TROKAMED GMBH ENDOSCOPES AND ACCESSORIES

K060233 · Trokamed GmbH · GCJ · Jul 19, 2006 · Gastroenterology, Urology

Device Facts

Record IDK060233
Device NameTROKAMED GMBH ENDOSCOPES AND ACCESSORIES
ApplicantTrokamed GmbH
Product CodeGCJ · Gastroenterology, Urology
Decision DateJul 19, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The laparoscopes and accessories are intended for use in providing access to and visualization of body cavities, organs, and canals to perform various diagnostic and therapeutic surgical procedures. The arthroscope is indicated for illumination during joint examinations, arthroscopies, biopsies and diagnosis of joint disease in minimally invasive procedures of the knee, shoulder, wrist (carpal tunnel syndrome), temporal mandibular joint, ankle and elbow. The cystoscopes and accessories are used for visualizing body cavities and organs (via natural passages). For examination, diagnosis and for therapy in conjunction with endoscopic accessories / auxiliary instruments through the schope's working channel. The instruments are reusable; they are intended to be cleaned, sterilized and reused.

Device Story

Trokamed GmbH Endoscopes and accessories include laparoscopes, arthroscopes, cystoscopes, trocar sleeves, Verres insufflation cannulas, and suction-irrigation systems. Devices provide access to and visualization of body cavities, organs, and canals for diagnostic and therapeutic surgical procedures. Operated by physicians in clinical settings. Devices are reusable and require manual cleaning and steam sterilization by the user. No electronic or software components described.

Clinical Evidence

No clinical data. Biocompatibility testing performed per ISO 10993 parts 1, 5, and 10.

Technological Characteristics

Reusable surgical endoscopes and accessories. Materials verified for biocompatibility per ISO 10993. Sterilization via steam-sterilizer (134°C/270°F, 3 bar, 10 minutes). Mechanical design for manual operation.

Indications for Use

Indicated for patients requiring diagnostic or therapeutic visualization of body cavities, organs, and canals via laparoscopy, arthroscopy (knee, shoulder, wrist, TMJ, ankle, elbow), or cystoscopy. Reusable instruments for use by clinicians.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ JUL 1 9 2006 Image /page/0/Picture/2 description: The image shows the logo for TROKAMED GmbH. The logo is black and white and features the word "TROKAMED" in bold letters. The "O" in TROKAMED is stylized as a circle with a break in the upper left quadrant. To the right of the word "TROKAMED" is the abbreviation "GmbH" in smaller letters. K .06.02 3.3. ........ Page 1 of 2 # 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS IN Accordance with SMDA of 1990 DATE OF APPLICATION: 2006-01-26 Submitted by: Trokamed GmbH Kleine Breite 17 78187 Geisingen Germany Tel.: +49 (0)7704/9244-0 Fax: +49 (0) 7704/9244-44 E-Mail: info@trokamed.de Homepage: www.trokamed.de # 1. Device Description Trade Name: Edoscopes and accessories Common Name: Endoscopes The Trokamed GmbH Endoscopes and accessories consist of: - A Various manually operated surgical instruments - A Several Trocar sleeves and accessories, Verres insufflation cannula and suction-firrigation systems - > Several Endoscopes (Arthroscope, Cystoscope) {1}------------------------------------------------ Trokamed GmbH Kob0233 Image /page/1/Picture/2 description: The image shows the logo for TROKAMED GmbH. The logo is black and white and features the company name in bold, sans-serif font. The "O" in TRO is stylized as a semi-circle, and the "GmbH" is written in a smaller font size to the right of the word "KAMED". # 2. Classification | Device: | Endoscope and accessories | |--------------------|---------------------------| | Panel: | 876 | | Product Code: | GCJ | | Device Class: | 2 | | Regulation Number: | 876.1500 | # 3. Intended Use The laparoscopes and accessories are intended for use in providing access to and visualization of body cavities, organs, and canals to perform various diagnostic and therapeutic surgical procedures. The arthroscope is indicated for illumination during joint examinations, arthroscopies, biopsies and diagnosis of joint disease in minimally invasive procedures of the knee, shoulder, wrist (carpal tunnel syndrome), temporal mandibular joint, ankle and elbow. The cystoscopes and accessories are used for visualizing body cavities and organs (via natural passages). For examination, diagnosis and for therapy in conjunction with endoscopic accessories / auxiliary instruments through the schope's working channel. The instruments are reusable; they are intended to be cleaned, sterilized and reused, ### 4. SUBSTANTIAL EQUIVALENCE | Manufacturer | 510(k) Number | Device name | |----------------|---------------|-------------------------------------------| | A Richard Wolf | K 991718 | Operating Laparoscope, 10-5mm, angeled... | | B Hans Hermann | K 051610 | Laparoscopes and accessories | | C Ackermann | K 974382 | Ackermann surgical instruments | #### 5. BIOCOMPATIBILITY The biocompatibility has been approved based on ISO 10993 parts one, five and ten by the accredited Laboratory Bioservice. # 6. STERILIZATION BY USER Trokamed GmbH delivers all endoscopes and accessories in Non-Sterile conditions. The user may sterilize these devices by using a validated and applicable sterilization process. Trokamed GmbH recommends to use a steam-sterilizer that uses a validated sterilization cycle of 134°C / 270°F, 3 bar, for 10 minutes. {2}------------------------------------------------ Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines representing its body and wings. The eagle is facing right. Encircling the eagle is the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" in a circular arrangement. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUL 1 9 2006 Trokamed GmbH % Karhlheinz Trondle President Kleine Breite 17 Geisingen, Germany 78187 Re: K060233 Trade/Device Name: Endoscopes and accessories Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: GCJ Dated: June 21, 2006 Received: June 28, 2006 Dear Karhlheinz Trondle: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Scctions 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 – Karhlheinz Trondle This letter will allow you to begin marketing your device as described in your Section 510(k) rremarket notification. The FDA finding of substantial equivalence of your device to a legally premaince no licate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you attent office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Ваваевичем Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ K060233 #### Indications for Use 510(k) Number (if known): кобог33 Device Name: Endoscopes and accessories Indications For Use: The laparoscopes and accessories are intended for use in providing access to and visualization of body cavities, organs, and canals to perform various diagnostic and therapeutic surgical nrocedures . The arthroscope is indicated for illumination during joint examinations, arthroscopies, biopsies and diagnosis of joint disease in minimally invasive procedures of the knee, shoulder, wrist (carpal tunnel syndrome), temporal mandibular joint, ankle and elbow. The cystoscopes and accessories are used for visualizing body cavities and organs (via natural pessages). For examination, diagnosis and for therapy in conjunction with endoscopic accessories / auxiliary instruments through the scope s working channel. The instruments are reusable; they are intended to be cleaned, sterilized and reused. Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) AND/OR Concurrence of CDRH, Office of Device Evaluation (ODE) Barbara Buehum for MKM and Neurological Devices Page 1 of _ 1 510(k) Number K060233
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...