Medeor™ Matrix is indicated for use in general surgical procedures for the reinforcement and repair of soft tissue where weakness exists including, but not limited to; defects of the thoracic wall, suture line reinforcement, and muscle flap reinforcement; urogynecological reinforcement (excluding transvaginal repair of pelvic organ prolapse) including but not limited to, rectal prolapse (excluding rectocele) using an abdominal approach, vaginal prolapse (excluding transvaginal repair of pelvic organ prolapse), reconstruction of the pelvic floor using an abdominal approach (excluding transvaginal repair of pelvic organ prolapse), hernia repair; soft tissue reconstructive procedures including plastic and reconstructive surgical applications, and for reinforcement of the soft tissues, which are repaired by suture or suture anchors, including but not limited to, rotator cuff, patellar, Achilles, biceps, quadriceps and other tendons. Medeor Matrix is not intended to replace normal body structure or provide the full mechanical strength to support tendon repair of the rotator cuff, patellar, Achilles, biceps, quadriceps, or other tendons. Sutures, used to repair the tear, and sutures or bone anchors used to attach the tissue to the bone, provide biomechanical strength for the tendon repair. Medeor Matrix is intended for one time use.
Device Story
Medeor Matrix is a resorbable, porcine-derived collagen surgical mesh; used for soft tissue reinforcement and repair. Supplied sterile in dry (lyophilized) or prehydrated (saline) forms. Applied by surgeons during general, urogynecological, or reconstructive procedures. Provides structural support to weakened tissues; used in conjunction with sutures or bone anchors for tendon repairs. Does not provide primary biomechanical strength for tendon repairs; serves as reinforcement. Intended for single-use only.
Clinical Evidence
Bench testing included tensile and suture retention testing. Biocompatibility testing performed per ISO 10993-1:2003 (cytotoxicity, sensitization, intracutaneous reactivity, systemic toxicity, sub-chronic toxicity, genotoxicity, implantation, pyrogenicity, hemolysis). In vivo studies conducted in sheep (defect repair) and rabbits (tissue reaction). All tests passed.
Technological Characteristics
Resorbable porcine-derived collagen surgical mesh. Supplied sterile in double-layer peel-open packaging. Available as lyophilized (dry) or prehydrated in saline. Biocompatibility evaluated per ISO 10993-1:2003.
Indications for Use
Indicated for patients requiring soft tissue reinforcement and repair in general surgery, including thoracic wall defects, muscle flaps, urogynecological procedures (abdominal approach only), hernia repair, and tendon reinforcement (rotator cuff, patellar, Achilles, biceps, quadriceps). Not for transvaginal pelvic organ prolapse repair.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for DSM, a company focused on health, nutrition, and materials. The logo features a stylized, circular graphic to the left of the company name, "DSM," which is written in a bold, sans-serif font. Below the company name is the tagline, "BRIGHT SCIENCE. BRIGHTER LIVING.", which is written in a smaller, sans-serif font.
## 510(k) Summary
| Submitted by: | Kensey Nash Corporation dba DSM Biomedical<br>735 Pennsylvania Drive<br>Exton, PA 19341 |
|-----------------------------|------------------------------------------------------------------------------------------------|
| Contact Person: | Lori Burns, MS, RAC<br>Manager Regulatory Affairs<br>Ph: (484) 713-2100<br>Fax: (484) 713-2903 |
| Date Prepared:<br>510(K) #: | June 13, 2014<br>K091499 |
| Device: | |
| Trade Name: | Medeor Matrix |
| Common/Usual Name: | Surgical Mesh |
| Proposed Classification: | 21 CFR 878.3300<br>FTM, OXH, OXE, OXB, OWY, PAJ |
Class II
## Device Description:
Medeor™ Matrix is a resorbable porcine-derived collagen surgical mesh intended for reinforcement of soft tissues. The device is supplied sterile in double-layer peel-open packages. The product is either packaged dry (lyophilized) to be hydrated prior to use, or can be supplied prehydrated, packaged and sterilized in saline.
## Predicate Devices:
| Manufacturer | Device | 510(k)# |
|-------------------------|-------------|---------|
| Kensey Nash Corporation | BioBlanket™ | K061030 |
| Lifecell | Strattice™ | K080353 |
#### Intended Use:
Medeor™ Matrix is indicated for use in general surgical procedures for the reinforcement and repair of soft tissue where weakness exists including, but not limited to; defects of the thoracic wall, suture line reinforcement, and muscle flap reinforcement; urogynecological reinforcement (excluding transvaginal repair of pelvic organ prolapse) including but not limited to, rectal prolapse (excluding rectocele) using an abdominal approach, vaginal prolapse (excluding transvaginal repair of pelvic organ prolapse), reconstruction of the pelvic floor using an abdominal approach (excluding transvaginal repair of pelvic organ prolapse), hernia repair; soft tissue reconstructive procedures including plastic and reconstructive surgical applications, and for reinforcement of the soft tissues, which are repaired by suture or suture anchors, including but not limited to, rotator cuff, patellar, Achilles, biceps, quadriceps and other tendons.
Medeor Matrix is not intended to replace normal body structure or provide the full mechanical strength to support tendon repair of the rotator cuff, patellar, Achilles, biceps, quadriceps, or other tendons. Sutures, used to repair the tear, and sutures or bone anchors used to attach the tissue to the bone, provide biomechanical strength for the tendon repair.
Medeor Matrix is intended for one time use.
{1}------------------------------------------------
## Performance Data:
Mechanical testing performed with Medeor Matrix demonstrated equivalence of the device to legally cleared predicate devices. Mechanical test reports were completed for the following:
- . Tensile Testing
- Suture Retention .
The following biocompatibility tests were conducted on the finished device according to the requirements of ISO 10993-1:2003, Biological evaluation of medical devices - Part 1:Evaluation and testing: Cytotoxicity, Sensitization, Intracutaneous Reactivity, Systemic Toxicity, Sub-Chronic Toxicity, Genotoxicity, Implantation, Pyrogenicity and Hemolysis.
The device was evaluated in two in vivo studies; defects were repaired and evaluated in a sheep model and a rabbit study was performed to evaluate tissue reactions.
Medeor Matrix passed the requirements of all tests.
## Substantial Equivalence:
Performance Testing has confirmed that Medeor Matrix is substantially equivalent to the predicate devices with regard to materials, intended use and technological characteristics, pursuant to section 510(k).
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized representation of an eagle or bird-like figure with three curved lines forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the bird-like figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 16, 2014
Kensey Nash Corporation % Ms. Alyssa J. Schwarts, MS, RAC Regulatory Affairs Affairs 735 Pennsylvania Drive Exton, Pennsylvania 19341
Re: K091499
Trade/Device Name: Medeor™ Matrix Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: FTM, OXB, OXE, OXH, OWY, PAJ Dated: October 12, 2009 Received: October 13, 2009
Dear Ms. Schwarts:
This letter corrects our substantially equivalent letter of October 22, 2009.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you: however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{3}------------------------------------------------
Page 2 - Ms. Alyssa J. Schwarts, MS. RAC
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## David Krause -S
- for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known) K091499
Device Name Medeor Matrix
## Indications for Use (Describe)
Medeor™ Matrix is indicated for use in general surgical procedures for the reinforcement and repair of soft tissue where weakness exists including, but not limited to: defects of the thoracic wall, suture line reinforcement, and muscle flap reinforcement; urogynecological reinforcement (excluding transvaginal repair of pelvic organ prolaps) including but not limited to, rectal prolapse (excluding rectocele) using an abdominal approach, vaginal prolapse (excluding transvaginal repair of pelvic organ prolapse), reconstruction of the pelvic floor using an abdominal approach (excluding transvaginal repair of pelvic organ prolapse), hemia repair; soft tissue reconstructive procedures including plastic and reconstructive surgical applications, and for reinforcement of the soft tissues, which are repaired by suture or suture anchors, including but not limited to, rotator cuff, patellar, Achilles, biceps, quadriceps and other tendons.
Medeor Matrix is not intended to replace normal body structure or provide the full mechanical strength to support tendon repair of the rotator cuff, patellar, Achilles, biceps, quadriceps, or other tendons. Sutures, used to repair the tear, and sutures or bone anchors used to attach the bone, provide biomechanical strength for the tendon repair.
Medeor Matrix is intended for one time use.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
## FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# I Krause
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.