ENDO VIP

K963323 · Richard Wolf Medical Instruments Corp. · GCJ · Nov 21, 1996 · Gastroenterology, Urology

Device Facts

Record IDK963323
Device NameENDO VIP
ApplicantRichard Wolf Medical Instruments Corp.
Product CodeGCJ · Gastroenterology, Urology
Decision DateNov 21, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

Video Image Processor ENDO VIP 5588 is a video contrast enhancer and is used in connection with a video system. It enhances the detail definition of medical video images. It is used as an accessory during diagnostic and therapeutic interventions in video endoscopy and video microscopy.

Device Story

ENDO VIP 5588; video contrast enhancer; connects between video camera and monitor; processes Y/C video signals (NTSC/PAL) in real-time; enhances detail definition of medical video images; manual 3-step enhancement control; used in video endoscopy and microscopy; assists clinicians in visualizing diagnostic/therapeutic interventions.

Clinical Evidence

No clinical tests performed. Bench testing conducted per IEC601-1, IEC601-1-2, and UL2601-1 standards.

Technological Characteristics

Digital video processor; Y/C video signal input/output; NTSC/PAL compatible; real-time processing; manual 3-step enhancement control; fixed algorithm with user-variable parameters; electrical safety per IEC601-1, IEC601-1-2, UL2601-1.

Indications for Use

Indicated for use as an accessory during diagnostic and therapeutic interventions in video endoscopy and video microscopy to enhance detail definition of medical video images.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 510(k) Summary of Safety and Effectiveness 15963323 NOV. 21, 1996 1.0 **Classification Name:** Endoscope and/or accessories 2.0 **Common/Usual Name:** Video Image Processor/Video Enhancer 3.0 **Trade Name:** ENDO VIP 4.0 **Equivalence:** The device is equivalent to devices sold by Karl Storz. 5.0 **Description:** The Video Image Processor ENDO VIP 5588 enhances the detail definition of medical video images. It is connected between video camera and video monitor. The enhancement level is manual controlled in 3 steps. 6.0 **Intended Use:** Video Image Processor ENDO VIP 5588 is a video contrast enhancer and is used in connection with a video system. It enhances the detail definition of medical video images. It is used as an accessory during diagnostic and therapeutic interventions in video endoscopy and video microscopy. 7.0 **Technological Characteristics:** The Video Image Processor ENDO VIP 5588 has a digital video processor and uses a fixed algorithm with user-variable parameters to improve detail visibility in video images. Y/C video signals, which are compatible with NTSC and PAL video standards, are processed in real time. 8.0 **Performance Data:** Independent laboratories tested the Video Image Processor ENDO VIP 5588 according to specified standards IEC601-1, IEC601-1-2 and UL2601-1. 9.0 **Clinical Tests:** No clinical tests performed. 10.0 **Data Conclusion:** Video Image Processor ENDO VIP 5588 is designed and tested to guarantee the safety and effectiveness, if the device is used according to the instruction manual. By: Robert L. Casarsa Date: Aug 22, 1996
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...