The OTIS 2.1 Optical Coherence Tomography System is indicated for use as an imaging tool in the evaluation of excised human tissue microstructure by providing two-dimensional, real-time depth visualization, with image review manipulation software for identifying and annotating regions of interest.
Device Story
OTIS 2.1 is an optical coherence tomography (OCT) imaging system for excised human tissue; uses non-ionizing, low-power optical radiation (super luminescent diode) to produce high-resolution, sub-surface images; utilizes an interferometer to compare reference light beam to backscattered light from tissue; features created by changes in optical properties (scattering, absorption, index of refraction). Device automates OCT scanning of specimen surface; provides two-dimensional, real-time depth visualization; includes software for image review, manipulation, and annotation of regions of interest. Operated by clinicians in a clinical setting; output displayed on touchscreen for review; assists in evaluating tissue microstructure. Benefits include standardized image collection and streamlined user experience for tissue assessment.
Clinical Evidence
Bench testing only. Verification and validation activities confirmed outputs met design inputs and fulfilled intended use. Testing included biocompatibility (ISO 10993-1, ISO 10993-5), basic safety, and electromagnetic compatibility (IEC 60601-1, IEC 60601-1-2, IEC 60825-1).
Technological Characteristics
OCT imaging system; super luminescent diode source; center wavelength 1325 ± 15 μm; lateral resolution 20 μm; axial resolution 10-15 μm; Class 1 laser. Designed per 21 CFR 820.30 and ISO 13485:2016. Standards: ANSI AAMI ISO 14971, AAMI ANSI ES 60601-1, IEC 60601-1-2, IEC 60825-1, ANSI AAMI ISO 10993-1/5, IEC 62304. Touchscreen interface; 108-132 V, 60 Hz power.
Indications for Use
Indicated for use as an imaging tool in the evaluation of excised human tissue microstructure by providing two-dimensional, cross-sectional, real-time depth visualization, with image review manipulation software for identifying and annotating regions of interest.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Perimeter Medical Imaging AI, Inc. % Cindy Domecus Regulatory Consultant Domecus Consulting Services, LLC 1171 Barroihet Avenue, Hillsborough, California 94010
February 25, 2021
Re: K203578
Trade/Device Name: OTIS 2.1 Optical Coherence Tomography System, THiA Optical Coherence Tomography System Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic Pulsed Echo Imaging System Regulatory Class: Class II Product Code: NQQ Dated: January 29, 2021 Received: February 1, 2021
Dear Cindy Domecus:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Neil R.P. Ogden Assistant Director, THT4A4 DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K203578
Device Name
OTIS 2.1 Optical Coherence Tomography Imaging System
Indications for Use (Describe)
The OTIS 2.1 Optical Coherence Tomography System is indicated for use as an imaging tool in the evaluation of excised human tissue microstructure by providing two-dimensional, real-time depth visualization, with image review manipulation software for identifying and annotating regions of interest.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K203578 510(k) Summary
### 1. Basic Information - 510(k) Owner
| 510(k) Owner: | Perimeter Medical Imaging AI, Inc. |
|------------------------|------------------------------------------------------------------------------------------------------------------|
| Address: | 359 Eastern Avenue, Suite 110<br>Toronto, Ontario, Canada<br>Phone: 647-360-0302 |
| Official Contact: | Cindy Domecus, R.A.C. (US & EU)<br>Principal<br>Domecus Consulting Services LLC<br>DomecusConsulting@comcast.net |
| Date Summary Prepared: | December 3, 2020 |
# 2. Device Name
Trade Name: Perimeter OTIS™ 2.1 and THiA™
Common Name: Optical Coherence Tomography System
Classification Name: Ultrasonic pulsed echo imaging system
Requlatory Classification: 21 CFR 892.1560
Product Code: NQQ
Classification: Class II
### 3. Legally Marketed Predicate Devices
Perimeter OTIS™ 2.0, K190404
### 4. Device Description
The Perimeter OTIS™ 2.1 is an imaging tool for use on excised human tissue. The Perimeter OTIS™ 2.1 is based on optical coherence tomography (OCT) imaging, which is similar to ultrasound, but uses non-ionizing, low-power optical radiation to produce high resolution, sub-surface images of a tissue sample. Due to the extremely high velocity of light, optical echoes (reflected and backscattered light
Perimeter OTIS™ 2.1 - Special 510(k) Submission
{4}------------------------------------------------
from the sample) cannot be measured directly using a photodetector. Instead, OCT devices use an interferometer to compare a reference beam of light to the backscattered light returning from the tissue sample. The features in an OCT image are created by changes in the optical properties (namely scattering, absorption, and index of refraction) of the sample.
The Perimeter OTIS™ 2.1 collects and displays OCT images of human tissue with comparable image quality to the predicate Perimeter OTIS 2.0 System (K190404). Like its predicate, the Perimeter OTIS™ 2.1 has automated the OCT scanning of the specimen surface, standardizing the image collection process and facilitating image manipulation for review and annotation. OTIS 2.1 allows for the concurrent image review and acquisition.
### 5. Indications for Use
The OTIS™ 2.1 is indicated for use as an imaging tool in the evaluation of excised human tissue microstructure by providing two-dimensional, cross-sectional, realtime depth visualization with image review manipulation software for identifying and annotating regions of interest.
| Characteristic | Perimeter OTIS™ 2.1<br>[Subject Device] | Perimeter OTIS™ 2.0<br>[Predicate Device] |
|--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------|
| Intended Use | Imaging tool in the evaluation of human<br>tissue microstructure by providing two-<br>dimensional, cross-sectional, real-time depth<br>visualization. | Same |
| Indications for Use | The OTIS 2.1 Optical Coherence Tomography<br>System is indicated for use as an imaging<br>tool in the evaluation of excised human tissue<br>microstructure, by providing two-<br>dimensional, cross-sectional, real-time depth<br>visualization, with image review manipulation<br>software for identifying and annotating<br>regions of interest. | Same |
| Measurement Technique | Optical Coherence Tomography | Same |
| Center Wavelength | $1325 \pm 15 \mu$ m | Same |
| Optical Source | Super Luminescent Diode | Same |
| Optical Radiation Safety | Safe for Indicated Use<br>Class 1 Laser | Same |
| Lateral Resolution | $20 \mu$ m | Same |
| Lateral Range | 870 mm | Same |
| Axial Resolution | 10 - 15 $\mu$ m in tissue | Same |
| Scan Acquisition Time | < 1 minute<br>[5 x 5 cm area] | Same |
| Input Devices | Touchscreen | Same |
## 6. Predicate Device Comparison
{5}------------------------------------------------
| Characteristic | Perimeter OTIS™ 2.1<br>[Subject Device] | Perimeter OTIS™ 2.0<br>[Predicate Device] |
|---------------------------------|--------------------------------------------|-------------------------------------------|
| Electrical Voltage<br>Frequency | 108 - 132 V, 60 Hz<br>[North American Use] | Same |
#### 7. Product and Quality Management Standards
The OTIS™ 2.1 was designed and developed under design controls per 21 CFR 820.30 and ISO 13485:2016. The following standards were followed:
- ANSI AAMI ISO 14971:2007/(R)2010 (Corrected 4 October 2007) Medical . Devices - Applications of Risk Management to Medical Devices
- AAMI ANSI ES 60601-1:2005/(R)2012, A1:2012, C1:2009/(R)2012 and . A2:2010/(R)2012 Medical Electrical Equipment – Part 1, Basic Safety and Essential Performance
- IEC 60601-1-2 Edition 4.0 2014-02 Medical Electrical Equipment Part 1-2: ● General Requirements for the Basic Safety and Essential Performance -Collateral Standard: Electromagnetic Disturbances - Requirements and Tests
- . IEC 60825-1:2007, Safety of Laser Products - Part 1 Equipment Classification and Requirements
- ANSI AAMI ISO 10993-1:2009/(R)2013 Biological Evaluation of Medical ● Devices – Part 1: Evaluation and Testing with a Risk Management Process
- ANSI AAMI ISO 10993-5:2009/(R)2014, Biological Evaluation of Medical ● Devices – Part 5: Tests for In Vitro Cytotoxicity
- IEC 62304:2006/A1:2015 Medical Device Software Software Life Cycle . Processes AAMI ANSI ES 60601-1:2005 - Basic Safety and Essential Performance
{6}------------------------------------------------
### 8. Technological Characteristics
The subject of this Special 510(k) submission is a device enhancement to provide an updated version of the predicate OTIS 2.0 device that has a smaller cart foot print, a faster PC, design improvements for manufacturability, component updates to address obsolescence and ensure longer useful life of the product and optimized image acquisition and processing, graphical user improvements for streamlined user experience with consistent screen displays.
These updates do not affect the intended use or alter the fundamental scientific technology of the predicate device.
## 9. Performance and Safety Testing
Perimeter completed verification and validation activities under Perimeter's Design Control procedure. This ensured that verification studies demonstrated that outputs met design input requirements, and that validation studies demonstrated that the Perimeter OTIS™ 2.1 fulfilled the intended use and met user needs.
Design verification and validation testing for OTIS 2.1 was carried out using the same test protocols/methodology as was used to test OTIS 2.0. Established testing aqainst the same ISO/IEC standards as applicable for OTIS 2.0, was conducted as follows:
- . Biocompatibility testing of the Consumable Set per ISO 10993-1:2009/(R)2013 and ISO 10993-5:2009/(R)2014.
- . External laser, basic safety and electromagnetic compatibility testing successfully demonstrated the safety of OTIS™ 2.1 in its intended environment.
#### 10. Conclusions
Perimeter's OTIS 2.1 has the same intended use, indications for use, principles of operation and same technological characteristics as compared to the predicate device. The minor differences implemented in the OTIS 2.0 platform to create OTIS 2.1 do not raise different questions of safety and effectiveness.
Design verification and validation testing demonstrates that OTIS 2.1 Optical Coherence Tomography Imaging System is as safe and effective as the predicate device and is substantially equivalent to OTIS 2.0.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.