The Kiierr Family of Devices, the 272 and 148 Lasers are indicated to promote hair growth in females who have Ludwig-Savin Classifications of I – II males with androgenetic alopecia who have Norwood-Hamilton Classifications of IIa - V, and for both, Fitzpatrick Skin Phototypes I to IV.
Device Story
Kiierr Family of Devices (Models 272 and 148) are laser hair growth caps; utilize low-level laser therapy (LLLT) via 272 or 148 laser diodes (650+/-5 nm) to stimulate hair follicles. Device is a helmet-style cap with protective liner; powered by lithium-ion battery with embedded controller. Intended for OTC home use by patients; hands-free operation. Treatment protocol involves 30-minute sessions, every other day, for 16 weeks. Output is non-heating red light; intended to promote hair growth in androgenetic alopecia. Benefit is non-invasive hair growth stimulation.
Clinical Evidence
No clinical performance data produced. Substantial equivalence is based on the device being identical to the predicate (K173678). Bench testing included electrical safety and electromagnetic compatibility (IEC 60601-1-2) and laser safety (IEC 60825-1). OTC usability study conducted with 40 subjects demonstrated a 90% pass rate for self-selection, assembly, and comprehension of hazards/maintenance.
Technological Characteristics
Helmet-style cap containing 148 or 272 laser diodes; wavelength 650+/-5 nm; optical output <= 5mW per diode. Powered by lithium-ion battery with embedded controller chip. Class 3R laser system. Conforms to IEC 60825-1 (laser safety) and IEC 60601-1-2 (EMC).
Indications for Use
Indicated for females with androgenetic alopecia (Ludwig-Savin I-II) and males with androgenetic alopecia (Norwood-Hamilton IIa-V), with Fitzpatrick Skin Phototypes I-IV, to promote hair growth.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font size below.
September 5, 2018
Kiierr International, LLC % Raymond Blanche Consultant NST Consultants, Inc 641 Shunpike Road, Suite 311 Chatham, New Jersey 07928
Re: K181878
Trade/Device Name: Kiierr Family of Devices, Model 272 Premier & Model 148 Pro Laser Hair Growth Caps Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared Lamp Regulatory Class: Class II Product Code: OAP Dated: May 9, 2018 Received: July 13, 2018
Dear Raymond Blanche:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmr/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Jennifer Stevenso For Binita S. Ashar, M.D., M.B.A., F.A.C.S
Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181878
#### Device Name
Kiierr Family of Devices, 272 and 148 Laser
#### Indications for Use (Describe)
The Kiierr Family of Devices, the 272 and 148 Laser Devices are indicated to promote hair growth in females with androgenetic alopecia who have Ludwig-Savin Classifications I - II, in males with androgenetic alopecia who have Norwood-Hamilton Classifications of IIa - V and for both, Fitzpatrick Classifications of Skin Phototypes I - IV.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# August 15, 2018 Revised
# Kiierr International, LLC
## Submitter's Contact Information
| Name: | Raymond R. Blanche |
|------------|----------------------------------------------------------------------------|
| Address | NST Consultants, Inc.<br>641 Shunpike Road, Suite 311<br>Chatham, NJ 07928 |
| Telephone: | (973) 539-7444 |
| Facsimile: | (973) 539-7445 |
## Name of Device and Name/Address of Sponsor
| Trade Name: | Kiierr Family of Devices, the 272 and 148 Laser |
|---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor Contact<br>Information: | Michael Anderson<br>Kiierr International, LLC<br>863 East Downington Avenue<br>Salt Lake City, Utah 84105-3211<br>Telephone: 435-840-8199 |
| Common or Usual Name: | Lamp, non-heating, for promotion of hair growth |
| Classification Name: | Infrared lamp per 21 CFR 890.5500 |
| Classification Code: | OAP (Laser, comb, hair) |
Predicate Devices:
| Device Trade Name | Manufacturer |
|-------------------------|-----------------------------------|
| Diode Laser Cap K173678 | Cosmo Far East Technology Limited |
{4}------------------------------------------------
### Reference Devices:
None
Date Prepared: May 9, 2018
### Intended Use / Indications for Use
The Kiierr Family of Devices, the 272 and 148 Lasers are indicated to promote hair growth in females who have Ludwig-Savin Classifications of I – II males with androgenetic alopecia who have Norwood-Hamilton Classifications of IIa - V, and for both, Fitzpatrick Skin Phototypes I to IV.
#### Technological Characteristics
The Kiierr Family of Devices, the 272 and 148 Lasers each contain the listed number of diodes lasers configured within an outer cap helmet and protective inner liner. The use of diode lasers provides for a full coverage of the upper 1/3 of the head i.e., the area commonly covered with stylized hair. The Kiierr Family of Devices, the 272 and 148 Lasers are powered by a lithium-ion battery pack that contains an embedded controller chip. In all other area of design, manufacturing, and aesthetic appearance, the devices are identical.
### Performance Data:
No clinical performance data was produced for this submission because the Kiierr Family of Devices, the 272 and 148 Lasers are the same devices as the predicate, the Diode Laser, cleared under K173678. All proposed devices and predicate devices are IDENTICAL and the same devices offered for PRIVATE LABEL by the manufacturer. They are the same device in optical, electronic, mechanical function and aesthetic appearance, as well as the same recommended clinical treatment regime. It is essential to understand that this class of device is manufactured by many different manufacturers, all using the same parts supplied by the same OEM suppliers. Hence, there are virtually no differences between the devices.
#### Substantial Equivalence
Both the Kiierr Family of Devices, the 272 and 148 Lasers and the Diode laser 272 use red light diode lasers, classified as class IIIa/3R laser systems by the IEC standard for allowable emission levels, which is a recognized standard by the FDA as well, and the adverse event profile is the same. The sponsor believes that there is no difference in the physical appearance or in the method of delivering the radiant energy of the systems and therefore, there are no variations in the therapeutic value or safety profile.
{5}------------------------------------------------
For these reasons, the Kiierr Family of Devices, the 272 and 148 Lasers overwhelmingly satisfies the FDA's substantial equivalence with respect to intended use, technological and design characteristics.
The Kiierr Family of Devices, the 272 and 148 Lasers are prospective devices that will be manufactured by the ISO 13485 compliant contract manufacturer, on the same platform as the Diode Laser, predicate devices. There are no safety concerns raised by this process of manufacturing because the model 272 and 148 are the same devices the predicate manufacturer sells under its proprietary label and name.
## Treatment Protocol
The Kiierr Family of Devices, the 272 and 148 Lasers and the identified predicates, possess the same treatment regime of 30 minutes, every other day, on non-consecutive days, for the initial treatment regime of 16 weeks.
The following Comparison Chart in support of substantial equivalence is provided:
| The Kiierr Family of Devices, the 272 and 148 Lasers | | Diode Laser 272 | Diode Laser 148 |
|---------------------------------------------------------------------------------------------|---------------|-----------------|-----------------|
| LLLT Device Type | Laser diode | Laser Diode | Laser Diode |
| Use Application | OTC | OTC | OTC |
| Intended Use - Androgenetic Alopecia | Yes | Yes | Yes |
| Contain Laser Diodes-Class 3R, | Yes 272 & 148 | Yes 272 | Yes 148 |
| Optical Output – 5mW or less | Yes | Yes | Yes |
| Helmet Design | Yes | Yes | Yes |
| 650+/-5 nm. | Yes | Yes | Yes |
| Marketing Clearance -Females & Males, OTC | Yes | Yes | Yes |
| Passive Use-Hands Free | Yes | Yes | Yes |
| OAP Classification | Yes | Yes | Yes |
| Classification Name -Infrared Lamp | Yes | Yes | Yes |
| Common Usage Name - Lamp, Non-Heating | Yes | Yes | Yes |
| General & Plastic Surgery Committee | Yes | Yes | Yes |
| Skin Phototypes - I- IV | Yes | Yes | Yes |
| Hamilton-Norwood IIa-V Hair Loss Classification | Yes | Yes | Yes |
| Ludwig-Savin I – II Hair Loss Classification | Yes | Yes | Yes |
| Treatment- 16 weeks, for 30 minute treatment times three times a week<br>on alternate days. | Yes | Yes | Yes |
| Device Class II | Yes | Yes | Yes |
The data presented in the Comparison Chart, demonstrates that all Kiierr models 272 and 148 devices are identical and equivalent, PRIVATE LABEL devices from the same manufacturer. The Models 272 and 148 will be manufactured on the same platform by the same manufacturer as the 272 and 148 Diode Lasers, by Cosmo Far East Technology, Ltd.
{6}------------------------------------------------
Based on this comparison and determination, the sponsor requests the FDA to clear the device via the 510(k) notice.
# Over – The – Counter Testing Program
To test volunteer subjects for the suitability of the The Kierr Family of Devices, the 272 and 148 Lasers 40 subjects were asked twenty-six questions, after being provided a standard retail package and a full owner's manual. The test subjects were given as long as they required to read and understand the product packaging and manual. No assistance was provided to them and they were not permitted to ask any questions of the interviewer. The interviewer then conducted the interview and filled in the responses from the subjects. The subjects were required to answer all questions correctly to be counted as PASS for the correct Self Selection or, to have made the correct decision to purchase the product or not; to assemble and use the product correctly and comprehend the hazards and maintenance procedures for the device. These decisions would be based upon their understanding of the Intended Use of the product and the manual.
If the questions were answered correctly, they were given a P for PASS. If any questions were answered incorrectly, they were given an F for FAIL. The number of subjects required to answer all questions correctly is 32 out of 40, for an 80% success rate.
The results of the Over-the-Counter testing demonstrate that the Kiierr Family of Devices, the 272 and 148 Lasers comply with the requirements the FDA determined to be applicable. The test revealed an overall 90% pass rate for the subject group of 40 male and female participants. The testing further demonstrates that age, education, socioeconomic group, race or medical hair loss status are not variants that prevent proper self-selection, usability and comprehension of hazards and maintenance procedures for the average consumer to successfully navigate the purchasing and use process of the The Kijerr Family of Devices, the 272 and 148 Lasers.
Based on this data, the sponsor believes that the The Kiierr Family of Devices, the 272 and 148 Lasers, for female and male users have met the requirements for OTC sale.
# Electrical Safety and Electromagnetic Compatibility Testing Performance
The The Kiierr Family of Devices, the 272 and 148 Lasers were evaluated for conformance to recognized international standards. The following is a list of these evaluations and tests that were found to be in conformance:
- 1. IEC 60825-1 Edition 2.0 2007-03 Safety of Laser Products Part 1: Equipment Classification and Requirements.
- 2. IEC 60601-1-2 Edition 3.0 2007-03 Medical Electrical Equipment Part 1-2 General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility - Part 1: Requirements and Tests.
# Conclusion
{7}------------------------------------------------
Based on the technical comparisons between the The Kiierr Family of Devices, the 272 and 148 Lasers and the identified predicates it can be concluded that the Kiierr models are identical to the predicate devices in technical specifications. Since the Kiierr models are built using the identical platform with no difference between models, it can be concluded that all of the Kiierr models are equally safe. Therefore, it can be determined that the The Kiierr Family of Devices, the 272 and 148 Lasers are Substantially Equivalent to the identified predicates.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.