← Product Code [GCJ](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCJ) · K172940

# ANCHOR Tissue Retrieval System (K172940)

_Conmed Corporation · GCJ · Oct 18, 2017 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/GCJ/K172940

## Device Facts

- **Applicant:** Conmed Corporation
- **Product Code:** [GCJ](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCJ.md)
- **Decision Date:** Oct 18, 2017
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Anchor Tissue retrieval system by CONMED is a sterile disposable pouch used with a dedicated introducer for the encapture and removal of an organ or tissue from the body cavity during laparoscopic surgery.

## Device Story

Sterile, disposable retrieval pouch with dedicated introducer; used in laparoscopic surgery. Device consists of rip-stop nylon pouch, stainless steel arms, internal pusher rod, and handle. Surgeon inserts device into body cavity; deploys pouch via pusher rod mechanism; captures tissue; removes pouch. Introducer handle features spring-loaded button and stop mechanism for bag separation. Subject device allows for redeployment within a single patient procedure, unlike the predicate. Used in operating room by surgeons to facilitate tissue extraction; benefits patient by enabling safe removal of tissue through small laparoscopic incisions.

## Clinical Evidence

Bench testing only. No clinical data. Performance verified through non-clinical bench and simulated use testing, including design verification and validation. Biocompatibility testing performed per ISO 10993-1:2009. Sterilization validated per ISO 11135:2014. Redeployment verification testing confirmed the device can be deployed multiple times within a single patient procedure.

## Technological Characteristics

Materials: rip-stop nylon, stainless steel, polycarbonate. Deployment: automatic via pusher rod compression. Separation: push-button mechanism. Sterilization: Ethylene Oxide (ISO 11135:2014). Biocompatibility: ISO 10993-1:2009. Packaging: ISO 11607-1:2006. Single-patient use. No electronic or software components.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Anchor Tissue Retrieval System ([K091930](/device/K091930.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 18, 2017

ConMed Corporation Ms. Rachelle Fitzgerald Senior Specialist, Regulatory Affairs 525 French Road Utica, New York 13502

Re: K172940

Trade/Device Name: ANCHOR Tissue Retrieval System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: GCJ Dated: September 25, 2017 Received: September 26, 2017

Dear Ms. Fitzgerald:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

> Sincerely, Jennifer R. Stevenson -S3

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

#### Indications for Use

510(k) Number (if known)

K172940

Device Name

Anchor Tissue Retrieval System by CONMED

Indications for Use (Describe)

The Anchor Tissue retrieval system by CONMED is a sterile disposable pouch used with a dedicated introducer for the encapture and removal of an organ or tissue from the body cavity during laparoscopic surgery.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

{3}------------------------------------------------

# 510(k) Summary of Safety and Effectiveness

Anchor Tissue Retrieval System™ by CONMED

In accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CFR 807.92. ConMed Corporation is hereby submitting the 510(k) Summary of Safety and Effectiveness for 510(k) number ______________________________________________________________________________________________________________________________________________

## A. Submitter

ConMed Corporation 525 French Road Utica, NY 13502

Establishment Registration: 1320894

## B. Company Contact

Rachelle Fitzgerald Senior Specialist, Regulatory Affairs T: (203) 799-2400 ext 8195 F: (203) 799-2401

### C. Device Name

| Proprietary Name:  | Anchor Tissue Retrieval System™ by CONMED |
|--------------------|-------------------------------------------|
| Common Name:       | Laparoscope, General & Plastic Surgery    |
| Panel:             | Gastroenterology/Urology                  |
| Product Code:      | GCJ                                       |
| Device Class:      | II                                        |
| Regulation Number: | 876.1500                                  |

### D. Predicate Device

| Primary Device Name: | Anchor Tissue Retrieval System |
|----------------------|--------------------------------|
| Company Name:        | CONMED Corporation             |
| 510(k):              | K091930                        |

This predicate has not been subject to a design-related recall.

### E. Device Description

The Anchor Tissue Retrieval System™ by CONMED is a sterile, disposable retrieval pouch for use with a dedicated introducer for the encapture and removal of an organ or tissue from the body cavity during laparoscopic surgery. The Anchor Tissue Retrieval System™ by CONMED consists of a rip stop nylon pouch, two stainless steel arms attached to internal pusher rod connected to a handle at the distal end to the patient. Around the internal pusher rod is the introducer and handle. The introducer handle

{4}------------------------------------------------

incorporates a spring loaded button and stop mechanism is built into the pusher rod. There is a braided drawstring which enters the introducer with the pusher rod and loops through the rip stop nylon pouch.

Differences between the predicate device and the Anchor Tissue Retrieval System™ by CONMED are limited to the additional feature that the Anchor Tissue Retrieval System™™ by CONMED is able to be redeployed within a single patient procedure.

#### F. Intended Use / Indications for Use

The Anchor Tissue retrieval system by CONMED is a sterile disposable pouch used with a dedicated introducer for the encapture and removal of an organ or tissue from the body cavity during laparoscopic surgery.

### G. Technological Characteristics

The Anchor Tissue Retrieval System™ by CONMED has the same technological characteristics as the predicate device in that the Anchor Tissue Retrieval System™ by CONMED has the same design, same deployment mechanism, and utilizes the same rip stop nylon pouch, introducer components, and external introducer as the predicate. As with the predicate, the Anchor Tissue Retrieval System™ by CONMED continues to function as a single patient/procedure use pouch with a dedicated introducer for the encapture and removal of the organ or tissue from the body cavity during laparoscopic surgery. The difference in the subject device from the predicate device is that the subject device has a new feature where, if the surgeon deems it appropriate, the Anchor Tissue Retrieval System™ by CONMED can be redeployed in the same single patient procedure. The Anchor Tissue Retrieval System™ by CONMED is safe and effective and substantially equivalent to the predicate as demonstrated by non-clinical performance testing for the same indications for use, target population, principles of operation, performance specifications, and standards for sterilization, packaging, and biocompatibility.

| Characteristic                               | Anchor Tissue Retrieval System™ by<br>CONMED                                                                                                                                                                                  | Predicate Device<br>K091930                                                                  |                                       |
|----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|---------------------------------------|
| Intended Use                                 | The Anchor Tissue retrieval system by<br>CONMED is a sterile disposable pouch used<br>with a dedicated introducer for the encapture<br>and removal of an organ or tissue from the<br>body cavity during laparoscopic surgery. | Same                                                                                         |                                       |
|                                              | Where Used                                                                                                                                                                                                                    | Operating room                                                                               | Same                                  |
|                                              | Prescription Only                                                                                                                                                                                                             | Yes                                                                                          | Same                                  |
|                                              | Design                                                                                                                                                                                                                        | Specimen bag and a delivery system<br>consisting of an introducer shaft deployment<br>handle | Same                                  |
|                                              |                                                                                                                                                                                                                               | Materials                                                                                    | Nylon, stainless steel, polycarbonate |
| Mechanism to deploy bag                      |                                                                                                                                                                                                                               | Automatic with compression of pusher rod                                                     | Same                                  |
| Mechanism to separate bag<br>from introducer | Push button                                                                                                                                                                                                                   | Same                                                                                         |                                       |

{5}------------------------------------------------

| Characteristic   | Anchor Tissue Retrieval System™ by<br>CONMED                                                                                                   | Predicate Device<br>K091930 |
|------------------|------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| Performance      | There are no FDA performance standards for<br>this device. Device performance was<br>adequately tested through bench testing<br>methodologies. | Same                        |
| Sterilization    | Ethylene Oxide per ISO 11135:2014                                                                                                              | Same                        |
| Biocompatibility | According to ISO 10993-1:2009                                                                                                                  | Same                        |

## H. Performance Testing

Non-clinical bench and simulated use testing demonstrate the Anchor Tissue Retrieval System™ by CONMED is substantially equivalent to the predicate device with regard to indication for use, materials, technology, and performance. Design verification testing demonstrates devices comply with design specifications and applicable sections of ISO 11607-1:2006. ISO 11135:2014. ISO 10993-7:2008. Results from design validation testing performed demonstrate that the Anchor Tissue Retrieval System™ by CONMED conforms to user needs and the intended use. Risk management activities in accordance with ISO 14971 demonstrate the risks associated with the use of the Anchor Tissue Retrieval System™ by CONMED, including redeployment, are mitigated to an acceptable level. No new issues of safety of effectiveness were identified and redeployment has no impact on current device safety and effectiveness. Analyses of these activities conclude the benefits associated with the use of the Anchor Tissue Retrieval System™ by CONMED outweigh the residual risks. Material analysis and testing demonstrate the patient contacting materials are biocompatible and comply with the requirements of ISO 10993-1:2009. Redeployment verification testing determined that the Anchor Tissue Retrieval System™ by CONMED can be deployed more than once within a single patient procedure. Performance testing demonstrates that the performance of the Anchor Tissue Retrieval System™ by CONMED is substantially equivalent to the predicate device.

#### Substantial Equivalence -

There are no differences between the predicate device and the proposed device regarding design, intended use, principals of operation and technical characteristics. Supporting information per this premarket submission confirms that the Anchor Tissue Retrieval System™ by CONMED does not raise any new risks of safety or effectiveness.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/GCJ/K172940](https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/GCJ/K172940)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
