This device is designed as an organ and tissue retractor for the use in minimally invasive surgical procedures to elevate organs and tissues to provide better access as well as visualization and stabilization of surgical sites.
Device Story
Surgical Retractor is a reusable manual instrument used during laparoscopic surgery. Device functions to mobilize, maneuver, elevate, and stabilize organs and tissues within the surgical cavity. Operated by surgeons to improve access and visualization of surgical sites. Device provides mechanical retraction; no electronic or automated components.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Reusable manual surgical instrument. Materials and dimensions are consistent with standard laparoscopic retractors. No energy source, software, or connectivity.
Indications for Use
Indicated for use in minimally invasive surgical procedures to elevate or retract organs and tissues to provide access, visualization, and stabilization of surgical sites.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
{0}------------------------------------------------
K112659
# 9 2012
MAR - 9 2012
# SECTION 5:
## 510(k) SUMMARY
| Submitter: | Surgical Innovations plc.<br>Clayton Wood House<br>6 Clayton Wood Bank<br>Leeds<br>LS16 6QZ<br>United Kingdom<br><br>Tel.: +44 (0)113 230-7597<br>Fax: +44 (0)113 230-7598 |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Stephen Seed<br>Quality Manager |
| Date Summary<br>Prepared: | September 2011 |
| Trade Name: | Surgical Retractor |
| Common Name: | Surgical Retractor |
| Classification Name: | Laparoscope, General and Plastic Surgery<br>(21 CFR 876.1500, Product Code GCJ) |
| Equivalent to: | Laparoscopic Retractors (Diamond-Flex) (K092684) Cardinal Health<br>Nathanson Liver Retractor (K942002) Automated Medical Products<br>Endo-Retract (K914190) Covidien |
Surgical Innovations Limited
Congical Innovations Limited
Corpor w
1
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{1}------------------------------------------------
Device Description: The Surgical Retractor is a Reusable device intended for mobilizing and maneuvering organ and tissue within the cavity during laparoscopic Surgical procedures.
Intended Use:
This device is designed as an organ and tissue retractor for the use in minimally invasive surgical procedures to elevate organs and tissues to provide better access as well as visualization and stabilization of surgical sites.
Substantial Laparoscopic Retractors- Cardinal Health, Inc. (K092684) Equivalence: Nathanson Liver Retractor-Automated Medical Products (K942002) Endo-Retract- Covidien (K914190)
# Surgical Innovations Limited
Clayton Wood House . 6 Clayton Wood Bank . Tel. +44(0)113 2307597 . Email. signing . Beg No. England 2298163
Leeds . LE E E E E ST reed? | 1216 602 Fax. +44(0)113 2307598 Web. www.surginno.com Wat Reg No. GB 519567317
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird with three wing-like shapes. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Surgical Innovations % Ms. Tracey Fearnley Clayton Wood House 6 Clayton Wood Bank Leeds LS166QZ United Kingdom
Re: K112659
Trade/Device Name: Surgical Retractor Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: GCJ Dated: March 2, 2012 Received: March 5, 2012
Dear Ms. Fearnley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r lease oe devised that I Dr. unation that your device complies with other requirements of the Act
MAR - 9 2012
{3}------------------------------------------------
Page 2 - Ms. Tracey Fearnley
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
for Mark N. Melkerson
Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### SECTION 4:
#### INDICATIONS FOR USE STATEMENT
n(k) Number (if known): Not known
1. NOT KNOWN
evice Name:
Surgical Retractor
dications for Use
This device is designed as an organ and tissue retractor for the use in minimally invasive surgical procedures to elevate or retract organs and tissues to provide better access as well as visualization and stabilization of surgical sites.
cription Use_ r 21 CFR 801 Subpart D) AND/ OR
Over-The-Counter Use (Per 21 CFR 801 Subpart C)
(Please DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil R.P. Ogden for mx
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number. K112659