Cyber BLADE is intended for use under endoscopic visualization for the morcellation and removal of dissected benign prostatic hyperplasia (BPH) tissue during endoscopic surgical procedures in urology.
Device Story
Cyber Blade is a single-use endoscopic morcellator used by urologists to remove dissected BPH tissue. The device consists of a sterile single-use blade handpiece, a reusable internal motor with a battery pack, and an AC-powered charger. It operates via a cutting window created by the relative movement of internal and external tubes, driven by the motor. Vacuum intensity and cutting speed are adjustable via controls on the handpiece. The device requires an external suction system and a collection canister. By enabling efficient tissue removal under endoscopic visualization, it facilitates minimally invasive urological surgery, potentially reducing procedure time and improving patient outcomes compared to traditional removal methods.
Clinical Evidence
Bench testing only. Performance testing included morcellation efficiency (approx. 17g/min), suction testing, dimensional verification, tensile strength of the blade, battery testing, and comparative testing against the Richard Wolf morcellator. Compliance with IEC 60601-1, IEC 60601-1-2, and various ISO standards for biocompatibility (ISO 10993 series), sterilization (ISO 11135, 11737), and packaging (ISO 11607) was verified.
Technological Characteristics
Materials: Stainless steel. Principle: Mechanical morcellation via relative movement of internal/external tubes. Power: Battery-powered handpiece with AC-powered charger. Dimensions: 38 cm length, 0.47 cm external diameter. Connectivity: None. Sterilization: Ethylene oxide (for single-use morcellator). Software: Embedded firmware for motor control.
Indications for Use
Indicated for morcellation and removal of dissected benign prostatic hyperplasia (BPH) tissue during endoscopic urological surgical procedures.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG ADMINISTRATION" in blue as well. The FDA is a federal agency responsible for regulating and supervising the safety of food, drugs, and other products.
July 31, 2020
SIDAM s.r.1. Andrea Tromba Regulatory Affairs Manager Via Statale Sud, 169 Mirandola (MO), 41037 Italy
Re: K192499 Trade/Device Name: Cyber Blade Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: GCJ Dated: June 18, 2020 Received: June 22, 2020
Dear Andrea Tromba:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Martha W. Betz, Ph.D. Acting Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192499
Device Name Cyber Blade
Indications for Use (Describe)
Cyber BLADE is intended for use under endoscopic visualization and removal of dissected benign prostatic hyperplasia (BPH) tissue during endoscopic surgical procedures in urology.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
### 5. 510(K) SUMMARY
| Applicant /<br>Manufacturer<br>Name and Address: | SIDAM s.r.l.<br>Via Statale Sud, 169<br>41037 San Giacomo Roncole<br>MIRANDOLA (MO) - Italy |
|-----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Contact Person: | Andrea Tromba<br>Regulatory Affairs Manager<br>SIDAM s.r.l.<br>Email: andrea@sidamit.it<br>Phone: +39 0535.25523<br>Fax:<br>+39 0535.25635 |
| Date Prepared: | 31/07/2020 |
| Device Name: | Cyber Blade |
| Classification: | Class II |
| Classification Name: | Endoscopes and accessories |
| Common name: | Laparoscope, General and Plastic Surgery |
| Regulation Number: | 21 CFR 876.1500 |
| Product Code: | GCJ |
| Main Predicate device<br>Secondary predicate device | Versacut + Tissue Morcellator (K133272), Lumenis Ltd<br>Morce Scope Set 8970 (K041610), Richard Wolf Medical<br>Instruments Corp. |
### Performance Standards:
There are no mandatory performance standards for this device.
### Description of the device:
Cyber Blade provides efficient and rapid morcellation under endoscopic visualization. The cutting action is driven by the motor in the handpiece.
Cyber Blade comprises:
- a single use sterile blade fixed to a handpiece where controls are placed
- a re-usable internal motor provided with integrated battery pack
- an AC powered charger for the battery pack
Cyber blade is also supplied with aspiration tubing and canister for the collection of evacuated tissue.
{4}------------------------------------------------
| REF | Description | notes |
|-----------|-----------------------------|------------------------|
| 20041001 | Cyber Blade Morcellator | sterile single use |
| 20042001 | Battery-powered motor | non sterile - reusable |
| 20049001 | Charging station | 100-220V, 50/60Hz |
| ABM000091 | Suction tube | sterile single use |
| 077.0562 | Canister (tissue collector) | |
Cyber Blade is comprehensive of the following parts:
| Key Specifications, mechanism of action, device function | |
|----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| Principle of operation | morcellation through the cutting effect of the cutting window, produced by the relative movement of the internal tube and the external tube. |
| Vacuum intensity regulation | Possible through the leverage on the morcellator |
| Speed regulation | Possible through the button on the morcellator |
| Morcellation velocity 1 | About 17 grams per minute |
| Handpiece dimension (external tube) | 38 cm length<br>0.47 cm external diameter |
| Materials in contact with the tissue | Stainless steel |
| Power supply | The morcellator is battery powered.<br>The battery charger is AC power supplied. |
| Supplied Sterile | Yes (only the morcellator) |
| Re-usable | No. The morcellator is single use only.<br>Its internal motor is reusable |
#### Indications for Use
Cyber BLADE is intended for use under endoscopic visualization for the morcellation and removal of dissected benign prostatic hyperplasia (BPH) tissue during endoscopic surgical procedures in urology.
#### Comparison of Technological Characteristics:
The subject device and the identified predicate devices have comparable general intended use and technological characteristics.
Any differences compared to the predicate devices do not present any new types of safety or effectiveness concern.
A comparison of the technical characteristics between the subject and predicate devices follows:
<sup>1</sup> The morcellated tissue coincides with the removed tissue, as all is aspired with the vacuum system
{5}------------------------------------------------
| Specification | Subject device | main predicate device | Secondary predicate device | Comparison to Predicate |
|-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|
| Device Name<br>(K #) | Cyber Blade | VERSACUT + TISSUE<br>MORCELLATOR (K133272) | MORCE SCOPE SET 8970<br>(K041610) | |
| Submitter | SIDAM s.r.l. | LUMENIS LTD. | RICHARD WOLF MEDICAL<br>INSTRUMENTS CORP. | |
| Principle of operation | morcellation through the cutting<br>effect of the cutting window,<br>produced by the relative<br>movement of the internal tube<br>and the external tube. | Morcellation through the cutting<br>effect of the cutting window,<br>produced by the relative<br>movement of the internal tube<br>and the external tube. | Morcellation through the cutting<br>effect of the cutting window,<br>produced by the relative<br>movement of the internal tube<br>and the external tube. | Same |
| Vacuum intensity<br>regulation | Possible through the leverage on<br>the morcellator | Possible from the vacuum<br>generator | Possible from the vacuum<br>generator | equivalent |
| Speed regulation | Possible through the button on<br>the morcellator | Possible from the vacuum<br>generator | Possible from the vacuum<br>generator | equivalent |
| Handpiece dimension<br>(external tube) | 38 cm length<br>0.47 cm external diameter | 39.5 cm length<br>0.47 cm external diameter | 35 or 38 cm length<br>0.475 cm external diameter | In the same range |
| Materials in contact<br>with the tissue | Stainless steel | Stainless steel | Stainless steel | equivalent |
| Power supply | The morcellator is battery<br>powered.<br>The battery charger is AC power<br>supplied. | AC power supplied | AC power supplied | A device battery powered<br>raises less concerns compared<br>to a device AC powered |
| Suction pump | Not part of the system, must use<br>an external suction system. | Part of the system | Part of the system | External FDA cleared suction<br>system to be used. |
| Supplied Sterile | Yes | No | Yes | A device provided sterile raises<br>less concerns compared to a<br>device intended to be sterilized<br>before use. |
| Re-usable | No. The morcellator is single use<br>only.<br>Its internal motor is reusable,<br>but it never comes in contact<br>with the patient | Yes | Available either single use and<br>reusable | A single use device raises less<br>concerns than a re-usable one. |
## Comparison of indications: as detailed in the comparison table below, the indications of the subject device are a sub-set of the indications of its predicate devices.
| Specification | Subject device | main predicate device | additional predicate device | Comparison to Predicate |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name<br>(K #) | Cyber Blade | VERSACUT + TISSUE<br>MORCELLATOR (K133272) | MORCE SCOPE SET 8970<br>(K041610) | |
| Submitter | SIDAM s.r.l. | LUMENIS LTD. | RICHARD WOLF MEDICAL<br>INSTRUMENTS CORP. | |
| Intended use | Cyber BLADE is intended for use<br>under endoscopic visualization<br>for the morcellation and<br>removal of dissected benign<br>prostatic hyperplasia (BPH)<br>tissue during endoscopic surgical<br>procedures in urology. | The VersaCut + Tissue<br>Morcellator is intended for use<br>under<br>direct or endoscopic<br>visualization for the<br>morcellation and removal<br>of dissected tissue during<br>pelviscopic, laparascopic,<br>percutaneous<br>and open surgical procedures<br>whenever access to the surgical<br>site is limited. | MORCE Scope Set 8970, in<br>conjunction with a<br>morcellation probe, and with<br>its sheaths and obturators, is<br>used in the cutting<br>(morcellation) and continuous<br>removal of large tissue masses.<br>In combination with the<br>corresponding auxiliary<br>instruments, it can be used as a<br>nephroscope in the<br>disintegration and<br>removal/aspiration of kidney<br>and bladder stones and<br>the removal of tumors via<br>percutaneous (kidney) or<br>transurethral (bladder)<br>passages, in conjunction with | Sub-set of the indications of<br>the predicate devices.<br>Endoscopy is a less invasive<br>access to the body compared<br>to laparoscopy or other<br>invasive techniques. |
{6}------------------------------------------------
| Specification | Subject device | main predicate device | additional predicate device | Comparison to Predicate |
|----------------------|----------------|--------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|
| Device Name<br>(K #) | Cyber Blade | VERSACUT + TISSUE<br>MORCELLATOR (K133272) | MORCE SCOPE SET 8970<br>(K041610) | |
| Submitter | SIDAM s.r.l. | LUMENIS LTD. | RICHARD WOLF MEDICAL<br>INSTRUMENTS CORP. | |
| | | | intracorporeal 5-4ithotripters<br>e.g. operated pneumatically,<br>by ultrasound, electro-<br>hydraulically or by laser, under<br>endoscopic control.<br>The POWER CONTROL 2303 in<br>conjunction with POWER STICK<br>M4 serves to drive WOLF<br>morcellators for the<br>continuous removal of ablated<br>tissue in endoscopic<br>operations.<br>The SUCTION PUMP is used for<br>aspirating irrigation fluid in<br>conjunction with a<br>resectoscope or a morcellator<br>following laser TURP. | |
# Summary of Non-Clinical Performance Testing :
Cyber Blade devices comply with the following recognized consensus standards or guidelines:
| STANDARD TITLE or guidelines | standard<br>recognized<br>by FDA? | FDA<br>recognition<br>number | exclusions<br>from the<br>standard |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------|------------------------------|------------------------------------|
| IEC 60601-1:2012, ed 3.1, Medical Electrical Equipment - Part 1:<br>General Requirements For Basic Safety And Essential Performance | Yes | 19-4 | No |
| IEC 60601-1-2 Edition 4: 2014, Medical Electrical Equipment - Part 1-<br>2: General Requirements For Basic Safety And Essential Performance<br>- Collateral Standard: Electromagnetic Compatibility - Requirements<br>And Tests | Yes | 19-8 | No |
| EN 556-1:2001/AC:2006 Sterilization of medical devices –<br>Requirements for medical devices to be designated "STERILE" – Part<br>1: requirements for terminally sterilized medical devices | No | N/A | No |
| ISO 15223-1 Third Edition 2016-11-01 Medical devices - Symbols to<br>be used with medical device labels, labelling, and information to be<br>supplied - Part 1: General requirements | Yes | 5-117 | No |
| EN 1041:2008 Information supplied by the manufacturer of medical<br>devices. | No | N/A | No |
| ISO 10993-5 Third edition 2009-06-01 Biological evaluation of<br>medical devices - Part 5: Tests for in vitro cytotoxicity | Yes | 2-245 | No |
| ISO 10993-1 Fourth edition 2009-10-15 Biological evaluation of<br>medical devices – Part 1: Evaluation and testing within a risk<br>management process [Including: Technical Corrigendum 1 (2010)] | Yes | 2-220 | No |
| STANDARD TITLE or guidelines | standard<br>recognized<br>by FDA? | FDA<br>recognition<br>number | exclusions<br>from the<br>standard |
| ISO 10993-7 Second edition 2008-10-15 Biological evaluation of<br>medical devices - Part 7: Ethylene oxide sterilization residuals<br>[Including: Technical Corrigendum 1 (2009)] | Yes | 14-408 | No |
| ISO 10993-10 Third Edition 2010-08-01 Biological evaluation of<br>medical devices - Part 10: Tests for irritation and skin sensitization | Yes | 2-174 | No |
| ISO 11135 Second edition 2014-07-15 Sterilization of health-<br>care products - Ethylene oxide - Requirements for the development,<br>validation and routine control of a sterilization process for medical<br>devices [Including: Amendment 1 (2018)] | Yes | 14-529 | No |
| ISO 11138-1 Third edition 2017-03 Sterilization of health care<br>products - Biological indicators - Part 1: General requirements | Yes | 14-502 | No |
| ISO 11607-1 First edition 2006-04-15 Packaging for<br>terminally sterilized medical devices - Part 1: Requirements for<br>materials, sterile barrier systems and packaging systems [Including:<br>Amendment 1 (2014)] | Yes | 14-454 | No |
| ISO 11607-2 First edition 2006-04-15 Packaging for<br>terminally sterilized medical devices - Part 2: Validation<br>requirements for forming, sealing and assembly processes<br>[Including: Amendment 1 (2014)] | Yes | 14-455 | No |
| ISO 11737-1 Third edition 2018-01 Sterilization of health care<br>products - Microbiological methods - Part 1: Determination of a<br>population of microorganisms on product | Yes | 14-514 | No |
| ISO 14971 Second edition 2007-03-01 Medical devices -<br>Application of risk management to medical devices | Yes | 5-40 | No |
| ISO 17664 Second edition 2017-10 Processing of health care<br>products - Information to be provided by the medical device<br>manufacturer for the processing of medical devices | Yes | 14-515 | No |
| ISO 17664 Second Edition 2017-10 Processing Of Health Care<br>Products - Information To Be Provided By The Medical Device<br>Manufacturer For The Processing Of Medical Devices | Yes | 14-515 | No |
| Software Verification and Validation Testing<br>Software verification and validation testing were conducted and<br>documentation was provided as recommended by FDA's Guidance<br>for Industry and FDA Staff, "Guidance for the Content of Premarket<br>Submissions for Software Contained in Medical Devices | - | - | - |
{7}------------------------------------------------
The following additional testing were performed: morcellation efficiency, suction test, dimensional specification verification, tensile strength of blade, comparative testing with Richard Wolf morcellator, risk analysis on the performance test, battery testing.
{8}------------------------------------------------
# Conclusion
As the indications for use are a sub-set of its predicate devices and the principle of operation and specifications are equivalent, Cyber Blade is substantially equivalent to its predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.