← Product Code [GCJ](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCJ) · K012771

# TROCAR SLEEVE AND ACCESSORIES, MODELS 1287-XX-XX (K012771)

_Pajunk GmbH · GCJ · Nov 14, 2001 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/GCJ/K012771

## Device Facts

- **Applicant:** Pajunk GmbH
- **Product Code:** [GCJ](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCJ.md)
- **Decision Date:** Nov 14, 2001
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic, 3rd-Party Reviewed

## Intended Use

Manually operated surgical device intended for making incisions into the patient's body to allow insertion of endoscopes and endoscopic accessories during general endoscopic and laparoscopic procedures. Manually operated surgical device intended for creating a surgical space by dissecting layers of connective tissue along natural planes to allow insertion of endoscopes and endoscopic accessories during general endoscopic and laparoscopic procedures. Manually operated surgical device intended for providing a well anchored and gas-tight entry port with an instrument channel to allow insertion of endoscopes and endoscopic accessories during general endoscopic and laparoscopic procedures.

## Device Story

Pajunk Trocars, Trocar Sleeves, and Balloon Systems are manually operated surgical instruments used by physicians in endoscopic and laparoscopic procedures. Trocars and sleeves facilitate body wall puncture and provide access for endoscopes/accessories. Safety-Trocars utilize a spring-loaded blunt obturator (Veress-concept) to cover the cutting tip upon abdominal entry, preventing internal injury. Distension and Ring-Anchor balloon systems provide tissue dissection and gas-tight, anchored entry ports for extraperitoneal surgery. Accessories include disposable valve tops (polymer), fixable slide cones for Hasson technique, reducer sleeves for instrument diameter adjustment, and trocar stops for fixation. Devices are used in clinical/surgical settings. Output is physical access to the surgical site; clinical benefit includes minimally invasive access and controlled tissue dissection.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Manual surgical instruments. Materials include polymer (valve tops) and metal (reusable tubes). Features include spring-loaded blunt obturator (Safety-Trocar), balloon-based distension/anchoring systems, and mechanical fixation components (slide cones, trocar stops).

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image contains a combination of letters and numbers. The characters "10F" are present, followed by the number "2". The characters appear to be handwritten in a dark ink or marker on a white background. The style of writing is somewhat stylized.

Image /page/0/Picture/2 description: The image shows the word "PAJUNK" in white letters against a black background. The letters are bold and sans-serif. The word is centered in the image and takes up most of the space.

MEDIZINTECHNOLOGIE

#### NOV 1 4 2001 510(k) Summary of Safety and Effectiveness

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR Part 807.92.

Applicant:

Pajunk GmbH Am Holzplatz 5 - 7 D-78187 Geisingen Germany 07704 9291-0 Tel.

Contact:

Martin Hauger Technical Director

Device Identification:

Common Name: Trocars, Trocar Sleeves

Trade Name: Trocar and Trocar Sleeves, Distension System, Anchoring System

# Indication and Device Description: General

The Pajunk Trocars together with the Trocar Sleeves, the Pajunk Balloon Systems and the according accessories are manually operated surgical devices used by physicians for making incisions into the patient's body to allow insertion of endoscopes and endoscopic accessories during general endoscopic and laparoscopic procedures. The devices follow the FDA Draft Guidance for the Content of Premarket Notifications for Endoscopes used in Gastroenterology and Urology, dated 3/17/95.

## Trocar Sleeves

The Paiunk Trocar Sleeves are rigid tubes placed with the help of Trocars or Obturators into the patient's body to allow insertion of endoscopes and endoscopic accessories.

## Trocars

The trocars are used together with the trocar sleeves for puncture of the patient's body. The trocar is then removed to allow insertion of endoscopes and endoscopic accessories.

# Safety -Trocars

The Pajunk Safety-Trocar features a sharp, beveled tip that is covered by an inside, spring loaded, blunt obturator (Veress-concept). During puncture, the obturator is pushed into the shaft against the force of the spring and thus releases the cutting tip which can now penetrate the skin layers. As soon as the abdominal space is reached, the blunt obturator is moved forward by the spring, covers the cutting tip and prevents injuries in the abdomen.

# Distension System

The PAJUNK distension balloon system is indicated for laparoscopic procedures specially in extraperitoneal surgery, where dissection of tissue in the extraperitoneal space under direct view is essential. Contraindications: The distention balloon system should not be used in cases where previous operations in the extraperitoneal space caused adhesions in the area of the extraperitoneal organs.

# Structural Balloon System

The PAJUNK structural balloon system is indicated for laparoscopic procedures specially in extraperitoneal surgery, where a solid anchored and gas-tight entry port with an instrument

{1}------------------------------------------------

K012771
2 of 2

Image /page/1/Picture/1 description: The image shows the word "PAJUNK" in white letters against a black background. The letters are bold and sans-serif. The word is centered in the image and takes up most of the space. The image is simple and high contrast.

MEDIZINTECHNOLOGIE

channel together with a mechanical support of the extraperitoneal work space is needed. The inflation of the balloon causes a continuing dissection and anchoring in the extraperitoneal work space. Contraindications: The structural balloon system should not be used in cases where previous operations in the extraperitoneal space caused adhesions in the area of the extraperitoneal organs.

# Ring-Anchor Balloon System

The PAJUNK Ring-Anchor balloon system is indicated for laparoscopic procedures specially in extraperitoneal surgery, where a solid anchored and gas-tight entry port with an instrument channel is needed. Contraindications: The Ring-Anchor balloon system should not be used in cases where previous operations in the extraperitoneal space caused adhesions in the area of the extraperitoneal organs.

#### Accessories:

#### Disposable Valve Top

The disposable valve top is used with the Pajunk trocar sleeves and provides the same intended use. The valve is used instead of the trap door valve. The top is made out of polymer instead out of chromated brass. The metal tube is reusable and identical to the ones of the Paiunk trocar sleeves. The advantage is the easier cleaning, sterilization and assembling procedure.

#### Fixable Slide Cones

The Pajunk fixable slide cones with or without a fixation thread and with suture holders on both sides have been designed for first puncture technique according to Hasson in laparoscopic applications.

#### Reducer Sleeve

The reducer sleeve is inserted into the actually used trocar sleeve to reduce the diameter for improved quide and gas-tight insertion of endoscopic surgical instruments with outer diameters smaller than the nominal inner diameter of the trocar sleeve.

#### Dilation Set

If the actually applied trocar sleeve does not provide sufficient space for needed endoscopes and endoscopic accessories or endoscopic surgical instrument it can be replaced with a larger trocar sleeve.

#### Trocar Stop

The trocar stop with fixation thread is used to fixate the insertion of the trocar sleeve. The trocar sleeve is fixed by hand with a screw.

#### Substantial Equivalence:

The Pajunk Trocar and Trocar Sleeves, Distension System, the Anchoring System and accessories are substantially equivalent to the predicate devices since the basic features and intended uses are similar. The minor differences between the Paiunk Trocar and Trocar Sleeves, Distension System, the Anchoring System and accessories and the predicate devices raise no new issues of safety and effectiveness, as these differences have no effect on the performance, function or intended use of these devices.

K. Haug

Signed:

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Pajunk GmbH c/o Mr. Mark Job 510(k) Program Manager TÜV Product Service, Inc. 1775 Old Highway 8 New Brighton, Minnesota 55112

NOV 1 4 2001

Re: K012771

Trade/Device Name: Trocars, Trocar Sleeves, Model 1287; Distension System, Ring-Anchor Balloon System

Regulation Number: 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: GCJ Dated: October 1, 2001 Received: October 3, 2001

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Mr. Mark Job

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely vours.

Sincerely yours,

for Mark McMullerson

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K01277/

# NOV 1 4 2001

Page ! of

510(k) Number (if known): KO12771

Trocar Sleeves Device Name:

Indications For Use:

Manually operated surgical device intended for making incisions into the patient's body to allow insertion of endoscopes and endoscopic accessories during general endoscopic and laparoscopic procedures

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Optional Format 3-10,98)

for Mark n Milkersen
(Division Sign-Off)
Division of General, Restorative
and Neurological Devices
510(k) Number K012771
Gil

{5}------------------------------------------------

Page 2 of 4

510(k) Number (if known): KO1277/

Device Name: Trocars

Indications For Use:

Manually operated surgical device intended for making incisions into the watient's body to allow insertion of endoscopes and endoscopic accessories during general endoscopic and laparoscopic procedures

(Please do not WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Optional Format 3-10-98)

Mark-N-Milburn

(Division Sign-Off) Division of General, Restorative and Neurological Devices

510(k) Number.

{6}------------------------------------------------

Page 3 of 4

510(k) Number (if known): KO12771

Device Name: Distension System

Indications For Use:

Manually operated surgical device intended for creating a surgical space by dissecting layers of connective tissue along natural planes to allow insertion of endoscopes and endoscopic accessories during general endoscopic and laparoscopic procedures

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Optional Format 3-10-98)
for Mark N Mikesen

(Division Sign-Off) Division of General, Restorative and Neurological Devices

**510(k) Number** K012

{7}------------------------------------------------

Page 4 of 4

# 510(k) Number (if known): KO12,771

Device Name: Ring-Anchor Balloon System

Indications For Use:

Manually operated surgical device intended for providing a well anchored Mandally operatod Calgannel to allow insertion of endoscopes and and light instrument onames to anoval endoscopic and laparoscopic procedures

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Optional Format 3-10-98)

Mark N-Mullenn

(Division Sign-Off) Division of General, Restorative and Neurological Devices

510(k) Number _

K0(2771

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/GCJ/K012771](https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/GCJ/K012771)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
