← Product Code [GCJ](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCJ) · K223593

# Inno-Port Disposable Bladeless Trocar, Inno-Port Disposable Optical Trocar (K223593)

_Taiwan Surgical Corporation · GCJ · Mar 22, 2023 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/GCJ/K223593

## Device Facts

- **Applicant:** Taiwan Surgical Corporation
- **Product Code:** [GCJ](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCJ.md)
- **Decision Date:** Mar 22, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The Inno-Port Disposable Bladeless Trocars are intended for use in a variety of gynecologic, general thoracic and urologic endoscopic procedures to create and maintain a port of entry. The Inno-Port Disposable Optical Trocars are intended for use in a variety of gynecologic, general thoracic and urologic endoscopic procedures to create and maintain a port of entry. The trocar may be used with or without visualization for primary and secondary insertions.

## Device Story

Disposable surgical trocar system; includes bladeless and optical variants. Used in gynecologic, general thoracic, and urologic endoscopic procedures to create/maintain access port. Consists of obturator and cannula with self-adjusting seal and stopcock valve for insufflation/desufflation. Optical variant includes scope retention mechanism for visualization during insertion. Operated by surgeons in clinical settings. Cannula seal prevents pneumoperitoneum loss during instrument exchange. Provides stable access for endoscopic instruments; facilitates minimally invasive surgical workflows.

## Clinical Evidence

Bench testing only. Performance metrics compared against predicates include puncture force, removal force, obturator insertion/withdrawal force, stopcock/duckbill/sealing airtightness, and cannula durability. All results met acceptance criteria and demonstrated substantial equivalence to predicate devices.

## Technological Characteristics

Disposable trocar; 5mm, 5-11mm, and 5-12mm configurations. Materials evaluated per ISO 10993-1. Features self-adjusting seal, stopcock valve, and optional scope retention mechanism. Sterilization via Gamma radiation. Mechanical operation; no software or electronic components.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- VeraPort™ V2 Bladeless Optical Trocar ([K130435](/device/K130435.md), [K112349](/device/K112349.md))
- VersaOne™ Bladeless Trocar ([K151548](/device/K151548.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 22, 2023

Taiwan Surgical Corporation Ken Chen Project Director 3F., No. 12, Sec. 2 Sheng Yi Rd. Zhubei City, Hsinchu County 30261 Taiwan

Re: K223593

Trade/Device Name: Inno-Port Disposable Bladeless Trocar, Inno-Port Disposable Optical Trocar Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: GCJ Dated: November 29, 2022 Received: December 1, 2022

Dear Ken Chen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

542 of the Act); 21 CFR 1000-1050.

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Digitally signed by Mark Mark Trumbore -S Trumbore -S Date: 2023.03.22 15:37:30 -04'00'

Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K223593

Device Name

Inno-Port Disposable Bladeless Trocar and Inno-Port Disposable Optical Trocar

### Indications for Use (Describe)

The Inno-Port Disposable Bladeless Trocars are in a variety of gynecologic, general thoracic and urologic endoscopic procedures to create and maintain a port of entry.

The Inno-Port Disposable Optical Trocars are intended for use in a variety of gynecologic, general thoracic and urologic endoscopic procedures to create and maintain a port of entry. The trocar may be used with or without visualization for primary and secondary insertions.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for TWSC, which is written in a blue sans-serif font. Below the logo is the text "Treat Patient As Family" in a smaller, gray sans-serif font. The logo is simple and modern, and the text below it conveys a message of care and compassion.

# 510(k) Summary

# The Assigned 510(k) Number: K223593 Date Prepared: 11/02/2022

This 510(k) summary of safety and effectiveness information is submitted as part of the Premarket Notification in compliance with requirements of CFR Part 807, Subpart E and Section 807.92

#### SUBMITTER: 1

| Submitter:                   | TAIWAN SURGICAL CORPORATION                                                      |
|------------------------------|----------------------------------------------------------------------------------|
| Mailing Address:             | 3F., No.12, Sec.2, ShengYi Rd., Zhubei City, Hsinchu County 302,<br>Taiwan       |
| Phone Number:                | +886-3-6588129                                                                   |
| Fax Number:                  | +886-3-6588355                                                                   |
| Contact Person:              | Ken Chen                                                                         |
| Title:                       | Project Director                                                                 |
| Address:                     | 3F., No.12, Sec.2, ShengYi Rd., Zhubei City, Hsinchu County 302<br>Taiwan        |
| Date of submission:          | 03/20/2023                                                                       |
| Email:                       | ra@twsc.com.tw                                                                   |
| Phone Number:                | +886-3-6588129 ext. 100                                                          |
| Fax Number:                  | +886-3-3588355                                                                   |
| DEVICE                       |                                                                                  |
| Trade Name:                  | Inno-Port Disposable Bladeless Trocar and Inno-Port Disposable<br>Optical Trocar |
| Common Name:                 | Surgical Trocar                                                                  |
| Panel Number:                | 78 Gastroenterology and Urology                                                  |
| Classification Name:         | 21 CFR Part 876.1500 Endoscope and accessories                                   |
| Classification Product Code: | GCJ                                                                              |

#### PREDICATE DEVICE 3

Device Class:

2

VeraPort™ V2 Bladeless Optical Trocar (11&1 2mm: K130435, 5 mm: K112349)

II

VersaOne™ Bladeless Trocar (K151548)

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for TWSC, which is written in a large, sans-serif, blue font. Below the logo is the text "Treat Patient As Family" in a smaller, gray, sans-serif font. The logo is simple and professional, and the text below it conveys the company's values.

#### DEVICE DESCRIPTION 4

## Inno-Port Disposable Bladeless Trocar

The Inno-Port Disposable Bladeless Trocars are available in the following configurations:

| Series            | Model No. | Product Description and Specification                                       |
|-------------------|-----------|-----------------------------------------------------------------------------|
| 5 mm<br>series    | DT501-7B  | Disposable Trocar, 5mm, Length 70mm, Bladeless (Obturator*1, Cannula*1)     |
|                   | DT501-XB  | Disposable Trocar, 5mm, Length 100mm, Bladeless (Obturator*1, Cannula*1)    |
|                   | DT501-7L  | Disposable Trocar, 5mm, Length 70mm, Bladeless (Obturator*1, Cannula*2)     |
|                   | DT501-XL  | Disposable Trocar, 5mm, Length 100mm, Bladeless (Obturator*1, Cannula*2)    |
|                   | DT501-71  | Disposable Trocar, 5mm, Length 70mm (Cannula*1)                             |
|                   | DT501-X1  | Disposable Trocar, 5mm, Length 100mm (Cannula*1)                            |
| 5-11 mm<br>series | DTW01-XB  | Disposable Trocar, 5-11mm, Length 100mm, Bladeless (Obturator*1, Cannula*1) |
|                   | DTW01-SB  | Disposable Trocar, 5-11mm, Length 150mm, Bladeless (Obturator*1, Cannula*1) |
|                   | DTW01-XL  | Disposable Trocar, 5-11mm, Length 100mm, Bladeless (Obturator*1, Cannula*2) |
|                   | DTW01-SL  | Disposable Trocar, 5-11mm, Length 150mm, Bladeless (Obturator*1, Cannula*2) |
|                   | DTW01-X1  | Disposable Trocar, 5-11mm, Length 100mm (Cannula*1)                         |
|                   | DTW01-S1  | Disposable Trocar, 5-11mm, Length 150mm (Cannula*1)                         |
| 5-12 mm<br>series | DTY01-XB  | Disposable Trocar, 5-12mm, Length 100mm, Bladeless (Obturator*1, Cannula*1) |
|                   | DTY01-SB  | Disposable Trocar, 5-12mm, Length 150mm, Bladeless (Obturator*1, Cannula*1) |
|                   | DTY01-XL  | Disposable Trocar, 5-12mm, Length 100mm, Bladeless (Obturator*1, Cannula*2) |
|                   | DTY01-SL  | Disposable Trocar, 5-12mm, Length 150mm, Bladeless (Obturator*1, Cannula*2) |
|                   | DTY01-X1  | Disposable Trocar, 5-12mm, Length 100mm (Cannula*1)                         |

The trocar cannula contains an internal seal to prevent loss of pneumoperitoneum when instruments are inserted or withdrawn. The seal system in the Inno-Port Disposable Bladeless Trocar is self-adjusting and accommodates instruments ranging from 5mm in diameter for trocars marked as 5mm; 5mm to 11mm in diameter for trocars marked as 11mm and 5mm to 12mm in diameter on trocars marked as 12mm. There is a stopcock valve for insufflation and rapid desufflation.

{5}------------------------------------------------

Traditional 510(k), K223593

Image /page/5/Picture/1 description: The image shows the logo for TWSC, which is written in large, blue, sans-serif font. Below the logo, in a smaller, light gray font, is the phrase "Treat Patient As Family". The logo is simple and professional, and the phrase conveys a message of care and compassion.

## Inno-Port Disposable Optical Trocar

| Series            | Model No. | Product Description and Specification                                     |
|-------------------|-----------|---------------------------------------------------------------------------|
| 5 mm series       | DT501-7Q  | Disposable Trocar, 5mm, Length 70mm, Optical (Obturator*1, Cannula*1)     |
|                   | DT501-XQ  | Disposable Trocar, 5mm, Length 100mm, Optical (Obturator*1, Cannula*1)    |
|                   | DT501-7P  | Disposable Trocar, 5mm, Length 70mm, Optical (Obturator*1, Cannula*2)     |
|                   | DT501-XP  | Disposable Trocar, 5mm, Length 100mm, Optical (Obturator*1, Cannula*2)    |
| 5-11 mm<br>series | DTW01-XQ  | Disposable Trocar, 5-11mm, Length 100mm, Optical (Obturator*1, Cannula*1  |
|                   | DTW01-SQ  | Disposable Trocar, 5-11mm, Length 150mm, Optical (Obturator*1, Cannula*1) |
|                   | DTW01-XP  | Disposable Trocar, 5-11mm, Length 100mm, Optical (Obturator*1, Cannula*2) |
|                   | DTW01-SP  | Disposable Trocar, 5-11mm, Length 150mm, Optical (Obturator*1, Cannula*2  |
| 5-12 mm<br>series | DTY01-XQ  | Disposable Trocar, 5-12mm, Length 100mm, Optical (Obturator*1, Cannula*1) |
|                   | DTY01-SQ  | Disposable Trocar, 5-12mm, Length 150mm, Optical (Obturator*1, Cannula*1  |
|                   | DTY01-XP  | Disposable Trocar, 5-12mm, Length 100mm, Optical (Obturator*1, Cannula*2) |
|                   | DTY01-SP  | Disposable Trocar, 5-12mm, Length 150mm, Optical (Obturator*1, Cannula*2) |

The trocar cannula contains an internal seal to prevent loss of pneumoperitoneum when instruments are inserted or withdrawn. The seal system in the Inno-Port Disposable Optical Trocar is self-adjusting and accommodates instruments ranging from in diameter for trocars marked as 5mm; 5mm to 11mm in diameter for trocars marked as 11mm and 5mm to 12mm in diameter on trocars marked as 12mm. The obturator contains a scope retention mechanism. There is a stopcock valve for insufflation and rapid desufflation.

#### INTENDED USE 5

### Inno-Port Disposable Bladeless Trocar

The Inno-Port Disposable Bladeless Trocars are intended for use in a variety of gynecologic, general thoracic and urologic endoscopic procedures to create and maintain a port of entry.

### Inno-Port Disposable Optical Trocar

The Inno-Port Disposable Optical Trocars are intended for use in a variety of gynecologic, general thoracic and urologic endoscopic procedures to create and maintain a port of entry. The trocar may be used with or without visualization for primary and secondary insertions.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for TWSC, which is written in blue, sans-serif font. Below the logo is the phrase "Treat Patient As Family" in a smaller, gray, sans-serif font. The logo is simple and modern, and the phrase below it conveys a message of care and compassion.

### 7 MATERIAL

All patient contacting materials have been evaluated according to ISO 10993-1 and the FDA's Guidance Use of International Standard ISO 10993-1, dated Sep. 04, 2020. All biocompatibility met the acceptance criteria.

### TECHNOLOGICAL CHARACTERISTICS 8

A comparison of the technological characteristics of the proposed Inno-Port Disposable Bladeless Trocar, Inno-Port Disposable Optical Trocar, and the predicate has been performed. The results of this comparison demonstrate that the Inno-Port Disposable Bladeless Trocar, Inno-Port Disposable Optical Trocar have similar technological characteristics compared to the marketed predicate device.

| Manufactures     | Taiwan Surgical Corp.                                                                                                                                                                                                                                                                                                                                                                        | Covidien                                                                                                                                                                                                                                                                                                                                                                                      |
|------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Name     | Inno-Port Disposable Bladeless Trocar and<br>Inno-Port Disposable Optical Trocar                                                                                                                                                                                                                                                                                                             | VersaOne™ Bladeless Trocar and<br>Versaport™ V2 Bladeless Optical Trocar                                                                                                                                                                                                                                                                                                                      |
|                  |                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                               |
| Certification    | --                                                                                                                                                                                                                                                                                                                                                                                           | K151548<br>K112349<br>K130435                                                                                                                                                                                                                                                                                                                                                                 |
| Device Class     | 2                                                                                                                                                                                                                                                                                                                                                                                            | 2                                                                                                                                                                                                                                                                                                                                                                                             |
| Intended use     | Inno-Port Disposable Bladeless Trocar:<br>The Inno-Port Disposable Bladeless<br>Trocars are intended for use in a variety of<br>gynecologic, general thoracic and<br>urologic endoscopic procedures to create<br>and maintain a port of entry.                                                                                                                                               | VersaOne™ Bladeless Trocar:<br>The VersaOne™ Bladeless Trocars are<br>intended for use in a variety of<br>gynecologic, general, thoracic and<br>urologic endoscopic procedures to create<br>and maintain a port of entry.                                                                                                                                                                     |
|                  | Inno-Port Disposable Optical Trocar:<br>The Inno-Port Disposable Optical Trocars<br>are intended for use in a variety of<br>gynecologic, general thoracic and<br>urologic endoscopic procedures to create<br>and maintain a port of entry. The trocar<br>may be used with or without visualization<br>for primary and secondary insertions.                                                  | Versaport™ V2 Bladeless Optical Trocar:<br>The Versaport™ V2 Bladeless Optical<br>Trocars are intended for use in a variety of<br>gynecologic, general, thoracic and<br>urologic endoscopic procedures to create<br>and maintain a port of entry. The trocar<br>may be used with or without visualization<br>for primary and secondary insertions.                                            |
| Components       | 5 mm: Obturator, Cannula with Self-<br>Adjusting Seal and Stopcock.                                                                                                                                                                                                                                                                                                                          | 5 mm: Obturator, Cannula with Self-<br>Adjusting Seal and Stopcock.                                                                                                                                                                                                                                                                                                                           |
| Manufactures     | Taiwan Surgical Corp.                                                                                                                                                                                                                                                                                                                                                                        | Covidien                                                                                                                                                                                                                                                                                                                                                                                      |
| Product Name     | Inno-Port Disposable Bladeless Trocar and<br>Inno-Port Disposable Optical Trocar                                                                                                                                                                                                                                                                                                             | VersaOne™ Bladeless Trocar and<br>Versaport™ V2 Bladeless Optical Trocar                                                                                                                                                                                                                                                                                                                      |
|                  | 5-11/12 mm: Obturator, Cannula with<br>Self-Adjusting Seal, Specimen Removal<br>Button, and Stopcock.                                                                                                                                                                                                                                                                                        | 5-11/12 mm: Obturator, Cannula with<br>Self-Adjusting Seal, Specimen Removal<br>Button, and Stopcock.                                                                                                                                                                                                                                                                                         |
| Specification    | 5 mm<br>Obturator: Ø 5.8 mm<br>Cannula:<br>(1) Outer diameter: Ø 8.2 mm<br>(2) Inside diameter: Ø 6 mm<br>5-11 mm<br>Obturator: Ø 11.2 mm<br>Cannula:<br>(1) Outer diameter: Ø 13.8 mm<br>(2) Inside diameter: Ø 11.5 mm<br>5-12 mm<br>Obturator: Ø 12.8 mm<br>Cannula:<br>(1) Outer diameter: Ø 15.35 mm<br>(2) Inside diameter: Ø 13 mm                                                    | 5 mm<br>Obturator: Ø 5.85 mm<br>Cannula:<br>(1) Outer diameter: Ø 8.2 mm<br>(2) Inside diameter: Ø 6 mm<br>5-11 mm<br>Obturator: Ø 11.2 mm<br>Cannula:<br>(1) Outer diameter: Ø 13.8 mm<br>(2) Inside diameter: Ø 11.5 mm<br>5-12 mm<br>Obturator: Ø 12.8 mm<br>Cannula:<br>(1) Outer diameter: Ø 15.35 mm<br>(2) Inside diameter: Ø 13 mm                                                    |
| Packaging        | Tyvek Pouch + Box                                                                                                                                                                                                                                                                                                                                                                            | Tyvek Pouch + Box                                                                                                                                                                                                                                                                                                                                                                             |
| Sterilization    | Gamma                                                                                                                                                                                                                                                                                                                                                                                        | EO                                                                                                                                                                                                                                                                                                                                                                                            |
| Material and its | In accordance with<br>ISO 10993-1                                                                                                                                                                                                                                                                                                                                                            | In accordance with<br>ISO 10993-1                                                                                                                                                                                                                                                                                                                                                             |
| Biocompatibility |                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                               |
| Performance      | Stability of Trocar:<br>1. Puncture force<br>5 mm: < 3,000 gf<br>5-11/12 mm: <5,000 gf<br>2. Removal force<br>5 mm: > 1,500 gf<br>5-11/12 mm: > 1,800 gf<br>Operation of Obturator:<br>1. Insertion Force<br>5 mm: < 450 gf<br>5-11/12 mm: < 1,200 gf<br>2. Withdrawal Force<br>5 mm: < 300 gf<br>5-11/12 mm: < 600 gf<br>Airtightness of Cannula:<br>1. Stopcock Airtightness: 30s not fall | Stability of Trocar:<br>1. Puncture force<br>5 mm: < 3,000 gf<br>5-11/12 mm: < 5,000 gf<br>2. Removal force<br>5 mm: > 1,500 gf<br>5-11/12 mm: > 1,800 gf<br>Operation of Obturator:<br>1. Insertion Force<br>5 mm: < 450 gf<br>5-11/12 mm: < 1,500 gf<br>2. Withdrawal Force<br>5 mm: < 300 gf<br>5-11/12 mm: < 600 gf<br>Airtightness of Cannula:<br>1. Stopcock Airtightness: 30s not fall |
| Manufactures     | Taiwan Surgical Corp.                                                                                                                                                                                                                                                                                                                                                                        | Covidien                                                                                                                                                                                                                                                                                                                                                                                      |
| Product Name     | Inno-Port Disposable Bladeless Trocar and<br>Inno-Port Disposable Optical Trocar                                                                                                                                                                                                                                                                                                             | VersaOne™ Bladeless Trocar and<br>Versaport™ V2 Bladeless Optical Trocar                                                                                                                                                                                                                                                                                                                      |
|                  | below 190 mmH2O (from 200 mmH2O)<br>1. Duckbill Airtightness: 30s not fall<br>below 190 mmH2O (from 200 mmH2O)<br>2. Sealing Airtightness: 30s not fall<br>below 150 mmH2O (from 160 mmH2O)                                                                                                                                                                                                  | below 190 mmH2O (from 200 mmH2O)<br>2. Duckbill Airtightness: 30s not fall<br>below 190 mmH2O (from 200 mmH2O)<br>3. Sea ling Airtightness: 30s not fall<br>below 150 mmH2O (from 160 mmH2O)                                                                                                                                                                                                  |
|                  | Durability of Cannula:<br>Pass GIA 15-times insert-out and then<br>pass the Airtightness test.                                                                                                                                                                                                                                                                                               | Durability of Cannula:<br>Pass GIA 15-times insert-out and then<br>pass the Airtightness test.                                                                                                                                                                                                                                                                                                |

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for TWSC, which is written in a large, blue, sans-serif font. Below the logo, in a smaller, light gray font, is the phrase "Treat Patient As Family". The logo is simple and modern, and the phrase conveys a message of care and compassion.

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the logo for TWSC, which is written in blue. Below the logo is the text "Treat Patient As Family" in a smaller, gray font. The logo is simple and modern, and the text is a clear statement of the company's values. The image is likely used on the company's website or marketing materials.

#### 9 PERFORMANCE

The bench performance tests, including Stability of Trocar, Operation of Obturator, Airtightness of Cannula, and Durability of Cannula as shown below were performed on Inno-Port Disposable Bladeless Trocar/Inno-Port Disposable Optical Trocar and its predicated device respectively. For the Additional Information (AI) Request from FDA, the performance testing of proposed product and predicated product for the evaluated representative models have been conducted, and the detailed comparison table and its differences have been updated in supplements. The test results showed that both proposed devices have the similar device performance compared to the predicate device.

- Stability of Trocar
	- -Puncture Force Test
	- Removal Force Test -
- Operation of Obturator
	- Insertion Force -
	- -Withdrawal Force
- Airtightness of Cannula
	- Stopcock Airtightness -
	- Duckbill Airtightness ।
	- Sealing Airtightness -
- Durability of Cannula

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for TWSC, which is written in blue. Below the logo is the phrase "Treat Patient As Family" in a smaller, gray font. The logo is simple and professional, and the phrase conveys a message of care and compassion.

## 10 CONCLUSIONS

Based on the intended use and/or indications for use, technological characteristics, performance testing and comparison to the predicate device, the Inno-Port Disposable Bladeless Trocar is substantially equivalent to the predicate device VeraPort™ V2 Bladeless Optical Trocar and the Inno-Port Disposable Optical Trocar is substantially equivalent to the predicate device VersaOne™ Bladeless Trocar.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/GCJ/K223593](https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/GCJ/K223593)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
