PERMARIDGE HYDROXYLAPATITE MATRIX

K960349 · Ceramed Corp. · FWP · Mar 13, 1996 · General, Plastic Surgery

Device Facts

Record IDK960349
Device NamePERMARIDGE HYDROXYLAPATITE MATRIX
ApplicantCeramed Corp.
Product CodeFWP · General, Plastic Surgery
Decision DateMar 13, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.3550
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Facial restoration and augmentation.

Device Story

Synthetic hydroxylapatite (HA) implant; woven sheet form. Input: high-purity, radiopaque, 1000-micron rounded HA particles. Transformation: particles organized into flexible, cloth-like matrix using absorbable suture. Output: implantable material for facial restoration/augmentation. Used by surgeons in clinical settings. Provides structural support for facial tissue; biocompatible mineral scaffold.

Clinical Evidence

Bench testing only. Material characterization via X-ray diffraction confirms 100% HA composition. Compliance with ASTM F1185 for trace elements and calcium-to-phosphorus ratio (1.69) verified.

Technological Characteristics

Material: 100% synthetic hydroxylapatite (HA) beads. Standards: ASTM F1185 (composition/trace elements). Geometry: 1000-micron diameter particles. Form: Woven sheet/cloth-like matrix via absorbable suture. Radiopaque. Non-resorbable ceramic component.

Indications for Use

Indicated for patients requiring facial restoration and augmentation.

Regulatory Classification

Identification

A chin prosthesis is a silicone rubber solid device intended to be implanted to augment or reconstruct the chin.

Predicate Devices

Submission Summary (Full Text)

{1} K960349 1 of 1 # 510(k) SUMMARY January 22, 1996 This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990. 1. **Submitter:** CeraMed Corporation 12860 West Cedar Drive Lakewood, CO 80228 (303) 985-0800 Contact Person: Barbara A. Watson 2. **Device Name:** PermaRidge Hydroxylapatite Matrix, 1000 microns Classification Name: Chin implant 3. **Predicate Device:** MEDPOR® Surgical Implant OsteoGraf/D-700 (Originally OsteoGraf/AR) 4. **Device Description:** PermaRidge is a synthetic form of hydroxylapatite, the major mineral component of tooth enamel and bone, produced in the form of a woven sheet. It is manufactured as high purity, radiopaque, rounded particles sized at 1000 microns diameter and organized into a flat, flexible clothlike form by means of absorbable suture. 5. **Intended Use:** Facial restoration and augmentation. 6. **Comparison of Product Characteristics:** PermaRidge consists of 100% synthetic hydroxylapatite beads strung on absorbable suture. X-ray diffraction shows PermaRidge beads to be 100% HA. The hydroxylapatite component of PermaRidge conforms to ASTM Standard # F1185, "Standard Specification for Composition of Ceramic Hydroxylapatite for Surgical Implants", for trace elements. Typical calcium to phosphorus mole ratio is 1.69.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...