← Product Code [GCJ](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCJ) · K200310

# 1688 Pendulum 4K Camera Head with Integrated Coupler, 1688 4K Camera Control Unit with Advanced Imaging Modality (K200310)

_Stryker · GCJ · Mar 6, 2020 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/GCJ/K200310

## Device Facts

- **Applicant:** Stryker
- **Product Code:** [GCJ](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCJ.md)
- **Decision Date:** Mar 6, 2020
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Intended Use

The 1688 Video Camera is indicated for use in general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy, and plastic surgery whenever a laparoscope/ arthroscope/ arthroscope is indicated for use. A few examples of the more common endoscope surgeries are Laparoscopic cholecystectomy. Laparoscopic hernia repair, Laparoscopic appendectomy, Laparoscopic pelvic lymph node detection, Laparoscopically assisted hysterectomy, Laparoscopic and thorascopic anterior spinal fusion, Anterior cruciate ligament reconstruction, Knee arthroscopy, Small joint arthroscopy, Decompression fixation, Wedge resection, Lung biopsy, Pleural biopsy, Dorsal sympathectomy, Pleurodesis, Internal mammary artery dissection for coronary artery bypass, Coronary artery bypass grafting where endoscopic visualization is indicated and Examination of the evacuated cardiac chamber during performance of valve replacement. The users of the 1688 Video Camera are general surgeons, gynecologists, cardiac surgeons, thoracic surgeons, plastic surgeons, orthopedic surgeons, ENT surgeons and urologists.

## Device Story

1688 4K Camera System provides live surgical video; captures visible and near-infrared (NIR) light via rigid/flexible scopes. System components: Camera Control Unit (CCU), camera head with integrated cable, and coupler. CMOS sensors acquire continuous image data; processed into 4K digital video stream (60 fps) for display. Used in OR by surgeons to visualize surgical field; supports white light, NIR-fluorescence, and NIR-transillumination modes. Output assists surgeon in real-time anatomical visualization and surgical guidance. Device is an evolution of K182160, adding a pendulum camera head configuration.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including performance testing (torque, optical, transmission), electrical safety (ANSI AAMI ES60601-1, IEC 60601-2-18), EMC (IEC 60601-1-2), software verification (IEC 62304), and cleaning/sterilization validation (AAMI TIR 12/30/17/58/77, ISO 17664/14937, ASTM E2314).

## Technological Characteristics

4K CMOS image sensor; 60 fps; digital video output. Supports white light, NIR-fluorescence, and NIR-transillumination. Components: CCU, camera head (standard, integrated, pendulum), and coupler. Standards: IEC 60601-1, IEC 60601-2-18, IEC 60601-1-2. Sterilization/disinfection per AAMI/ISO standards.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- 1688 4K Camera System with Advance Imaging Modality ([K182160](/device/K182160.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Stryker April Malmborg Director, Regulatory Affairs 5900 Optical Ct San Jose, California 95138

### Re: K200310

Trade/Device Name: 1688 Pendulum 4K Camera Head with Integrated Coupler, 1688 4K Camera Control Unit with Advanced Imaging Modality Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: GCJ Dated: February 4, 2020 Received: February 6, 2020

### Dear April Malmborg:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica Mavadia-Shukla Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

K200310

Device Name

1688 4K Camera System with Advance Imaging Modality

#### Indications for Use (Describe)

The 1688 Video Camera is indicated for use in general laparoscopy, ear endoscopy, ear endoscopy, sinuscopy, and plastic surgery whenever a laparoscope/ endoscope/ arthroscope is indicated for use. A few examples of the more common endoscope surgeries are Laparoscopic cholecystectomy, Laparoscopic hernia repair, Laparoscopic appendectomy, Laparoscopic pelvic lymph node detection, Laparoscopically assisted hysterectorny, Laparoscopic and thorascopic anterior spinal fusion, Anterior cruciate ligament reconstruction, Knee arthroscopy, Decompression fixation, Wedge resection, Lung biopsy, Pleural biopsy, Dorsal sympathectomy, Pleurodesis, Internal mammary artery dissection for coronary artery bypass, Coronary artery bypass grafting where endoscopic visualization is indicated and Examination of the evacuated cardiac chamber during performance of valve replacement. The users of the 1688 Video Camera are general surgeons, gynecologists, cardiac surgeons, plastic surgeons, orthopedic surgeons, ENT surgeons and urologists.

Type of Use (Select one or both, as applicable)

| <div> <span> <span style="font-size: 16px;">☑</span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </span> </div> |
|-------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☐</span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </span> </div>  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 C.F.R Part 807.92(c).

### Submitter:

| Applicant:      | Stryker Endoscopy<br>5900 Optical Court<br>San Jose, CA 95138                                                                                  |
|-----------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | April Malmborg, RAC<br>Director, Regulatory Affairs<br>Phone: (408) 754-2473<br>Facsimile: (408) 754-2598<br>Email: april.malmborg@stryker.com |
| Date Prepared:  | February 4, 2020                                                                                                                               |

### Subject Device:

| Name of Device:         | 1688 4K Camera System with Advance Imaging Modality            |
|-------------------------|----------------------------------------------------------------|
| Common or<br>Usual Name | 3-chip Video Camera                                            |
| Classification<br>Name: | Laparoscope, General and Plastic Surgery (21 C.F.R. §876.1500) |
| Regulatory Class:       | II                                                             |
| Product Code:           | GCJ                                                            |
| 510(k) Review<br>Panel: | General & Plastic Surgery                                      |

#### Predicate Device(s):

| 1688 4K Camera System with Advance Imaging Modality | K182160 |
|-----------------------------------------------------|---------|
|-----------------------------------------------------|---------|

NOTE: The predicate devices have not been subject to a design-related recall.

## Device Description:

The 1688 4K Camera System with Advanced Imaging Modality is an endoscopic camera system that produces live video in the surgical field during surgical endoscopic procedures. The system is sensitive in the visible and infrared spectrum. The optical image is transferred from the surgical site to the camera head by a variety of rigid and flexible scopes, which are attached to the camera head. The 1688 4K Camera System consists of three main components: (1) a camera control unit (CCU); (2) a camera head with an integral cable that connects to the CCU (integrated or standard); and (3) a coupler for attaching an endoscope to the camera head.

{4}------------------------------------------------

### Indications for Use:

The 1688 Video Camera is indicated for use in general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy, and plastic surgery whenever a laparoscope/ arthroscope/ arthroscope is indicated for use. A few examples of the more common endoscope surgeries are Laparoscopic cholecystectomy. Laparoscopic hernia repair, Laparoscopic appendectomy, Laparoscopic pelvic lymph node detection, Laparoscopically assisted hysterectomy, Laparoscopic and thorascopic anterior spinal fusion, Anterior cruciate ligament reconstruction, Knee arthroscopy, Small joint arthroscopy, Decompression fixation, Wedge resection, Lung biopsy, Pleural biopsy, Dorsal sympathectomy, Pleurodesis, Internal mammary artery dissection for coronary artery bypass, Coronary artery bypass grafting where endoscopic visualization is indicated and Examination of the evacuated cardiac chamber during performance of valve replacement. The users of the 1688 Video Camera are general surgeons, gynecologists, cardiac surgeons, thoracic surgeons, plastic surgeons, orthopedic surgeons, ENT surgeons and urologists.

| Item                              | Subject Device                                                                                                                                                                                                                                                                                         | Predicate Device                                                        |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|
|                                   | 1688 4K Camera System                                                                                                                                                                                                                                                                                  | 1688 4K Camera System                                                   |
| Manufacturer                      | Stryker                                                                                                                                                                                                                                                                                                | Same as subject device                                                  |
| Submission Reference              | Current Submission                                                                                                                                                                                                                                                                                     | K182160                                                                 |
| Intended Use                      | Endoscopic visible and near-<br>infrared light illumination and<br>imaging during surgical endoscopic<br>procedures                                                                                                                                                                                    | Same as subject device                                                  |
| Indications for Use               | NOTE 1                                                                                                                                                                                                                                                                                                 | Same as subject device                                                  |
| Imaging Modes                     | White Light<br>Near-infrared – fluorescence<br>Near-infrared - transillumination                                                                                                                                                                                                                       | Same as subject device                                                  |
| Safety Standards                  | IEC 60601-1<br>IEC 60601-2-18<br>IEC 60601-1-2                                                                                                                                                                                                                                                         | Same as subject device                                                  |
| Principles of Operation           | Via an optical scope and coupler,<br>light is projected onto one or more<br>complementary metal oxide<br>semiconductor image sensors which<br>acquire a continuous stream of<br>image data. The image data is<br>processed to provide a video stream<br>that is then sent to a display for<br>viewing. | Same as subject device                                                  |
| Image Sensor                      | CMOS image sensor                                                                                                                                                                                                                                                                                      | Same as subject device                                                  |
| Image Processing/ Video<br>Output | Digital                                                                                                                                                                                                                                                                                                | Same as subject device                                                  |
| Resolution                        | 4K (up to3840 x 2160)                                                                                                                                                                                                                                                                                  | Same as subject device                                                  |
| Frame Rate                        | 60 frames per second                                                                                                                                                                                                                                                                                   | Same as subject device                                                  |
| System Components                 | Camera Control Unit<br>Camera Head(s): Standard,<br>Integrated, and Pendulum<br>Coupler                                                                                                                                                                                                                | Camera Control Unit<br>Camera Head(s): Standard & Integrated<br>Coupler |

#### Comparison of Technological Characteristics with the Predicate Device:

{5}------------------------------------------------

NOTE 1: The 1688 Video Camera is indicated for use in general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy, and plastic surgery whenever a laparoscope/ arthroscope is indicated for use. A few examples of the more common endoscope surgeries are Laparoscopic cholecystectomy, Laparoscopic hernia repair, Laparoscopic appendectomy, Laparoscopic pelvic lymph node detection, Laparoscopically assisted hysterectomy, Laparoscopic and thorascopic anterior spinal fusion, Anterior cruciate ligament reconstruction, Knee arthroscopy, Small joint arthroscopy, Decompression fixation, Wedge resection, Lung biopsy, Pleural biopsy, Dorsal sympathectomy, Pleurodesis, Internal mammary artery dissection for coronary artery bypass grafting where endoscopic visualization is indicated and Examination of the evacuated cardiac chamber during performance of valve replacement. The users of the 1688 Video Camera are general surgeons, gynecologists, cardiac surgeons, thoracic surgeons, plastic surgeons, orthopedic surgeons, ENT surgeons and urologists.

## Performance Data:

The following performance data were provided in support of the substantial equivalence determination.:

| Test                    | Method/ Rationale                                                                                                                     | Results |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------|---------|
| Performance Testing     | Locking Knob Actuation Torque<br>Rotating Endobody Torque<br>Pendulum Focus Ring Torque<br>Optical Compliance<br>Coupler Transmission | Pass    |
| Electrical              | ANSI AAMI ES60601-1:2005 + A1:2012<br>IEC 60601-1-6:2010 + A1:2013<br>IEC 60601-2-18:2009                                             | Pass    |
| EMC                     | IEC60601-1-2:2014                                                                                                                     | Pass    |
| Software Verification   | IEC 62304:2015                                                                                                                        | Pass    |
| Cleaning & Disinfection | AAMI TIR 12:2010<br>AAMI TIR 30:2011<br>ASTM E2314:2014<br>ISO 17664:2017                                                             | Pass    |
| Sterilization           | AAMI TIR 12:2010<br>AAMI TIR 17:2008<br>AAMI ST58:2013<br>AAMI ST77:2013<br>ISO 17664:2017<br>ISO 14937:2009                          | Pass    |

NOTE: The 1688 4K Camera System does not require clinical studies to support the determination of substantial equivalence.

# Conclusions:

The 1688 4K Camera System with Advance Imaging Modality is substantially equivalent in design, intended use, principles of operation, technological characteristics and safety features to the predicate device. There are no new issues of safety and/or effectiveness introduced by the 1688 4K Camera System with Advance Imaging Modality when used as instructed.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/GCJ/K200310](https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/GCJ/K200310)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
