K223562 · Xpan, Inc. · GCJ · Jun 2, 2023 · Gastroenterology, Urology
Device Facts
Record ID
K223562
Device Name
Xpan Universal Trocar System
Applicant
Xpan, Inc.
Product Code
GCJ · Gastroenterology, Urology
Decision Date
Jun 2, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Indications for Use
The Xpan Universal Trocar System is intended to provide dilation access with or without visualization to the abdominal and thoracic cavities for performing diagnostic and operative abdominal and thoracic procedures. - Laparoscopic access to the abdominal cavity, both primary and secondary punctures. - Thoracoscopic access to the thoracic cavity, both primary and secondary punctures.
Device Story
Xpan Universal Trocar System is a single-use, radially expandable trocar system for abdominal and thoracic access. System comprises a 3mm expandable port (expandable to 5mm or 12mm) and optical bladeless obturators. Principle of operation involves tissue dilation via rigid members surrounded by an elastomeric cover; expansion occurs upon insertion of 5mm or 12mm cannula/obturator. Device features include one-way valves, backup seals, and detachable valves for specimen removal or desufflation. Used in clinical settings (OR) by surgeons. Provides a universal port solution; allows 3mm instrument use prior to expansion. Benefits include reduced need for Veress needle, optical visualization, and versatile instrument sizing. Healthcare providers use the device to establish and maintain pneumoperitoneum for endoscopic procedures.
Clinical Evidence
Bench testing only. No clinical or animal studies were performed. Performance characteristics evaluated included insertion/retention/expansion forces, instrument compatibility, light transmission, visualization, and maintenance of pneumoperitoneum. Biocompatibility testing conducted per ISO 10993-1.
Technological Characteristics
Single-use, radially expandable trocar system. 3mm port expands to 5mm or 12mm. Features optical bladeless tips, one-way valves, and detachable valve assemblies. Materials are biocompatible per ISO 10993. Sterilization via Ethylene Oxide (ISO 11135).
Indications for Use
Indicated for patients requiring laparoscopic or thoracoscopic access to the abdominal or thoracic cavities for diagnostic or operative procedures, including both primary and secondary punctures.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
VersaOne 5mm and 12mm Optical Trocars (K112349, K130435)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 2, 2023
Xpan Inc. % Michele Lucey Consultant Lakeshore Medical Device Consulting, LLC 128 Blye Hill Landing Newbury, New Hampshire 03255
Re: K223562
Trade/Device Name: Xpan Universal Trocar System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: GCJ Dated: [NOTE: Use date of most recent supplement] Received: May 4, 2023
Dear Michele Lucey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
| Mark<br>Trumbore -S | Digitally signed by<br>Mark Trumbore -S<br>Date: 2023.06.02<br>08:15:43 -04'00' |
|---------------------|---------------------------------------------------------------------------------|
|---------------------|---------------------------------------------------------------------------------|
Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Traditional 510(k) Premarket Notification Xoan Universal Trocal System DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# Indications for Use
510(k) Number (if known)
Device Name Xpan Universal Trocar System
Indications for Use (Describe)
The Xpan Universal Trocar System is a single use system that is intended to provide dilation access with or without visualization to the abdominal and thoracic cavities for performing diagnostic and operative abdominal and thoracic procedures. The system is indicated for the following uses:
- · Laparoscopic access to the abdominal cavity, both primary and secondary punctures.
- · Thoracoscopic access to the thoracic cavity, both primary and secondary punctures.
Type of Use (Select one or both, as applicable)
| <div> <span style="font-size:10pt"> <span style="font-family:Wingdings">X</span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span style="font-size:10pt"> <span style="font-family:Wingdings">☐</span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for XPAN. The logo consists of a stylized letter X, with one side in black and the other in blue, followed by the letters PAN in black. The blue side of the X has an arrow pointing upwards, while the black side has an arrow pointing downwards.
## Traditional
510(k) SUMMARY
Page 1 of 8
## Submitter:
| Applicant | XPAN Inc. |
|---------------------------|-----------------------------------------|
| Contact Person | Michele Lucey |
| | Lakeshore Medical Device Consulting LLC |
| | 603-748-1374 |
| | Email: lucey_m@msn.com |
| Date Prepared | November 17, 2022 |
| Subject Device: | |
| Proprietary/Trade Name | Xpan Universal Trocar System |
| Common or Usual Name | Surgical Trocar |
| Classification Name | Endoscope and Accessories |
| Regulatory Class | Class II |
| Classification Code | GCJ |
| Regulation Number | 876.1500 |
| Review Panel | General and Plastic Surgery |
| Predicate Device: | |
| Proprietary/Trade<br>Name | VersaStep™ Bladeless Trocar |
| 510(k) Number | K012539 |
| Common or Usual Name | Surgical Trocar |
| Classification Name | Endoscope and Accessories |
| Regulatory Class | Class II |
| Classification Code | GCJ |
| Regulation Number | 876.1500 |
| Review Panel | General and Plastic Surgery |
| Reference Devices: | |
| Proprietary/Trade Name | GeniCon NanoPort 3mm Trocar |
| 510(k) Number | K982472 |
| Common or Usual Name | Surgical Trocar |
| Classification Name | Cannula, Surgical |
| Regulatory Class | Class I (reclassified 2001) |
| Classification Code | GEA |
| Regulation Number | 876.4800 |
| Review Panel | General and Plastic Surgery |
| Proprietary/Trade Name | VersaOne 5mm and 12mm Optical Trocars |
| 510(k) Number | K112349, K130435 |
| Common or Usual Name | Surgical Trocar |
| Classification Name | Endoscope and Accessories |
| Regulatory Class | Class II |
| Classification Code | GCJ |
| Regulation Number | 876.1500 |
| Review Panel | General and Plastic Surgery |
{4}------------------------------------------------
#### Intended Use:
The Xpan Universal Trocar System is intended to provide dilation access with or without visualization to the abdominal and thoracic cavities for performing diagnostic and operative abdominal and thoracic procedures.
- Laparoscopic access to the abdominal cavity, both primary and secondary punctures. .
- Thoracoscopic access to the thoracic cavity, both primary and secondary punctures.
#### Device Description/Technological Characteristics:
The Xpan Universal Trocar System provides a universal port solution wherein the 3mm expandable port is radially expandable from 3mm up to 12mm. Xpan is offered in three sizes and is a single use device.
{5}------------------------------------------------
The 5mm and 12mm cannulas are supplied with an optical bladeless tip and include a detachable valve to allow for specimen removal or rapid desufflation. The 3mm expandable trocar can be assembled with the 5mm or 12mm Xpan obturators after tissue insertion or before tissue insertion if desired. The exterior shaft of the 3mm expandable trocar is ribbed to provide security in tissue once inserted in tissue.
| Device Description |
|-----------------------------------------------------------------|
| Description |
| 3mm expandable (up to 12mm) trocar with a 3mm obturator |
| 5mm Cannula with a 5mm Obturator with an optical bladeless tip |
| 12mm Cannula and a 12mm Obturator with an optical bladeless tip |
#### Performance Data:
Nonclinical performance data has been conducted for the Xpan Universal Trocar System was conducted to evaluate safety and performance, where appropriate device performance was compared against the predicate, reference, or legally marketed comparator devices.
#### Biocompatibility
Biocompatibility evaluation of the subject device was conducted in accordance with ISO 10993-1, Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process and 2020 FDA Guidance, Use of International Standard ISO 10993-1, Biological evaluation of medical devices – Part 1''; Evaluation and testing within a risk management process. The subject device contacts tissue for less than 24 hours, so the following endpoints were considered, and all materials were found to be biocompatible:
Cytotoxicity
Sensitization
Intracutaneous Irritation
Acute Systemic Toxicity
Material Mediated Pyrogenicity
#### Bench Testing
Design Verification testing was performed with the subject device. The following
performance characteristics were evaluated:
Trocar Insertion Force
Trocar Retention Force
Expansion Force
Instrument Insertion/Removal Force
Light Transmission and Tissue Visualization
Insufflation and Desufflation
Maintenance of Pneumoperitoneum
Instrument Compatibility
{6}------------------------------------------------
#### Sterilization /Shelf Life
Sterilization validation was conducted in accordance with ISO11135, Sterilization of health-care products – Ethylene oxide – Requirements for the development, validation, and routine control of a sterilization process for medical devices.
Shelf life testing was performed to evaluate package integrity and device functionality following artificial aging and simulated transit conditioning.
### Animal Studies
No animal studies were performed
## Clinical Studies
No clinical studies were performed
## Comparison with the Predicate Device:
Substantial equivalence of the Xpan Universal Trocar System is based on comparing the intended use and indications for use and technological characteristics of the predicate device. The following table compares the intended use/indications and key technological characteristics between the Xpan Universal Trocar System and the predicate device. Where technological characteristics are different, reference devices have been identified. These differences are minor and do not raise new questions of safety and effectiveness.
| Feature | Subject Device | Predicate<br>device | Reference<br>Device #1 | Reference Devices<br>#2 and #3 | Comparison |
|----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Xpan Universal<br>Trocar System | VersaStep<br>(K012539) | GeniCon<br>NanoPort 3mm<br>Trocar<br>(K982472) | VersaOne 5mm and<br>12mm Bladeless<br>Optical Trocars<br>(K112349 and<br>K130435<br>respectively). | |
| FDA<br>Classification | Class II | Class II | Class I | Class II | Same |
| Product Code(s) | GCJ | GCJ | GEA<br>(reclassified 2001) | GCJ | Same as the<br>predicate device |
| Regulation<br>Number | 876.1500 | 876.1500 | 878.4800 | 876.1500 | Same as the<br>predicate device |
| Anatomical<br>Location | Abdominal and<br>Thoracic | Abdominal and<br>Thoracic | Abdominal and<br>Thoracic | Abdominal and<br>Thoracic | Same |
| Feature | Subject Device | Predicate<br>device | Reference<br>Device #1 | Reference Devices<br>#2 and #3 | Comparison |
| | Xpan Universal<br>Trocar System | VersaStep<br>(K012539) | GeniCon<br>NanoPort 3mm<br>Trocar<br>(K982472) | VersaOne 5mm and<br>12mm Bladeless<br>Optical Trocars<br>(K112349 and<br>K130435<br>respectively). | |
| Insertion | Through an<br>incision | Through an<br>incision | Through an<br>incision | Through an incision | Same |
| Tissue<br>anchoring | Dilation and<br>ridges of<br>expandable<br>members | The dilation of<br>the mesh sheath | Bumps on<br>cannula | Ridges on cannula | Differences do<br>not raise<br>questions of<br>safety and<br>effectiveness |
| Initial Entry<br>Device (Sheath)<br>Outer Diameter | 6.6 mm | 4 mm | 8.4mm | 5mm device: 8.25 mm<br>12mm device: 15.4<br>mm | Difference does<br>not raise<br>questions of<br>safety and<br>effectiveness |
| Expandable<br>Range | 5mm and 12mm | 5mm and 12mm | N/A | N/A | Same |
| Method of<br>expansion | Three rigid<br>members form a<br>3mm cannula and<br>are surrounded by<br>an elastomeric<br>cover. The rigid<br>members and<br>elastomeric cover<br>move radially<br>outwards and<br>expand upon<br>insertion of 5mm<br>or 12mm cannula<br>and obturator | A collapsed<br>mesh<br>constrained by<br>an outer sheath<br>is expanded<br>upon insertion<br>of 5mm or<br>12mm cannula<br>and blunt<br>obturator<br>through its<br>lumen which<br>disrupts the<br>outer sheath to<br>allow the mesh<br>to expand. | N/A | N/A | Same, both<br>devices<br>accomplish<br>expansion by<br>increasing the<br>radial diameter<br>through a flexible<br>mechanism.<br>Structural<br>differences do<br>not raise new<br>questions of<br>safety and<br>effectiveness |
| Feature | Subject Device | Predicate device | Reference Device #1 | Reference Devices #2 and #3 | Comparison |
| 3mm<br>Trocar/Cannula<br>capability | Xpan Universal<br>Trocar System | VersaStep<br>(K012539) | GeniCon<br>NanoPort 3mm<br>Trocar<br>(K982472) | VersaOne 5mm and<br>12mm Bladeless<br>Optical Trocars<br>(K112349 and<br>K130435<br>respectively). | |
| 3mm<br>Trocar/Cannula<br>capability | The Xpan 3mm<br>Expandable<br>Trocar (equivalent<br>of the predicate's<br>sheath) has long,<br>rigid members<br>that form a rigid<br>pathway and<br>lumen for 3mm<br>interventional<br>endoscopic<br>instruments to be<br>used. Can be used<br>standalone 3mm<br>conventional<br>trocar | The sheath is<br>made of flexible<br>material and has<br>no one-way<br>valve, it only<br>creates an initial<br>pathway in the<br>tissue and<br>cannot be used<br>as standalone<br>trocar/cannula<br>for 3mm<br>instruments. | The trocar<br>comprises a<br>cannula that<br>allows entry of<br>3mm<br>instruments. It<br>also comprises a<br>one-way valve<br>and backup valve<br>that provide<br>sealing | N/A | The capability of<br>the Xpan device<br>as a 3mm trocar<br>does not raise<br>new questions of<br>safety and<br>effectiveness |
| One Way Valve | Yes<br>In 3mm, 5mm,<br>and 12mm | Yes<br>In 5mm, and<br>12mm only | Yes | Yes | The capability of<br>the Xpan device<br>at 3mm does not<br>raise new<br>questions of<br>safety and<br>effectiveness |
| Backup valve<br>for instruments<br>seal | Yes<br>In 3mm, 5mm,<br>and 12mm | Yes<br>In 5mm, and<br>12mm only | Yes | Yes | The capability of<br>the Xpan device<br>at 3mm does not<br>raise new<br>questions of<br>safety and<br>effectiveness |
| Range of<br>instrument sizes | Can be used with<br>3mm instruments<br>in 3mm<br>Expandable | No instruments<br>before<br>expansion of<br>the sheath | 3mm instruments | 5mm instruments in<br>the 5mm device | The capability of<br>the Xpan device<br>at 3mm does not<br>raise new |
| Feature | Subject Device | Predicate device | Reference Device #1 | Reference Devices #2 and #3 | Comparison |
| | Xpan Universal Trocar System | VersaStep (K012539) | GeniCon NanoPort 3mm Trocar (K982472) | VersaOne 5mm and 12mm Bladeless Optical Trocars (K112349 and K130435 respectively). | |
| | Device prior to expansion<br>5mm instruments in 5mm Expanded Device<br>5mm-12mm instruments in 12mm Expanded Device | 5mm instruments in the 5mm Expanded Device<br>5mm-12mm instruments in the 12mm Expanded Device | | 5mm-12mm instruments in the 12mm device | questions of safety and effectiveness |
| Veress Needle Required for insertion | Not required | Required | Not required | Not required | The integrated bladeless 3mm obturator tip in the Xpan device is similar to 3mm reference device and precludes the need for a Veress needle. This difference does not raise new questions of safety and effectiveness |
| Optical Insertion | Yes<br>For 5mm and 12mm insertions only | No | No | Yes | This difference does not raise new questions of safety and effectiveness |
| Detachable valve (for Specimen Retrieval) | Yes<br>In 5mm and 12mm | Yes<br>In 12mm only | No | No in the 5mm device<br>Yes in the 12mm device | The additional capability of a detachable valve in both the 5mm and 12mm sizes, |
| Feature | Subject Device | Predicate<br>device | Reference<br>Device #1 | Reference Devices<br>#2 and #3 | Comparison |
| | Xpan Universal<br>Trocar System | VersaStep<br>(K012539) | GeniCon<br>NanoPort 3mm<br>Trocar<br>(K982472) | VersaOne 5mm and<br>12mm Bladeless<br>Optical Trocars<br>(K112349 and<br>K130435<br>respectively). | |
| | | | | | does not raise<br>new questions of<br>safety and<br>effectiveness |
| Biocompatibility | ISO 10993 | ISO 10993 | ISO 10993 | ISO 10993 | Same |
| How Supplied | Sterile, single use | Sterile, single<br>use | Sterile, single use | Sterile, single use | Same |
| Sterilization<br>Method | Ethylene Oxide | Ethylene Oxide | Ethylene Oxide | Ethylene Oxide | Same |
| Sterility<br>Assurance Level | 10-6 | 10-6 | 10-6 | 10-6 | Same |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
## Conclusion:
Substantial equivalence has been demonstrated based upon the supporting data in this submission demonstrating that the Xpan Universal Trocar System is as safe and effective as the legally marketed predicate device (K012539) and does not raise different questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.