ConnectedOR Hub with Device and Voice Control

K181258 · Stryker · GCJ · Nov 5, 2018 · Gastroenterology, Urology

Device Facts

Record IDK181258
Device NameConnectedOR Hub with Device and Voice Control
ApplicantStryker
Product CodeGCJ · Gastroenterology, Urology
Decision DateNov 5, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The intended use of the Connected OR Hub with Device and Voice Control system is to allow for voice control and remote control of medical device settings by the surgeons or operating room personnel, thereby eliminating the need of manual operation of those devices compatible with the Connected OR Hub, or relying upon verbal communications between the surgeon and other personnel in the operating room in order to adjust the surgical equipment. It also has an additional digital documentation functionality to electronically capture, transfer, store and display medical device data (Class I device function), which is independent of the functions or parameters of any attached device.

Device Story

ConnectedOR Hub acts as a centralized control console for compatible endoscopic and general surgery equipment. Input includes voice commands via headset/base station and manual inputs via handheld IR remote. Device transforms these inputs into control signals to adjust settings of connected surgical devices, eliminating manual adjustment or verbal communication between OR staff. System also performs Class I documentation functions: capturing, transferring, storing, and displaying medical device data independently of connected device parameters. Used in ORs by surgeons and clinical staff. Output allows surgeons to manage equipment settings directly, improving workflow efficiency and reducing reliance on OR personnel. Hardware/software upgrades support higher processing speeds and video resolution compared to predicate.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including device/voice control performance, wireless coexistence (AAMI TIR 69:2017, ANSI C63.27:2017), electrical safety (ANSI/AAMI ES60601-1), EMC (IEC 60601-1-2), usability (IEC 62366-1), and software validation (IEC 62304).

Technological Characteristics

Console-based system with IR remote and voice control headset. Upgraded internal hardware and OS for higher processing/video resolution. Connectivity includes wireless device control. Electrical safety per ANSI/AAMI ES60601-1; EMC per IEC 60601-1-2. No patient-contacting materials.

Indications for Use

Indicated for use with compatible endoscopic and general surgery devices in general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy, and plastic surgery wherever a laparoscope or arthroscope is indicated. Patient population includes those undergoing procedures such as cholecystectomy, hernia repair, appendectomy, pelvic lymph node dissection, hysterectomy, spinal fusion, ACL reconstruction, knee/shoulder/small joint arthroscopy, decompression fixation, wedge resection, lung/pleural biopsy, dorsal sympathectomy, pleurodesis, coronary artery bypass, and cardiac valve replacement. Users include general, gynecological, cardiac, thoracic, plastic, ENT, orthopedic, and urological surgeons.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that. November 5, 2018 Stryker Jay Nayar Staff Regulatory Affairs Specialist 5900 Optical Court San Jose, California 95138 Re: K181258 Trade/Device Name: ConnectedOR Hub with Device and Voice Control Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: GCJ, HRX Dated: September 26, 2018 Received: September 28, 2018 Dear Jay Nayar: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Image /page/1/Picture/7 description: The image shows a digital signature. The signature is from Long H. Chen -S. The date of the signature is 2018.11.05, and the time is 14:30:52 -05'00'. Date: 2018.11.05 for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K181258 #### Device Name ConnectedOR Hub with Device and Voice Control #### Indications for Use (Describe) The ConnectedOR Hub is indicated for use with compatible endoscopic and general surgery devices. The ConnectedOR Hub can be used in general laparoscopy, nasopharyngoscopy, sinuscopy, and plastic surgery wherever a laparoscope, endoscope, or an arthroscope is indicated for use. A few examples of the more common endoscopic surgeries are laparoscopic cholecystectomy, laparoscopic hernia repair, laparoscopic appendectomy, laparoscopic pelvic lymph node dissection, laparoscopically assisted hysterectory, laparoscopic anterior spinal fusion, anterior cruciate ligament reconstruction, knee arthroscopy, small joint arthroscopy, decompression fixation, wedge resection, lung biopsy, pleural biopsy, dorsal sympathectomy, pleurodesis, internal mammary artery dissection for coronary artery bypass, coronary artery bypass grafting where endoscopic visualization is indicated and examination of the evacuated cardiac chamber during performance of valve replacement. ConnectedOR Hub users are general surgeons, gynecologists, cardiac surgeons, thoracic surgeons, orthopedic surgeons, ENT surgeons, and urologists. X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the word "stryker" in a bold, sans-serif font. The letters are black and the background is white. The word is written in all lowercase letters. # Section 5. 510(k) Summary This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92(c). #### General Information: | 510(k) Sponsor: | Stryker Endoscopy | |-------------------------|------------------------------------------------------------------------------------------------------------| | Address | 5900 Optical Court<br>San Jose, CA 95138 | | FDA Registration Number | 2936485 | | Contact Information | April Malmborg<br>Director, Regulatory Affairs<br>Phone: 408-754-2473<br>Email: april.malmborg@stryker.com | | Date Prepared | 11/02/2018 | #### Device Identification: #### Proposed Device: | Trade Name | Connected OR Hub with Device and Voice Control | |-------------------------|------------------------------------------------| | Common Name | SDC 4K Information Management System | | Regulatory Class | II | | Classification Name | Laparoscope, General and Plastic Surgery | | Regulation Number | 21 CFR 876.1500 | | Product Code | GCJ | | Subsequent Product Code | HRX | #### Predicate Device: | Trade Name | Stryker SDC3 HD Information Management System with<br>Wireless Device Control Capability | |---------------------------------------------------------------|------------------------------------------------------------------------------------------| | Common Name | SDC3 HD Information Management System | | Regulatory Class | II | | Classification Name | Laparoscope, General and Plastic Surgery | | Regulation Number | 21 CFR 876.1500 | | Product Code | GCJ | | Subsequent Product Code | HRX | | This device has not been subjected to a design-related recall | | {4}------------------------------------------------ #### Device Description: The Connected OR Hub with Device and Voice Control is a medical device that allows the surgeon to control the state, selection, and settings of any compatible device attached to it. It also has operating room documentation functionalities (Class I device function) to electronically capture, transfer, store and display medical device data independently of the functions or parameters of the connected medical device. The Connected OR Hub with Device and Voice Control (also referred to as proposed device in the following sections) consists of the following: - a. A Connected OR Hub console - b. A Device control package (contains an optional software upgrade and a handheld Infrared (IR) remote control) - c. A Voice control package (contains an optional software upgrade and a headset and base station) The proposed device is the latest iteration of the Stryker SDC3 HD Information Management System with wireless device control (hereafter referred to as "predicate device", cleared under K160332). The device and voice control features of the proposed device are identical to the predicate device. The operating room documentation functionalities of electronically capturing, transferring, storing and displaying medical device data (Class I device functions) independently of the functions or parameters of any connected medical device remain similar to the predicate device. The proposed device accesses existing controls within each device and is secondary to the built-in control interface that is already on each device which is identical to the predicate device Although the device and voice control characteristics of the proposed device remain the same as compared to the predicate device, the operating system and the hardware were altered significantly leading to the need for a traditional submission per the FDA guidance Deciding When to Submit a 510(k) for a Change to an Existing Device and Deciding When to Submit a 510(k) for a Software Change to an Existing Device. This submission hence focuses on providing data regarding substantial equivalence of the proposed device (Stryker Connected OR Hub with Device and Voice Control) to the predicate device (Stryker SDC3 HD Information Management System with wireless device and voice control) #### Indications for Use: #### This submission is not proposing any changes from the indications cleared for the predicate device. The Connected OR Hub is indicated for use with compatible endoscopic and general surgery devices. The Connected OR Hub can be used in general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy, and plastic surgery wherever a laparoscope, or an arthroscope is indicated for use. A few examples of the more common endoscopic surgeries are laparoscopic cholecystectomy, laparoscopic hernia repair, laparoscopic appendectomy, laparoscopic pelvic {5}------------------------------------------------ lymph node dissection, laparoscopically assisted hysterectomy, laparoscopic and thorascopic anterior spinal fusion, anterior cruciate ligament reconstruction, knee arthroscopy, shoulder arthroscopy, small joint arthroscopy, decompression fixation, wedge resection, lung biopsy, pleural biopsy, dorsal sympathectomy, pleurodesis, internal mammary artery dissection for coronary artery bypass, coronary artery bypass grafting where endoscopic visualization is indicated and examination of the evacuated cardiac chamber during performance of valve replacement. Connected OR Hub users are general surgeons, gynecologists, cardiac surgeons, thoracic surgeons, plastic surgeons, orthopedic surgeons, ENT surgeons, and urologists. ### Intended Use: # This submission is not proposing any changes from the intended uses cleared for the predicate device. The intended use of the Connected OR Hub with Device and Voice Control system is to allow for voice control and remote control of medical device settings by the surgeons or operating room personnel, thereby eliminating the need of manual operation of those devices compatible with the Connected OR Hub, or relying upon verbal communications between the surgeon and other personnel in the operating room in order to adjust the surgical equipment. It also has an additional digital documentation functionality to electronically capture, transfer, store and display medical device data (Class I device function), which is independent of the functions or parameters of any attached device. #### Comparison of Technological Characteristics with the Predicate Device: The Connected OR Hub with Device and Voice Control has the same technological characteristics and design as the predicate with the exception of the software operating system and internal hardware components. The software and hardware were upgraded to support product sustainment, higher processing speeds and higher video resolution. All other technological characteristics of the proposed device are the predicate. The proposed device uses the same device control communication protocols as the predicate device, employs the same voice recognition software engine and controls the same types of connected devices as the predicate device. The Connected OR Hub has similar technological characteristics as the predicate device in the following areas: - Operating principles ● - Software architecture ● - Electrical characteristics - Mechanical characteristics - Communication characteristics - Performance characteristics ● - . Compatibility with controllable devices as listed in the product labeling - Energy source - Materials (no patient contacting materials) ● {6}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the word "Stryker" in a bold, sans-serif font. The letters are black and the background is white. The word is the logo for Stryker Corporation, a medical technology company. A complete comparison of technological characteristics between the proposed and predicate devices is provided in Section 14- Substantial Equivalence Discussion. # Summary of Non-Clinical Testing Performed: | Test | Method | Results | |-------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------| | Device and<br>Voice Control<br>performance | Verified if the performance of device and voice control<br>functionalities with compatible devices is in accordance with<br>device performance specifications. | Pass | | Compatibility<br>with Connected<br>OR Spoke | Verified if proposed device was compatible with Connected<br>OR Spoke in accordance with device performance<br>specification. | Pass | | Environmental<br>performance | The functionality of the product was verified for the specified<br>operating temperature and humidity range. | Pass | | Video<br>Compatibility | Verified if the Connected OR Hub is compatible with<br>supported Cameras, Monitors, and peripherals that the Stryker<br>customers use in the field. | Pass | | Memory<br>Profiling and<br>Performance | Verified if the performance and system response times of the<br>Class II Medical Device functionality (Voice and Device<br>Control functionality) while the Connected OR Hub's Class I<br>documentation functionalities is in use are in accordance with<br>device performance specifications. | Pass | | Voice<br>Recognition<br>Performance | Verified if the voice recognition performance of the proposed<br>system when voice commands are issued to the system in<br>accordance with device performance specification. | Pass | | Voice Accuracy<br>Benchmark | Verified if the voice recognition accuracy of the proposed<br>system when voice commands are issued to the proposed<br>system are equivalent or better to the predicate system. | Pass | | Wireless<br>technology<br>performance | Verified the performance and coexistence of the wireless<br>technology in the proposed device in accordance with AAMI<br>TIR 69:2017, ANSI C63.27:2017 and FDA guidance<br>document Radio Frequency Wireless Technology in Medical<br>Devices. | Pass | | Electrical Safety | In accordance with ANSI/AAMI ES60601-1:2005/(R)2012<br>and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012<br>(Consolidated Text) | Pass | | EMC Testing | In accordance with IEC60601-1-2 Edition 4.0 2014-02 | Pass | | Usability | Product was designed and tested per IEC 60601-1-6 Edition<br>3.1 2013-10 and ANSI/AAMI/IEC 62366-1:2015 | Pass | | System<br>validation and<br>simulated user<br>testing | Validated system performance in accordance with defined user<br>needs and device performance specifications | Pass | | Software<br>Validation &<br>Verification | In accordance with IEC 62304 and FDA Guidance Document<br>Principles of Software Validation | Pass | | Test | Method | Results | | Cybersecurity<br>verification and<br>Network scan | In accordance with device performance specifications, cybersecurity risk assessment and FDA guidance documents -<br><i>Content of Premarket Submissions for Management of<br/>Cybersecurity in Medical Devices</i> | Pass | {7}------------------------------------------------ # sti NOTE: The Connected OR Hub with Device and Voice Control does not require clinical studies to support the determination of substantial equivalence. ## Conclusion: Based upon the indications for use, intended use, technological characteristic, performance testing and comparison to the predicate device, the Stryker Connected OR Hub with Device and Voice Control raises no new questions of safety and effectiveness as compared to the predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...