DEXTERITY PNEUMO SLEEVE SET

K962147 · Medical Creative Technologies, Inc. · GCJ · Jul 9, 1996 · Gastroenterology, Urology

Device Facts

Record IDK962147
Device NameDEXTERITY PNEUMO SLEEVE SET
ApplicantMedical Creative Technologies, Inc.
Product CodeGCJ · Gastroenterology, Urology
Decision DateJul 9, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Dexterity® Pneumo Sleeve is intended to provide abdominal access for the surgeon’s hand while preserving pneumoperitoneum during laparoscopic surgery. The Protector® Retractor is intended to provide incision retraction and to protect it against wound contamination during both laparoscopic and open surgery. This device allows the surgeon to use his hand while performing corrective surgery in the abdominal cavity under pneumoperitoneum.

Device Story

Device consists of Dexterity® Pneumo Sleeve and Protector® Retractor; used in laparoscopic surgery. Pneumo Sleeve: sterile, disposable plastic film sleeve with adhesive flange for abdominal attachment and velcro strap for sealing around surgeon's arm; maintains pneumoperitoneum during hand-assisted access. Protector® Retractor: sterile, disposable open-ended cylinder with flexible rings; inserted into incision to retract wound and prevent contamination. Operated by surgeons in OR. Device enables hand-assisted laparoscopic procedures; allows tactile feedback while maintaining gas seal. Benefits include potential for reduced conversion to open surgery and improved surgical access compared to standard laparoscopic instruments.

Clinical Evidence

Clinical evidence includes a pilot study (n=40) and a pivotal prospective, randomized, controlled trial across 10 centers in the US and Europe comparing hand-assisted laparoscopic colon resections to routine laparoscopic surgery. Endpoints included procedure time, instrument changes, conversion rates, costs, recovery time, morbidity, and mortality. Results confirmed safety and effectiveness.

Technological Characteristics

Materials: 3 mil plastic film, flexible plastic, acrylate adhesive. Components: adhesive flange, velcro strap, flexible rings. Sterile, disposable. No electronic or software components.

Indications for Use

Indicated for patients undergoing laparoscopic surgery requiring hand-assisted access to the abdominal cavity while maintaining pneumoperitoneum.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{1} JUL 9 1996 APPENDIX H K962147 # 510(k) SUMMARY This summary is organized per 21 CFR 807.92. Medical Creative Technologies, Inc. 2950-N Advance Lane Colmar, PA 18915 Telephone Number: 215-997-9689 Fax Number: 215-997-2371 Name of Contact: Perk Crook Date Summary Prepared: April 3, 1996 Name of Device: Trade Name: Dexterity® Pneumo Sleeve and Protector® Retractor Common Name: Extended Laparoscopy Device Classification Name: Extended Laparoscopy Device Identification of Predicate or Legally Marketed Device(s): 1. Trocar/Cannula by Storz 2. Protector® Retractor by Medical Creative Technologies, Inc. The Dexterity® Pneumo Sleeve and Protector® Retractor are substantially equivalent to a Trocar/Cannula from an access to the peritoneal cavity standpoint. Both devices allow access to the peritoneal cavity while preventing CO₂ leakage during insertion and withdrawal of instruments. In the case of the Trocar/Cannula, the instruments are laparoscopic instruments and in the case of Dexterity® Pneumo Sleeve Set 510(k) 5/31/96 {2} the Dexterity® Pneumo Sleeve and Protector® Retractor, the instrument is the surgeon’s hand. The Trocar/Cannula protects the wound as does the Protector® Retractor from trauma and/or cross-contamination. The Protector® Retractor which is a component of this device is the same device as the Protector® Retractor (second predicate device named above) approved by 510(k) Number K954824. The Dexterity® Pneumo Sleeve is made from the same 3 mil thick plastic film used to make the sleeve of the Protector® Retractor. ## Description of the Device: The Dexterity® Pneumo Sleeve is a sterile, disposable device. It prevents leakage of $\mathrm{CO}_{2}$ and provides an extension of the pneumoperitoneum. It has an adhesive flange at one end that is placed over the incision and adhered to the patient’s abdomen. A velcro strap at the other end is used to tighten the Sleeve around the surgeon’s arm. The Protector® Retractor is a sterile, disposable device. The Retractor consists of an open-ended cylinder with a flexible ring at each end. One ring is inserted through the incision into the peritoneal cavity while the other remains outside the incision. The Retractor holds the incision open and protects it against wound contamination. ## Statement of Intended Use: The Dexterity® Pneumo Sleeve is intended to provide abdominal access for the surgeon’s hand while preserving pneumoperitoneum during laparoscopic surgery. The Protector® Retractor is intended to provide incision retraction and to protect it against wound contamination during both laparoscopic and open surgery. This device allows the surgeon to use his hand while performing corrective surgery in the abdominal cavity under pneumoperitoneum. ## Summary of Technical Characteristics of this Device Compared to the Predicate Device: The Dexterity® Pneumo Sleeve and Protector® Retractor are technically the same as the Trocar/Cannula (one of the predicate devices) in that the Trocar/Cannula Dexterity® Pneumo Sleeve Set 510(k) 5/31/96 {3} allows access to the abdominal cavity for laparoscopic instruments and the Dexterity® Pneumo Sleeve and Protector® Retractor allow access to the abdominal cavity for the surgeon’s hand. The Trocar/Cannula is made out of stainless steel or plastic while the Dexterity® Pneumo Sleeve and Protector® Retractor are made out of a plastic film, flexible plastic and an acrylate adhesive. The Protector® Retractor component of the Dexterity® Pneumo Sleeve and Protector® Retractor is the "Protector® Retractor" shown as the second predicate device above. The material making up the sleeve of the Protector® Retractor is the same material as used to make the Dexterity® Pneumo Sleeve. ## Summary of Performance Data: ### Non-clinical Test Results: Bench testing was performed on each component of the Dexterity® Pneumo Sleeve and Protector® Retractor to assure that they would function as intended during use. Based on the data collected in the bench testing, it was concluded that all of the materials were proper for their intended use. ### Animal Test Results: #### Functional testing on pigs: Several studies were accomplished using pigs to further verify the function of the device and its components. The results of these studies were effective in fine-tuning the design and working out surgeons’ preferences. #### Clinical Results: Two clinical studies were conducted with the Dexterity® Pneumo Sleeve. An initial pilot study involving ten surgeons and forty patients validated the operating principles of the device and allowed surgeons to gain additional device experience. As a result of this original study, minor device changes were implemented in the device design. Dexterity® Pneumo Sleeve Set 510(k) 5/31/96 {4} A second, pivotal clinical study comparing extended laparoscopic surgery using the Dexterity® Pneumo Sleeve and Protector® Retractor with routine laparoscopic surgery for colon resections, was conducted at 10 clinical centers in the United States and Europe. Safety and effectiveness endpoints considered in this trial included: overall procedure time, number of instrument changes required, conversion rates, i.e., the number of times a case could not be completed laparoscopically and had to be converted to an open procedure, direct operation costs, time for patients to return to normal activity, morbidity and mortality. The results of this prospective, randomized, controlled trial confirmed the safety and effectiveness of the Dexterity® Pneumo Sleeve Set for its intended use as an adjunct to laparoscopic surgery. Dexterity® Pneumo Sleeve Set 510(k) 5/31/96
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...