The CORE E3 Suction/Irrigator is designed to be used in conjunction with the CORE Trumpet Handpiece and probes to provide controlled powered irrigation during laparoscopic surgical procedures (e.g., laparoscopic cholecystectory and laparoscopic gynecological procedures). It may also be used for resection of filmy adhesions (i.e., hydrodissection) and peritoneal lavage.
Device Story
CORE E3 Suction/Irrigator is a single-use, battery-powered device for laparoscopic surgery. It consists of a handpiece, dual tubing, and optional probe. The device connects to a standard suction apparatus. The operator controls suction and irrigation flow via color-coded buttons on the handpiece. A mechanical pump, powered by six AAA alkaline batteries, generates fluid output. Used in hospitals/healthcare facilities by surgeons. The device provides controlled irrigation and aspiration to assist in surgical visualization and tissue management. Benefits include improved surgical site access and fluid management during minimally invasive procedures.
Clinical Evidence
Bench testing only. Testing included simulated use, functionality verification (flow rate 1.52-2.08 L/min), electrical safety, EMC, and biocompatibility (cytotoxicity, sensitization, irritation, pyrogenicity, acute systemic toxicity, hemocompatibility). Packaging and sterilization (ISO 11135) were also validated.
Technological Characteristics
Materials: 304/420 Stainless Steel, Polycarbonate, EPDM Rubber, ABS, PVC. Power: 6 AAA alkaline batteries. Principle: Mechanical pumping system. Form factor: Handpiece with dual tubing and probe. Connectivity: None. Sterilization: Ethylene Oxide (SAL 10^-6). Standards: IEC 60601-1 (Electrical Safety), IEC 60601-1-2 (EMC), ISO 10993-1 (Biocompatibility), ISO 11135 (Sterilization), ISO 11607-1 (Packaging).
Indications for Use
Indicated for patients undergoing laparoscopic surgical procedures, including cholecystectomy and gynecological procedures, requiring controlled powered irrigation, hydrodissection of filmy adhesions, or peritoneal lavage.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square, the words "U.S. FOOD & DRUG" are written in blue, with the word "ADMINISTRATION" written in a smaller font size below.
February 19, 2021
ConMed Corporation Tina Mornak Principal Regulatory Affairs Specialist 525 French Road Utica, New York 13502
Re: K202303
Trade/Device Name: CORE E3 Suction/irrigator Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: GCJ Dated: August 13, 2020 Received: August 14, 2020
Dear Tina Mornak:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K202303
Device Name CORE E3 Suction/Irrigator
Indications for Use (Describe)
The CORE E3 Suction/Irrigator is designed to be used in conjunction with the CORE Trumpet Handpiece and probes to provide controlled powered irrigation during laparoscopic surgical procedures (e.g., laparoscopic cholecystectory and laparoscopic gynecological procedures). It may also be used for resection of filmy adhesions (i.e., hydrodissection) and peritoneal lavage.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
# CORE E3™ Suction/Irrigator
In accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CER 807.92, ConMed Corporation is hereby submitting the 510(k) Summary of Safety and Effectiveness for 510(k) number K202303.
#### l. SUBMITTER
ConMed Corporation 525 French Road Utica, NY 13502 Company Contact: Tina Mornak Principal Regulatory Affairs Specialist t: 303-699-7600 x135086 christinamornak@conmed.com Date Prepared: January 21, 2021
#### II. DEVICE NAME
| Proprietary Name: | CORE E3™ Suction/Irrigator |
|--------------------|--------------------------------------------------------|
| Model Numbers: | CD8185-B – CORE E3 Suction/Irrigator, 5mm x 32cm Probe |
| | CD81300-B - CORE E3 Suction/Irrigator, Handpiece Only |
| Common Name: | Laparoscope, General & Plastic Surgery |
| Panel: | Gastroenterology/Urology |
| Product Code: | GCJ |
| Device Class: | II |
| Regulation Number: | 876.1500 |
#### lll. PREDICATE/LEGALLY MARKETED DEVICE
| Primary Device Name: | Hydro-Surg® Laparoscopic Irrigator |
|------------------------|------------------------------------|
| Company Name: | Davol Inc. |
| 510(k): | K961492 |
| Reference Device Name: | StrykeFlow |
| Company Name: | Stryker |
| 510(k): | K963646 |
#### IV. DEVICE DESCRIPTION
The CORE E3TM Suction/Irrigator consists of a suction/irrigation handpiece and dual tubing that connects to a battery-powered mechanical pumping system to generate fluid output and can include a suction/irrigation probe. The device connects to a standard suction apparatus to deliver suction. Control of the flow (suction/irrigation) is generated by depressing the color-coded buttons on the handpiece. The handpiece can also be attached to separate compatible electrodes or probes, including reusable probes. The mechanical pumping system is powered with six (6) standard AAA alkaline batteries.
{4}------------------------------------------------
#### V. INTENDED USE / INDICATIONS FOR USE
The CORE E3TM Suction/Irrigator is designed to be used in conjunction with the CORE Trumpet Handpiece and probes to provide controlled powered irrigation during laparoscopic surgical procedures (e.g., laparoscopic cholecystectomy and laparoscopic gynecological procedures). It may also be used for resection of filmy adhesions (i.e., hydrodissection) and peritoneal lavage.
#### VI. COMPARISON OF THE TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The CORE E3™ Suction/Irrigator is similar to the predicate device in that the design of both include a battery-powered pump, dual tubing, trumpet handpiece, and optional probe. As with the predicate, the CORE E3™ Suction/Irrigator continues to function as a single use suction/irrigation device to provide controlled powered irrigation during laparoscopic surgical procedures. The main difference in the subject device from the predicate device is the number and size of batteries that powers the device, and the average delivered irrigation flow rates. The CORE E3™ Suction/Irrigator is safe and effective and substantially equivalent to the predicate as demonstrated by non-clinical performance testing.
| | Subject Device | Predicate Device |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | CORE E3™ Suction/Irrigator | Hydro-Surg® Laparoscopic<br>Irrigator |
| Indications for Use | The CORE E3 Suction/Irrigator is<br>designed to be used in conjunction<br>with the CORE Trumpet Handpiece<br>and probes to provide controlled<br>powered irrigation/aspiration during<br>laparoscopic surgical procedures<br>(e.g., laparoscopic cholecystectomy<br>and laparoscopic gynecological<br>procedures). It may also be used for<br>resection of filmy adhesions (i.e.,<br>hydrodissection) and peritoneal<br>lavage. | The Hydro-Surg Irrigator is designed<br>to be used in conjunction with a<br>laparoscopic probe handle and tip to<br>provide controlled powered<br>irrigation/aspiration during<br>laparoscopic surgical procedures<br>(e.g., laparoscopic cholecystectomy<br>and laparoscopic gynecological<br>procedures). It may also be used for<br>resection of filmy adhesions (i.e.,<br>hydrodissection) and peritoneal<br>lavage. |
| Design | Battery-powered pump, dual tubing,<br>trumpet handpiece, and optional<br>probe. | Battery-powered pump, dual tubing,<br>trumpet handpiece, and optional<br>probe. |
| Materials | 304 Stainless Steel, Polycarbonate,<br>420 Stainless Steel, EPDM Rubber,<br>ABS, and PVC. Six (6) AAA alkaline<br>batteries. | Plastic and metal components,<br>material composition unknown. Eight<br>(8) AA alkaline batteries. |
| Performance Testing | Device performance was<br>tested through bench testing<br>methodologies to demonstrate<br>functionality and that the<br>device met the minimum flow<br>rate of 1.2 L/min. Flow Rate range approx..1.52-<br>2.08 L/min Electrical safety per IEC<br>60601-1 Electromagnetic Compatibility<br>(EMC) per IEC 60601-1-2 | Device performance was tested<br>through bench testing<br>methodologies to demonstrate<br>the device met the minimum<br>flow rate of 1.2 L/min. Flow Rate range approx..1.42-<br>1.5 L/min Electromagnetic Compatibility<br>(EMC) per IEC 60601-1-2 |
| Single Use / Reusable | Single Use | Single Use |
{5}------------------------------------------------
| | Ethylene Oxide | Ethylene Oxide |
|------------------------|----------------------------------------------|----------------------------------------------|
| Sterilization | Sterility Assurance Level (SAL) of<br>10-6 | Sterility Assurance Level (SAL) of<br>10-6 |
| Environment Use | Healthcare facility/hospital | Healthcare facility/hospital |
| Principle of Operation | Battery powered mechanical<br>pumping system | Battery powered mechanical<br>pumping system |
#### VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
## Bench Testing
Bench testing was conducted on the CORE E3 device, including:
- Simulated use testing ●
- Functionality verification testing, including flow rate testing and accessory testing
## Electrical safety and electromagnetic compatibility
Electrical safety and EMC testing were conducted on the CORE E3 device. The system complies with the IEC 60601-1 standard for electrical safety and IEC 60601-1-2 standard for EMC.
## Biocompatibility
The biocompatibility evaluation for the CORE E3 device was conducted in accordance with ISO 10993-1:2018 Biological evaluation of medical devices - Part-1: Evaluation and testing within a risk management process and application of applicable principles detailed in FDA Guidance Use of International Standard ISO 10993-1. " Biological evaluation of medical devices – Part 1: Evaluation and Testing within a Risk Manaqement Process" (June 16, 2016). The following tests were completed:
- . Cytotoxicity
- Sensitization
- Irritation or Intracutaneous Reactivity ●
- Material-Mediated Pyrogenicity ●
- Acute Systemic Toxicity ●
- Hemocompatibility
The CORE E3 device is categorized as external communicating with a body contact of tissue/bone/dentin. The duration of body contact is limited (<24 hours).
## Packaqing and Sterilization
Packaging and sterilization were conducted on the CORE E3 device, including:
- Ethylene Oxide Sterilization Validation per ISO 11135 ●
- . Packaging Validation per ISO 11607-1
#### CONCLUSION VIII.
The subject CORE E3 device is substantially equivalent in design, materials, indications for use, principles of operation and technological characteristics to the predicate Hydro-Surg® Laparoscopic Irrigator by Davol Incorporated. Based upon the findings of non-clinical testing, the differences present no issues of safety and efficacy and the subject CORE E3 device is substantially equivalent to the Hydro-Surg® Laparoscopic Irrigator (K961492).