DYONICS VISION 635 DIGITAL CAPTURE SYSTEM, MODEL 7209088

K011944 · Smith & Nephew, Inc. · GCJ · Sep 19, 2001 · Gastroenterology, Urology

Device Facts

Record IDK011944
Device NameDYONICS VISION 635 DIGITAL CAPTURE SYSTEM, MODEL 7209088
ApplicantSmith & Nephew, Inc.
Product CodeGCJ · Gastroenterology, Urology
Decision DateSep 19, 2001
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The Dyonics Vision 635 Digital Capture System is used in the operating room to digitally capture intra-operative still and motion images using the camera head, the foot switch, the front panel of the capture unit or the keyboard. Images are then stored in one of several standard image file formats on transportable media or to an Ethernet network for long term archival, retrieval or printing using third party image application software.

Device Story

Dyonics Vision 635 Digital Capture System (DV 635) captures intra-operative still images and motion video; connects to standard video output devices; accepts user inputs via keyboard, front panel switches, camera head buttons, or footswitch. System stores data on internal hard disk; transfers to removable media (CD-R, ZIP disks) or network drives via Ethernet; supports direct printing via print server. Used in operating rooms by surgeons/staff to archive, retrieve, and present surgical imagery. Output formats include JPEG, bitmap, DICOM, postscript, MPEG1, and MPEG2. Device facilitates documentation and presentation of surgical procedures.

Clinical Evidence

No clinical data; bench testing only.

Technological Characteristics

System utilizes standard video connections for input/throughput. Storage via internal hard disk, removable media (CD-R, ZIP), and Ethernet network connectivity. Image processing uses standard JPEG compression algorithms from the independent JPEG Group's software library. Supports multiple file formats including JPEG, bitmap, DICOM, postscript, MPEG1, and MPEG2.

Indications for Use

Indicated for use in the operating room to digitally capture intra-operative still and motion images during surgical procedures.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## SECTION B: ADMINISTRATIVE INFORMATION # l. 510(k) Summary Dyonics Vision 635 Digital Capture System Date Prepared: K011944 This 510(k) summary is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR \$807.92. #### Submitter A. Smith & Nephew, Inc. Endoscopy Division 160 Dascomb Road Andover, MA 01810 #### Company Contact B. Steven Jackson Manager, Regulatory Affairs Smith & Nephew, Inc. Endoscopy Division 3600 NW 138th St. Oklahoma City, OK 73134 Phone (405) 936-3085 Fax (405) 936-3059 #### C. Device Name Dyonics Vision 635 Digital Capture System Trade Name: Common Name: Picture Archiving and Communications System (PACS) Classification Name: Image Processing System Class II, per 21 CFR 892.2050 Classification: Product Code: LLZ #### D. Predicate Device Fotomaster™ Hamilton Thorn Research 100 Cummings Center Suite 102-C Beverly, MA 01915 Substantial Equivalence - K000380, May 3, 2000 #### E. Description of Device The Dyonics Vision 635 Digital Capture System (DV 635) is designed to provide surgeons the ability to capture still images and motion video during surgical procedures in various file formats for archival and presentation purposes. The DV 635 connects to any device with standard video outputs via standard video connections, and provides video throughput to video monitors, and other video peripheral devices. The DV 635 utilizes an external keyboard for input of basic patient and case > Dyonics Vision 635 Digital Capture System Page 11 of 227 {1}------------------------------------------------ information, and to set the system configuration. To capture images, the DV 635 utilizes user inputs from front panel switches, camera head buttons, the keyboard or optional footswitch. Images and motion video clips are stored on an internal hard disk drive, and later transferred to a removable storage media, including CD-R and ZIP™ disks, or to an network drive or printer via a Ethernet connection. The DV 635 may also print images directly to a postscript printer via a print server. #### F. Intended Use The Dyonics Vision 635 Digital Capture System is used in the operating room to digitally capture intra-operative still and motion images using the camera head, the foot switch, the front panel of the capture unit or the keyboard. Images are then stored in one of several standard image file formats on transportable media or to an Ethernet network for long term archival, retrieval or printing using third party image application software. #### G. Comparison of Technological Characteristics The Dyonics Vision 635 Digital Capture System (DV 635) is equivalent to the Hamilton Thorn Research Fotomaster™ in that both process, digitize, store, and save images obtained during surgical procedures using remote triggers and use standard publicly available JPEG image compression algorithms from the independent JPEG Group's software library. The DV 635 can also store generate images in bitmap, DICOM and postscript formats, and motion video clips MPEG1 and MPEG2 formats . The DV 635 can also connect directly to a network via an Ethernet connection and download stored images to a network storage device or networked printer or may be connected to a local printer via a local print server. These changes do not impact the safety or efficacy of the device. Steve Jack 6-20-01 Steven Jackson Manager, Regulatory Affairs {2}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its body and wings. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" surrounding it. ### Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 SEP 1 9 2001 Mr. Steven Jackson Manager, Regulatory Affairs Smith & Nephew. Inc. Endoscopy Division 3600 NW 138th Street Oklahoma City, Oklahoma 73134 Re: K011944 Trade/Device Name: Dyonics Vision 635 Digital Capture System Regulation Number: 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: GCJ Dated: June 20, 2001 Received: June 21, 2001 Dear Mr. Jackson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ # Page 2 - Mr. Steven Jackson This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally premated predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of additionaly 21 011 - 11 - 11 - 11 - 11 - 11 - 11 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Susan Walker, us Celia M. Witten, Ph.D., M.D. Dir Div Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ### KO11944 510(k) Number : Unknown Device Name : Dyonics Vision 635 Digital Capture System Indications for Use : The Dyonics Vision 635 Digital Capture System is indicated to be used in the operating the from to digitally I he Dyonce Vision 055 Digital Capture Bystein is marcase witch, the front switch, the front panel of the front panel of the fi capture unit or the keyboard. Images are then stored in one of several standard image file formats on capture unit of the Keyodate. Thinges are there to the receivel, retrieval or printing using third party image application software. # (PLEASE DO WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Zl Division Sign Off (Division Sign-Off) Division of General, Restorative and Neurological Devices 510(k) Number *K011944* (Optional Format 3-10-98) Dyonics Vision 635 Digital Capture System Page 13 of 227
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...