← Product Code [GCJ](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCJ) · K193315

# Xenocor Articulating Xenoscope Laparoscope (K193315)

_Xenocor, Inc. · GCJ · Jan 27, 2020 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/GCJ/K193315

## Device Facts

- **Applicant:** Xenocor, Inc.
- **Product Code:** [GCJ](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCJ.md)
- **Decision Date:** Jan 27, 2020
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The Articulating Xenoscope is intended to be used in diagnostic and therapeutic procedures for endoscopy and endoscopic surgery within the thoracic and peritoneal cavities including the female reproductive organs.

## Device Story

System comprises single-use, sterile articulating laparoscope and XenoBox signal converter. Laparoscope features 5mm carbon fiber shaft with ±30° articulating tip and 6-LED ring light source. XenoBox converts digital camera signal to HDMI for display on HD monitors. Used in OR/clinic by surgeons for visualization during endoscopic surgery. Surgeon views real-time video on external monitor to guide diagnostic or therapeutic interventions. Benefits include improved visualization via articulation and smaller 5mm access port compared to rigid predecessors.

## Clinical Evidence

Bench testing only. Included biocompatibility (ISO 10993-1), shaft bend strength, accelerated aging (1-year shelf life), and design validation. Design validation involved 10 trained surgeons evaluating the device; all functional tests met pre-determined acceptance criteria.

## Technological Characteristics

5mm diameter rigid carbon fiber shaft with heat shrink sheathing; ±30° articulating tip; 6-LED ring light source; Type CF applied part; EO gas sterilization (SAL 10^-6); HDMI output via XenoBox; 1-year shelf life.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Xenocor, Xenoscope Laparoscopic System ([K161838](/device/K161838.md))

## Reference Devices

- Xenoscope Laparoscope ([K171344](/device/K171344.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 27, 2020

Xenocor, Inc % Spencer Walker Director of Regulatory Affairs 630 Komas Dr., Suite 200 Salt Lake City, Utah 84108

Re: K193315

Trade/Device Name: Xenocor Articulating Xenoscope Laparoscope Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: GCJ, HET, GCQ Dated: November 26, 2019 Received: November 29, 2019

Dear Spencer Walker:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K193315

Device Name

Xenocor Articulating Xenoscope Laparoscope

Indications for Use (Describe)

The Articulating Xenoscope is intended to be used in diagnostic and therapeutic procedures for endoscopy and endoscopic surgery within the thoracic and peritoneal cavities including the female reproductive organs.

Type of Use (Select one or both, as applicable)

| <span style="font-size: 10pt;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|-----------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 O)</span>                         |

| X Prescription Use (Part 21 CFR 801 Subpart D)

__ Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

| 510(k) Summary       |   |                                                                                                                                                            |
|----------------------|---|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter:           |   | Xenocor®, Inc                                                                                                                                              |
| Contact Person:      |   | Spencer Walker, MSC - Director Regulatory Affairs<br>630 Komas Dr. Suite 200<br>Salt Lake City, UT 84108<br>(801) 581-5080                                 |
| Date Prepared:       |   | Jan. 4, 2020                                                                                                                                               |
| Trade Name:          |   | Xenocor® Articulating Xenoscope™ Laparoscope                                                                                                               |
| Classification Name: |   | Endoscope and Accessories<br>21 CFR §876.1500, Product Code(s) GCJ, GCQ,<br>Gynecologic Laparoscope and Accessories<br>21 CFR §884.1720, Product Code HET, |
| Device Class:        |   | Class II                                                                                                                                                   |
| Predicate Device(s): | ● | K161838 - Xenocor®, Xenoscope™ Laparoscopic System                                                                                                         |

### Device Description:

The Articulating Xenoscope™ System contains two separate functioning components. First is the single-use, sterile Articulating Xenoscope™ Laparoscope Device, which includes a 0° camera on 5 mm rigid shaft with a ± 30° articulating tip, 10 cm or 36 cm long shaft, and high-definition video image. For certain procedures the shorter 10 cm laparoscope is preferred. Likewise, for other procedures, the longer 36 cm laparoscope is preferred. Except for the length difference, the scientific principles, materials of construction and design are otherwise identical. The second is the XenoBox™, which converts the digital signal from the camera to HDMI signal for display onto the HD video screen for the surgeon to view. Together, the Xenoscope™ and the Xenobox™ comprise the Xenoscope™ Laparoscopic System and work synergistically together.

### Indications for Use:

The Indications for Use is the same as the predicate device, which is:

The Articulating Xenoscope™ is intended to be used in diagnostic and therapeutic procedures for endoscopy and endoscopic surgery within the thoracic and peritoneal cavities including the female reproductive organs.

{4}------------------------------------------------

## Technological Characteristics and Substantial Equivalence Comparative Analysis:

Modifications in design and materials of the previously 510(k) cleared Xenoscope Laparoscope device (K161838) resulted in two (2) additional models (XSA-0-0510, XSA-0-0536) which include a smaller 5mm diameter rigid carbon fiber shaft with articulating tip and handle features. The camera is the same as the predicate, and the combined 6 smaller LED's of the subject device equal the 1 larger LED of the predicate in energy and light output.

It has been demonstrated that the modified Xenoscope™ Laparoscopic device is comparable to the predicate and reference device in intended use, fundamental scientific technology, design, principles of operation and functional performance evaluations and is substantially equivalent as summarized in Table 1. The Articulating Xenoscope™ Laparoscopic System has been fully assessed within the Xenocor® Risk Management and Design Controls systems. The differences raise no additional or different questions of safety or effectiveness from that already identified for the predicate and reference devices.

| Table 1: Xenoscope System Table of Substantial Equivalence |                                                                                                                                                                                                                                                                                          |                                                              |                                                                                                                                                                                                                                                                                                |
|------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product<br>Attribute                                       | Predicate Device:<br>Xenoscope Laparoscope<br>(K161838)                                                                                                                                                                                                                                  | Reference Device:<br>Xenoscope Laparoscope<br>(K171344)      | Subject Device:<br>Articulating Xenoscope<br>Laparoscope (K193315)                                                                                                                                                                                                                             |
| Indications for<br>Use                                     | The Xenoscope is<br>intended to be used in<br>diagnostic and therapeutic<br>procedures for endoscopy<br>and endoscopic surgery<br>within the thoracic and<br>peritoneal cavities<br>including the female<br>reproductive organs.                                                         | Same                                                         | Same                                                                                                                                                                                                                                                                                           |
| Classification<br>Info.                                    | Gastroenterology-Urology<br>Devices - Endoscope and<br>Accessories<br>(21 CFR §876.1500),<br>Product code: GCJ<br><br>Obstetrical and<br>Gynecological Diagnostic<br>Devices - Gynecologic<br>Laparoscope and<br>Accessories<br>(21 CFR §884.1720),<br>Product code: HET<br><br>Class II | Same                                                         | Gastroenterology-Urology<br>Devices - Endoscope and<br>Accessories<br>(21 CFR §876.1500),<br>Product code: GCJ & GCQ<br><br>Obstetrical and<br>Gynecological Diagnostic<br>Devices - Gynecologic<br>Laparoscope and<br>Accessories<br>(21 CFR §884.1720),<br>Product code: HET<br><br>Class II |
| Single Use                                                 | Yes                                                                                                                                                                                                                                                                                      | Same                                                         | Same                                                                                                                                                                                                                                                                                           |
| Sterile                                                    | Provided Sterile (EO gas)                                                                                                                                                                                                                                                                | Same                                                         | Same                                                                                                                                                                                                                                                                                           |
| Table 1: Xenoscope System Table of Substantial Equivalence |                                                                                                                                                                                                                                                                                          |                                                              |                                                                                                                                                                                                                                                                                                |
| Product<br>Attribute                                       | Predicate Device:<br>Xenoscope Laparoscope<br>(K161838)                                                                                                                                                                                                                                  | Reference Device:<br>Xenoscope Laparoscope<br>(K171344)      | Subject Device:<br>Articulating Xenoscope<br>Laparoscope (K193315)                                                                                                                                                                                                                             |
| Prescription                                               | Yes                                                                                                                                                                                                                                                                                      | Same                                                         | Same                                                                                                                                                                                                                                                                                           |
| Anatomical<br>Access                                       | Thoracic and abdominal<br>body cavities, hollow<br>organs, and canals,<br>including female<br>reproductive organs.                                                                                                                                                                       | Same                                                         | Same                                                                                                                                                                                                                                                                                           |
| Fundamental<br>Scientific<br>Technology                    | The tip has an imaging<br>sensor that sends digital<br>video information, which<br>the Xenobox coverts to an<br>HDMI output for display on<br>commonly used HD<br>monitors.                                                                                                              | Same                                                         | Same                                                                                                                                                                                                                                                                                           |
| Part No.                                                   | XS-0-1010, XS-0-1036,<br>XD-A1                                                                                                                                                                                                                                                           | XS-30-1010, XS-30-1036                                       | XSA-0-0510, XSA-0-0536,                                                                                                                                                                                                                                                                        |
| Design                                                     | 0° camera angle, 10 mm<br>rigid shaft                                                                                                                                                                                                                                                    | 30° camera angle, 10 mm<br>rigid shaft                       | 0° camera angle, 5 mm<br>rigid shaft with articulating<br>tip                                                                                                                                                                                                                                  |
| Shaft Diameter<br>(OD)                                     | 10 mm                                                                                                                                                                                                                                                                                    | 10 mm                                                        | 5 mm                                                                                                                                                                                                                                                                                           |
| Shaft Tip                                                  | Rigid (0°)                                                                                                                                                                                                                                                                               | Rigid (30°)                                                  | Articulating Tip (≥ 30°)                                                                                                                                                                                                                                                                       |
| Shaft Lengths                                              | 10 cm,<br>36 cm                                                                                                                                                                                                                                                                          | Same                                                         | Same                                                                                                                                                                                                                                                                                           |
| Shaft Material                                             | Gray Polycarbonate Lexan<br>shaft                                                                                                                                                                                                                                                        | Gray Fiber reinforced<br>polymer Polygon PG-F-<br>5400 shaft | Carbon Fiber, covered with<br>heat shrink sheathing                                                                                                                                                                                                                                            |
| Camera Angle                                               | Image: 0º camera angle                                                                                                                                                                                                                                                                   | Image: 30° camera angle with 30° view when rotated           | Image: 0º camera angle with ≥ 30º view when articulated                                                                                                                                                                                                                                        |
|                                                            | 0º camera angle                                                                                                                                                                                                                                                                          | 30° camera angle with 30°<br>view when rotated               | 0º camera angle with ≥ 30º<br>view when articulated                                                                                                                                                                                                                                            |
| Light<br>Source                                            | 1 LED                                                                                                                                                                                                                                                                                    | 1 LED                                                        | 6 LEDs (ring)                                                                                                                                                                                                                                                                                  |
| Handle                                                     | Basic Ergonomic Handle                                                                                                                                                                                                                                                                   | Handle with rotation<br>mechanism                            | Handle with articulation<br>mechanism                                                                                                                                                                                                                                                          |
|                                                            | Image: Basic Ergonomic Handle                                                                                                                                                                                                                                                            | Image: Handle with rotation mechanism                        | Image: Handle with articulation mechanism                                                                                                                                                                                                                                                      |
| Table 1: Xenoscope System Table of Substantial Equivalence |                                                                                                                                                                                                                                                                                          |                                                              |                                                                                                                                                                                                                                                                                                |
| Product<br>Attribute                                       | Predicate Device:<br>Xenoscope Laparoscope<br>(K161838)                                                                                                                                                                                                                                  | Reference Device:<br>Xenoscope Laparoscope<br>(K171344)      | Subject Device:<br>Articulating Xenoscope<br>Laparoscope (K193315)                                                                                                                                                                                                                             |
| Packaging                                                  | Double Pouched Tyvek/<br>Mylar Pouch, and backer<br>card                                                                                                                                                                                                                                 | Same                                                         | Same                                                                                                                                                                                                                                                                                           |
| Sterilization                                              | EO Sterile (SAL 10-6)                                                                                                                                                                                                                                                                    | Same                                                         | Same                                                                                                                                                                                                                                                                                           |
| Shelf Life                                                 | 1 years                                                                                                                                                                                                                                                                                  | Same                                                         | Same                                                                                                                                                                                                                                                                                           |
| Biocompati-<br>bility                                      | Patient contacting<br>components meet ISO<br>10993 standard                                                                                                                                                                                                                              | Same                                                         | Same                                                                                                                                                                                                                                                                                           |
| Applied Part<br>Class (per IEC<br>60601-2-18;<br>(2009))   | Type CF                                                                                                                                                                                                                                                                                  | Same                                                         | Same                                                                                                                                                                                                                                                                                           |

{5}------------------------------------------------

{6}------------------------------------------------

### Verification and Functional/Safety Testing:

The following functional tests were performed. All data met pre-determined acceptance criteria.

- Biocompatibility - Biocompatibility testing was done due to the new shaft materials. All tests were done per the same protocol and acceptance criteria as the predicate device, and per ISO 10993-1 and in accordance with the 2016 FDA guidance document "Use of International Standard ISO 10993-1 for External communicating device, direct tissue contact, duration ≤ 24 hours, the following tests were performed and passed:
	- Cytotoxicity o
	- Sensitization o
	- Irritation/ Intracutaneous Toxicity O
	- Systemic Injection O
	- Material Mediated Pyrogenicity о
- . Shaft Bend Strength Testing - Due to the smaller 5 mm shaft diameter and shaft material change to carbon fiber with addition of the heat shrink sheath, per internal requirements, and based on the risk evaluation for this change, the inspection of the laparoscope shaft to confirm functionality after bending the shaft was done. The protocol and acceptance criteria were the same as the predicate device. All devices tested passed.
- . Accelerated Aging - Per internal requirements and based on the risk evaluation for this device, all devices were mechanically functional after 1-year shelf life testing. The protocol and acceptance criteria were the same as the predicate

{7}------------------------------------------------

device with the added visual examination of the heat shrink materials and the ≥ 30° articulation in both the X and Y axis.

- . Software Validation – The software for the Xenobox and the subject device has not changed. The subject and predicate devices use the same software.
- . Packaging - The proposed changes to the Xenoscope Laparoscope did not affect the packaging or its configuration.
- Design Validation - The modified design was validated by intended users (trained surgeons) through an evaluation of the Articulating Xenoscope™. The study protocol & acceptance criteria were the same as the predicate, with the only deviation being the predicate was evaluated by 3 physicians, and the subject device was evaluated by 10 physicians).

## Conclusion:

The modifications to the Xenoscope™ were made per Xenocor's quality procedures and design control systems. The Articulating Xenoscope™ Laparoscopic System is substantially equivalent to the cited predicate device. Additionally, the Articulating Xenoscope™ Laparoscopic System met all acceptance criteria to confirm safety and effectiveness.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/GCJ/K193315](https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/GCJ/K193315)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).