VIDEO CAMERA/LIGHTSOURCE COMBINATION UNIT MODEL VCL-150

K980350 · Cuda Products Co. · GCJ · Apr 28, 1998 · Gastroenterology, Urology

Device Facts

Record IDK980350
Device NameVIDEO CAMERA/LIGHTSOURCE COMBINATION UNIT MODEL VCL-150
ApplicantCuda Products Co.
Product CodeGCJ · Gastroenterology, Urology
Decision DateApr 28, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

To provide light for fiberoptic cables and provide a camera for procedure video capabilities. Provide light for instrumentation via fiberoptic cables For for use in. surgical fields or in office procedures. lightsource attaches to fiberoptic cable which is then attached to a video scope, the camera also attaches to the video scope (such as a video otoscope, or an intra oral dental scope).

Device Story

Device is a combination video camera and light source (Model VCL-150). Light source provides illumination via fiberoptic cables connected to video scopes; camera captures video signals from the scope. Used in surgical or office settings by clinicians. Output is a video feed for visualization of the procedure site. Benefits include enhanced visualization during endoscopic or scope-based examinations.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Combination unit consisting of a light source and video camera. Interfaces with fiberoptic cables and video scopes (e.g., otoscopes, dental scopes).

Indications for Use

Indicated for use in surgical fields or office procedures requiring illumination and video visualization via fiberoptic cables and video scopes (e.g., video otoscopes, intra-oral dental scopes).

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an eagle with its wings spread, facing to the left. The eagle is stylized and appears to be made up of three separate lines. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 APR 28 1998 Ms. Kimberly Reed Cuda Products Corporation 6000 Powers Avenue Jacksonville, Florida 32217-2279 Re: K980350 Video Camera/Lightsource Model VCL-150 Trade Name: Requlatory Class: II Product Code: GCJ Dated: December 10, 1997 Received: January 28, 1998 Dear Ms. Reed: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਮ substantially equivalent determination assumes compliance with the current Good Manufacturinq Practice requirement, as set . . forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {1}------------------------------------------------ Page 2 - Ms. Reed This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Celia M. Witten, Ph.D., a M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ Page_of_of_ 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ Video Camera/Lightsource Combination Unit VCL-150 Device Name: ## Indications For Use: To provide light for fiberoptic cables and provide a camera for procedure video capabilities. Provide light for instrumentation via fiberoptic cables For for use in. surgical fields or in office procedures. lightsource attaches to fiberoptic cable which is then attached to a video scope, the camera also attaches to the video scope (such as a video otoscope, or an intra oral dental scope). (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use ----- (Optional Format 1-2-96)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...