LEKSELL GAMMAPLAN 4C WITH MULTIVIEW

K042269 · Elekta Instrument AB · MUJ · Sep 8, 2004 · Radiology

Device Facts

Record IDK042269
Device NameLEKSELL GAMMAPLAN 4C WITH MULTIVIEW
ApplicantElekta Instrument AB
Product CodeMUJ · Radiology
Decision DateSep 8, 2004
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.5050
Device ClassClass 2

Indications for Use

The Leksell GammaPlan 4C with MultiView is a computer-based dose planning system specifically designed for use with the Leksell Gamma Knife®. The Leksell GammaPlan 4C with MultiView is intended to be used for planning the dosimetry of treatments in stereotactic radiosurgey and stereotactic radio therapy. It processes the inputs from the healt care professionals such that the desired radiation dose is provided by the Leksell Gamma Knife® to a precisely defined target area within the cranium.

Device Story

Computer-based dose planning system for Leksell Gamma Knife®; used by clinicians for stereotactic radiosurgery/radiotherapy planning. Inputs include medical images (CT/MRI/PET); system performs co-registration of image series to support frameless imaging; enables visualization of brain atlas structures and color-mapped PET images. Output is a calculated dosimetry plan for radiation delivery to intracranial targets. Facilitates precise radiation targeting; reduces need for multiple frame-based image acquisitions.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Computer-based software system for radiation dose planning. Features image co-registration, brain atlas integration, and PET image visualization. Operates as an adjunct to the Leksell Gamma Knife® system.

Indications for Use

Indicated for patients requiring stereotactic radiosurgery or stereotactic radiotherapy planning for intracranial target areas. Used by healthcare professionals to calculate dosimetry for delivery via Leksell Gamma Knife®.

Regulatory Classification

Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## ELEKTA INSTRUMENT AB Dokumentnamn/Nome of document Special 510(k) 12 | Utfärdare/Issuer<br>Anders Skoglund | Ref nr/Dok nr/Ref no/Doc no | Utgava / Edition | | |--------------------------------------------------------|-----------------------------|------------------|--| | Avser/Regarding<br>Leksell GammaPlan 4C with MultiView | | Directory | | GammaPlan 4C with MultiView Kø4 2269 ## Section 4- 510(k) Summary As Required by 21 CFR 807.87(k)510 (k) Summary ### Subscribers Name & Address 1. Elekta Instrument AB Kungstensgatan 18, P:O Box 7593 SE-103 93 Stockholm, Sweden Tel: (011) 46 8 587 254 00 Fax: (011) 46 8 587 255 00 Contact Person for this submission: Mr Anders Skoglund Official Correspondent: Mr Peter Löwendahl #### Trade Name 2. Leksell GammaPlan 4C with MultiView #### Device Classification 3. | Common Name | Product Code | Class | Regulation Number | |---------------------------------------|--------------|-------|-------------------| | Radionuclide radiation therapy system | IWB | II | 892.5750 | ### Regulatory History (Unmodified Predicate Device) 4. | Devices | 510(k) # | |-------------------|----------| | Leksell GammaPlan | K973441 | #### 5. Other relevant submissions | Devices | 510(K) # | |-----------------------------------|----------| | Leksell Surgiplan with Atlaspace | K013861 | | Leksell Surgiplan with Imagemerge | K033340 | {1}------------------------------------------------ | ELEKTA INSTRUMENT AB | Ky4 2264 | Dokumentnamn/Name of document | 13 | |----------------------|-----------------------------|-------------------------------|----| | Special 510(k) | | | | | Utfärdare/Issuer | Ref nr/Dok nr/Ref no/Doc no | Utgåva /Edition | | | Anders Skoglund | - | 1 | | Avser/Regarding Leksell GammaPlan 4C with MultiView ### Device Description (for detailed description see Section "Device Description") 6. The Leksell GammaPlan 4C with MultiView adds support for frameless images by means of co-registration of image series. This simplifies stereotactic radiosurgery procedures as only one image series needs to be acquired with the Leksell stereotactic frame. Leksell GammaPlan 4C with MultiView also adds support for brain altas structures and visualization of color mapped PET images for use in preoperative planning. #### Intended Use 7. The Leksell GammaPlan 4C with MultiView is a computer-based dose planning system specifically designed for use with the Leksell Gamma Knife®. The Leksell GammaPlan 4C with MultiView is intended to be used for planning the dosimetry of treatments in stereotactic radiosurgey and stereotactic radio therapy. It processes the inputs from the healt care professionals such that the desired radiation dose is provided by the Leksell Gamma Knife® to a precisely defined target area within the cranium. #### Substantial Equivalence 8 The functionality for the Lekscll GammaPlan 4C with MultiView is equivalent to its predicate device Leksell GammaPlan (K973441) in safety and effectiveness. The fundamental technical characteristics are similar to those of the predicate device. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines representing its wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 ## SEP - 8 2004 Mr. Peter Löwendahl Quality and Regulatory Affairs Manager Elekta Instrument AB P.O. Box 7593 SE-103 93 Stockholm SWEDEN Re: K042269 Trade/Device Name: Leksell GammaPlan 4C with MultiView Regulation Number: 21 CFR 892.5750 Regulation Name: Radionuclide radiation therapy system Regulatory Class: II Product Code: 90 MUJ Dated: August 20, 2004 Received: August 23, 2004 Dear Mr. Löwendahl: We have reviewed your Section 510(k) premarket notification of intent to market the device we neve roviewed your coreemined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreated 9776, the enactment date of the Medical Device Amendments, or to eoniner of process that have been recordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Oosmotion For (100) the ) the device, subject to the general controls provisions of the Act. The r ou may, the crowisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can may or subject to Sateral Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act that I Dr. I has had regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act), 21 CFR 1000-1050. {3}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will anow you to oght hands.org of substantial equivalence of your device to a legally premarket notification. The PDA maily gour device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your de to the following numbers, based on the regulation number at the top of the letter: | 8xx.1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Additionally, for questions on and personation entitled, "Miso, please note the regulation entitled, "Misbranding Other of Complanos at (201) 22ation" (21CFR Part 807.97) you may obtain. Other general by relection on your responsibilities under the Act may be obtained from the Division of Small mornation on your responsionities and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Nancy C. Brydon Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ | ELEKTA INSTRUMENT AB | | |----------------------|--| |----------------------|--| Dokumentnamn/Name of document 16 ELEKTA INSTRUMENT AB Special 510(k) | Utfärdare/Issuer | Anders Skoglund | Utgåva /Edition | 1 | |-----------------------------|-------------------------------------|-----------------|---| | Ref nr/Dok nr/Ref no/Doc no | - | Directory | | | Avser/Regarding | Leksell GammaPlan 4C with MultiView | | | # Section 7- Indications for Use Statement | 510(k) Number | To be defined<br>K\$\phi\$42269 | |---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name | Leksell GammaPlan 4C with MultiView | | Indications for Use | The Leksell GammaPlan 4C with MultiView is a computer-based dose<br>planning system specifically designed for use with the Leksell Gamma<br>Knife®. The Leksell GammaPlan 4C with MultiView is intended to be<br>used for planning the dosimetry of treatments in stereotactic<br>radiosurgey and stereotactic radio therapy. It processes the inputs<br>from the healt care professionals such that the desired radiation dose<br>is provided by the Leksell Gamma Knife® to a precisely defined target<br>area within the cranium. | X Prescription Use (Part 21 CFR 801 Subpart D) -AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY) Concurrence of CDRH, Office of Device Evaluation (ODE) David A. Layman (Division Sign-Off) Division of Reproductive, Abdomis and Radiological Devices 510(k) Number J
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...