The Bard® 3DMAX™ Light Mesh is a sterile, single use device indicated for use in the reinforcement of soft tissue where weakness exists, such as the repair of hernia defects.
Device Story
Bard® 3DMAX™ Light Mesh is a sterile, single-use, anatomically shaped surgical mesh designed for laparoscopic inguinal hernia repair. Constructed from knitted polypropylene monofilaments (0.0048-inch diameter), the device features a preformed shape with sealed edges to facilitate easier positioning compared to flat mesh sheets. It includes an 'M'-shaped orientation marker and an arrow, both composed of polypropylene monofilament dyed with Phthalocyaninato(2-) copper, to assist surgeons in identifying the medial edge during placement. The mesh is designed to be stretched in two directions to accommodate and reinforce tissue defects. Used by surgeons in an operating room setting, the device provides a structural scaffold for tissue reinforcement. The anatomical design and orientation markers aim to improve surgical efficiency and accuracy during hernia repair procedures, potentially reducing operative time and improving patient outcomes.
Clinical Evidence
Bench testing only. No clinical data provided. Laboratory testing compared the subject device to predicates (Soft Mesh, 3DMAX™ Mesh, Mersilene Mesh) evaluating thickness, pore size, density, stiffness, tensile strength, percent elongation, suture pullout strength, burst strength, and tear resistance per FDA guidance. Biocompatibility testing for polypropylene and the copper-dyed orientation marker yielded acceptable results.
Indicated for reinforcement of soft tissue where weakness exists, such as hernia defect repair in patients requiring surgical intervention.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
Predicate Devices
Bard® Soft Mesh (Davol Inc.)
Bard® 3DMAX™ Mesh (Davol Inc.)
Mersilene Mesh (Ethicon, Inc.)
Submission Summary (Full Text)
{0}------------------------------------------------
K091659 page 43
## SECTION 7.0 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
#### A. Submitter Information
AUG 0 8 0009
Submitter's Name:
Address:
Davol Inc.
Subsidiary of C.R. Bard, Inc 100 Crossings Boulevard Warwick, R1 02886
- Telephone: Fax: Contact Person: Date of Preparation:
(401) 825-8582 (401) 825-8765 Gail Dow June 3, 2009
#### B. Device Name
Trade Name: Common/Usual Name: Classification Name:
Bard® 3DMAX™ Light Mesh Surgical Mesh Surgical Mesh, Polymeric
#### C. Predicate Device Name
Bard® Soft Mesh (Davol Inc.) Trade name: Trade name: Bard® 3DMAX™ Mesh (Davol Inc.) Mersilene Mesh (Ethicon, Inc.) Trade name:
#### D. Device Description
The Proposed Bard® 3DMAX™ Light Mesh, is designed to fit the inguinal anatomy. The device is anatomically shaped and preformed with sealed edges that allow for easier positioning of the device than a traditional flat sheet of mesh in a laparoscopic inguinal hernia repair. The device is constructed of knitted polypropylene monofilaments 0.0048 inches in diameter. The knit construction allows the mesh to be stretched in both directions in order to accommodate and reinforce tissue defects. The device contains an orientation marker in an "M" shape to designate the medial edge of the mesh and an arrow pointing to the medial aspect. Both the arrow and "M" will aid the user during positioning and placement of the device. The orientation marker is composed of polypropylene monofilament dyed with Phthalocyaninato(2-) copper colorant.
# PREMARKET NOTIFICATION FOR THE BARD® 3DMAX™ Light Mesh
### -CONFIDENTIAL-
{1}------------------------------------------------
#### E. Intended Use
The Bard® 3DMAX™ Light Mesh is a sterile, single use device indicated for use in the reinforcement of soft tissue where weakness exists, such as the repair of hernia defects.
#### ட் Summary of Similarities and Differences in Technological Characteristics, Performance and Intended Use
Bard® 3DMAX™ Light Mesh, the currently marketed Bard® 3DMAX™ Mesh, Bard® Soft Mesh and Ethicon Mersilene Mesh devices are all indicated for repair of hernia defects. The intended use for the 3DMAX™ Light Mesh is the same as the Predicate devices, Soft Mesh and 3DMAX™ Mesh, with the exception of the indication for repair of chest wall defects.
The Predicate Bard® Soft Mesh is a major component of the Proposed 3DMAX™ Light Mesh and both are made from the same polypropylene monofilaments with a diameter of approximately 0.0048 inches. The Proposed 3DMAX™ Light Mesh has similar physical attributes and performance characteristics as the Predicates 3DMAX™ Mesh, Soft Mesh, and Mersilene Mesh. Furthermore, it has similar materials and manufacturing methods as the Predicate 3DMAX™ Mesh.
The key differences in the Proposed 3DMAX™ Light Mesh compared to the currently marketed 3DMAX™ Mesh are the diameter of the polypropylene monofilament and the knit pattern of the mesh. In order to create a thin lightweight mesh and reduce the amount of material used in the mesh compared to traditional monofilament polypropylene meshes, the Proposed 3DMAX™ Light Mesh is constructed from monofilaments with a diameter of approximately 0.0048 inches. However, the weights (grams/square inch) of Mersilene Mesh (0.0274) and Soft Mesh (0.0282) are comparable to the 3DMAX™ Light Mesh (0.0272). In addition, the Proposed 3DMAX™ Light Mesh has larger pore knit construction than the Predicate Bard 3D Max™ Mesh.
The 3DMAX™ Light Mesh, and the Predicate 3DMAX™ Mesh are similar as both are designed to fit the inquinal anatomy. Unlike the Predicates. Mersilene Mesh and Soft Mesh, which are flat mesh sheets, the Proposed 3DMAX™ Light Mesh is anatomically shaped and preformed with sealed edges that allow for easier positioning in a laparoscopic inguinal hernia repair. The edge seal design will be the same as the currently marketed 3DMAX™ Mesh, however a layer of polypropylene mesh will be added during the edge seal lamination process to provide sufficient material to create a smooth, non-porous edge.
Unlike the Predicates, Mersilene Mesh and Soft Mesh, the Proposed 3DMAX™ Light Mesh and the currently marketed 3DMAX™ Mesh contain a medial marker composed of polypropylene monofilament and Phthalocyaninato(2-) copper colorant.
### PREMARKET NOTIFICATION FOR THE BARD® 3DMAX™ Light Mesh
#### -CONFIDENTIAL-
{2}------------------------------------------------
K091659 Page 3/3
#### G. Performance Data
Laboratory bench testing was performed to assess the effects of the new characteristics of the Proposed 3DMAX™ Light Mesh. These tests compared the Proposed Product against the Predicate Products Soft Mesh, 3DMAX™ Mesh and Mersilene Mesh. In accordance with FDA's "Guidance for the Preparation of a Premarket Notification for a Surqical Mesh" (March 2. 1999), tests included physical and performance evaluations of the products in terms of thickness, pore size, density, stiffness, tensile strength, percent elongation, suture pullout strength, burst strength and tear resistance. Laboratory testing discussed in this submission demonstrates that the material chosen and the design utilized in manufacturing 3DMAX™ Light Mesh is substantially equivalent to the referenced Predicate Products. The results of this testing are in Section 15.
Additionally, biocompatibility testing covering the polypropylene material used in the Proposed 3DMAX™ Light Mesh, as well as the orientation marker composed of polypropylene monofilament and Phthalocyaninato(2-) copper colorant have received acceptable results. Results are summarized in Section 14.0 and copies of the test reports are in Exhibit 4.
The results show that the 3DMAX™ Light Mesh is substantially equivalent to the currently marketed Predicate devices. Therefore, based on laboratory testing and biocompatibility data, the Proposed 3DMAX™ Light Mesh, is safe and effective for its intended use.
## PREMARKET NOTIFICATION FOR THE BARD® 3DMAX™ Light Mesh
### -CONFIDENTIAL-
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 0 3 2009
C.R. Bard, Inc. % Ms. Gail Dow Regulatory Affairs Associate 100 Crossings Boulevard Warwick, Rhode Island 02886
Re: K091659
Trade/Device Name: Bard® 3DMax"" Light Mesh Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: Class II Product Code: FTL Dated: June 3, 2009 Received: June 9, 2009
Dear Ms. Dow:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Ms. Gail Dow
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
sincerely yours,
Mark N. Melkerson
Director Division of Surgical. Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# INDICATION FOR USE STATEMENT
510(k) Number (if known):
----
Not Known
Bard® 3DMAx™ Light Mesh Device Name:
The Bard® 3DMAX™ Light Mesh will be indicated for use in Indications for Use: the reinforcement of soft tissue where weakness exists, such as the repair of hernia defects.
Over-The-Counter AND/OR Prescription Use _____________________________________________________________________________________________________________________________________________________________ X Use (21 CFR 801 (Part 21 CFR 801 Subpart D) Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel Keane for mxm
(Division Sign Off)
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K091659
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.