Biodesign Sling, Biodesign Plastic Surgery Matrix, Biodesign Anal Fistula Plug
K161221 · Cook Biotech Incorporated · FTM · May 26, 2016 · General, Plastic Surgery
Device Facts
Record ID
K161221
Device Name
Biodesign Sling, Biodesign Plastic Surgery Matrix, Biodesign Anal Fistula Plug
Applicant
Cook Biotech Incorporated
Product Code
FTM · General, Plastic Surgery
Decision Date
May 26, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Biodesign Sling: For implantation to reinforce soft tissues where weakness exists in the urological, gynecological and gasteroenterological anatomy including but not limited to the following procedures: transvaginal repair of stress urinary incontinence, such as pubourethral support and bladder support, and transabdominal repair of apical vaginal prolapse, colon and rectal prolapse, and sacrocolposuspension. By providing pubourethral support, the sling may be used for the treatment of urinary incontinence resulting from urethral hypermobility or intrinsic sphincter deficiency. Biodesign Plastic Surgery Matrix: For implantation to reinforce soft tissue where weakness exists in patients requiring soft tissue repair or reinforcement in plastic or reconstructive surgery. Biodesign Anal Fistula Plug: For implantation to reinforce soft tissue where a rolled configuration is required, for repair of anal, rectal and enterocutaneous fistulas.
Device Story
Biodesign devices are porcine small intestinal submucosa (SIS) implants used to reinforce soft tissue. Devices are implanted by surgeons in clinical settings to treat conditions like stress urinary incontinence, prolapse, and fistulas. The SIS material acts as a scaffold for tissue repair. Modifications include additional wash steps in the depuration process to remove impurities, plus minor design changes (perforations in sling/matrix, drying method change for sling). Devices are supplied sterile in Tyvek/PE pouches. Clinical benefit is derived from the reinforcement of weakened anatomical structures, facilitating natural tissue repair.
Clinical Evidence
Bench testing only. Evidence includes biocompatibility (ISO 10993), viral inactivation (prEN 12442-3), mechanical testing (burst force and suture retention strength), chemical residuals testing, and endotoxin level measurements. Design controls and risk analysis (ISO 14971) were performed.
Technological Characteristics
Material: Porcine small intestinal submucosa (SIS). Sterilization: Ethylene oxide. Packaging: Tyvek/PE pouch. Shelf-life: 18 months. Form factors: Rectangular sheets (Sling/Matrix) or cone (Plug). Modifications: Additional depuration wash steps; perforations added to Sling and Matrix; modified drying method for Sling.
Indications for Use
Indicated for patients requiring soft tissue reinforcement or repair in urological, gynecological, gastroenterological, plastic, or reconstructive procedures, including stress urinary incontinence, prolapse, and anal/rectal/enterocutaneous fistulas.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, connected by a flowing line that resembles a ribbon or banner. The profiles are arranged in a way that suggests unity and collaboration. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 26, 2016
Cook Biotech, Inc. Mr. Perry W. Guinn VP, Regulatory Affairs and Quality Assurance 1425 Innovation Place West Lafayette, IN 47906
Re: K161221
Trade/Device Name: Biodesign Sling, Biodesign Plastic Surgery Matrix,
Biodesign Anal Fistula Plug Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: Class II
Product Code: FTM Dated: April 27, 2016
Received: April 29, 2016
Dear Mr. Guinn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Jennifer R. Stevenson -A
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161221, Page 1 of 3
Device Name
Biodesign Sling
Indications for Use (Describe)
For implantation to reinforce soft tissues where weakness exists in the urological and gasteroenterological anatomy including but not limited to the following procedures: transvaginal repair of stress urinary incontinence, such as pubourethral support and bladder support, and transabdominal repair of apical prolapse, colon and rectal prolapse, and sacrocolposuspension. By providing pubourethral support, the sling may be used for the treatment of urinary incontinence resulting from urethral hypermobility or intrinsic sphincter deficiency.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------|-------------------------------------------------------------------------|
| <div> <span>☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K161221, Page 2 of 3
Device Name
Biodesign Plastic Surgery Matrix
Indications for Use (Describe)
For implantation to reinforce soft tissue where weakness exists in patients requiring soft tissue repair or reinforcement in plastic or reconstructive surgery.
Type of Use (Select one or both, as applicable)
| <span style="font-size:100%;">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-------------------------------------------------------------------------------------|
| <span style="font-size:100%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K161221, Page 3 of 3
Device Name
Biodesign Anal Fistula Plug
Indications for Use (Describe)
For implantation to reinforce soft tissue where a rolled configuration is required, for repair of anal, rectal and enterocutaneous fistulas.
Type of Use (Select one or both, as applicable)
| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 510(k) Summary
Submitted by: Perry Guinn Vice President, Quality Assurance and Regulatory Affairs Cook Biotech Incorporated 1425 Innovation Place West Lafayette, IN 47906 (765) 497-3355 27 April 2016
This 510(k) is a bundled submission pertaining to three devices manufactured by Cook Biotech Incorporated (CBI):
- 1. Biodesign Sling
- 2. Biodesign Plastic Surgery Matrix
- 3. Biodesign Anal Fistula Plug
This submission is presented as a bundle because the main technological modification under consideration is the same for all three devices. Table 5-1 below compiles the key regulatory elements for each subject device, which remain unchanged from their respective predicates.
| Proprietary Name | Biodesign Sling | Biodesign Plastic<br>Surgery Matrix | Biodesign Anal<br>Fistula Plug |
|------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Predicate Device | Surgisis Sling<br>(K992159) | SIS Plastic Surgery<br>Matrix (K034039) | SIS Fistula Plug<br>(K050337) |
| Intended Use | Intended to be used for<br>implantation to reinforce soft<br>tissues where weakness exists in<br>the urological, gynecological and<br>gasteroenterological anatomy<br>including but not limited to the<br>following procedures: transvaginal<br>repair of stress urinary<br>incontinence, such as pubourethral<br>support and bladder support, and<br>transabdominal repair of apical<br>vaginal prolapse, colon and rectal<br>prolapse, and<br>sacrocolposuspension. By<br>providing pubourethral support,<br>the sling may be used for the<br>treatment of urinary incontinence<br>resulting from urethral<br>hypermobility or intrinsic<br>sphincter deficiency. | Intended for<br>implantation to<br>reinforce soft tissue<br>where weakness<br>exists in patients<br>requiring soft tissue<br>repair or<br>reinforcement in<br>plastic or<br>reconstructive<br>surgery. | Intended for<br>implantation to<br>reinforce soft tissue<br>where a rolled<br>configuration is<br>required, for repair of<br>anal, rectal and<br>enterocutaneous<br>fistulas. |
### Table 5-1. Device names, predicates, intended uses and classifications:
{6}------------------------------------------------
| Proprietary Name | Biodesign Sling | Biodesign Plastic<br>Surgery Matrix | Biodesign Anal<br>Fistula Plug |
|------------------------------------|-----------------|-------------------------------------|--------------------------------|
| Common/Usual<br>Name | Surgical Mesh | Surgical Mesh | Surgical Mesh |
| Proposed<br>Classification<br>Name | Mesh, Surgical | Mesh, Surgical | Mesh, Surgical |
| Product Code | FTM, PAJ, PAG | FTM | FTM |
| Device Class | II | II | II |
| Regulation<br>Number | 878.3300 | 878.3300 | 878.3300 |
## Device Descriptions:
The three subject devices of this bundled submission share many of the same technological characteristics:
- Composed of porcine small intestinal submucosa (SIS) ●
- Packaged in a Tyvek/PE pouch
- Labeled with a shelf-life of 18 months
- Sterilized using ethylene oxide
The only differences between the three devices are the indications (and associated labeling) and the dimensional specifications (specific to the indication and anatomic requirement for each device). Both the indications and the dimensional specifications of each subject device, however, are unchanged from the corresponding predicate device. The indications have been described in Table 5-1 and the dimensional specifications are presented in Table 5-2, below:
| | Biodesign Sling | Biodesign Plastic<br>Surgery Matrix | Biodesign Anal Fistula<br>Plug |
|-------------------------------------------------------------------|--------------------------------------------------------|--------------------------------------------------------------------------|-------------------------------------------------------------------|
| Product Dimensions<br>(range) – as cleared<br>in predicate 510(k) | $20 cm^2$ to $140 cm^2$<br>$70$ to $600 \u03bcm$ thick | 2 to 70 mm width<br>up to 200 mm length<br>$100$ to $1500 \u03bcm$ thick | 2 mm minor diameter to<br>7 mm major diameter by<br>100 mm length |
| Shape | Rectangular | Rectangular | Cone |
### Table 5-2. Dimensional Specifications
### Description of Technological Modifications:
A series of wash steps was added to the depuration process of the base material for all three subject devices. This process modification serves to further remove impurities from the base material.
{7}------------------------------------------------
In addition, two minor technological changes were made to two of the subject devices:
- . Addition of perforations to the Biodesign Sling and Biodesign Plastic Surgery Matrix
- . Change in drying method for the Biodesign Sling
The incorporation of the additional wash steps and the minor technological changes do not result in a change to the fundamental technology of the devices.
## Summary of Supporting Evidence for Substantial Equivalence:
The following items are provided to demonstrate substantial equivalence to the predicate devices:
- . Analysis of 510(k) Substantial Equivalence Decision-Making Process as outlined in FDA's Guidance Document, The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)]
- . Biocompatibility analysis performed in accordance with ISO 10993
- Viral Inactivation verification performed in accordance with prEN 12442-3: 2000: . Viral inactivation
- Full design control as required by 21CFR 820.30 including review of risk analysis . performed in compliance with ISO 14971. Risk analysis methods used include:
- Failure mode, effects and criticality analysis (FMECA) o
- Residual risk review o
- o Post production review
- Mechanical testing for burst force strength and suture retention strength ●
- Chemical residuals testing .
- Measurement of endotoxin levels ●
## Substantial Equivalence:
Table 5-3 below provides a comparison of the subject devices and their respective predicates.
## Conclusion:
CBI modified three of its devices by adding a series of wash steps to the depuration process of SIS. The additional wash steps remove impurities from the base material without affecting the fundamental technology of the devices. Any potential new risks associated with the changes to the predicate devices have been identified by appropriate risk analysis techniques. These potential new risks have been addressed with verification and validation activities that meet the pre-determined acceptance criteria to ensure that no change to device safety has occurred.
{8}------------------------------------------------
| | Biodesign Sling,<br>Biodesign Plastic Surgery Matrix,<br>Biodesign Anal Fistula Plug<br>(Subject Devices) | Surgisis Sling,<br>SIS Plastic Surgery Matrix,<br>SIS Fistula Plug<br>(Predicate Devices) |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| 510(k) | K161221 | K992159 (Surgisis Sling)<br>K034039 (SIS Plastic<br>Surgery Matrix)<br>K050337 (SIS Fistula Plug) |
| Indication for Use<br>(The indications for<br>use of each device<br>in the bundled<br>510(k) are<br>unchanged from<br>that of its respective<br>predicate device.) | <i>Biodesign Sling:</i> For implantation to reinforce soft<br>tissues where weakness exists in the urological,<br>gynecological and gasteroenterological anatomy<br>including but not limited to the following<br>procedures: transvaginal repair of stress urinary<br>incontinence, such as pubourethral support and<br>bladder support, and transabdominal repair of<br>apical vaginal prolapse, colon and rectal prolapse,<br>and sacrocolposuspension. By providing<br>pubourethral support, the sling may be used for the<br>treatment of urinary incontinence resulting from<br>urethral hypermobility or intrinsic sphincter<br>deficiency.<br><br><i>Biodesign Plastic Surgery Matrix:</i> For implantation | <i>Surgisis Sling:</i><br>Same |
| | to reinforce soft tissue where weakness exists in<br>patients requiring soft tissue repair or<br>reinforcement in plastic or reconstructive surgery. | <i>SIS Plastic Surgery Matrix:</i><br>Same |
| | <i>Biodesign Anal Fistula Plug:</i> For implantation to<br>reinforce soft tissue where a rolled configuration is<br>required, for repair of anal, rectal and<br>enterocutaneous fistulas. | <i>SIS Fistula Plug:</i><br>Same |
| Material | Porcine small intestinal submucosa; (constituents<br>of the extracellular matrix) | Same |
| Technological<br>Characteristics | Additional wash steps included in base material<br>processing | Original base material<br>processing |
| Supplied sterile? | Yes | Same |
| Sterilization<br>method | Ethylene Oxide | Same |
| Packaging | Tyvek/PE Pouch | Same |
| Shelf-Life | 18 months | Same |
| Intended for single<br>use? | Yes | Same |
# Table 5-3. Substantial Equivalence Information
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.