K182008 · C.R. Bard, Inc. · FTL · Dec 14, 2018 · General, Plastic Surgery
Device Facts
Record ID
K182008
Device Name
Phasix ST Mesh with Echo 2 Positioning System
Applicant
C.R. Bard, Inc.
Product Code
FTL · General, Plastic Surgery
Decision Date
Dec 14, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Phasix™ ST Mesh is indicated for use in the reinforcement of soft tissue, where weakness involving soft tissue repair, such as for the repair of hernias. The Echo 2™ Positioning System is intended to facilitate the delivery and positioning of the mest during laparoscopic hernia repair.
Device Story
Phasix ST Mesh with Echo 2 Positioning System is a sterile, single-use surgical mesh pre-attached to a positioning system; used in laparoscopic hernia repair. Mesh consists of co-knitted Poly-4-hydroxybutyrate (P4HB) and Polyglycolic Acid (PGA) fibers with a bioresorbable hydrogel coating (sodium hyaluronate, carboxymethylcellulose, polyethylene glycol). Positioning system comprises nitinol wire, nylon frame, and PET hoisting suture. Device is inserted via trocar using an introducer tool; positioning system facilitates deployment and placement. Once mesh is fixed, positioning system is manually detached and removed. Hydrogel coating minimizes visceral tissue attachment; P4HB fibers provide long-term reinforcement. Benefits include facilitated laparoscopic delivery and reduced tissue adhesion.
Clinical Evidence
No clinical data. Evidence includes bench testing (mesh integrity, connector strength, ball burst), simulated use evaluations (delivery, support, fixation, removal), and porcine model performance testing. Biocompatibility testing (cytotoxicity, pyrogenicity, chemical characterization) performed.
Indicated for reinforcement of soft tissue where weakness exists, including hernia repair. Used in laparoscopic hernia repair procedures.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
Predicate Devices
Ventralight TM ST Mesh with Echo 2TM Positioning System (K170294)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services-USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in three lines, with "FDA" being the largest and boldest, followed by "U.S. FOOD & DRUG" and then "ADMINISTRATION".
December 14, 2018
CR Bard Inc. Shirin Mate Regulatory Affairs Specialist 100 Crossing Boulevard Warwick, Rhode Island 02886
Re: K182008
Trade/Device Name: Phasix ST Mesh with Echo 2 Positioning System Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: FTL, OWT, OOD, ORQ, GCJ Dated: November 13, 2018 Received: November 14, 2018
Dear Ms. Mate:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Cynthia Chang -S
for
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K182008
Device Name
Phasix™ ST Mesh with Echo 2TM Positioning System
Indications for Use (Describe)
Indications for Use (Describe)
Phasix™ ST Mesh is indicated for use in the reinforcement of soft tissue, where weakness involving soft tissue repair, such as for the repair of hernias.
The Echo 2™ Positioning System is intended to facilitate the delivery and positioning of the mest during laparoscopic hernia repair.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the The burden time for this collection of internation to and maintain the deter to the most time to review the collections, search existing data sources, gamer and his burden estimate or any other aspect and review the collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Phasix™ ST Mesh with Echo 2TM Positioning System
#### 510(k) Summary
#### 21 CFR 807.92
As required by the Safe Medical Devices Act of 1990, coded under Section 513, Part (I)(3)(A) of the Food, Drug and Cosmetic Act, a 510(k) summary upon which substantial equivalence determination is based is as follows:
#### l. Submitter Information:
Davol Inc., C. R. Bard, Inc. 100 Crossings Boulevard Warwick, RI 02886 Phone: (401) 825-8569 Fax: (401) 825-8765
Contact Person: Shirin Mate Title: Regulatory Affairs Specialist Email: shirin.s.mate@crbard.com
Date of submission: July 26th, 2018
#### II. Subject Device
| 510(k) Type: | Traditional 510(k) |
|--------------------------------------------|-------------------------------------------------|
| Trade Name: | Phasix™ ST Mesh with Echo 2™ Positioning System |
| Common or Usual Name: | Surgical Mesh |
| Classification Name: | Mesh, Surgical, Absorbable, Abdominal Hernia |
| Classification Panel:<br>Regulatory Class: | General and Plastic Surgery<br>Class II |
| Regulation Number: | 21 CFR § 878.3300 |
| Product Code: | OWT, OOD, FTL, ORQ, GCJ |
{4}------------------------------------------------
# III, Primary Predicate Device:
| Name of Device: | Ventralight TM ST Mesh with Echo 2TM Positioning System (K170294)<br>cleared on May 1st 2017 |
|-----------------------|----------------------------------------------------------------------------------------------|
| Common or Usual Name: | Surgical Mesh |
| Classification Name: | Mesh, Surgical, Polymeric |
| Classification Panel: | General & Plastic Surgery |
| Regulatory Class: | Class II |
| Regulation Number: | 21 CFR § 878.3300 |
| Product Code: | FTL, ORQ, GCJ |
### IV. Secondary Predicate Device:
| Name of Device: | Phasix™ ST Mesh (K173143) cleared on April 25th, 2018 |
|-----------------------|-------------------------------------------------------|
| Common or Usual Name: | Surgical Mesh |
| Classification Name: | Mesh, Surgical, Absorbable, Abdominal Hernia |
| Classification Panel: | General and Plastic Surgery |
| Regulatory Class: | Class II |
| Regulation Number: | 21 CFR § 878.3300 |
| Product Code: | OWT, OOD, FTL |
# V. Device Description:
The Phasix™ ST Mesh with Echo 2™ Positioning System is a sterile single use mesh preattached to the Echo 2™ Positioning System. The subject device is a combination of the FDA cleared Phasix™ ST Mesh (K173143) and the Echo 2™ Positioning System that is cleared by the FDA in the Ventralight™ ST Mesh with Echo 2™ Positioning System (K170294). The Echo 2™ Positioning System is removed following the mesh placement and fixation. As with the Ventralight™ ST Mesh with Echo 2™ Positioning System (K170294), many sizes of the subject device include an introducer tool accessory.
The subject device is intended for use in the reinforcement of soft tissue, where weakness exists, in procedures involving soft tissue repair, such as for the repair of hernias and will utilize a positioning system to facilitate the delivery and positioning of the mesh during laparoscopic hernia repair.
{5}------------------------------------------------
Procedures involving "laparoscopic hernia repair" also pertain to those which are conducted under laparoscopic visualization through a trocar, such as robotic assisted laparoscopic ventral hernia repair (LVHR).
The Phasix™ ST Mesh in the subject device is a fully resorbable mesh with a resorbable hydrogel coating and is identical to the secondary predicate device, Phasix™ ST Mesh (K173143). It is co-knitted using Poly-4-hydroxybutyrate (P4HB) and Polyglycolic Acid (PGA) fibers to form two sided mesh. The PGA side of the mesh is coated with a bioresorbable, chemically modified sodium hyaluronate (HA), Carboxymethylcellulose (CMC) and Polyethylene Glycol (PEG) based hydrogel. The fascial P4HB side of the mesh allows for a prompt fibroblastic response through its interstices, allowing for complete tissue ingrowth, similar to P4HB mesh alone. The visceral PGA side of the mesh coated with a bioresorbable hydrogel separates the mesh from underlying tissues and organ surfaces to help minimize tissue attachment to the mesh. Shortly after hydration, the biopolymer coating becomes a hydrated gel that is resorbed from the site in less than 30 days.
Preclinical implantation studies indicate that resorption of the P4HB fibers is minimal throughout the 12 week expected healing period and up to 26 weeks post implantation. Significant degradation of the mesh fibers is observed in preclinical studies within 12 to 18 months, indicating loss in mechanical integrity and strength of the mesh. While fiber segments were observed at 18 months, they continued to degrade.
This Phasix™ ST Mesh (K173143) is preattached to the Echo 2™ positioning system on the hydrogel side of the mesh using nylon connectors. The Echo 2™ Positioning System is composed of nitinol wire, encased in a sealed nylon frame. A Polyethylene Terephthalate (PET) center hoisting suture is attached to the nylon frame. The Echo 2™ positioning system utilized in the subject device is identical in materials and construction to the Ventralight™ ST Mesh with Echo 2™ Positioning System (K170294).Once inserted into the abdomen, the positioning system facilitates both deployment and positioning of the mesh. After initial fixation is complete, the Echo 2™ Positioning System is manually detached from the mesh and then removed from the body through a trocar or skin incision.
#### Device Accessories
The sterile stainless steel introducer tool (subject device accessory) is identical to the introducer tool utilized in Ventralight™ ST Mesh with Echo 2™ Positioning System (K170294)
{6}------------------------------------------------
and will be included with the several sizes of the subject device. The introducer tool assists in rolling and introducing the subject device via the trocar into abdominal cavity.
#### VI.Indications for Use of Device:
The Phasix™ ST Mesh with Echo 2TM Positioning System is indicated as below:
Phasix™ ST Mesh is indicated for use in the reinforcement of soft tissue, where weakness exists, in procedures involving soft tissue repair, such as for the repair of hernias.
The Echo 2™ Positioning System is intended to facilitate the delivery and positioning of the mesh during laparoscopic hernia repair.
# VII. Technological Comparison to Predicate Devices:
# Comparison of the technological characteristics with the predicate devices:
- Indications for use: The indication for use of Phasix™ ST Mesh in subject . device is identical to the indications for use of Phasix™ ST Mesh (K173143 with the exception of specific indication for hiatal hernia) and identical to the Ventralight™ ST Mesh with Echo 2™ Positioning System (K170294 with exception of term "soft tissue prosthesis"). The indications for use of Echo 2™ Positioning System of subject device is similar to the Ventralight™ ST Mesh with Echo 2™ Positioning System (K170294). The subject and predicate devices all have similar intended use i.e. reinforcement/ reconstruction of soft tissue deficiencies, such as repair of hernias.
- Mesh component: The Phasix™ ST Mesh utilized in subject device is identical to the . PhasixTM ST Mesh (K173143). The subject and both predicate devices consists of identical PGA fibers in their mesh construct and an identical HA/CMC PEG based hydrogel layer adhered on the PGA side of the mesh. The placement of Echo 2™ positioning system for both the subject and primary predicate devices is on the hydrogel side of mesh.
- Mesh Size: The mesh sizes of the subject device, Phasix™ ST Mesh with Echo . 2™ Positioning System are identical to the marketed mesh sizes of the Ventralight™ ST Mesh with Echo 2TM Positioning System. Similar mesh sizes are also currently marketed for the Phasix™ ST Mesh (K173143).
{7}------------------------------------------------
- Positioning system: The materials and construction of the Echo 2™ Positioning . System utilized in the Phasix™ ST Mesh with Echo 2™ Positioning System are identical to the Ventralight™ ST Mesh with Echo 2™ Positioning System (K170294). The Echo 2™ Positioning System sizes are similar between the subject and predicate device, as the subject device Echo 2™ Positioning System has been optimized to fit various sizes of Phasix™ ST Mesh.
- Introducer tool: Introducer tool utilized in Phasix™ ST Mesh with Echo 2™ . Positioning System is identical to the introducer tool utilized in Ventralight™ ST Mesh with Echo 2 TM Positioning System (K170294).
- Principle of Operation: The principle of operation of Phasix™ ST Mesh with . Echo 2™ Positioning System with respect to delivery, support, access/ fixation and removal is identical to the Ventralight TM ST Mesh with Echo 2™ Positioning System (K170294).
The subject device has similar intended use, indications for use, principle of operation and similar technological characteristics with regards to design, materials, manufacturing processes, packaging and sterilization (EtO) as the primary and secondary predicate devices. The subject and the predicate devices all are manufactured at the same manufacturing facility.
#### Performance Data: VIII.
The following performance data is provided in support of substantial equivalence determination.
#### Performance Standards
No performance standards have been established for this device under Section 514 of the Federal Food, Drug, and Cosmetic Act.
### Biocompatibility Testing
The biocompatibility evaluation for the Phasix™ ST Mesh with Echo 2™ Positioning System was conducted in accordance with the Guidance for Industry and Food and Drug Administration Staff Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" June 16, 2016, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process," as recognized by the FDA.
{8}------------------------------------------------
The materials used in construction of the Phasix™ ST Mesh, Echo 2™ Positioning System and introducer tool are known to be biocompatible as demonstrated by their safe clinical use and successful clearance by the FDA under K173143 and K170294 respectively. Therefore, these tests were not repeated. However, in order to address any risks associated with the interactions between the Phasix™ ST Mesh and the Echo 2™ Positioning System, the subject device was evaluated using the following tests.
Following tests support the biocompatibility of the subject device:
- 1. Cytotoxicity
- 2. Pyrogenicity
- 3. Chemical Characterization.
The biocompatibility test results demonstrate that the subject device is biocompatible and there are no interactions between the Phasix™ ST Mesh and the Echo 2™ Positioning System to affect its safety and effectiveness.
# Electrical safety and electromagnetic compatibility (EMC)
The Phasix™ ST Mesh with Echo 2™ Positioning System is not an electro-mechanical medical device nor is it medical system; therefore this section does not apply.
#### Software Verification and Validation Testing
The Phasix™ ST Mesh with Echo 2™ Positioning System does not contain software therefore this section does not apply.
#### Product Testing
Phasix™ ST Mesh with Echo 2™ Positioning System passed all the test requirements and its test results demonstrated the substantial equivalence to the test results of the predicate devices. Performance and mechanical testing on the subject device included:
- 1. Mechanical testing of subject device:
- Mesh Integrity (Gel Disruption) .
- Connector to mesh attachment strength .
- Ball Burst .
- 2. Simulated use evaluations using bench top simulator:
- . Delivery
- Support .
- Access/ Fixation .
{9}------------------------------------------------
- Removal .
- 3. Performance testing in porcine model
- Delivery .
- Support .
- Access/ Fixation .
- Removal .
All the samples of the subject device successfully met the established acceptance criteria and performed as intended and similar to the predicate devices.
# Animal studies:
Since the Phasix™ ST Mesh in subject device is identical to the secondary predicate, Phasix™ ST Mesh (K173143), animal studies were not required to support the subject device.
# Clinical Study:
Clinical study was not required in support of the subject device, Phasix™ ST Mesh with Echo 2™ Positioning System.
#### Conclusions:
The subject and predicate devices have similar intended use, indications for use, principle of operation and technological characteristics. All test results provided in support of the subject device demonstrate safety, effectiveness and performance similar to the predicate devices. Therefore, the subject device, Phasix™ ST Mesh with Echo 2™ Positioning System, is substantially equivalent to the primary predicate device, Ventralight™ ST Mesh with Echo 2™ Positioning System (K170294), and secondary predicate device, Phasix™ ST Mesh (K173143).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.