K182087 · Feg Textiltechnik Forschungs- Und Entwicklungsgesellschaft · FTL · Oct 31, 2018 · General, Plastic Surgery
Device Facts
Record ID
K182087
Device Name
DynaMesh-POSTERIOR
Applicant
Feg Textiltechnik Forschungs- Und Entwicklungsgesellschaft
Product Code
FTL · General, Plastic Surgery
Decision Date
Oct 31, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
DynaMesh®- POSTERIOR serves to support the tissue and stabilize the fascial structures of the abdominal wall. DynaMesh®- POSTERIOR is indicated for repairing hernias in all current extraperitoneal surgical techniques.
Device Story
DynaMesh-POSTERIOR is a sterile, non-absorbable, bio-stable PVDF monofilament knitted mesh; used for hernia repair in extraperitoneal surgical techniques. Device provides tissue support and fascial structure stabilization. Features green-colored PVDF monofilament zigzag pattern as visual aid for clinicians to ensure tension-free positioning. Implanted by surgeons during hernia repair procedures. Benefits include mechanical reinforcement of abdominal wall structures.
Clinical Evidence
Bench testing only. Tests included burst strength, tensile strength, maximum elongation, suture pullout strength, tear resistance, porosity, and mesh stiffness to demonstrate performance and substantial equivalence.
Technological Characteristics
Knitted mesh composed of 100% non-absorbable, bio-stable Polyvinylidene fluoride (PVDF) monofilament (SOLEF®1008). Includes Phtalocyanine green (CAS 1328-53-6) dye for visual orientation. Supplied as a sterile, flat sheet. Sterilized via Ethylene Oxide.
Indications for Use
Indicated for hernia repair in all current extraperitoneal surgical techniques for patients requiring tissue support and fascial structure stabilization of the abdominal wall.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services (HHS) and the Food and Drug Administration (FDA). The HHS logo is on the left and consists of a stylized caduceus symbol. The FDA logo is on the right and features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font size below.
October 31, 2018
FEG Textiltechnik Forschungs- und Entwicklungsgesellschaft % Neil Armstrong RA Advisor to FEG Textiltechnik MeddiOuest Regulatory Affairs Ltd Unit 73B Hebron Industrial Estate Kilkenny, R95 CRK2 Ireland
Re: K182087
Trade/Device Name: DynaMesh-POSTERIOR Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: FTL Dated: July 11, 2018 Received: August 2, 2018
Dear Mr. Armstrong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/5 description: The image shows the name "Joseph Nielsen" in a large, bold font. To the right of the name, there is some smaller text that reads "Digitally s DN: c=US ou=FDA, S, 0.9.2342.1 Date: 201". The text appears to be part of a digital signature or certificate.
Digitally signed by Joseph Nielsen -S
DN: c=US, o=U.S. Government, ou=HHS,
ou=FDA, ou=People, cn=Joseph Nielsen -
S,
0.9.2342.19200300.100.1.1=2000367505
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
See PRA Statement below.
07/11/2018
510(k) Number (if known) K182087
Device Name DynaMesh®- POSTERIOR
Indications for Use (Describe)
DynaMesh®- POSTERIOR serves to support the tissue and stabilize the fascial structures of the abdominal wall. DynaMesh®- POSTERIOR is indicated for repairing hernias in all current extraperitoneal surgical techniques.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <span style="color: black;"> <span style="font-weight: normal;"> <span style="font-style: normal;"> <span style="text-decoration: none;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> </span> </span> </span> </span> </span> | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K182087 Page 1 of 3
FEG Textiltechnik Forschungs- und Entwicklungsgesellschaft mbH Prager Ring 70 D-52070 Aachen Germany Phone: 011 49 241 1892374-0 011 49 241 1892374-59 Fax:
# 510(k) Summary
(as required by 21 CFR 807.92(c))
#### Owner's Name:
FEG Textiltechnik Forschungs- und Entwicklungsgesellschaft mbH Prager Ring 70 D - 52070 Aachen Germany Tel: +49 241 189 23 74 - 0 +49 241 189 23 74 - 59 Fax: Stefan Schneemelcher Contact Person: e-mail: schneeme@feg-textiltechnik.de
Date this Summary was Prepared: July 11, 2018
#### Classification:
Name: Regulation: Product Code: Regulatory Class: Mesh, Surgical, Polymeric 21 CFR 878.3300 79 FTL II
#### Common/Usual Name:
Mesh Implant used in Hernia Surgery
#### Proprietary Name:
DynaMesh®-POSTERIOR
{4}------------------------------------------------
## Predicate Devices used to Demonstrate Substantial Equivalence:
The FEG Textiltechnik DynaMesh®-POSTERIOR is substantially equivalent to:
| Product Name: | DynaMesh®-VENTRAL, K161550 |
|---------------|------------------------------------|
| Manufacturer: | FEG Textiltechnik, Aachen, Germany |
### Description, including Intended Use
DynaMesh®- POSTERIOR serves to support the tissue and stabilize the fascial structures of the abdominal wall.
DynaMesh®- POSTERIOR is indicated for repairing hernias in all current extraperitoneal surgical techniques.
The FEG Textiltechnik Forschungs- und Entwicklungsgesellschaft mbH DynaMesh®-POSTERIOR is a textile structure knitted from non-absorbable, bio-stable Polyvinylidene fluoride (PVDF) monofilament fiber.
The grade of Polyvinylidene Fluoride (PVDF) used is SOLEF®1008 from Solvay Solexis.
The knitted mesh structure consists of uncoloured PVDF Monofilaments with a ziqzag pattern of green-coloured PVDF Monofilaments as visual aid to the clinician that facilitates a rapid orientation and visual monitoring of whether the mesh is positioned tension-free.
The used green dye is Phtalocyanine green, CAS1328-53-6, PG7 from BASF.
DynaMesh®-POSTERIOR is supplied as a sterile, flat sheet of material without any coatings. The FEG Textiltechnik's DynaMesh®-POSTERIOR will be packed in a cardboard box. Within the cardboard box the device will be doublepacked in two paper-poly pouches suitable for Ethylene Oxide sterilization; the labeled outer pouch contains the inner pouch and four adhesive labels for use on patient records.
#### Technological Characteristics Compared to Predicate Devices
The FEG Textiltechnik DynaMesh®-POSTERIOR is:
- for the same intended use, ●
- . knitted with the same equipment by the same personnel in the same facilities,
- packaged in same materials by same processes in the same facilities
- . labeled with the same label format by the same processes in the same facilities
- . contract sterilized in the same chamber with the same cycle at the same subcontract facility
as the FEG Textiltechnik DynaMesh®-VENTRAL K161550.
The FEG Textiltechnik DynaMesh®-POSTERIOR has a slightly different structure to give more uniform properties in both horizontal and vertical directions, as a clinician
{5}------------------------------------------------
| Feature | FEG Textiltechnik<br>DynaMesh®-POSTERIOR | FEG Textiltechnik<br>DynaMesh®-VENTRAL |
|-----------------------|------------------------------------------|----------------------------------------|
| Clearance/Approval | Subject Device | K161550 |
| Structure | knitted mesh | same |
| Material | 100% PVDF Monofilament | same |
| Manufacturing Process | Knitted by FEG | same |
| Packaging Materials | poly, paper, cardboard | same |
| Packaging Process | Packed by FEG | same |
| Sterilization Process | Sterilized by Rose GmbH | same |
#### preference option for situations where there is no clear preferred orientation.
## Summary of Non-Clinical and Clinical Data
Various Laboratory Bench Tests have been conducted on DynaMesh®-POSTERIOR to demonstrate performance to an accepted norm and/or substantial equivalence to the predicate products:
- . Burst Strength Test
- Tensile Strength ●
- Maximum elongation ●
- Suture Pullout Strength Test ●
- . Tear Resistance Testing
- Porosity Test ●
- . Mesh stiffness
#### Conclusion
From review of the non-clinical and clinical data, FEG Textiltechnik conclude that the DynaMesh®-POSTERIOR is substantially equivalent in terms of safety, effectiveness and performance to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.