← Product Code [GCJ](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCJ) · K141592

# UNIMICRO SUCTION IRRIATION TUBING SET (K141592)

_Unimicro Medical Systems (Shenzhen) Co., Ltd. · GCJ · Aug 4, 2014 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/GCJ/K141592

## Device Facts

- **Applicant:** Unimicro Medical Systems (Shenzhen) Co., Ltd.
- **Product Code:** [GCJ](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCJ.md)
- **Decision Date:** Aug 4, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The Unimicro Suction Irrigation Tubing set is available with an array of probe designs to facilitate lavage during laparoscopic surgery. This device has applications in laparoscopic, general, thoracic and urology procedures to provide suction and irrigation functions to help flush blood and tissue debris from the operative site during laparoscopy to aid visualization.

## Device Story

Single-use, sterile suction irrigation tubing set for laparoscopic surgery. Device consists of handpiece with two trumpet-style valves, probe, and tubing lines. One line connects to irrigation fluid supply; other connects to aspiration pump. Valves provide manual control of irrigation and aspiration. Handpiece includes adjustable instrument adapter allowing introduction of surgical instruments through probe to operative site. Used in OR by surgeons to evacuate blood, tissue debris, and smoke from surgical site. Benefits include improved visualization and site clearance during minimally invasive procedures.

## Clinical Evidence

Bench testing only. Includes in vitro and in vivo preclinical physical, mechanical, and biocompatibility testing (cytotoxicity, intracutaneous reactivity, maximization sensitization). Results met pre-defined acceptance criteria.

## Technological Characteristics

Disposable, sterile, Class II suction/irrigation tubing set. Features handpiece with two trumpet-style valves, probe, and tubing lines. Adjustable instrument adapter for multi-diameter instrument compatibility. Biocompatibility testing performed per standard requirements. No software or electronic components.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Unimax Suction Irrigation Set ([K103509](/device/K103509.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, overlapping to create a sense of depth and connection. The profiles are rendered in a simple, line-art style.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 4, 2014

Unimicro Medical Systems Company, Ltd. % Mr. Long Yang Shenshen Hlongmed Biotech Company Ltd. R15-08, East Building, Yihai Plaza, Chuangye Road, Nanshan District Shenzhen, Guangdong 518054 People's Republic China

Re: K141592

Trade/Device Name: Unimicro Suction Trigation Tubing set, models: Suction Irrigation Tubing Set A, Suction Irrigation Tubing Set B Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: GCJ Dated: June 20, 2014 Received: June 25, 2014

Dear Mr. Yang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{1}------------------------------------------------

Page 2 - Mr. Long Yang

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# David Krause -S

- for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

#### Indications for Use

510(k) Number (if known) K141592

#### Device Name

Unimicro Suction Irrigation Tubing set, models : Suction Trigation Tubing Set A , Suction Irrigation Tubing Set B

#### Indications for Use (Describe)

The Unimicro Suction Irrigation Tubing set is available with an array of probe designs to facilitate lavage during laparoscopic surgery. This device has applications in laparoscopic, general, thoracic and urology procedures to provide suction and irrigation functions to help flush blood and tissue debris from the operative site during laparoscopy to aid visualization.

Type of Use (Select one or both, as applicable)

|× Prescription Use (Part 21 CFR 801 Subpart D)

_ | Over-The-Counter Use (21 CFR 801 Subpart C)

#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

FORM FDA 3881 (1/14)

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# □ Unimicro Medical Systems (ShenZhen) Co., Ltd. 510(K) SUMMARY

This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR \$807.92.

Type of submission :Traditional

The assigned 510(K) number is:

## 1. Submitter information:

Manufacturer Name: Unimicro Medical Systems (ShenZhen) Co.,Ltd. Address: 2/F, Bldg 31, The 3rd Industrial Area, Mashantou, Gongming Street, Guangming New District, ShenZhen City,Guangdong Province, China Tel : 0086-755-27111581 Fax : 0086-755-27111580 Establishment Registration Number:3010806467

## 2. Contact person:

Long Yang (COO)

Shenzhen Hlongmed Biotech Co., Ltd.

R1508, East Building, Yihai Plaza, Chuangye Road, Nanshan District,Shenzhen, P.R. China

Tel: 0086-755-86664986

Fax: 0086-755-86664933

E-mail: yanglong@hlongmed.com

# 3. Identification of the Device :

Trade Name: Unimicro Suction Irrigation Tubing set Model: Suction Irrigation Tubing Set A. Suction Irrigation Tubing Set B. Common Name: Suction Irrigation Classification Name: Laparoscope, General & Plastic Surgery Regulation Number:876.1500 Device Classification: II Product Code:GCJ

{5}------------------------------------------------

# □ Unimicro Medical Systems (ShenZhen) Co., Ltd.

#### 4. Identification of the Predicative Device

| Device Name                         | Common<br>Name        | Manufacturer                      | Classification<br>and Code | Classification<br>regulation | 510(k)<br>number |
|-------------------------------------|-----------------------|-----------------------------------|----------------------------|------------------------------|------------------|
| Unimax<br>Suction<br>Irrigation Set | Suction<br>Irrigation | Unimax<br>Medical<br>Systems Inc. | Class II,<br>GCJ           | 21CFR<br>876.1500            | K103509          |

Table 1: Predicative Device Information

### 5. Intended Use and Indications for Use of the subject device

The Unimicro Suction Irrigation Tubing set is available with an array of probe designs to facilitate lavage during laparoscopic surgery. This device has applications in laparoscopic gynecologic, general, thoracic and urology procedures to provide suction and irrigation functions to help flush blood and tissue debris from the operative site during laparoscopy to aid visualization.

### 6. Device Description

The Unimicro Suction Irrigation Tubing set is indicated for use in patients undergoing a laparoscopic surgical procedure. It is designed to deliver-sterile irrigation fluids to surgical sites during laparoscopic procedures and to evacuate blood, tissue debris, and smoke from the surgical site.

The suction irrigation set consists of a hand piece equipped with two trumpet style valves, a probe, and connecting lines of tubing, one set designed to attach to a supply of irrigation fluid, and the other designed to attach to an aspiration pump. The valves allow controlled irrigation and aspiration during a surgical procedure.

The hand piece of the suction irrigator is designed to allow instruments to be introduced through the suction irrigation probe to reach the operative site. The instrument adapter is adjustable to allow a variety of instruments and diameters. It is a single use, disposable device and is sold sterile.

### 7. Non-clinical Testing

A series of in vitro and in vivo preclinical physical, mechanical and biocompatibility tests

{6}------------------------------------------------

# UMilllifer

were performed to assess the safety and effectiveness of the Unimicro Suction Irrigation Tubing set.

All the test results demonstrate the performance of Unimicro Suction Irrigation Tubing set meets the requirements of its pre-defined acceptance criteria and intended uses.

The results of the non-clinical testing demonstrate that the Unimicro Suction Irrigation Tubing set is as safe and effective as the predicate devices.

### 8. Substantial Equivalence Determination

The Unimicro Suction Irrigation Tubing set submitted in this 5 10(k) file is substantially equivalent in intended use, design, technology/principles of operation, materials and performance to the cleared Unimax Suction Irrigation Set which is the subject of K103509.

Differences between the devices cited in this section do not raise any new issues of safety or effectiveness.

The comparison to predicate device as below Table 2.

| Item                             | Proposed Device                                                                                                                                                                                                                                                                                                                      | Predicate Device                                                                                                                                                                                                                                                                                                                     |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name                       | Unimicro Suction<br>Irrigation Tubing set                                                                                                                                                                                                                                                                                            | Unimax Suction Irrigation Set                                                                                                                                                                                                                                                                                                        |
| 510(K)<br>Submitter              | Unimicro Medical<br>Systems (ShenZhen)<br>Co.,Ltd.                                                                                                                                                                                                                                                                                   | Unimax Medical Systems<br>Inc.                                                                                                                                                                                                                                                                                                       |
| 510(K) Number                    | ----------                                                                                                                                                                                                                                                                                                                           | K103509                                                                                                                                                                                                                                                                                                                              |
| Classification<br>regulation     | 21 CFR 876.1500                                                                                                                                                                                                                                                                                                                      | 21 CFR 876.1500                                                                                                                                                                                                                                                                                                                      |
| Classification<br>and Code       | Class II ,<br>GCJ                                                                                                                                                                                                                                                                                                                    | Class II ,<br>GCJ                                                                                                                                                                                                                                                                                                                    |
| Device<br>Classification<br>Name | Laparoscope, General &<br>Plastic Surgery                                                                                                                                                                                                                                                                                            | Laparoscope, General &<br>Plastic Surgery                                                                                                                                                                                                                                                                                            |
| Indications for                  | This device is available with                                                                                                                                                                                                                                                                                                        | This device is available with                                                                                                                                                                                                                                                                                                        |
| Use                              | an array of probe designs to facilitate lavage during laparoscopic surgery.<br>This device has applications in laparoscopic gynecologic, general, thoracic and urology procedures to provide suction and irrigation functions to help flush blood and tissue debris from the operative site during laparoscopy to aid visualization. | an array of probe designs to facilitate lavage during laparoscopic surgery.<br>This device has applications in laparoscopic gynecologic, general, thoracic and urology procedures to provide suction and irrigation functions to help flush blood and tissue debris from the operative site during laparoscopy to aid visualization. |
| Sterile                          | Yes                                                                                                                                                                                                                                                                                                                                  | Yes                                                                                                                                                                                                                                                                                                                                  |
| Disposable                       | Yes                                                                                                                                                                                                                                                                                                                                  | Yes                                                                                                                                                                                                                                                                                                                                  |
| Biocompatibility                 | Cytotoxicity Test;<br>Intracutaneous Reactivity Test;<br>Maximization Sensitization Test;                                                                                                                                                                                                                                            | Cytotoxicity Test;<br>Intracutaneous Reactivity Test;<br>Maximization Sensitization Test;                                                                                                                                                                                                                                            |
| Specification                    | The device consists of a hand piece equipped with two trumpet style,valves, a probe, and connecting lines of tubing                                                                                                                                                                                                                  | The device consists of a hand piece equipped with two trumpet style,valves, a probe, and connecting lines of tubing                                                                                                                                                                                                                  |
| Function                         | The tubing one set designed to attach to a supply of irrigation fluid, and the other designed to attach to an aspiration pump.<br><br>The valves allow controlled irrigation and aspiration during a surgical procedure                                                                                                              | The tubing one set designed to attach to a supply of irrigation fluid, and the other designed to attach to an aspiration pump.<br><br>The valves allow controlled irrigation and aspiration during a surgical procedure                                                                                                              |

Table 2 : Comparison to Predicate Device

{7}------------------------------------------------

# பார்!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!

{8}------------------------------------------------

# പ്രസിച്ചിത്രത്തി Unimicro Medical Systems (ShenZhen) Co., Ltd.

| Additional<br>accessories | / | Reusable Monopolar Probe:<br>Description Size: |
|---------------------------|---|------------------------------------------------|
|                           |   | Spatula Probe:5mm×33mm                         |
|                           |   | J-Hook Probe:5mm×33mm                          |
|                           |   | L-Hook Probe:5mm×33mm                          |

### 9. Conclusion

After analyzing bench tests, safety testing data, it can be concluded that: Unimicro Suction Irrigation Tubing set is safe and effective as the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/GCJ/K141592](https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/GCJ/K141592)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
