← Product Code [GCJ](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCJ) · K123280

# NEOCLOSE HASSON, NEOCLOSE UNIVERSAL (K123280)

_Neosurgical, Ltd. · GCJ · Apr 18, 2013 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/GCJ/K123280

## Device Facts

- **Applicant:** Neosurgical, Ltd.
- **Product Code:** [GCJ](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCJ.md)
- **Decision Date:** Apr 18, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The neoClose device is intended to facilitate the delivery of absorbable AutoAnchors through soft tissues of the body during endoscopic/ laparoscopic surgery.

## Device Story

Suture passer device for endoscopic/laparoscopic surgery; facilitates delivery of absorbable AutoAnchors through soft tissues for approximation. Two variants: neoClose Hasson (includes sealed trocar anchoring) and neoClose Universal (standard soft tissue approximation). System consists of AccuGuide (delivery guide) and AutoAnchors loaded on drivers. Operated by surgeons in clinical settings. Device delivers anchors to secure tissue; enables closure of surgical sites. Benefits include efficient soft tissue approximation during minimally invasive procedures.

## Clinical Evidence

Evidence includes design verification bench testing and an acute GLP animal study. The animal study compared the neoClose device to the Carter-Thomason CloseSure System®, demonstrating substantial equivalence in safety and effectiveness for soft tissue approximation.

## Technological Characteristics

Suture passer system; includes AccuGuide (Hasson or Universal variants), drivers, and absorbable AutoAnchors. Mechanical operation for suture/anchor delivery. Materials and sterilization methods are equivalent to predicate. No software or electronic components.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Carter-Thomason CloseSure System®

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

page 1/2

# APR 1 8 2013

# 510(k) Summary For neoClose Device

In accordance with 21 CFR 807.92 of the Federal Code of Regulations the following 510(k) summary is submitted for the neoClose Device

| 510(k) Number         | K123280                                                                                                              |
|-----------------------|----------------------------------------------------------------------------------------------------------------------|
| Date Prepared         | 19 October 2012                                                                                                      |
| Proprietary Name      | neoClose Hasson and neoClose Universal                                                                               |
| Common Name           | Suture passer                                                                                                        |
| Classification Name   | §876.1500 Product code GCJ<br>Endoscope and Accessories.                                                             |
| Device Classification | Class II                                                                                                             |
| Device Panel          | General and Plastic Surgery Devices                                                                                  |
| Predicate Device      | Carter-Thomason CloseSure System®                                                                                    |
| Submitter             | neoSurgical Ltd.<br>Block 12<br>Galway Technology Park,<br>Parkmore,<br>Galway,<br>Ireland<br>Ph: +353 (0)91 421 000 |
| Contact Person        | Orla Brennan<br>Block 12<br>Galway Technology Park,<br>Parkmore,<br>Galway,<br>Ireland<br>Ph: +353 (0)91 421 000     |

### Device Description

The neoClose device is intended to facilitate the delivery of absorbable AutoAnchors through soft tissues of the body during endoscopic surgery. There are two neoClose product codes, neoClose Hasson and neoClose Universal, each consisting of an AccuGuide and two AutoAnchors loaded onto two Drivers.

The neoClose Hasson is designed to provide sealed anchoring of a laparoscopic trocar and suture placement for subsequent soft tissue approximation. The neoClose Universal is equivalent to the neoClose Hasson design in its method of soft tissue approximation but does not provide for sealed anchoring of a trocar. Soft tissue approximation is facilitated by delivering two AutoAnchors, through an AccuGuide with a Driver. The AutoAnchors and Drivers for both the neoClose

{1}------------------------------------------------

K123280 Page 2/2

Hasson and Universal are identical but the AccuGuides differ in construction. The Hasson AccuGuide features Suture Locks, a flexible Sleeve, and two Guide Channels while the Universal AccuGuide features a Thumb Grip and one Guide Channel.

### Intended Use

The neoClose device is intended to facilitate the delivery of absorbable AutoAnchors through soft tissues of the body during endoscopic/ laparoscopic surgery.

## Summary of Technological Characteristics of Current Device Compared to the Predicate Device

The technical characteristics of the applicant device are substantially equivalent to the predicate device(s) with respect to indications for use, product design, materials, packaging, labeling and sterilization methods.

### Support of Substantial Equivalence

neoSurgical has submitted information on indication for use, design and principle of operation, biocompatibility and performance characteristics to establish that the neoClose device is substantially equivalent to the currently marketed predicate devices. A detailed justification for substantial equivalence was completed which includes a side-by-side comparison of the product attributes of the applicant device and predicate devices. Design Verification bench testing and a GLP Animal Study support the performance of the neoClose device and demonstrates that it is at least as safe and effective as the predicate devices.

- Intended Use . The neoClose components have the same intended use as the predicates.
- . Physical Characteristics

There are no significant technological differences between the neoClose device and predicate devices affecting intended use or safety and effectiveness.

#### . Anatomical Sites

The neoClose device and its predicates may be utilized in the same anatomical site.

- . Performance
Design Verification bench testing and an acute GLP Animal Study support the use of the neoClose device and demonstrate that it is at least as safe and effective as the predicate devices.

#### Safety Characteristics .

The GLP Animal Study demonstrated that the safety and effectiveness of the neoClose and the primary predicate device, the Carter-Thomason CloseSure System®, are substantially equivalent.

The neoClose device is substantially equivalent to the predicate devices since it has the same intended use, does not raise new concerns regarding safety and effectiveness and is at least as safe and effective as the predicate devices when used in accordance with the Instructions for Use.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The caduceus is a symbol of medicine and health. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol.

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

April 18, 2013

neoSurgical, Ltd. % Orla Brennan Director, Quality Assurance & Regulatory Affairs Block 12, Galway Technology Park Parkmore, Galway, Ireland

Re: K123280

Trade/Device Name: neoClose Hasson, neoClose Universal Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class 2 Product Code: GCJ Dated: March 13, 2013 Received: March 25, 2013

Dear Orla Brennan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{3}------------------------------------------------

Page 2 - Orla Brennan

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours, FOR

# Peter Di Rumm -S

Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use Statement

| 510(k) Number:       | K123280                                                                                                                                                        |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name:         | neoClose                                                                                                                                                       |
| Indications for Use: | The neoClose device is intended to facilitate the delivery of absorbable AutoAnchors through soft tissues of the body during endoscopic/ laparoscopic surgery. |

Prescription Use__X Use_ (Per 21 CFR 801.109)

OR

Over-The-Counter

# PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE)

# David Krause

(Division Sign-Off) Division of Surgical Devices 510(k) Number: K123280

1 of 1

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/GCJ/K123280](https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/GCJ/K123280)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
