MRCAT brain

K193109 · Philips Medical Systems MR Finland · MUJ · Jan 24, 2020 · Radiology

Device Facts

Record IDK193109
Device NameMRCAT brain
ApplicantPhilips Medical Systems MR Finland
Product CodeMUJ · Radiology
Decision DateJan 24, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5050
Device ClassClass 2
AttributesAI/ML

AI Performance

OutputAcceptanceObservedDev DSTest DS
Radiotherapy Dose Calculation Accuracy (Brain)Gamma analysis criterion 2%/2mm realized in 98% of voxels within the PTV or exceeding 75% of the maximum dose; average simulated dose deviation <= 5% or 1 Gy (whichever is greater) in 99% of indicated patients in the volume of sensitive organsNot explicitly stated as a single numerical value, but noted as meeting acceptance criteriaCNN trained using matched pairs of CT and MRCAT source images; training is locked and not adapted during use.Clinical performance study: MRCAT images from patients used to calculate dose and compare against CT-based plans to demonstrate robustness and equivalence.

Indications for Use

MRCAT imaging is intended to provide the operator with information of tissue properties for radiation attenuation estimation purposes in photon external beam radiotherapy treatment planning.

Device Story

MRCAT Brain is a software add-on for Philips Ingenia, Ingenia Ambition, and Ingenia Elition MR systems. It processes T1-weighted 3D mDixon (in-phase and water) MR images acquired during a simulation session. A convolutional neural network (CNN) segments bone and body outline, assigning Hounsfield Unit (HU) values to voxels based on fat/water intensities to create CT-like density maps. The software runs on the MR console in the background. Clinicians use the generated MRCAT images for radiation attenuation estimation in external beam radiotherapy treatment planning. The device enables 'MR-only' simulation, potentially reducing patient motion artifacts by eliminating the need for separate CT scans. The algorithm is locked post-release.

Clinical Evidence

No clinical trials; evidence based on non-clinical performance testing. Robustness demonstrated by comparing dose plans generated from MRCAT images against CT-based plans using gamma analysis (criterion 2%/2mm, 98% pass rate in PTV). Geometric accuracy verified to ±1mm for 200mm sphere and ±5mm for 500mm sphere.

Technological Characteristics

Software plug-in for Ingenia MR systems. Uses 3D mDixon FFE imaging sequence. CNN-based segmentation and HU assignment. Locked algorithm. Complies with IEC 62304 (software lifecycle), IEC 60601-1 (safety), and ISO 14971 (risk management). DICOM RT export.

Indications for Use

Indicated for radiotherapy treatment planning for primary and metastatic brain tumor patients.

Regulatory Classification

Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. January 24, 2020 Philips Medical Systems MR Finland % Janne Marvola, Ph.D. Regulatory Engineer Ayritie 4 01510 Vantaa FINLAND Re: K193109 Trade/Device Name: MRCAT brain Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: Class II Product Code: MUJ Dated: November 4, 2019 Received: November 8, 2019 Dear Dr. Marvola: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for {1}------------------------------------------------ devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K193109 Device Name MRCAT Brain Indications for Use (Describe) MRCAT Brain is a software add-on for Ingenia, Ingenia Ambition, and Ingenia Elition MR systems. Intended Use: MRCAT imaging is intended to provide the operator with information of tissue properties for radiation attenuation estimation purposes in photon external beam radiotherapy treatment planning. Indications for use: MRCAT Brain is indicated for radiotherapy treatment planning for primary and metastatic brain tumor patients. | Type of Use (Select <i>one</i> or <i>both</i> , as applicable) | | |----------------------------------------------------------------|--| |----------------------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the word "PHILIPS" in a bold, sans-serif font. The color of the text is a bright, saturated blue. The letters are evenly spaced and appear to be a logo or brand name. # 6. 510(k) Summary ## 510(k) Summary of Safety and Effectiveness This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR §807.92. | Date Prepared: | January 20, 2020 | |-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------| | Manufacturer: | Philips Medical Systems MR Finland<br>Äyritie 4<br>01510 Vantaa, Finland | | | Establishment Registration Number: 9680194 | | Primary Contact<br>Person: | Janne Marvola<br>Regulatory Engineer<br>Phone: +358-40-126 1214<br>Fax: +358-9-3487 2406<br>E-mail: regulatory.mr.therapy@philips.com | | Secondary<br>Contact Person | Osku Ilvonen<br>Head of Q&R<br>Phone: +358-40-552 6197<br>Fax: +358-9-3487 2406<br>E-mail: osku.ilvonen@philips.com | | Device Name: | MRCAT brain | | Classification: | Classification Name: Medical charged-particle radiation therapy<br>system | | | Classification<br>Regulation: 21 CFR §892.5050 | | | Classification Panel: Radiology | | | Device Class: Class II | | | Product code: MUJ (System, Planning, Radiation Therapy<br>Treatment) | {4}------------------------------------------------ | Predicate Device: | Trade Name: | MRCAT Pelvis | | |------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|--| | | Manufacturer: | Philips Medical Systems MR Finland | | | | 510(k) Clearance: | K182888 (April 30, 2019) | | | | Classification<br>Regulation: | 21 CFR, Part 892.5050 | | | | Classification Name: | Radiation Therapy Planning System | | | | Classification Panel: | Radiology | | | | Device Class: | Class II | | | | Product Code: | MUJ (System, planning, radiation therapy<br>treatment) | | | Reference Device: | Trade Name: | AcQPlan 5.0 | | | | Manufacturer: | Philips Medical Systems MR Finland | | | | 510(k) Clearance: | K013644 (September 12, 2002) | | | | Classification<br>Regulation: | 21 CFR, Part 892.5840 | | | | Classification Name: | Radiation Therapy Planning System | | | | Classification Panel: | Radiology | | | | Device Class: | Class II | | | | Product Code: | MUJ (System, planning, radiation therapy<br>treatment) | | | Device<br>description: | MRCAT brain is a software application to Ingenia, Ingenia Ambition,<br>and Ingenia Elition MR systems. MRCAT brain is available to the<br>customer as an option to Ingenia MR-RT package, which is a set of<br>accessories for Ingenia systems. | | | | | Automated generation of MRCAT images takes place at the MR<br>console of Ingenia. The embedded image post-processing runs in the<br>background parallel to image acquisition. MRCAT algorithm enables<br>automatic tissue characterization: Bones are segmented from mDixon<br>in-phase and water images using machine learning based segmentation.<br>Body outline is segmented using in-phase and water images. Tissues<br>are then assigned a continuum of HU values depending on the fat and | | | {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame. | water intensities of the voxels. The HU assignment provides MRCAT images with CT-like density information. | | |------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications for Use: | MRCAT Brain is a software add-on for Ingenia, Ingenia Ambition, and Ingenia Elition MR systems. | | | Intended Use: | | | MRCAT imaging is intended to provide the operator with information of tissue properties for radiation attenuation estimation purposes in photon external beam radiotherapy treatment planning. | | | Indications for use: | | | MRCAT brain is indicated for radiotherapy treatment planning for primary and metastatic brain tumor patients. | | Fundamental Scientific Technology: | MRCAT brain functionality is implemented as a software plug-in for the MR main software and it contains the following main features: | | | 1) Automatic post-processing tool delivering MRCAT images<br>2) Examcard with mDixon imaging protocol<br>3) DICOM export of MRCAT image. | | | MRCAT Image Generation | | | MRCAT images are generated with an ExamCard post-processing step, which uses the images from the previous mDixon scan. | | | The post-processing logic takes care of launching MRCAT algorithm executable calculating a new 3D MRCAT image. The post-processing is started once the acquired mDixon MR images have been reconstructed. The first step of MRCAT generation is to pre-process the images to ensure that the MRCAT source images have consistent intensities. The intensity normalized images are then used as input in a convolutional neural network (CNN). The CNN is trained using matched pairs of CT and MRCAT source images. The training of the CNN is locked and is not adapted during use. The output of the CNN is post-processed to create images in CT values. The generated MRCAT images are checked for correctness to ensure validity of the generated MRCAT for radiation treatment. The sanity checks ensure that the imaging field of view has been positioned correctly and that the MRCAT body outline matches that of the MR. The HU values for the | {6}------------------------------------------------ MRCAT brain are calibrated using registered CT images. Once the process is running, post-processing logic exchanges information with the algorithm: - Image source data to algorithm, and image output data back to . the post-processing step - Progress notifications ● - Error and warning notifications ● The 3D MRCAT image from the post-processing step is stored into the MR image database. ### mDIXON scan A T1-weighted Fast Field Echo (FFE) 3D mDixon dual echo imaging protocol, with imaging parameters optimized for MRCAT image postprocessing and for geometric accuracy, is delivered as a part of MRCAT brain option. The mDixon imaging sequence provides two image contrasts for the MRCAT algorithm: inphase and water images. MRCAT brain uses fixed parameters for the mDixon scan, only the image stack location is configurable. An mDixon imaging protocol, with imaging parameters optimized for MRCAT image post-processing and for geometric accuracy, is delivered as a part of MRCAT brain option. MRCAT brain uses fixed parameters for mDixon scan, only the image stack location is configurable. ## DICOM Export The MRCAT post-processing step stores the image data returned by the MRCAT algorithm into MR database. MRCAT images can be exported in DICOM format enabling the use as primary images in the treatment planning systems ### Hardware platform description The new software extensions introduced by MRCAT brain run on the MR console of Ingenia. Based on the information provided above, the MRCAT brain is considered substantially equivalent to the primary currently marketed and predicate device (K182888, April 30, 2019) in terms of fundamental scientific technology. {7}------------------------------------------------ | Summary of Non-Clinical Performance Data: | The MRCAT brain complies with the following international and FDA-recognized consensus standards: | |-------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | International and FDA-recognized consensus standards:<br>ANSI/AAMI ES60601-1: 2012, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance IEC 60601-1-6:2010, Medical electrical equipment - Part 1-6: General requirements for safety - Collateral standard: Usability. IEC 60601-2-33:2015, Medical electrical equipment - Part 2-33: Particular requirements for the safety of magnetic resonance equipment for medical diagnosis. IEC 62304:2016, Medical device software - Software life-cycle processes IEC 62366-1:2015, Medical devices – Application of usability engineering to medical devices ISO 14971:2007, Medical devices – Application of risk management to medical devices | | | Non-Clinical verification and or validation tests have been performed with regards to the intended use, the technical claims, the requirement specifications and the risk management results. | | | Non Clinical verification and or validation test results demonstrate that the MRCAT brain:<br>Complies with the aforementioned international and FDA-recognized consensus standards Meets the acceptance criteria and is adequate for its intended | Therefore, the MRCAT brain is substantially equivalent to the primary currently marketed and predicate device (K182888, April 30, 2019) in terms of safety and effectiveness. Detailed comparison for selected features is presented in Table 6-1 below. {8}------------------------------------------------ | Table 6-1 | | | | |--------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------| | Comparison of the primary currently marketed and predicate<br>device, MRCAT Pelvis versus the proposed MRCAT Brain | | | | | Device | <i>MRCAT Pelvis</i> | <i>MRCAT Brain</i> | Similarities<br>and<br>Differences | | Manufacturer | Philips Medical<br>Systems MR<br>Finland | Philips Medical<br>Systems MR<br>Finland | - | | 510(k)<br>Number | K182888 | K193109 | N/A | | Product Code | MUJ | MUJ | Identical | | Regulation<br>Number | 892.5050 | 892.5050 | Identical | | Regulation<br>Name | Accelerator,<br>Linear, Medical | Accelerator,<br>Linear, Medical | Identical | | Intended use | MRCAT imaging is<br>intended to<br>provide the<br>operator with<br>information of<br>tissue properties<br>for radiation<br>attenuation<br>estimation<br>purposes<br>in photon external<br>beam radiotherapy<br>treatment<br>planning. | MRCAT imaging is<br>intended to<br>provide the<br>operator with<br>information of<br>tissue properties<br>for radiation<br>attenuation<br>estimation<br>purposes<br>in photon external<br>beam radiotherapy<br>treatment<br>planning. | Identical | {9}------------------------------------------------ Image /page/9/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame. The background is plain white. | Indications<br>for use | MRCAT Pelvis is<br>indicated for<br>radiotherapy<br>treatment<br>planning of soft<br>tissue cancers in<br>the pelvic region. | MRCAT Brain is<br>indicated for<br>radiotherapy<br>treatment<br>planning for<br>primary and<br>metastatic brain<br>tumor patients. | No significant<br>difference.<br>MRCAT Pelvis<br>and MRCAT<br>Brain are<br>both<br>indicated for<br>radiotherapy<br>treatment<br>planning in a<br>defined<br>region.<br>Brain tumors<br>are soft tissue<br>tumors. | |------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Primary<br>image<br>dataset | MRCAT | MRCAT | No significant<br>difference | | Secondary<br>image<br>dataset | mDixon, MRI | mDixon, MRI | No significant<br>difference<br>MR images<br>obtained in<br>the same<br>imaging<br>session are<br>inherently in<br>the same<br>frame of<br>reference. | | Registration<br>between<br>primary and<br>secondary<br>image<br>datasets | Secondary mDixon<br>MR image, source<br>data to MRCAT, is<br>inherently<br>registered as part<br>of MRCAT<br>algorithm with<br>MRCAT image,<br>which simplifies<br>workflow.<br>Other MR images,<br>like T2w and<br>fiducial marker<br>detection images<br>are registered<br>using tools<br>available in RTP<br>system | Secondary mDixon<br>MR image, source<br>data to MRCAT, is<br>inherently<br>registered as part<br>of MRCAT<br>algorithm with<br>MRCAT image,<br>which simplifies<br>workflow.<br>Other MR images,<br>like T2w images<br>are registered<br>using tools<br>available in RTP<br>system | No significant<br>difference<br>Secondary<br>MR images<br>are obtained<br>in the same<br>imaging<br>session<br>reducing the<br>possibility of<br>patient<br>motion<br>between<br>images. | | Primary<br>image<br>density<br>information | MRCAT image<br>intensity<br>information is<br>provided in<br>Hounsfield Unit<br>(HU) values. | MRCAT image<br>intensity<br>information is<br>provided in<br>Hounsfield Unit<br>(HU) values. | No significant<br>difference.<br>MRCAT Pelvis<br>and MRCAT<br>Brain both<br>have<br>continuous<br>HU value<br>approach. | | Conversion<br>from primary<br>image to<br>density<br>values used<br>in dose<br>calculation | Primary image HU<br>values are<br>converted to<br>densities through<br>density table<br>specific for the<br>MRCAT. | Primary image HU<br>values are<br>converted to<br>densities through<br>density table<br>specific for the<br>MRCAT. | No significant<br>difference<br>MRCAT has<br>specific<br>density table<br>that is used in<br>a similar<br>manner to CT<br>specific<br>density<br>tables. | | MRCAT<br>algorithm | Bones are<br>segmented from<br>mDixon inphase<br>and water images<br>using model based | Bones are<br>segmented from<br>mDixon inphase<br>and water images<br>using machine | No significant<br>difference<br>Segmentation<br>is done for<br>both MRCAT | | | | | | | segmentation. The<br>segmented bones<br>are the femurs,<br>pelvic bones and<br>lumbar vertebrae<br>L5 and L1.<br><br>Body outline is<br>segmented using<br>inphase and water<br>images. | learning based<br>segmentation. The<br>segmented bones<br>are in skull, upper<br>C-spine and jaw.<br><br>Body outline is<br>segmented using<br>inphase and water<br>images. | Pelvis and<br>Brain using<br>the mDIXON<br>image<br>contrasts.<br><br>HU value<br>assignment is<br>done based<br>on mDixon<br>image<br>intensities. | | | Bones are assigned<br>a continuum of HU<br>values between<br>dense cortical<br>bone and light<br>spongy bone<br>depending on the<br>fat and water<br>intensities of the<br>voxels.<br><br>Soft tissue are<br>assigned a<br>continuum of HU<br>values between fat<br>and muscle tissue<br>depending on the<br>fat and water<br>intensities of the<br>voxels.<br><br>The HU values for<br>the MRCAT Pelvis<br>are calibrated<br>using registered CT<br>images from<br>several sites. | Bones are assigned<br>a continuum of HU<br>values between<br>dense cortical<br>bone and light<br>spongy bone<br>depending on the<br>fat and water<br>intensities of the<br>voxels.<br><br>Soft tissue are<br>assigned a<br>continuum of HU<br>values depending<br>on the fat and<br>water intensities of<br>the voxels.<br><br>The HU values for<br>the MRCAT Brain<br>are calibrated<br>using registered CT<br>images from<br>several sites.<br><br>MRCAT Brain<br>algorithm is fully<br>trained before<br>product release,<br>after which the<br>algorithm is<br>locked. | The models<br>used are<br>equivalent in<br>relation to<br>dose and<br>positioning<br>accuracy.<br><br>Both<br>algorithms<br>are locked;<br>they do not<br>change after<br>installation<br>based on new<br>data during<br>the use. | | | Patient<br>positioning | Ingenia MR-RT<br>with MRCAT Pelvis<br>supports MR Only<br>simulation with<br>relative patient<br>marking. | Ingenia MR-RT<br>with MRCAT Brain<br>supports MR Only<br>simulation with<br>relative patient<br>marking. | No significant<br>difference<br>The visibility<br>of bone<br>structures is<br>equivalent for<br>both<br>products.<br>Internal<br>markers are<br>not used for<br>brain tumors. | | | Patient positioning<br>in the treatment<br>machine must be<br>checked either<br>with cone beam<br>computed<br>tomography<br>(CBCT) or plain<br>radiographs by<br>registering bone<br>structures or<br>internal fiducial<br>markers. | Patient positioning<br>in the treatment<br>machine must be<br>checked either<br>with cone beam<br>computed<br>tomography<br>(CBCT) or plain<br>radiographs by<br>registering bone<br>structures. | | | Dose | The simulated | The simulated | No significant | | accuracy | dose based on<br>MRCAT images<br>shall not differ in<br>95% of prostate<br>cancer patients<br>(gamma analysis<br>criterion 3%/3mm<br>realized in 99% of<br>voxels within the<br>PTV or exceeding<br>75% of the<br>maximum dose)<br>when compared<br>with CT based<br>plan. | dose based on<br>MRCAT Brain<br>images shall not<br>differ in 95% of the<br>indicated patients<br>(gamma analysis<br>criterion 2%/2mm<br>realized in 98% of<br>voxels within the<br>PTV or exceeding<br>75% of the<br>maximum dose)<br>when compared<br>with CT-based<br>plan. | difference.<br>The same<br>dose<br>evaluation<br>methodology<br>is used for<br>both<br>products. The<br>criteria are<br>selected<br>based on the<br>needs of the<br>application. | | | The average<br>simulated dose<br>based on MRCAT<br>images shall not<br>deviate more than<br>10% for voxels<br>exceeding 5Gy in<br>99% of the<br>indicated patients<br>in the volume of<br>sensitive organs<br>when compared<br>with CT based<br>plan. | The average<br>simulated dose<br>based on MRCAT<br>Brain shall not<br>deviate more than<br>5% or 1 Gy, which<br>ever is greater, in<br>99% of the<br>indicated patients<br>in the volume of<br>sensitive organs<br>when compared<br>with CT based<br>plan. | | | Geometric<br>accuracy | MRCAT accuracy:<br>± 1 mm accuracy:<br>200 mm diameter<br>sphere<br>± 5 mm accuracy:<br>500 mm diameter<br>sphere (limited in<br>the bore direction<br>by +/- 160 mm<br>from the z=0 mm<br>plane ) | MRCAT accuracy:<br>± 1 mm accuracy:<br>200 mm diameter<br>sphere<br>± 5 mm accuracy:<br>500 mm diameter<br>sphere (limited in<br>the bore direction<br>by +/- 160 mm<br>from the z=0 mm<br>plane ) | No significant<br>difference | | MRCAT<br>source<br>imaging<br>sequence | mDIXON 3D scan<br>with acquired<br>voxel size of<br>1.40/1.40/1.40<br>mm, and<br>bandwidth/pixel<br>430Hz (1.5T) or<br>860Hz (3T).<br>Most scanning<br>parameters locked | mDIXON 3D scan<br>with acquired<br>voxel size of<br>1.10/1.10/1.40<br>mm (1.5 T) and<br>1.1/1.1/1.1 mm<br>(3T), and<br>bandwidth/pixel<br>481Hz (1.5T) or<br>868Hz (3T).<br>Most scanning<br>parameters locked | No significant<br>difference<br>Especially,<br>the sensitivity<br>to B0 induced<br>distortion in<br>the read<br>direction is<br>about 0.2<br>mm/ppm in<br>both MRCAT<br>Pelvis and<br>MRCAT Brain.<br>In both<br>solutions the<br>essential<br>parameters<br>are locked to<br>avoid user<br>errors<br>potentially<br>affecting the<br>accuracy and<br>reliability of<br>the method. | | DICOM RT<br>export | yes | yes | No significant<br>difference | {10}------------------------------------------------ Image /page/10/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame. The background is plain white. {11}------------------------------------------------ Image /page/11/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is white, providing a strong contrast that makes the word stand out. {12}------------------------------------------------ Image /page/12/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is white, which makes the blue letters stand out. # Philips Medical Systems MR Finland {13}------------------------------------------------ Image /page/13/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame. The background is white. # Philips Medical Systems MR Finland {14}------------------------------------------------ Image /page/14/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame. The background is plain white. | Summary of<br>Clinical Data: | The robustness of the MRCAT brain algorithm for producing<br>equivalent dose plans to CT using gamma analysis with criterion of<br>1%/1mm is shown by post-processing MRCAT images from patients,<br>and calculating dose using the MRCAT images. | |-------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | In summary, the MRCAT brain images are spatially accurate radiation<br>attenuation estimates that can aid in the EBRT planning of primary and<br>metastatic brain tumors. | | Substantial<br>Equivalence<br>Conclusion: | The MRCAT brain is substantially equivalent to the primary currently<br>marketed and predicate device (K182888, April 30, 2019) in terms of<br>design features, fundamental scientific technology, indications for use,<br>and safety an…
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...