RayStation 11B

K220141 · RaySearch Laboratories AB (PUBL) · MUJ · Apr 15, 2022 · Radiology

Device Facts

Record IDK220141
Device NameRayStation 11B
ApplicantRaySearch Laboratories AB (PUBL)
Product CodeMUJ · Radiology
Decision DateApr 15, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5050
Device ClassClass 2
AttributesSoftware as a Medical Device

Intended Use

RayStation is a software system for radiation therapy and medical oncology. Based on user input, RayStation proposes treatment plans. After a proposed treatment plan is reviewed and approved by authorized intended users, RayStation may also be used to administer treatments. The system functionality can be configured based on user needs.

Device Story

RayStation 11B is a radiation therapy treatment planning system (TPS) used by clinicians to plan, analyze, and administer radiation therapy and medical oncology treatments. Input data includes patient imaging (CT, CBCT) and treatment machine parameters. The system utilizes dose and optimization engines to propose treatment plans based on user input. New features include EQD2 dose computation using biological linear quadratic models; synthetic CT generation from CBCT by combining patient-specific CT data to improve photon dose calculation accuracy; and computation of dose-weighted Linear Energy Transfer (LET) for proton/light ion plans to assess radiobiological effects. The system comprises three applications: RayStation (planning), RayPhysics (commissioning), and RayTreat (plan delivery/record keeping). A simplified version, RayPlan 11B, offers a limited license set. Clinicians review and approve plans generated by the system before treatment administration. The device benefits patients by providing accurate, optimized radiation dose delivery and advanced radiobiological evaluation tools.

Clinical Evidence

Bench testing only. Verification and validation included system tests, risk analysis-based testing for use error mitigation, unit/subsystem testing, and dose engine validation. Clinical evaluation and user site validation were performed for new features (EQD2, synthetic CT, and LET/RBE components) to confirm adequacy for clinical use. No clinical trial data provided.

Technological Characteristics

Software-based radiation therapy treatment planning system. Operates on a shared software platform with the predicate. Features include dose optimization engines, EQD2 biological modeling, synthetic CT generation algorithms, and LET/RBE calculation modules. Software is classified as a device of major concern. Connectivity includes plan transfer to treatment delivery devices.

Indications for Use

Indicated for radiation therapy and medical oncology treatment planning, analysis, and administration. Intended for use by authorized clinicians to propose, review, and approve treatment plans for patients requiring radiation therapy.

Regulatory Classification

Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. RaySearch Laboratories AB (publ) % David Hedfors Quality and Regulatory Affairs Director Eugeniavägen 18 Stockholm. 113 68 Sweden Re: K220141 Trade/Device Name: RayStation 11B Regulation Number: 21 CFR 892.5050 Regulation Name: Medical Charged-Particle Radiation Therapy System Regulatory Class: Class II Product Code: MUJ Dated: January 10, 2022 Received: January 18, 2022 Dear David Hedfors: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Julie Sullivan, Ph.D. Assistant Director Nuclear Medicine and Radiation Therapy Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K220141 Device Name RayStation 11B Indications for Use (Describe) RayStation is a software system for radiation therapy and medical oncology. Based on user input, RayStation proposes treatment plans. After a proposed treatment plan is reviewed and approved by authorized intended users, RayStation may also be used to administer treatments. The system functionality can be configured based on user needs. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| | <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | |-------------------------------------------------------------------------------------------------| | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ K220141 Image /page/3/Picture/1 description: The image contains the logo for RaySearch Laboratories. The logo consists of the text "RaySearch" in blue, with the word "Laboratories" underneath in gray. To the right of the text is a graphic of several gray lines intersecting at a central blue point, resembling rays emanating from a source. # 510(k) Summary RayStation 11B {4}------------------------------------------------ | Document ID and Title | Version: | |-------------------------------|----------| | 510(k) Summary RayStation 11B | 1.0 | ## Table of Contents | 1. | 510(k) Summary | 3 | |----|--------------------------------------------------|---| | | 1.1 510(k) owner | 3 | | | 1.2 Contact person | 3 | | | 1.3 Preparation date | 3 | | | 1.4 Trade name | 3 | | | 1.5 Common name | 3 | | | 1.6 Classification name | 3 | | | 1.7 Predicate devices | 3 | | | 1.8 Device description | 3 | | | 1.9 Intended use | 4 | | | 1.10 Technological characteristics summary | 4 | | | 1.11 Assessment of non-clinical performance data | 4 | | | 1.12 Test conclusion | 4 | {5}------------------------------------------------ Document ID and Title 510(k) Summary RayStation 11B Version: 1.0 ## 510(k) Summary #### 1.1 510(k) owner 1. RaySearch Laboratories AB (publ) Eugeniavägen 18 113 68 Stockholm Sweden Tel: +46 8 510 530 00 #### 1.2 Contact person David Hedfors Quality and Regulatory Affairs Director RaySearch Laboratories AB (publ) quality@raysearchlabs.com Email: +46 722 366 110 Tel: #### Preparation date 1.3 January 10th, 2022 #### 1.4 Trade name The trade name is RayStation. The marketing name is RayStation 11B and RayPlan 11B. #### 1.5 Common name Radiation therapy treatment planning system #### 1.6 Classification name Medical charged-particle radiation therapy system (21 CFR 892.5050, Product Code MUJ) #### 1.7 Predicate devices K211867 RayStation 11.0 #### 1.8 Device description RayStation is a treatment planning system for planning, analysis and administration of radiation therapy and medical oncology treatment plans. It has a modern user interface and is equipped with fast and accurate dose and optimization engines. RayStation consists of multiple applications: - . The main RayStation application is used for treatment planning. - . The RayPhysics application is used for commissioning of treatment machines to make them available for treatment planning and used for commissioning of imaging systems. - The RayTreat application is used for sending plans to treatment delivery devices for treatment and receiving records of performed treatments. The device to be marketed, RayStation 11B, contains modified features compared to RayStation 11.0 as indicated below: A simplified license configuration of RayStation is marketed as RayPlan. RayPlan has a limited set of purchasable licenses and some modules will not be accessible. RayPlan is marketed as RayPlan 11B. EQD2 dose computation (new) - From photon and/or braction doses, it is possible to compute, deform and accumulate the two Gray equivalent (EQD2) dose. The computation of the EQD2 dose uses the biological linear quadratic model, which is also the basis for the already released biological optimization and evaluation functionality. {6}------------------------------------------------ | Document ID and Title | Version: | |-------------------------------|----------| | 510(k) Summary RayStation 11B | 1.0 | Generation of synthetic CT from CBCT (new) - Two new methods (algorithms) for synthetic CT generation will be included. The synthetic CT images are created by combining information in the CBCT image and a CT image for the specific patient to allow for dose computation using the HU values in the image, as for regular CT images. In RayStation 11.0 it is possible to compute dose on CBCT images for photons using bulk density assignments. The added functionality will improve the photon dose calculation accuracy on CBCT images. Handling of LET and other RBE components (new) - This functionality enables possibility to compute and evaluate the dose weighted LET (Linear Energy Transfer) for proton and light ion plans. LET is an additional dosimetric measure that can be used to assess the radiobiological effect of the proton and light ion radiation. Radiobiological equivalent (RBE) dose is a derived quantity with dependence on both the physical dose and the LET. In RayStation 11.0, it is possible to compute and evaluate RBE doses. #### 1.9 Intended use RayStation is a software system for radiation therapy and medical oncology. Based on user input, RayStation proposes treatment plans. After a proposed treatment plan is reviewed and approved by authorized intended users, RayStation may also be used to administer treatments. The system functionality can be configured based on user needs. #### 1.10 Technological characteristics summary The technological characteristics are the same for RayStation 11B as for the predicate device RayStation 11.0. Both versions are built on the same software platform and share design to a high degree. Both versions have been developed under the same quality system, by the same development teams, meeting the same requirements for safety and effectiveness. The device to be marketed, RayStation 11B contains new features as indicated below: - . EQD2 dose computation (new) - . Generation of synthetic CT from CBCT (new) - Handling of LET and other RBE components (new) Related to machine learning, there is no change compared to the predicate device. #### 1.11 Assessment of non-clinical performance data The test specification of RayStation 11B is a further developed version of the test specification of RayStation 11.0. This is supported by the requirements specification, for which the same is true. The successful verification and validation of RayStation 11B therefore support the substantial equivalence of the above RayStation versions. EOD2 dose computation (new) – Validation for photon and/or braction doses were validated as part of the Clinical Evaluation for Brachy and User Site Validation demonstrates that the dose computation adequate for clinical use. Generation of synthetic CT from CBCT (new) – Validation of improved photon dose calculation accuracy on CBCT images was performed in CBCT Conversion validation demonstrates that the dose computation adequate for clinical use. Handling of LET and other RBE components (new) - Validation of dose weighted LET (Linear Energy Transfer) for proton and light ion plans were performed as part of the Proton PBS Monte Carlo validation. The validation demonstrates that the dose computation adequate for clinical use. #### 1.12 Test conclusion The determination of substantial equivalence is not based on an assessment of non-clinical performance data. However, the entire system verification and validation specifications and reports are included in the submission as required by a software device of major concern. A number of different types of verification activities have been performed: - . System Tests of RayStation {7}------------------------------------------------ - Risk analysis-based tests for use error mitigation verification ● - . Unit and subsystem testing for low-level testing - . Dose engine validation including internal testing - User validation in cooperation with cancer clinics - . Reviews of design, code and Master Labeling The data obtained from the verification show that system tests, unit and subsystem tests have passed, and the validations been completed successfully. The reviews of design, code and labeling are also passed. From the successful verification and validation activities, the conclusion can be drawn that RayStation 11B has met specifications and is as safe, as effective and performs as well as or better than the legally marketed predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...