# EASYGRIP FLO-41 Precision MIS Delivery System (K193137)

_Baxter Healthcare Corportation · GCJ · Jan 8, 2020 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/GCJ/K193137

## Device Facts

- **Applicant:** Baxter Healthcare Corportation
- **Product Code:** [GCJ](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCJ.md)
- **Decision Date:** Jan 8, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The EASYGRIP FLO-41 Precision MIS Delivery System is indicated for delivering compatible hemostatic agents to bleeding sites through a 5 mm or larger trocar.

## Device Story

The EASYGRIP FLO-41 Precision MIS Delivery System is a sterile, single-use endoscopic applicator designed for one-handed delivery of hemostatic agents to surgical bleeding sites. The device consists of a 41 cm cannula (5 mm OD), a 5 mL reservoir syringe, and a dual check valve. The clinician fills the reservoir by attaching a hemostatic agent syringe via Luer lock and depressing the plunger. The device is inserted through a 5 mm or larger trocar; the clinician actuates a trigger handle to dispense the agent. An optional 1.5 mL syringe is provided for saline flushing of residual agent from the cannula. The device enables precise, controlled application of hemostatic matrix, assisting in surgical hemostasis and potentially reducing blood loss during minimally invasive procedures.

## Clinical Evidence

Bench testing only. No clinical data presented. Performance testing included trigger force, malleable tip bend/robustness, interface leak, saline retainment, cannula bend, tensile strength, consecutive/intermittent delivery, and dual check valve functionality. All tests met pre-defined acceptance criteria with 95% confidence. Biocompatibility testing conducted per ISO 10993-1 (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity).

## Technological Characteristics

Sterile, single-use endoscopic applicator. Materials: Stainless steel (collar, wire), polyurethane (tubing), polycarbonate (syringe/valve), thermoplastic elastomer (seals), epoxy adhesive. Features: 41 cm cannula, 5 mL reservoir, dual check valve, trigger-actuated delivery. Connectivity: None. Sterilization: Not specified. Standards: ISO 80369-7 (Luer), ISO 594-1/2 (valve), ISO 10993 (biocompatibility).

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Malleable Tip Endoscopic Applicator ([K123847](/device/K123847.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

January 8, 2020

Baxter Healthcare Corporation Ms. Jeanette Licata Sr. Associate, Regulatory Affairs 32650 N. Wilson Road Round Lake, Illinois 60073

Re: K193137

Trade/Device Name: EASYGRIP FLO-41 Precision MIS Delivery System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: GCJ Dated: November 11, 2019 Received: November 13, 2019

Dear Ms. Licata:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809; medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K193137

Device Name

EASYGRIP FLO-41 Precision MIS Delivery System

Indications for Use (Describe)

The EASYGRIP FLO-41 Precision MIS Delivery System is indicated for delivering compatible hemostatic agents to bleeding sites through a 5 mm or larger trocar.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995,

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the text "Page 1 of 9" at the top. Below that is the word "Baxter" in a stylized, bold, blue font. The word "Baxter" appears to be a logo or brand name. The page number indicates that this is the first page of a nine-page document.

# Section 5. 510(k) Summary

November 11, 2019

#### SPONSOR:

Baxter Healthcare Corporation 32650 N. Wilson Road Round Lake, Illinois 60073

Phone: (224) 948 3996 Fax: (224) 270 4119

# SPONSOR CONTACT:

Jeanette Licata Sr. Regulatory Associate

Phone: (224) 948 3996 Fax: (224) 270 4119 Email: jeanette licata@baxter.com

# IDENTIFICATION OF THE DEVICE:

Common Name: Endoscopic Applicator Trade/Device Name: EASYGRIP FLO-41 Precision MIS Delivery System Classification Panel: 79 General and Plastic Surgery Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: GCJ

|  |  |  |  | Table 1. Code Number for EasyGrip FLO-41 Precision MIS Delivery System |  |
|--|--|--|--|------------------------------------------------------------------------|--|
|  |  |  |  |                                                                        |  |

| Code Number | Name                                          |
|-------------|-----------------------------------------------|
| ADS201897   | EASYGRIP FLO-41 Precision MIS Delivery System |

{4}------------------------------------------------

### PREDICATE DEVICE:

The predicate device, the Malleable Tip Endoscopic Applicator, is manufactured by Micromedics Inc., dba Nordson Medical.

| Trade Name                             | Company                                  | Predicate<br>510(k) | Clearance<br>Date |
|----------------------------------------|------------------------------------------|---------------------|-------------------|
| Malleable Tip Endoscopic<br>Applicator | Micromedics, Inc. dba Nordson<br>Medical | K123847             | 01/08/2013        |

#### REASON FOR SUBMISSION:

The basis for this premarket notification is a modification to the design of the Malleable Tip Endoscopic Applicator to allow for one handed application of compatible hemostatic agents to bleeding sites.

#### INDICATIONS FOR USE:

The EASYGRIP FLO-41 Precision MIS Delivery System is indicated for delivering compatible hemostatic agents to bleeding sites through a 5mm or larger trocar.

### DESCRIPTION OF THE DEVICE:

The EASYGRIP FLO-41 Precision MIS Delivery System (EASYGRIP FLO-41 System) is a sterile, single-use device that consists of two components: (1) one applicator device with a 41 cm long cannula (5 mm outer diameter) and (2) one empty 1.5 mL syringe.

The EASYGRIP FLO-41 System is used to deliver compatible hemostatic agents to bleeding sites. After preparation of compatible hemostatic agent per the instructions, the proposed device is filled by first attaching the compatible hemostatic agent syringe to the syringe port by a Luer lock connection and depressing the syringe plunger, which then fills the enclosed 5 mL reservoir syringe with the hemostatic matrix. The delivery of hemostatic agent to the bleeding site would be performed by actuating the device trigger handle, which can be performed using one hand. As an optional step, if application of the residual hemostatic agent that remains within the cannula is needed, an empty, disposable 1.5 mL syringe is provided and would be filled with non-heparinized saline. Following the same steps as with the hemostatic agent, the reservoir syringe would be filled with 1.5 mL of saline and the residual hemostatic agent would be expelled from the cannula by actuating the device trigger handle.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the text "Page 3 of 9" at the top, indicating it is a page number within a larger document. Below the page number, the word "Baxter" is displayed in a bold, blue font. The word "Baxter" appears to be a logo or brand name.

### TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE:

The proposed device is substantially equivalent to the predicate device, previously cleared under 510(k) premarket notification K123847 on January 8, 2013. The intended use and function of the proposed device are equivalent to the predicate device.

Table 3 is a device comparison table outlining the differences between the predicate and proposed devices.

{6}------------------------------------------------

Cannula

Malleable Tip Tubing:

Unknown

Image /page/6/Picture/2 description: The image shows the word "Baxter" in a bold, blue font. The font is sans-serif and appears to be a corporate logo. The word is horizontally oriented and takes up most of the image space.

Additional testing has confirmed the

biocompatibility of the materials. The different

technological characteristics of the proposed

device do not raise different questions of

safety and effectiveness.

| Features               | Predicate Device<br>Cleared under K123847                                                                                                                      | Proposed Device<br>EasyGrip FLO-41 Precision MIS<br>Delivery System                                                                                                      | Assessment of Differences                                                                                                             |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use           | The Malleable Tip Endoscopic Applicator<br>is intended for use in delivering hemostatic<br>agents to bleeding surgical sites through a<br>5mm or larger trocar | The EASYGRIP FLO-41 Precision<br>MIS Delivery System is intended<br>for delivering compatible<br>hemostatic agents to bleeding sites<br>through a 5mm or larger trocar.  | Not Applicable                                                                                                                        |
| Indications for<br>Use | The Malleable Tip Endoscopic Applicator<br>is intended for use in delivering hemostatic<br>agents to bleeding surgical sites through a<br>5mm or larger trocar | The EASYGRIP FLO-41 Precision<br>MIS Delivery System is indicated<br>for delivering compatible<br>hemostatic agents to bleeding sites<br>through a 5mm or larger trocar. | Not Applicable                                                                                                                        |
| Sterile                | Yes                                                                                                                                                            | Same                                                                                                                                                                     | Not Applicable                                                                                                                        |
| Non-Pyrogenic          | Yes                                                                                                                                                            | Same                                                                                                                                                                     | Not Applicable                                                                                                                        |
| Single Use             | Yes                                                                                                                                                            | Same                                                                                                                                                                     | Not Applicable                                                                                                                        |
| Materials (Fluid Path) | Sheath (Tube):<br>PolyMed® High strength fiber with vinyl<br>ester resin                                                                                       | Sheath (Tube):<br>Same                                                                                                                                                   | The proposed device's collar, malleable tip<br>tubing and adhesive material have passed<br>functional testing for their intended use. |
|                        | Collar:<br>Unknown                                                                                                                                             | Collar:<br>Stainless Steel                                                                                                                                               |                                                                                                                                       |

Stainless Steel

Malleable Tip Tubing:

Polyurethane tubing with

encapsulated 304 Stainless Steel

wire

Adhesive Epoxy

{7}------------------------------------------------

Image /page/7/Picture/2 description: The image contains the word "Baxter" in a bold, blue font. The word is the primary focus of the image and is centrally located. The font appears to be sans-serif and has a slightly italicized appearance.

| Features                                             | Predicate Device<br>Cleared under K123847                                                                              | Proposed Device<br>EasyGrip FLO-41 Precision MIS<br>Delivery System<br>(cannula/collar and collar/malleable<br>tip)<br>Loctite M-31CL | Assessment of Differences                                                                                                                                                                                                                                                         |
|------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Luer Lock<br>Connector<br>(overmolded on<br>Cannula) | Acrylonitrile Butadiene Styrene (ABS)                                                                                  | Same                                                                                                                                  | Not Applicable                                                                                                                                                                                                                                                                    |
| Stylet                                               | PolyMed® composite rod<br>Acrylonitrile Butadiene Styrene (ABS) tip<br>Acrylonitrile Butadiene Styrene (ABS)<br>handle | Proposed device doesn't have a<br>stylet                                                                                              | The proposed device doesn't have a stylet.<br>Design control activities have been conducted<br>and have confirmed that the different<br>technological characteristics of the new device<br>do not raise different questions of safety and<br>effectiveness.                       |
| 1.5 mL empty<br>flush syringe                        | Not applicable                                                                                                         | Barrel: Polycarbonate                                                                                                                 | The predicate device does not have a 1.5 mL<br>empty flush syringe. Design control activities<br>have been conducted and have confirmed that<br>the different technological characteristics of<br>the new device do not raise different questions<br>of safety and effectiveness. |
|                                                      |                                                                                                                        | Plunger:Polycarbonate + 10% Glass<br>Filled<br>Piston Seal:Thermoplastic elastomer                                                    |                                                                                                                                                                                                                                                                                   |

# Table 3. Device Comparison

{8}------------------------------------------------

Image /page/8/Picture/2 description: The image shows the word "Baxter" in a bold, blue font. The word is written in a sans-serif typeface. The letters are evenly spaced and the word is centered in the image. The background is white.

# Table 3. Device Comparison

| Features          | Predicate Device<br>Cleared under K123847 | Proposed Device<br>EasyGrip FLO-41 Precision MIS<br>Delivery System                                       | Assessment of Differences                                                                                                                                                                                                                                                |
|-------------------|-------------------------------------------|-----------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Dual check valve  | Not applicable                            | Body: Polycarbonate<br>Swivel Nut: Copolyester<br>Female Luer Ports: Polycarbonate<br>Valve: Silicone     | The predicate device does not have a dual<br>check valve. Design control activities have<br>been conducted and have confirmed that the<br>different technological characteristics of the<br>new device do not raise different questions of<br>safety and effectiveness.  |
| Reservoir Syringe | Not applicable                            | Plunger: Polypropylene<br>homopolymer<br>Barrel: Polypropylene<br>Piston Seal: Thermoplastic<br>elastomer | The predicate device does not have a reservoir<br>syringe. Design control activities have been<br>conducted and have confirmed that the<br>different technological characteristics of the<br>new device do not raise different questions of<br>safety and effectiveness. |

{9}------------------------------------------------

### SUMMARY OF SUPPORTING DATA:

Baxter Healthcare Corporation conducts risk analyses and design verification tests based on the result of these analyses. All test results meet their acceptance criteria and support that the proposed device is appropriately designed for its intended use.

#### Performance Data:

The following bench tests (Table 4) were conducted to evaluate the functional performance of the EASYGRIP FLO-41 Precision MIS Delivery System. All tests met the acceptance criteria.

| Test                                                                    | Acceptance Criteria                                                                                                                                                                                                                                                                                                                            |
|-------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trigger Force to Deliver Hemostatic Matrix                              | The Trigger Force Test evaluated with<br>95% confidence that for 95% of the<br>devices the force required to deliver<br>hemostatic matrix is less than 30 lbf.                                                                                                                                                                                 |
| Malleable Tip Bend and Robustness Test                                  | The Malleable Tip Bend and Robustness<br>Test demonstrated with 95% confidence<br>that 95% of the malleable tips shall not<br>kink when bent to 90 degrees and will be<br>able to deliver the product.                                                                                                                                         |
| Device Interface Leak test                                              | The Device Interface Leak Test<br>demonstrated with 95% confidence that<br>95% of the device interfaces between the<br>cannula, dual check valve and syringe<br>will not leak during the use of the device.                                                                                                                                    |
| Saline Retainment inside the Device after Residual Flush                | The Saline Retainment Test<br>demonstrated with 95% confidence that<br>90% of the devices would retain the 1.5<br>ml Saline within the device after the<br>residual hemostatic matrix is delivered.                                                                                                                                            |
| Cannula Bend Test at Various Trocar Positions                           | The Cannula Bend Test demonstrated<br>with 95% confidence that 95% of the<br>devices would not break and leak at the<br>device connections after being subjected<br>to the bend test.                                                                                                                                                          |
| Device Tensile Strength                                                 | The Device Tensile Strength Test<br>demonstrated with 95% confidence that<br>95% of the device connections can<br>withstand the tensile pull force of 5lb<br>without failure.                                                                                                                                                                  |
| Consecutive Delivery of Hemostatic Agent                                | The Consecutive Delivery Test<br>demonstrated with 95% confidence that<br>95% of the devices will be able to                                                                                                                                                                                                                                   |
| Test                                                                    | Acceptance Criteria                                                                                                                                                                                                                                                                                                                            |
|                                                                         | successfully deliver 3 consecutive<br>applications of the hemostatic matrix<br>without device failure (Failure is defined<br>as unable to deliver the product,<br>mechanical component failure, syringe<br>damage leak at connections, clogged tip).                                                                                           |
| Intermittent Delivery of Hemostatic Agent                               | The Intermittent Delivery Test<br>demonstrated with 95% confidence that<br>95% of the devices were able to deliver<br>the hemostatic matrix over the range of 2<br>hours without device failure (Failure is<br>defined as unable to deliver the product,<br>mechanical component failure, syringe<br>damage leak at connections, clogged tip). |
| Device Functionality to Start and Stop Delivery of<br>Hemostatic Matrix | The Device Functionality Test<br>demonstrated with 95% confidence that<br>95% of the devices were able to stop the<br>delivery of the hemostatic matrix<br>through the device within 5 seconds,<br>when the force on the trigger was<br>released from the device.                                                                              |
| Dual Check Valve Function Test                                          | The Dual Check Valve Test<br>demonstrated with 95% confidence that<br>90% of the devices would require less<br>than 17.3lbf force to fill the reservoir<br>syringe with hemostatic matrix through<br>the check valve.                                                                                                                          |
| Material Compatibility with Thrombin                                    | The Material Compatibility with<br>Thrombin Test demonstrated with 95%<br>confidence that for 90% of the devices<br>there is no statistical significance in the<br>                                                                                                                                                                            |
| Cannula Luer Test for compliance with ISO 80369-7                       | Tested in accordance to ISO 80369-7                                                                                                                                                                                                                                                                                                            |
| 1.5 ml Syringe Test for compliance to ISO 80369-7                       | Tested in accordance to ISO 80369-7                                                                                                                                                                                                                                                                                                            |
| Dual Check Valve Test for compliance to ISO 594-1 and<br>ISO 594-2      | Tested in accordance to ISO 594-1 and<br>594-2                                                                                                                                                                                                                                                                                                 |
| Reservoir Syringe Fill Capacity                                         | The Reservoir Syringe Fill Capacity Test<br>demonstrated with 95% confidence that<br>90% of the reservoir syringes can<br>accommodate at least 5ml of product.                                                                                                                                                                                 |

### Table 4. Performance Data Summary for Proposed Device

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image shows the text "Page 8 of 9" at the top. Below that is the Baxter logo in blue, with a stylized font. The logo is the main focus of the image.

# Table 4. Performance Data Summary for Proposed Device

{11}------------------------------------------------

#### Biocompatibility:

For the device fluid path, biocompatibility assessments were conducted to meet an ISO 10993-1 Qualification of External Communicating Devices; Tissue, Bone, Dentin; Limited (<24 hrs) contact. Biocompatibility assessments were conducted on final, finished and sterilized devices for all fluid path materials of EASYGRIP FLO-41 Precision MIS Delivery System. The following tests were conducted as part of the biocompatibility testing for the device fluid path of the EASYGRIP FLO-41 Precision MIS Delivery System:

- Cytotoxicity ISO 10993-5
- Sensitization ISO 10993-10
- Intracutaneous (Irritation) Reactivity ISO 10993-10
- Systemic Toxicity (acute) ISO 10993-11
- Materials Mediated Pyrogen ISO 10993-11

Based upon the results of this limited duration, external communicating, tissue, bone, dentin testing, the EASYGRIP FLO-41 Precision MIS Delivery System has been shown to be biocompatible and appropriate for its intended use.

#### CONCLUSION:

The performance testing, (e.g., Interface Leak Test, Tensile Strength Test, Consecutive Delivery of Hemostatic Agent Test), summarized in Table 4, support that the EASYGRIP FLO-41 Precision MIS Delivery System performs comparably to the predicate device. Additional supporting data, such as the biocompatibility testing, demonstrate the safety of the EASYGRIP FLO-41 Precision MIS Delivery System. The performance and safety data support that the EASYGRIP FLO-41 Precision MIS Delivery System does not raise different questions of safety and effectiveness and is substantially equivalent to the predicate device that is currently marketed for the same intended use.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/GCJ/K193137](https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/GCJ/K193137)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).