K040688 · Standard Imaging, Inc. · MUJ · May 11, 2004 · Radiology
Device Facts
Record ID
K040688
Device Name
STANDARD IMAGING E.IMRT CALCULATOR
Applicant
Standard Imaging, Inc.
Product Code
MUJ · Radiology
Decision Date
May 11, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Software as a Medical Device
Intended Use
The e.IMRT Calculator is a simple, stand-alone software program provided on a CD-rom for use on an appropriately configured personal computer. It is intended to assist the oncologist or medical physicist in creating an optimum electron treatment plan based on the treatment objective for the patient. Using previously established mathematical equations, the e.IMRT Calculator suggests several potential electron beam treatment energy solutions, within user selected parameters, by combining several discrete energies typically available on the linear accelerator(s). It uses previously user gathered or generated depth dose data sets for each electron energy available (4, 6, 9, 12, 15, 16, 18, 20 or 22 MeV) on the user's specific liner accelerator(s) as its primary input. Other inputs involve the identification of the facility, specific linear accelerator(s) and bolus used. The output of the e.IMRT calculator is a hardcopy printout of the suggested electron beam treatment energies, in both numeric and graphical formats. The site, linear accelerator and bolus information are also provided on this printout are the site. The e.IMRT Calculator does not, however, electronically store any patient identification data or information. Inputs involving patient information are only used for hardcopy printouts. Specific electron beam parameters are selected from the solution options provided by the e.IMRT Calculator for non-direct (manual) input into the user's radiation treatment planning system. The treatment planning system then creates a treatment plan based on the treatment objective for the patient, which is reviewed by the attending oncologist for acceptability prior to implementation with a linear accelerator.
Device Story
Software tool for radiation oncology; assists oncologists/physicists in electron beam treatment planning. Inputs: user-provided depth dose data sets for specific linear accelerator energies (4-22 MeV), facility ID, linear accelerator model, and bolus type. Operation: uses established mathematical equations to combine discrete electron energies to suggest optimal treatment energy solutions within user-defined parameters. Output: hardcopy printout (numeric/graphical) of suggested beam energies. Usage: standalone PC software; non-direct input of suggested parameters into primary radiation treatment planning system. Benefits: facilitates optimization of electron treatment plans; does not store patient data. Clinical decision-making: oncologist reviews planning system output for acceptability before implementation on linear accelerator.
Clinical Evidence
Bench testing only. Validation performed at Standard Imaging and University of Wisconsin-Madison. Testing included algorithm transfer, results presentation/graphing, absolute dose depth measurements, interface/compatibility assessment, and treatment planning system comparison analysis. Device met predetermined design specifications and risk analysis objectives.
Technological Characteristics
Standalone software program for PC; utilizes established mathematical equations for electron beam energy calculation. Complies with IEC 60601-1-4 (collateral standard for programmable medical systems). No patient data storage. Input: user-generated depth dose data sets. Output: hardcopy printout.
Indications for Use
Indicated for use by oncologists or medical physicists to assist in creating optimum electron treatment plans for patients requiring radiation therapy.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for "Standard Imaging". The logo is in black and white and features the company name in a bold, sans-serif font. Below the name is a stylized graphic that resembles a burst or star, with lines radiating outward from a central point. The overall design is simple and modern.
MAY 11 2004
KD40668
## March 11, 2004
| Subject: | 510(k) Summary of Safety and Effectiveness Information for the<br>Standard Imaging e.IMRT Calculator |
|----------------------|------------------------------------------------------------------------------------------------------|
| Proprietary Name: | Standard Imaging e.IMRT Calculator |
| Common Name: | Electron Beam Planning Calculator |
| Classification Name: | Medical Charged-Particle Radiation Therapy System, Treatment Planning<br>Computer Program |
| Classification: | Class II -- 21CFR892.5050, 90MUJ |
| Panel: | Radiology |
| Contact Person: | Raymond Riddle, Vice President, Regulatory Affairs |
The 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 1992.
The Standard Imaging e.IMRT Calculator is substantially equivalent to the electron portion of the ADAC Laboratories (Philips) Pinnacle 3 Radiation Therapy Planning Software covered by 510(k) Numbers K951581, K993923, K032724.
The e IMRT Calculator is a simple, stand-alone software program provided on a CD-rom for use on an appropriately configured personal computer. It is intended to assist the oncologist or medical physicist in creating an optimum electron treatment plan based on the treatment objective for the patient.
Using previously established mathematical equations, the e.IMRT Calculator suggests several potential electron beam treatment energy solutions, within user selected parameters, by combining several discrete energies typically available on the linear accelerator(s). It uses previously user gathered or generated depth dose data sets for each electron energy available (4. 6, 9, 12, 15, 16, 18, 20 or 22 MeV) on the user's specific liner accelerator(s) as its primary input. Other inputs involve the identification of the facility, specific linear accelerator(s) and bolus used.
Image /page/0/Picture/9 description: The image shows the text "Radiation Calibration and QA Instruments". The text is white and is on a black background. The text is centered in the image.
COMPUTILL SOLUTIONS PROVIDENCE
for the best care
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a logo for "Standard Imaging". The logo is in black and white and features the words "Standard Imaging" in a stylized font. The word "Standard" is above the word "Imaging", and there is a registered trademark symbol next to the word "Imaging". Below the words is a design that looks like rays emanating from a central point.
The output of the e.IMRT calculator is a hardcopy printout of the suggested electron beam The butput of the c.HAN's numeric and graphical formats. The site, linear accelerator and bolus information are also provided on this printout are the site. The e.IMRT Calculator does not, however, electronically store any patient identification data or information. Inputs involving patient information are only used for hardcopy printouts.
Specific electron beam parameters are selected from the solution options provided by the e. IMRT Calculator for non-direct (manual) input into the user's radiation treatment planning system. The treatment planning system then creates a treatment plan based on the treatment objective for the patient, which is reviewed by the attending oncologist for acceptability prior to implementation with a linear accelerator.
The Standard Imaging e.IMRT Calculator was designed to comply with the applicable portions of the following voluntary standard:
- IEC 60601-1-4 (Edition 1 1 2000-04) -- Collateral standard for programmable medical 1. systems
The Standard Imaging e.IMRT Calculator has been validated at Standard Imaging and at the University of Wisconsin - Madison. This validation included the following:
- Algorithm transfer l .
- Results presentation and graphing 2.
- Absolute dose depth measurements 3.
- Interface, compatibility, use and misuse 4.
- Treatment planning system comparison analysis 5.
The Standard Imaging e.IMRT Calculator has met its predetermined design specifications, risk analysis and validation objectives.
10
Radiation Calibration and OA Instrum
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three wavy lines extending from its body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 11 2004
Mr. Raymond T. Riddle, PE, RAC Vice President, Regulatory Affairs Standard Imaging, Inc. 7601 Murphy Drive MIDDLETON WI 53562-2532
Re: K040688 Trade/Device Name: Standard Imaging e.IMRT Calculator Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle Radiation therapy system Regulatory Class: II Product Code: 90 MUJ Dated: March 11, 2004 Received: March 16, 2004
Dear Mr. Riddle:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act), 21 CFR 1000-1050
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally promated predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please conact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx. 1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brydon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K040688
## Indications for Use
## 510(k) Number (if known): に040688
Device Name:
Standard Imaging e.IMRT Calculator
Indications For Use:
The e.IMRT Calculator is a simple, stand-alone software program provided on a CD-rom for use on an appropriately configured personal computer. It is intended to assist the oncologist or medical physicist in creating an optimum electron treatment plan based on the treatment objective for the patient.
Using previously established mathematical equations, the e.IMRT Calculator suggests several potential electron beam treatment energy solutions, within user selected parameters, by combining several discrete energies typically available on the linear accelerator(s). It uses previously user gathered or generated depth dose data sets for each electron energy available (4, 6, 9, 12, 15, 16, 18, 20 or 22 MeV) on the user's specific liner accelerator(s) as its primary input. Other inputs involve the identification of the facility, specific linear accelerator(s) and bolus used.
The output of the e.IMRT calculator is a hardcopy printout of the suggested electron beam treatment energies, in both numeric and graphical formats. The site, linear accelerator and bolus information are also provided on this printout are the site. The e.IMRT Calculator does not, however, electronically store any patient identification data or information. Inputs involving patient information are only used for hardcopy printouts.
Specific electron beam parameters are selected from the solution options provided by the e.IMRT Calculator for non-direct (manual) input into the user's radiation treatment planning system. The treatment planning system then creates a treatment plan based on the treatment objective for the patient, which is reviewed by the attending oncologist for acceptability prior to implementation with a linear accelerator.
Prescription Use v (Part 21 CFR 801 Subpart D)
and/or
Over-the-Counter-Use (Part 21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel A. Lynn
Division Sir 510(k) Number
Page 1 of 1
\$\delta\$
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.