← Product Code [GCJ](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCJ) · K223718

# HandX Self-Righting Needle Holder (K223718)

_Human Xtensions , Ltd. · GCJ · Jun 5, 2023 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/GCJ/K223718

## Device Facts

- **Applicant:** Human Xtensions , Ltd.
- **Product Code:** [GCJ](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCJ.md)
- **Decision Date:** Jun 5, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The HandX™ is intended to assist in the control of Human Xtensions laparoscopic instruments including needle holder and grasper, for endoscopic manipulation of tissue, including grasping, approximation, ligation, suturing, during laparoscopic surgical procedures. It is intended to be used by trained physicians in an operating room environment in accordance with its Instructions for Use.

## Device Story

HandX™ Self-Righting Needle Holder is a single-use, disposable, ETO-sterilized laparoscopic instrument; connects to HandX™ electromechanical system. System translates motor rotations into end-effector articulation; specifically orients suturing needles to 90° vertical position. Used in operating rooms by trained physicians; assists in tissue manipulation and suturing. Surgeon controls device via HandX™ interface; output is mechanical articulation of the needle holder tip. Benefits include improved needle positioning and control during minimally invasive suturing tasks.

## Clinical Evidence

Bench testing only. Testing included articulation angle measurement, dimensional verification, suture holding force, needle pulling moment, needle holding force, and needle righting transition ability. All tests met predefined acceptance criteria.

## Technological Characteristics

Stainless steel shaft and electrode tip; 330mm length; 5.5mm diameter. Electromechanically operated via HandX™ system; software-controlled. Features 360° axial rotation and 0-80° articulation. Single-use; ETO sterilized (SAL 10^-6).

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- HX Device, Needle Holder, Human Xtensions Ltd. ([K173919](/device/K173919.md))

## Reference Devices

- ENDOPATH® Self-Righting Needle Holder ([K972679](/device/K972679.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

June 5, 2023

Human Xtensions Ltd. Orit Yaniv VP QA/RA Human Xtensions Ltd. Grand Netter Building 4 Meir Ariel Street Netanya, 4250574, Israel Telephone: + (972) 77 36 30 300 Email: orit@human-x.com

Re: K223718

Trade/Device Name: HandXTM Self-Righting Needle Holder Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: GCJ Dated: December 11, 2022 Received: December 12, 2022

Dear Orit Yaniv:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Digitally signed by Mark Mark Trumbore -S Trumbore -S Date: 2023.06.05
11:23:01 -04'00'

Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

Device Name HandX(TM) Self-Righting Needle Holder

Indications for Use (Describe)

The HandX is intended to assist in the control of Human Xtensions laparoscopic instruments including needle holder and grasper, for endoscopic manipulation of tissue, including grasping, approximation, ligation, suturing, during laparoscopic surgical procedures. It is intended to be used by trained physicians in an operating room environment in accordance with its Instructions for Use.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

{3}------------------------------------------------

#### Premarket Notification 510(k) Summary

| Submitter:                                                                                                                         | Human Xtensions Ltd.<br>Grand Netter Building<br>4 Meir Ariel Street<br>Netanya, 4250574, Israel<br>Telephone: + (972) 77 36 30 300                                          |
|------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact:                                                                                                                           | Orit Yaniv<br>Human Xtensions Ltd.<br>Grand Netter Building<br>4 Meir Ariel Street<br>Netanya, 4250574, Israel<br>Telephone: + (972) 77 36 30 300<br>Email: orit@human-x.com |
| Date Prepared:                                                                                                                     | May 24, 2023                                                                                                                                                                 |
| Subject Device:                                                                                                                    |                                                                                                                                                                              |
| Trade/Device Name:<br>Common/Usual Name:<br>Device Classification Name:<br>Product Codes:<br>Product Class:<br>Regulation Numbers: | HandX™ Self-Righting Needle Holder<br>Articulated Laparoscopic Instruments<br>Endoscope and Accessories<br>GCJ<br>Class II<br>21 CFR Parts 876.1500                          |
| Predicate Devices:                                                                                                                 | HX Device, Needle Holder, Human Xtensions Ltd.<br>(K173919)                                                                                                                  |
| Reference Device:                                                                                                                  | ENDOPATH® Self-Righting Needle Holder (K972679)<br>Ethicon. Used as a reference device for Self-Righting<br>function.                                                        |

#### Device Description:

The Self-Righting Needle Holder is a single-use, disposable, ETO sterilized instrument for use with the HandX™ device. The Self-Righting Needle Holder is connected to the HandX™ and transmits the motors' rotations to articulate the movement of the Self-Righting Needle Holder's end effector. Self-Righting Needle Holder orients the suturing needle vertically to 90° mounting. It is designed to address the surgeons' needs relating to suturing for various surgical purposes.

Indications for Use:

{4}------------------------------------------------

The HandX™ is intended to assist in the control of Human Xtensions laparoscopic instruments including needle holder and grasper, for endoscopic manipulation of tissue, including grasping, approximation, ligation, suturing, during laparoscopic surgical procedures. It is intended to be used by trained physicians in an operating room environment in accordance with its Instructions for Use.

# Comparison of Technological Characteristics with the Predicate Devices:

The Self-Righting Needle Holder is substantially equivalent to the predicate device, Human Xtensions' HX Device (K173919), in its intended use, contraindications, fundamental technology, packaging, sterilization method and general characteristics. Its Self-Righting function is the same as Ethicon's ENDOPATH® Self-Righting Needle Holder (K972679).

A summary of the technological characteristics of the Self-Righting Needle Holder device in comparison to those of the predicate device and reference device is presented in the table below.

| Description         | HandX™ Self-Righting Needle<br>Holder<br>(Subject device)                                                                                                                                                                                                                                                                                                                                             | HX Device, Needle Holder<br>(Predicate device)                                                                                                                                                                                                                                                                                                                                                       | ENDOPATH® Self-<br>Righting Needle<br>Holder<br>(Reference device)                                                                                                                             |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number       | K223718                                                                                                                                                                                                                                                                                                                                                                                               | K173919                                                                                                                                                                                                                                                                                                                                                                                              | K972679                                                                                                                                                                                        |
| Product Code        | GCJ                                                                                                                                                                                                                                                                                                                                                                                                   | GCJ                                                                                                                                                                                                                                                                                                                                                                                                  | GAT                                                                                                                                                                                            |
| CFR                 | 21 CFR 876.1500                                                                                                                                                                                                                                                                                                                                                                                       | 21 CFR 876.1500                                                                                                                                                                                                                                                                                                                                                                                      | 21 CFR 876.4730                                                                                                                                                                                |
| Indications for Use | The HandX™ is intended to assist in the control of Human Xtensions laparoscopic instruments including needle holder and grasper, for endoscopic manipulation of tissue, including grasping, approximation, ligation, suturing, during laparoscopic surgical procedures. It is intended to be used by trained physicians in an operating room environment in accordance with its Instructions for Use. | The HX Device is intended to assist in the accurate control of HX laparoscopic instruments including needle holder and grasper, for endoscopic manipulation of tissue, including grasping, approximation, ligation, suturing, during laparoscopic surgical procedures. It is intended to be used by trained physicians in an operating room environment in accordance with its Instructions for Use. | The ENDOPATH® Endoscopic Tissue Fastening System (ETFS) is intended for use in minimally invasive surgical applications where soft tissue is being approximated with the interrupted stitches. |
| Contra- indications | This device is intended for use only as indicated.                                                                                                                                                                                                                                                                                                                                                    | This device is intended for use only as indicated.                                                                                                                                                                                                                                                                                                                                                   | NA                                                                                                                                                                                             |
| Specialty           | Laparoscopic                                                                                                                                                                                                                                                                                                                                                                                          | Laparoscopic                                                                                                                                                                                                                                                                                                                                                                                         | Minimally invasive surgical application                                                                                                                                                        |
| Mode of operation   | Electromechanically operated, software controlled                                                                                                                                                                                                                                                                                                                                                     | Electromechanically operated, software controlled                                                                                                                                                                                                                                                                                                                                                    | Manually                                                                                                                                                                                       |
| Insertion Method    | Via 5mm Trocar                                                                                                                                                                                                                                                                                                                                                                                        | Via 5mm Trocar                                                                                                                                                                                                                                                                                                                                                                                       | Via 5mm Trocar                                                                                                                                                                                 |
| Articulation        | Yes                                                                                                                                                                                                                                                                                                                                                                                                   | Yes                                                                                                                                                                                                                                                                                                                                                                                                  | No                                                                                                                                                                                             |

{5}------------------------------------------------

| Description                     | HandX™ Self-Righting Needle<br>Holder<br>(Subject device) | HX Device, Needle Holder<br>(Predicate device) | ENDOPATH® Self-<br>Righting Needle<br>Holder<br>(Reference device) |
|---------------------------------|-----------------------------------------------------------|------------------------------------------------|--------------------------------------------------------------------|
| Articulation<br>range           | 0 to 80°±10°                                              | 0 to 85°±5°                                    | NA                                                                 |
| Articulation<br>locking feature | Yes                                                       | Yes                                            | NA                                                                 |
| Tip axial<br>rotation feature   | 360°                                                      | 360°                                           | NA                                                                 |
| Shaft material                  | Stainless steel                                           | Stainless steel                                | Stainless steel                                                    |
| Instrument type                 | Self-Righting Needle Holder                               | Needle Holder                                  | Self-Righting Needle<br>Holder                                     |
| Material for<br>electrode tip   | Stainless steel                                           | Stainless steel                                | Stainless steel                                                    |
| Shaft length                    | 330 mm ±0.1 mm                                            | 330 mm ±0.1 mm                                 | NA                                                                 |
| Shaft diameter                  | 5.5 mm ±0.2 mm                                            | 5.5 mm ±0.1 mm                                 | NA                                                                 |
| Single Use /<br>Reusable        | Single use instrument                                     | Single use instrument                          | Re-usable                                                          |
| Packaging                       | Blister                                                   | Blister                                        | NA                                                                 |
| Provided<br>sterilized          | Yes                                                       | Yes                                            | NA                                                                 |
| Sterilization<br>method         | EtO                                                       | EtO                                            | NA                                                                 |
| Sterility<br>assurance level    | 10-6                                                      | 10-6                                           | NA                                                                 |

lt can be concluded that the Self-Righting Needle Holder is substantially equivalent to its predicate. No new questions of safety and effectiveness were raised.

# Performance Data:

The following performance data were provided in support of the substantial equivalence determination:

#### Biocompatibility Testing:

The materials of the Self-Righting Needle Holder are identical to those of the predicate device. The predicate device has been tested and is considered biocompatible for its intended use. Therefore, from a biocompatibility and toxicological perspective, the device remains unchanged, and no additional 

{6}------------------------------------------------

testing is required for the HandX™ Self-righting Needle Holder device since only the jaws geometry of the needle was changed.

# Bench Testing:

The Self-Righting Needle Holder was subjected to bench testing to evaluate its performance and demonstrate that the design outputs meet the design input requirements, and that the device is safe and effective for its intended use.

The verification bench testing is described in the table below, all tests met the predefined acceptance criteria. The testing identified no new questions of safety and effectiveness.

| Test                                      | Test Summary                                                                                                                          | Conclusions                                                                                                                                 |
|-------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|
| Self-Righting Needle Holder<br>Angle Test | The purpose of the test was<br>to measure the subject<br>device articulation angles<br>while connected to the<br>HandX device.        | The subject device tip<br>articulation ability is<br>comparable to the predicate<br>device.                                                 |
| SRNH instrument<br>Dimensional Attribute  | The purpose of the test was<br>to confirm the subject<br>device dimensions to the<br>specifications                                   | The subject device<br>dimensions are within the<br>defined requirements.                                                                    |
| SRNH Suture Holding Force                 | The purpose of the test was<br>to measure the subject<br>device suture holding force.                                                 | The subject device suture<br>holding force withstands the<br>defined requirements as the<br>predicate device.                               |
| SRNH Needle Pulling<br>Moment             | The purpose of this test was<br>to determine whether the<br>subject device maintains<br>the needle in place under<br>external moment. | The subject device complies<br>with the needle pulling<br>moment requirement.                                                               |
| SRNH Needle Holding Force                 | The purpose of this test was<br>to determine whether the<br>subject device maintains<br>the needle in place under<br>external force.  | The subject device complies<br>with the needle holder<br>holding force requirement                                                          |
| SRNH Needle Righting                      | The purpose of the test was<br>to verify the subject device<br>transition ability of a<br>surgical needle to a right<br>position.     | The subject device<br>demonstrated transition<br>ability of surgical needles to<br>right position, thus met the<br>predefined requirements. |

# Conclusion:

Based upon the intended use, indications, technological characteristics and performance testing, as well as comparison to the predicate device, it can be concluded that the Self-Righting Needle Holder is substantially equivalent to its predicate device. The minor differences between the subject and predicate devices do not raise any questions of safety and effectiveness.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/GCJ/K223718](https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/GCJ/K223718)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
