K013535 · 3D Line USA, Inc. · MUJ · Jan 18, 2002 · Radiology
Device Facts
Record ID
K013535
Device Name
EVTOOL MODEL # V. 1.0
Applicant
3D Line USA, Inc.
Product Code
MUJ · Radiology
Decision Date
Jan 18, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Intended Use
EVTOOL software module provides the ability to compare and evaluate the radiation dose delivery to the tumor volume and surrounding organs from generated dose volume histograms and radiobiological values. It expands the DMLC IV-ERGO system, which is a combination of a radiation collimator with multiple tungsten leaves that move during delivery of radiation therapy and a computer based treatment planning and control system. The EVTOOL is intended to be used for medical procedures on patients to be prescribed and performed by a suitably trained and certified medical professional.
Device Story
EVTOOL is a software module integrated into the DMLC IV-ERGO radiation therapy planning system. It inputs stored treatment plans and patient data from the ERGO database. The software calculates and displays dose volume histograms (DVH) and organ dose statistics (mean, modal, maximum, minimum). It incorporates radiobiological models—specifically Tumor Control Probability (TCP), Normal Tissue Complication Probability (NTCP), and Uncomplicated Tumor Control Probability (UTCP)—to assist physicians in optimizing dose prescriptions and fractionation. The output is displayed in a 4-window interface with zoom capabilities, allowing side-by-side comparison of multiple treatment plans. Used in clinical settings by radiation oncologists or medical physicists, the device does not deliver radiation directly but provides quantitative analysis to support clinical decision-making regarding the most appropriate treatment plan, ensuring effective control of the actual dose delivered to the patient.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Software module running on a Silicon Graphics workstation using the Irix operating system. Integrates with DMLC IV-ERGO radiation collimator system. Provides quantitative analysis of dose volume histograms and radiobiological models (TCP, NTCP, UTCP).
Indications for Use
Indicated for use in radiation therapy planning for patients requiring evaluation of radiation dose delivery to tumor volumes and surrounding organs.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
K250963 — PlanOne 1 · Cosylab Jsc, Control System Laboratory · Nov 20, 2025
K013644 — ACQPLAN 5.0 · Philips Medical Systems, Inc. · Sep 12, 2002
K142142 — QUALITY REPORTS · Sun Nuclear Corporation · Dec 5, 2014
K133247 — ECLIPSE TREATMENT PLANNING SYSTEM · Varian Medical Systems, Inc. · Feb 7, 2014
Submission Summary (Full Text)
{0}------------------------------------------------
01353.5
Premarket Notification 3D Line: EVTOOL Date : 20 October 2001
JAN 1 8 2002
Image /page/0/Picture/3 description: The image shows a logo with the text "3D LINE". The "3D" is stylized to appear three-dimensional, with shading and depth. The word "LINE" is written below "3D" in a simpler, blocky font. The overall design has a slightly pixelated or low-resolution appearance.
Date: October 20, 2001
Department of Health and Human Services Center of Devices and Radiological Health Office of Device Evaluation Pre-Market Notification section
3D Line USA 2807 Old Court Rd Baltimore MD 21208 Phone: 410-580-1730 Fax: 410-580-1732
# 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION
### as required by section 807.92(c)
#### Submitter of 510(k) a.
| Company name: | 3D Line USA |
|-----------------|--------------------------------------------|
| Registration # | 9041925 |
| Address: | 2807 Old Court Road<br>Baltimore, MD 21208 |
| Contact Person: | Karen Rigamonti<br>President |
| Phone: | 410-580-1730 |
| Fax: | 410-580-1732 |
### b. Device Name:
| Trade/Proprietary Name: | EVTOOL |
|-------------------------|----------------------------------------------------------------------|
| Common/Usual Name: | Radiation Therapy Planning System |
| Classification Name: | Accelerator, Linear, Medical, Accessory<br>21 CFR 892.5050 Class II. |
### C. Legally Marketed Predicate Device(s)
Our device is substantially equivalent to the legally marketed predicate devices cited in the table below.
| Manufacturer | Device | 510(k) # |
|--------------|--------------|----------|
| 3D Line | DMLC IV-ERGO | K001163 |
### Description d.
EVTOOL is a software module that has been designed to expand the features and functions of the DMLC IV- ERGO treatment planning software. The software runs on a silicon graphics workstation and Irix operating system.
{1}------------------------------------------------
## Premarket Notification 3D Line: EVTOOL Date : 20 October 2001
EVTOOL software is used for comparing and evaluating radiation therapy treatment plans from a quantitative and radiobiological point of view. The software displays one or more stored plans for comparison and analysis. It can display the dose volume histograms of the various plans and organ dose statistics (mean, modal, maximum and minimum dose). It allows the user to input radiobiological corrections in order to enhance the evaluation of the treatment plan. It also optimizes the dose prescription by displaying reference dose and number of treatment fractions.
EVTOOL is fully integrated into the ERGO treatment planning environment in order to easily exchange information from the patients' database. The patient information is selected and automatically loaded into the EVTOOL software. From the patient database the user can select one or multiple treatment plans to display and calculate the dose statistics, including dose volume histograms and radiobiological models. The analytical information is displayed in 4 windows with a zoom capability. These windows can display the overall plan and individual organs of interest. EVTOOL provides options for incorporating radiobiological corrections to assist the physician in determining the optimal dose prescription and fractionation based on TCP (Tumor Control Probability), NTCP (Normal Tissue Complication Probability) and UTCP (Uncomplicated Tumor Control Probability) values.
This software is a useful tool in evaluating the entire volumetric treatment plan, it is not used to directly treat the patient. It provides the physician with a tool to compare various plan versions side by side in order to determine the most appropriate treatment plan to be used with the DMLC IV equipment. The display of the treatment plans within EVTOOL provides a permanent record and effective control on the actual dose delivered to the patient.
#### Intended use e.
EVTOOL software module provides the ability to compare and evaluate the radiation dose delivery to the tumor volume and surrounding organs from generated dose volume histograms and radiobiological values. It expands the DMLC IV-ERGO system, which is a combination of a radiation collimator with multiple tungsten leaves that move during delivery of radiation therapy and a computer based treatment planning and control system.
### Summary of technological considerations f.
The EVTOOL software is substantially equivalent to the predicate devices. It enhances the functionality of the defined predicate device to provide evaluation tools for radiation therapy treatment planning.
Karen Rigamonti
Name: Karen Rigamonti, MD, MBA, M.P.H. Title President 3D Line USA. Inc. Baltimore, MD
10/18/01
Date
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle with three lines representing its wings and three lines representing its legs.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 1 8 2002
Karen Rigamonti, M.D. President 3D Line USA, Inc. 2807 Old Court Road BALTIMORE MD 21208 Re: K013535
Trade/Device Name: EVTOOL Model V 1.0 Accessory to RTP System Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II
Product Code: 90 MUJ Dated: October 20, 2001 Received: October 23, 2001
Dear Dr. Rigamonti:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Premarket Notification 3D Line: EVTOOL Date : 20 October 2001
Image /page/4/Picture/1 description: The image shows a logo with the text "3D LINE" in a stylized, three-dimensional design. The "3D" is stacked on top of the word "LINE", with the "3" and "D" forming a single, connected shape. The letters have a textured appearance, giving them a sense of depth and dimension. The overall design is simple yet modern, suggesting a focus on technology or innovation.
October 20, 2001
Department of Health and Human Services Center of Devices and Radiological Health Office of Device Evaluation Pre-Market Notification section
3D Line USA
2807 Old Court Rd Baltimore MD 21208 Phone: 410-580-1730 Fax: 410-580-1732
## Statement of Intended Use
Device Name:
EVTOOL
### Intended use
EVTOOL software module provides the ability to compare and evaluate the radiation dose delivery to the tumor volume and surrounding organs from generated dose volume histograms and radiobiological values. It expands the DMLC IV-ERGO system, which is a combination of a radiation collimator with multiple tungsten leaves that move during delivery of radiation therapy and a computer based treatment planning and control system.
## Prescription use
The EVTOOL is intended to be used for medical procedures on patients to be prescribed and performed by a suitably trained and certified medical professional.
Karen Rigamonti
10/18/01
---
Date
Name: Karen Rigamonti, MD, MBA, M.P.H. Title: President > 3D Line USA, Inc. Baltimore, MD
Nancy C Broadon
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K013535
**Prescription Use**
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.