                                                                                                                                                                                                                                                  | Stainless Steel                                                                                                                                                                                                                                                   |
| Needle diameter<br>(Gauge)          | 16G and 18G                                                                                                                                                                                                                                                      | 14G, 16G, 18G and 20G                                                                                                                                                                                                                                             |
| Needle length                       | 8.5, 9.5, 13.5, 14.5, 18.5 and 19.5 cm                                                                                                                                                                                                                           | 6, 9, 15 and 20 cm                                                                                                                                                                                                                                                |
| Specimen notch size                 | 10 mm and 20 mm                                                                                                                                                                                                                                                  | 10 mm and 20 mm                                                                                                                                                                                                                                                   |

## Table 5-1: Substantial equivalence table

{6}------------------------------------------------

#### 5.7.1 Indication for Use

AprioMed has determined that the intended use of AprioCore plus are substantially equivalent to the predicate device.

#### 5.7.2 Materials

AprioCore plus and the predicate device is constructed with medical grade stainless steel and the needle material fulfill the, by FDA, recognized consensus standard, ISO 9626 Stainless Steel Needle Tubing for the Manufacture of Medical Devices. AprioMed has concluded that the subject device is substantial equivalent to the predicate device.

#### 5.7.3 Use Environment

AprioCore plus and the predicate device are used in the same clinical setting using conventional imaging guidance equipment excluding MRI. The users are physicians trained in percutaneous image-guided biopsy sampling of soft tissue. AprioMed has concluded that the subject device is substantial equivalent to the predicate device.

#### 5.7.4 Size

AprioCore plus needle gauge size fulfill the, by FDA, recognized consensus standard, ISO 9626 Stainless Steel Needle Tubing for the Manufacture of Medical Devices. AprioMed has determined that the gauge, length and specimen notch size of AprioCore plus is substantially equivalent to the predicate device.

#### 5.7.5 Activation force

Subject and predicate device was tested side-by-side to compare activation force. The activation force shall be equivalent or higher compared to the predicate device to prevent accidentally activation. AprioMed concluded that the subject device is substantial equivalent to the predicate device.

#### 5.7.6 Penetration force

Subject and predicate device was tested side-by-side to compare penetration force. AprioCore plus shall have a penetration force that is equal or lower than corresponding penetration force of the predicate device, in order to cut through tissue easily. AprioMed concluded that the subject device is substantial equivalent to the predicate device.

#### 5.7.7 Specimen size

Subject and predicate device was tested side-by-side to compare specimen extraction size by weight. AprioMed concluded that the subject device is substantial equivalent to the predicate device.

#### Summary of non-clinical and performance testing 5.8

Bench testing was performed to evaluate the performance and functionality of the subject device against requirement specification. The subject device has been subjected to compliance testing according to, by FDA, recognized consensus standards ISO 9626, ISO 10993-7, ISO 10993-1, ISO 11607-1. Results from testing performed confirms that the design requirement specification and user needs have been met. The subject device is confirmed to be safe and effective for the intended use.

{7}------------------------------------------------

#### 5.8.1 Sterilization and shelf life

AprioCore plus is delivered sterile and have successfully been tested according to ISO 11607-1. The label shelf life is 3 years.

#### 5.8.2 Biocompatibility testing

AprioCore plus has successfully been tested for cytotoxicity, sensitization, intracutaneously irritation, acute systemic toxicity and material medicated pyrogenicity. The test results verify that the biocompatibility criteria given in ISO 10993 are fulfilled. AprioMed concludes that AprioCore plus is non-toxic and biocompatible.

#### 5.8.3 Performance testing - Bench

The performance of AprioCore plus has been verified. Tests as described in table 5-2 have been completed.

| Test:                                       | Description:                                                                                                                                                                                                                                                   |
|---------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Depth projection                            | To confirm that subject device will not extend over the stylet tip (over-throw) during use. Conformity has been demonstrated.                                                                                                                                  |
| Penetration force                           | To confirm that the penetration force of the subject device is equivalent to predicate device. Conformity has been demonstrated.                                                                                                                               |
| Loading force                               | To confirm that the loading force of the subject device is equivalent to predicate device and that the user is able to load the subject device (force required to compress the spring). Conformity has been demonstrated.                                      |
| Activation force                            | To confirm that the force to activate the subject device (release the spring) is not too low, resulting in unintentional activation during use. Conformity has been demonstrated.                                                                              |
| Mechanical durability                       | To confirm that subject device withstands the forces applied to the subject device during normal use. Conformity has been demonstrated.                                                                                                                        |
| Tissue sample extraction test               | To confirm that the subject device can successfully retrieve biopsy specimen multiple times. Conformity has been demonstrated.                                                                                                                                 |
| Tip configuration                           | To confirm that it is possible to take a biopsy specimen closer to nearby sensitive anatomy/structure compared to predicate device. Subject device shall have a tip dead-space that is shorter compared to predicate device. Conformity has been demonstrated. |
| Visibility                                  | To confirm that the invasive part of subject device is visible in CT, Fluoroscopy and Ultrasound. Conformity has been demonstrated.                                                                                                                            |
| Compatibility                               | To confirm that the device is compatible with coaxial needle standardized sizes as per ISO 9626 has been applied. AprioCore plus has been demonstrated to be compatible with AprioMed Gangi-SoftGuard and Gangi-HydroGuard.                                    |
| Qualification metal tubing/needle component | The stainless-steel tubing fulfills the requirement in ISO 9626 Stainless steel needle tubing for the manufacture of medical devices – Requirements and test methods. Conformity has been demonstrated.                                                        |

Table 5-2: Performance testing summary

# 5.9 Summary and Conclusion

{8}------------------------------------------------

Conclusion drawn from testing completed is that AprioCore plus is equivalent to the predicate. AprioCore plus is as safe, as effective, and performs as well as or better than the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/KNW/K193268](https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/KNW/K193268)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
