K161052 · Oxford Performance Materials, Inc. · KKY · Jul 20, 2016 · General, Plastic Surgery
Device Facts
Record ID
K161052
Device Name
OsteoFab Patient Specific Facial Device
Applicant
Oxford Performance Materials, Inc.
Product Code
KKY · General, Plastic Surgery
Decision Date
Jul 20, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.3500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The OsteoFab® Patient Specific Facial Device (OPSFD) is designed individually for each patient for enhancement, to correct trauma, and/or to correct defects in facial bone. The OPSFD is also designed individually for non-load bearing enhancement of mandibular bone.
Device Story
Patient-specific facial implant; manufactured via laser sintering using patient CT imaging data and CAD. Device serves as non-load bearing implant for facial bone defects, trauma, or enhancement. Surgeon-operated; attached to native bone using commercially available fixation systems. Shipped non-sterile. Modifications include updated thickness specifications and fixation guidelines (plating/lagging) to ensure structural integrity and safe surgical contouring. Benefits include patient-specific anatomical fit for reconstructive procedures.
Clinical Evidence
Bench testing only. No clinical data provided. Performance data included non-clinical verification and validation of implant thickness and fixation (plating/lagging) parameters based on risk analysis.
Technological Characteristics
Material: HTR-PEKK. Manufacturing: Laser sintering. Form factor: Patient-specific custom implant. Dimensions: Max diameter 20cm, thickness 1mm-20mm. Fixation: Commercially available systems. Connectivity: None. Sterilization: Non-sterile. Software: CAD-based design.
Indications for Use
Indicated for patients requiring facial bone enhancement, trauma correction, or defect repair, and non-load bearing mandibular bone enhancement.
Regulatory Classification
Identification
A polytetrafluoroethylene with carbon fibers composite implant material is a porous device material intended to be implanted during surgery of the chin, jaw, nose, or bones or tissue near the eye or ear. The device material serves as a space-occupying substance and is shaped and formed by the surgeon to conform to the patient's need.
Predicate Devices
OsteoFab® Patient Specific Facial Device (K133809)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around an emblem. The emblem consists of a stylized human figure with three faces, representing the department's focus on health and well-being. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 20, 2016
Oxford Performance Materials, Inc. Ms. Leigh Ayres SRA Director 30 S. Satellite Road South Windsor, CT 06074
Re: K161052
Trade/Device Name: OsteoFab Patient Specific Facial Device Regulation Number: 21 CFR 878.3500 Regulation Name: Polytetrafluoroethylene with carbon fibers composite implant material Regulatory Class: Class II Product Code: KKY Dated: June 22, 2016 Received: June 23, 2016
Dear Ms. Ayres:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
- Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K161052
Device Name
OsteoFab® Patient Specific Facial Device
Indications for Use (Describe)
The OsteoFab® Patient Specific Facial Device (OPSFD) is designed individually for each patient for enhancement, to correct trauma, and/or to correct defects in facial bone. The OPSFD is also designed individually for non-load bearing enhancement of mandibular bone.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------|---------------------------------------------------------------------|
| <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CER 801 Subpart D) | <span style="">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows a logo design with the letters O, P, and M. Each letter is constructed within a geometric framework of squares, circles, and lines. The letter O is orange and is inside a diamond shape, the letter P is black and is inside a square, and the letter M is orange and is inside a square.
## OPSFD 510(k) Summary as Required by Title 21 CFR 807.92
510(k) Submitter: Oxford Performance Materials, Inc. P.O. Box 585 30 South Satellite Road South Windsor, CT 06074 1-860-698-9300
Contact Person: Leigh Ayres, Director of Scientific and Regulatory Affairs Date of 510(k) summary statement preparation: July 12, 2016 Proprietary name: OsteoFab® Patient Specific Facial Device Common or Usual Name: Polytetrafluoroethylene (PTFE) with Carbon Fibers Classification: 878.3500 General/Plastic Surgery Review Panel: General/Plastic Surgery Medical Device Classification: Class II Product Code: KKY Predicate Device: Oxford Performance Materials, Inc. OsteoFab® Patient Specific Facial Device (K133809)
## Description of the Device
HTR-PEKK is a custom implant and the shapes and sizes vary within the following specifications: (1) maximum diameter is 20cm (2) minimum thickness is 1mm (2mm in areas of fixation), (3) maximum thickness is 20mm and the maximum thickness for holes is 10mm, (4) maximum open density is 25%, (5) minimum as designed through-hole diameter is 3mm, (6) maximum as designed though-hole size must meet these specifications (7) minimum distance from the edge of an as designed through-hole (for a cluster of perfusion-holes) to the edge of a device is 15mm, (8) minimum distance from the center of an as designed dimple to the edge for plating is 2.5mm for a 1.5mm diameter screw, (9) minimum distance from the center of an as designed dimple to the edge for lagging is 2.5mm for a 1.5mm diameter screw, (10) minimum distance between two lag holes is 3.25mm for a 1.5mm diameter screw, (11) minimum distance from the center of an as designed dimple to the edge for lagging is 3.75mm for a 2mm diameter screw, and (12) minimum distance between two lag holes is 3.75mm for a 2mm diameter screw.
Note: number 7 regarding as designed holes was changed to add a qualifier based on the context of that configuration of through-holes.
{4}------------------------------------------------
The OPSFD is constructed with the use of the patient's CT imaging data and computer aided design to determine the dimensions of each implant. The OPSFD is built by a LASER sintering machine. The OPSFD is attached to native bone with commercially available fixation systems. The OPSFD is a non-load bearing single use device and it is shipped non-sterile.
## Intended Use Statement
The OsteoFab® Patient Specific Facial Device (OPSFD) is designed individually for each patient for enhancement, to correct trauma, and/or to correct defects in facial bone. The OPSFD is also designed individually for non-load bearing enhancement of mandibular bone.
## Description of the Modifications
The description in the labeling is the same as the description that was submitted in K133809 except the following changes were made:
- 1. Implant minimum thickness was amended (number 2).
- 2. Implant maximum thickness was increased (number 3).
- 3. The minimum through-hole diameter was defined in terms of an as designed instead of an as built specification (number 5).
- 4. The maximum through-hole diameter was defined in terms of an as designed instead of an as built specification that must meet all of the specifications in the · · device description section (number 6).
- ട. A qualification statement for the minimum edge distance regarding a cluster of perfusion-holes was added (number 7).
- 6. The minimum as designed edge distance for plating was added (number 8).
- 7. The minimum as designed edge and through-hole center to center distances for lagging were added (numbers 9, 10, 11, and 12).
The following additions were made to the Warnings section of the IFU for this Special 510(k):
- 8. Screw fixation for plating or lagging should be placed only in areas of an implant that has minimum thickness of a 2mm (number 1).
- 9. The Surgeon should take special care if contouring is needed in areas where there is going to be fixation regarding implant thickness and distance to the edge (number 5).
The warnings regarding fixation and contouring were derived from the results of the performance testing.
## Performance Data and Risk Analysis
New performance data was not required for changing the specifications from "as built" to "as designed" because the change was a correction. New performance data was not needed for
{5}------------------------------------------------
the change to the maximum though-hole specification because it was a correction - the 5mm test specification had been included in the K133809 evaluation as an aid to determine the minimum though-hole size.
New performance data was not required regarding the qualifier that was added for the 15mm edge specification. Performance data was not required because the qualifying statement regarding a cluster of through-holes provided clarity to the device description to insure safer or more effective use. The statement about the cluster of though-holes provided clarity because the intent of that specification is to provide a margin of safety for the location of though-holes when they could be near to the location of dimples that are utilized for fixation procedures.
The changes in the package insert for implant thickness and fixation (plating and lagging) required new non-clinical performance data based on the results obtained from risk analysis. Risks were calculated according to the severity, probability, and detectability of failure modes.
The new performance data was submitted in the Special 510(k) for implant thickness and fixation in order to verify and validate the changes to the package insert. The data obtained was proof of performance.
### Substantial Equivalence Conclusion
The modifications to the description and warnings of the labeling for the subject device are substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.