K982821 · Multimedia Medical Systems, Inc. · MUJ · Nov 5, 1998 · Radiology
Device Facts
Record ID
K982821
Device Name
MMS THERPACPLUS B3DTUI MODEL NUMBER 6.7
Applicant
Multimedia Medical Systems, Inc.
Product Code
MUJ · Radiology
Decision Date
Nov 5, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The MMS·TherpacPLUS B3DTUI is a software application for pre-operative planning and post-operative evaluation of permanent implant brachytherapy procedures for the treatment of prostate cancer. It assists physicians and medical physicists in pre-planning the patient's implant procedure and evaluating the actual post-operative results of the procedure.
Device Story
MMS-TherpacPLUS B3DTUI is a software-based treatment planning system for prostate brachytherapy. It accepts inputs from clinical imaging modalities via DICOM, video capture, or data drives. The system supports structure contouring, template registration, and seed placement. It features automated routines for seed placement, dose optimization, and post-procedure seed identification. The device provides 2D and 3D visualization and generates reports for clinical review. Used in clinical settings by physicians and medical physicists, the system assists in pre-planning implant procedures and evaluating post-operative results, facilitating clinical decision-making regarding radiation dose delivery and treatment efficacy.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Software application running on standard Intel PCs with Microsoft Windows 32-bit OS. Supports OR-safe hardware platforms. Interfaces with electromagnetic digitizers, video capture cards, printers, film scanners, and DICOM-compliant imaging sources. Features include structure contouring, template registration, automated seed placement, dose optimization, and automatic seed identification for post-procedure evaluation. Provides 2D and 3D visualization.
Indications for Use
Indicated for pre-operative planning and post-operative evaluation of permanent implant brachytherapy procedures for patients diagnosed with prostate cancer. Intended for use by physicians and medical physicists.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
Predicate Devices
Multimedia Medical Systems “Therpac Radiotherapy Treatment Planning System” (K930506)
K150636 — VariSeed 9.0 · Varian Medical Systems, Inc. · May 8, 2015
K984241 — ROCS ULTRASEED · Radiation Oncology Computer Systems · May 12, 1999
Submission Summary (Full Text)
{0}------------------------------------------------
# Summary of 510(k) Safety and Effectiveness Data for MMS-TherpacPLUS B3DTUI
K982821
| Prepared: | August 6, 1998 |
|----------------------|-------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Multimedia Medical Systems, Inc. |
| Address: | 400 Ray C. Hunt Drive, Suite 380<br>Charlottesville, VA 22903 |
| Phone: | 804/977-8710 |
| Contact: | Alexander Crosby<br>Vice President of Quality |
| Common Name: | Transperineal ultrasound-guided implant prostate<br>brachytherapy planning system |
| Device Name: | MMS-TherpacPLUS B3DTUI, version 6.7 |
| Classification Name: | Treatment planning computer, CFR 892.5050 |
| Predicate Devices: | Multimedia Medical Systems “Therpac Radiotherapy<br>Treatment Planning System”, K930506<br>SSGL “Prowess 2000”. K924240 |
## Device Description:
Hardware Platform and Operating System: The application runs on standard Intel PCs under Microsoft Windows® 32-bit operating systems.
The application is available on an OR-safe hardware platform. This hardware platform has electrical safety, emission control and cleaning provisions.
## Peripherals and Accessories:
The application interfaces with an electromagnetic digitizer, video capture cards, printers, film scanners, as well as DICOM and other image importer and data drives and external sources.
## Software Features:
The application supports image capture and importation from clinical imaging modalities sources, structure contouring and template registration, seed placement and dosimetry display for procedure planning and post-plan evaluation. Planning functions are supported by auto seed placement and dose optimization routines. The application supports post-procedure evaluation
> Section 6 Page 1 of 2
{1}------------------------------------------------
through automatic seed identification. Both planning and evaluation functions allow plans to be printed on reports, support visualization in 2D and 3D views. The applications has two user support modules - seed source specification and database manager.
### Intended Use
The MMS·TherpacPLUS B3DTUI is a software application for pre-operative planning and post-operative evaluation of permanent implant brachytherapy procedures for the treatment of prostate cancer. It assists physicians and medical physicists in pre-planning the patient's implant procedure and evaluating the actual post-operative results of the procedure.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized design featuring three profiles that appear to be connected, possibly representing people or the concept of health and human services.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
5 1998 NOV
Alexandria Crosby Vice President of Quality Multimedia Medical Systems 400 Rav C. Hunt Drive, S380 Charlottesville, VA 22903
Re:
K982821
MMS-TherapacPLUS B3DTUI Dated: August 7, 1998 Received: August 11, 1998 Regulatory class: II 21 CFR 892,5730/Procode: 90 MUJ
Dear Mr. Crosby:
We have reviewed your Section 510%) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Ouality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in ritto diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdr/dsmarnain.html".
Sincerely yours,
Lillian Yin, Ph.D.
Lillian Yin, Ph.D. Director, Division of Reproductive Abdominal, Ear, Nose and Throa and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
| | - 3- 3- 2 - 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- 2- | |
|--|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| | | |
| | | |
Page of
| 5 0(k) Number (if known): | K982821 |
|---------------------------|--------------------------|
| Device Name: | MMS THERPAC PLUS B3D TUI |
| Indications For Use: | |
··
The MMS . TherpacPLUS B3DTUI is a software application for pre-operative planning and post-operative evaluation of permanent implant brachytherapy procedures for the treatment of prostate cancer. It assists physicians and medical physicists in pre-olanning the patient's implant procedure and evaluating the actual post-operative results of the procedure.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Segerson
---
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K982821
Prescription Use
(Per 21 CFR 801.109)
O
OR
Over-The-Counter Use
بخ
(Optional Format 1-2-96)
ﺎﺋﮯ
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.