K211250 · New View Surgical, Inc. · GCJ · Jun 21, 2021 · Gastroenterology, Urology
Device Facts
Record ID
K211250
Device Name
VisionPort System
Applicant
New View Surgical, Inc.
Product Code
GCJ · Gastroenterology, Urology
Decision Date
Jun 21, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The VisionPort System is intended to be used in diagnostic and therapeutic procedures for endoscopic surgery within the thoracic and peritoneal cavities including the female reproductive organs to establish a path of entry for endoscopic instruments and to provide an image of the surgical field. The trocar may be used with or without visualization for primary and secondary insertions.
Device Story
VisionPort System provides endoscopic access and visualization. Inputs: optical images from integrated CMOS camera/LED module in single-use handpiece. Processing: Control unit performs image processing (no patient data retention); outputs video to OR monitors. Operation: Handpiece includes 12.9mm access port, lens cleaning (saline wash/wiper), and obturator for visible entry; control pad manages zoom, pan, and LED intensity. Used in hospitals/surgery centers by healthcare professionals. Benefits: Integrated visualization during trocar insertion; facilitates surgical field access and monitoring; electronic zoom/pan capabilities; lens cleaning maintains image clarity during procedures.
Clinical Evidence
Bench testing only. No clinical data. Testing included sterilization validation (ISO 11135), biocompatibility (ISO 10993-1), electrical safety/EMC (IEC 60601 series), and functional testing per ISO 8600. Simulated use testing compared performance against ENDOPATH XCEL Bladeless Trocar, demonstrating equivalent performance in surgical access and visualization.
Technological Characteristics
Single-use handpiece with CMOS sensor, 2 LEDs, 0° lens, and integrated saline/wiper cleaning system. 12.9mm internal diameter access port. Control unit provides DVI output to monitors. Materials comply with ISO 10993-1. Electrical safety per IEC 60601-1; EMC per IEC 60601-1-2. Sterilized via Ethylene Oxide (EtO) per ISO 11135.
Indications for Use
Indicated for patients undergoing diagnostic and therapeutic endoscopic surgery within the thoracic and peritoneal cavities, including female reproductive organs. Used to establish entry for endoscopic instruments and provide surgical field visualization. No specific age or gender contraindications stated.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a sans-serif font.
June 21, 2021
New View Surgical, Inc. % Maureen O'Connell President O'Connell Regulatory Consultant, Inc. 44 Oak Street Stoneham, Massachusetts 02180
Re: K211250
Trade/Device Name: VisionPort System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: GCJ, HET Dated: April 23, 2021 Received: April 26, 2021
Dear Maureen O'Connell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531 -542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K211250
Device Name VisionPort™ System
Indications for Use (Describe)
The VisionPort System is intended to be used in diagnostic and therapeutic for endoscopic surgery within the thoracic and peritoneal cavities including the female reproductive organs to establish a path of entry for endoscopic instruments and to provide an image of the surgical field. The trocar may be used with or without visualization for primary and secondary insertions.
Type of Use (Select one or both, as applicable)
|X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) SUMMARY
# (K211250)
#### New View Surgical, Inc. VisionPort System
#### SUBMITTER
New View Surgical, Inc. 555 Massachusetts Ave. Ste. 5 Boston, MA 02118
#### Submission Correspondent
Maureen O'Connell O'Connell Regulatory Consultants, Inc. 44 Oak Street Stoneham. MA 02180 Phone: 978-207-1245
Date Prepared: June 2, 2021
#### DEVICE
Trade Name of Device: VisionPort System Common or Usual Name: Laparoscope Classification Name(s): Laparoscope, General & Plastic Surgery; 21 C.F.R. §876.1500 and Laparoscope, Gynecologic (and Accessories); 21 C.F.R. §884.1720 Regulatory Class: II Product Code(s): GCJ and HET
# PREDICATE DEVICE(S)
Primary: Xenocor Xenoscope Laparoscopic System cleared in K161838 ENDOPATH XCEL Bladeless Trocar cleared in K122511 Secondary:
# DEVICE DESCRIPTION
The VisionPort System consists of an access port with an integrated camera and light source, the handpiece, a wired control pad to control the camera functions, and a control unit, specifically:
- 1. VisionPort System Handpiece single use
- HD camera and lighting module a.
- b. 12.9mm internal diameter access port with 5-12mm instrument seal
- Lens cleaning system (saline wash and wiper blade) C.
- d. Connections for saline wash source and insufflation tubing
- e. Removable obturator designed for visible entry
- 2. VisionPort System Control Pad single use
- a. Camera control functions (zoom, pan)
{5}------------------------------------------------
- b. Light emitting diode (LED) adjustment
- c. Image capture and display control buttons
- 3. VisionPort System Control Unit
- a. System power
- b. Image processing (no patient data recording or retention)
- Outputs to operating room monitor(s) and/or third-party Data Control Unit (for image C. capture and printing options)
# INDICATIONS FOR USE
The VisionPort System is intended to be used in diagnostic and therapeutic procedures for endoscopic surgery within the thoracic and peritoneal cavities including the female reproductive organs to establish a path of entry for endoscopic instruments and to provide an image of the surgical field. The trocar may be used with or without visualization for primary and secondary insertions.
# SUBSTANTIAL EQUIVALENCE
The VisionPort System described in this notification and for use under the conditions of the proposed labeling is substantially equivalent to a legally marketed predicate device that is a Class II medical device. The VisionPort System is substantially equivalent to the Xenosope Laparoscopic System cleared in K161838 and a secondary predicate device, the ENDOPATH XCEL Bladeless Trocar cleared in K122511. A comparison to the primary predicate device is provided in Table 1 and to the secondary predicate device in Table 2. The combined intended use of the primary predicate and secondary predicate is the same as the intended use of the VisionPort System.
| VisionPort System Substantial Equivalence: Primary Predicate Device | | |
|---------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristic | VisionPort System | Xenocor Xenoscope<br>Laparoscopic System |
| 510(k) Number | K211250 | K161838/K171344 |
| Product Code | GCJ, HET | GCJ, HET |
| Indications for Use | Intended to be used in<br>diagnostic and therapeutic<br>procedures for endoscopic<br>surgery within the thoracic<br>and peritoneal cavities<br>including the female<br>reproductive organs to<br>establish a path of entry for<br>endoscopic instruments and<br>to provide an image of the<br>surgical field. The trocar may<br>be used with or without<br>visualization for primary and<br>secondary insertions. | Intended to be used in diagnostic<br>and therapeutic procedures for<br>endoscopy and endoscopic<br>surgery within the thoracic and<br>peritoneal cavities including the<br>female reproductive organs. |
| Clearance Type | Prescription | Prescription |
| User | Healthcare Professional | Healthcare Professional |
| Where Used | Hospital, Surgery Center | Hospital, Surgery Center |
| System Description | The system includes the following components:<br>-VisionPort System<br>Handpiece<br>-VisionPort System Control Pad<br>-VisionPort System Control Unit | The system includes the following components:<br>-Xenoscope Laparoscope Device<br>-Xenoscope Dongle |
| Component<br>Comparison | VisionPort System<br>Handpiece | Xenoscope Laparoscope |
| Image Sensor | CMOS | CMOS |
| Lens | 0° | 0° |
| LEDs | 2 | 1 |
| Video Image | High-definition video | High-definition video |
| Image Transfer | Video cable from camera/scope to Control Unit | Video cable from camera/scope to Control Unit (Dongle) |
| Placement into<br>Body Cavity | Incorporates an access port/trocar | Inserted through an access port/trocar |
| Lens Cleaning | Incorporates a lens cleaning system (saline flush and wiper) | No lens cleaning system |
| Camera Focus | Fixed Focus | Manual Focus |
| Camera Zoom | Electronic Zoom | Manual Zoom |
| Camera Light | Light (LED) Attenuation | Light (LED) Attenuation |
| Camera Monitor | Monitor Display Control | Monitor Display Control |
| Camera Image<br>Capture Relay | Image Capture Relay using third party image capture system | NA |
| Sterility | Single use sterile | Single use sterile |
| Component<br>Comparison | VisionPort System Control Unit | Xenoscope Dongle |
| Use | Multi-use | Multi-use |
| Image Transfer<br>from<br>Camera/Scope to<br>Control Unit | Video Cable | Micro-HDMI |
| Image Transfer<br>from Control Unit<br>to Monitor | DVI | HDMI v1.3 |
| USB | 2.0 service port (manufacturing) | 3.0 service port (manufacturing) |
| Standards<br>Compliance | | |
| Electrical Safety | Per IEC 60601-1 | Per IEC 60601-1 |
| EMC | Per IEC 60601-1-2 | Per IEC 60601-1-2 |
| Biocompatibility | Per ISO 10993-1 | Per ISO 10993-1 |
Table 1 VisionPort System Sub ce
{6}------------------------------------------------
| Characteristic | VisionPort System | ENDOPATH XCEL Bladeless<br>Trocar with OPTIVIEW<br>Technology |
|----------------|-------------------|---------------------------------------------------------------|
| 510(k) Number | K211250 | K122511 |
{7}------------------------------------------------
| Product Code | GCJ, HET | GCJ |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | Intended to be used in<br>diagnostic and therapeutic<br>procedures for endoscopic<br>surgery within the thoracic and<br>peritoneal cavities including the<br>female reproductive organs to<br>establish a path of entry for<br>endoscopic instruments and to<br>provide an image of the<br>surgical field. The trocar may<br>be used with or without<br>visualization for primary and<br>secondary insertions. | Has applications in abdominal,<br>thoracic, and gynecological<br>minimally invasive surgical<br>procedures to establish a path of<br>entry for endoscopic instruments.<br>The trocar may be used with or<br>without visualization for primary<br>and secondary insertions. |
| Clearance Type | Prescription | Prescription |
| User | Healthcare Professional | Healthcare Professional |
| Where Used | Hospital, Surgery Center | Hospital, Surgery Center |
| Components | Sleeve and obturator | Sleeve and obturator |
| Cannula Length | 100 mm | 100 mm |
| Cannula Inner<br>Diameter | 12.9 mm | 12.9 mm |
| Cannula Outer<br>Diameter | 17.8 mm | 15.5 mm |
| Endoscopic View | Incorporates a 0-degree<br>endoscopic view of the trocar<br>tip and provides visibility of<br>individual tissue layers during<br>insertion using an image<br>capture chip, light source and<br>mirror | Accommodates a 0-degree or 30-<br>degree endoscope to view the<br>trocar tip and provides visibility<br>of individual tissue layers during<br>insertion. |
| Seal System | The sleeve has a two-seal<br>system for maintaining<br>pneumoperitoneum. The trocar<br>sleeves contain two seals, an<br>outer integrated non-removable<br>self-adjusting seal that<br>accommodates instruments<br>ranging from 5 to 12 mm in<br>diameter and an internal seal. | The sleeve has an integrated two-<br>seal system for maintaining<br>pneumoperitoneum. The trocar<br>sleeves contain two seals, an<br>outer integrated removable self-<br>adjusting seal that accommodates<br>instruments ranging from 5 to 12<br>mm in diameter and an internal<br>seal. |
| Stopcocks | The sleeve has two welded<br>stopcocks on the side of the<br>cannula assembly, both<br>compatible with standard luer<br>lock fittings which provide for<br>attachment of insufflation<br>tubing and a saline flush for<br>lens cleaning. | The sleeve has one welded<br>stopcock assembly, which is<br>compatible with a standard luer<br>lock fitting which provides for<br>attachment of insufflation tubing. |
| Stability Threads | Sleeve cannula contains<br>integrated stability threads for | Sleeve cannula contains<br>integrated stability threads for |
The VisionPort System has the same technological characteristics as the combination of the primary and secondary predicate devices. The different technological characteristics of the
{8}------------------------------------------------
devices do not raise different questions of safety and effectiveness. Performance data is provided which supports substantial equivalence.
### PERFORMANCE DATA
Non-clinical testing was performed to verify that the proposed device met all design specifications and is substantially equivalent to the predicate device.
### Sterilization and Shelf Life
Sterilization validation was performed as described in ISO 11135:2014, Sterilization of healthcare products -Ethylene oxide - Requirements for the development. validation and routine control of a sterilization process for medical devices. The sterilization assurance level (SAL) is 10-6.
Testing of EtO residuals was conducted per ISO 10993-7 Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals. VisionPort System Handpiece samples met the requirements of the standard.
Testing of package seal strength was performed in accordance with ASTM F88 - Standard Test Method for Seal Strength of Flexible Barrier Materials. All samples demonstrated a seal strength greater than 1.0 lbf/in specification. Visual inspection shows that all seals met specifications with no anomalies noted. The shelf life is confirmed by age testing both accelerated and real time in accordance with ASTM 1980-02. Reference Standard Guide for Accelerated Aging of Sterile Medical Device Packages.
#### Biocompatibility Testing
The materials used for the VisionPort System comply with biocompatibility requirements outlined in ISO 10993-1:2009 and the Guidance for Industry and Food and Drug Administration Staff, Use of International Standard ISO 10993 Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a Risk Management Process and are considered to be biocompatible. The contact type of the VisionPort System is: external communicating, tissue contact with limited contact duration and testing was provided to show compliance with ISO 10993-1. The battery of tests included:
- . Cytotoxicity
- . Sensitization
- Intracutaneous Irritation
- Systemic Toxicity ●
- . Pyrogenicity
# Electrical safety and electromagnetic compatibility (EMC)
The VisionPort System was tested for electrical safety and found to be in compliance with IEC 60601-1. IEC 60601-1-6. IEC 62366-1 and IEC 60601-2-18, and for electromagnetic compatibility to IEC 60601-1-2.
# Functional Testing
Testing was performed to verify that the VisionPort Handpiece functions as intended:
{9}------------------------------------------------
- When tested in accordance with ISO 8600-1 through ISO 8600-6. All test results met . acceptance criteria and were performed in accordance ISO 8600.
- In accordance with design input requirements. The test data demonstrates that the ● VisionPort System functions as intended and in accordance with design input requirements.
Testing was provided which compared performance of the VisionPort System compared to the secondary predicate device (ENDOPATH® XCEL® Bladeless Trocar with OPTIVIEW Technology (K122511)) and to document substantial equivalence in performance related to recommended use. The testing showed that the VisionPort System can be operated in simulated use conditions when used in accordance with the Manual and IFU and is equivalent to the predicate systems for the attributes tested in this protocol.
# CONCLUSIONS
The performance data supports the safety of the device and shows that the device performs as intended. The substantial equivalence discussion supports the finding of substantial equivalence to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.