The Implant Sciences 125 Iodine Brachytherapy Seeds with activities from 0.1 to 5.0 mCi are indicated for temporary or permanent interstitial or intracavity implantation or surface application to treat selected localized tumors. They can be used either as primary treatment (such as for prostate cancer or unresectable tumors) or as treatment for residual disease after excision of primary or recurrent tumors. The 125 Iodine Seeds may be used concurrently with or following treatment with other interventions, such as external beam therapy, hyperthermia or chemotherapy. Tumors of the head, neck, lung, pancreas, prostate and other accessible tumors are commonly treated.
Device Story
Implant Sciences 125 Iodine Brachytherapy Seeds (Model 3000) consist of laser-welded titanium capsule containing aluminum-clad copper rod; rod serves as radioactive iodine source substrate and radiopaque x-ray marker for localization. Seeds provided non-sterile; require hospital-based sterilization via autoclave or ethylene oxide. Physician implants seeds into localized tumors (prostate, head, neck, lung, pancreas) for interstitial/intracavity radiation therapy. Radiation emission provides localized treatment; can be used as primary therapy or for residual disease; compatible with external beam therapy, hyperthermia, or chemotherapy. Marker allows visualization of seed orientation/location during clinical procedures.
Clinical Evidence
Bench testing only. Data provided demonstrates compliance with ANSI/HPS N43.6-1997 standards for sealed radioactive sources, including temperature resistance, external pressure resistance, impact resistance, vibration integrity, and puncture resistance. Capsule leak testing performed per ANSI standards. Radiation profiles compared to existing Iodine-125 seeds confirm consistency.
Technological Characteristics
Titanium capsule; aluminum-clad copper rod substrate/marker. Dimensions identical to predicates. Energy source: Iodine-125 radionuclide. Sterilization: Hospital-based autoclave or ethylene oxide. Testing per ANSI/HPS N43.6-1997.
Indications for Use
Indicated for patients with localized tumors, including prostate cancer, unresectable tumors, or residual disease following excision of primary/recurrent tumors. Applicable to tumors of the head, neck, lung, pancreas, prostate, and other accessible sites. Used for temporary or permanent interstitial, intracavity, or surface application.
Regulatory Classification
Identification
A radionuclide brachytherapy source is a device that consists of a radionuclide which may be enclosed in a sealed container made of gold, titanium, stainless steel, or platinum and intended for medical purposes to be placed onto a body surface or into a body cavity or tissue as a source of nuclear radiation for therapy.
Special Controls
*Classification.* Class II (special controls). A prostate seeding kit intended for use with a radionuclide brachytherapy source only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
North American Scientific, Inc. (NASI) loGold™ I-125 Brachytherapy Source (Model MED3631-A) (K972271)
Medi-Physics, Inc. (Amersham) I-125 Seeds (No. 6711) (K914281)
International Isotopes, Inc. Imagyn 125 Iodine Brachytherapy Seed (K982421)
K994317 — I-PLANT, MODEL 3500 · Implant Sciences Corp. · Mar 21, 2000
K023242 — I-PLANT MODEL 3500 (125IODINE BRACHYTHERAPY SEEDS)-STERILE · Implant Sciences Corp. · Dec 4, 2002
K081066 — BRACHYTHERAPY SOURCE DEVICE, MODEL 9011 · Medi-Physics, Inc., Dba GE Healthcare · May 5, 2008
K992963 — I3RAD 125 I BRACHYTHERAPY SEED, MODEL I125-0401 · International Isotopes, Inc. · Sep 21, 1999
K991526 — ISOSTAR I-125 INTERSTITIAL SEED, MODEL IS-12501 · Imagyn Medical Technologies, Inc. · Jun 17, 1999
Submission Summary (Full Text)
{0}------------------------------------------------
# SUMMARY OF SAFETY AND EFFECTIVENESS
# General Company Information
K990193
| Name: | Implant Sciences Corporation |
|----------------------------|--------------------------------------------------|
| Address: | 107 Audubon Road, #5<br>Wakefield, MA 01880-1246 |
| Telephone: | (781) 246 - 0700 |
| Fax: | (781) 246 - 1167 |
| General Device Information | |
| Product Name: | Implant Sciences "I-plant" 125Iodine Brachytherapy Seeds (Model 3000) |
|-----------------|-----------------------------------------------------------------------|
| Classification: | Brachytherapy Radionuclide Source, 21 CFR 892.5730 - Class II |
# Predicate Devices
North American Scientific, Inc. (NASI) loGold™ I-125 Brachytherapy Source (Model MED3631-A) [510(k) K9722711
Medi-Physics. Inc. (Amersham) |-125 Seeds (No. 6711) |510(k) K9142811
International Isotopes, Inc. Imagyn 125 lodine Brachytherapy Seed [510(k) K982421]
Theragenics Corp. (Nuclear Medicine, Inc.) TheraSeed® Palladium 103 Implants (Model 100) [510(k) K852542 and K874787]
# Description
The Implant Sciences 125 Iodine Brachytherapy Seeds consist of a laser welded titanium capsule containing an aluminum clad copper rod that serves as both a substrate for the radioactive iodine source and as a radiopaque x-ray marker. The marker identifies the source location and orientation. The seeds are provided non-sterile and must be sterilized prior to use.
# Intended Use
The Implant Sciences 125 lodine Brachytherapy Seeds with activities from 0.1 to 5.0 mCi are indicated for temporary or permanent interstitial or intracavity implantation or surface application to treat selected localized tumors. They can be used either as primary treatment (such as for prostate cancer or unresectable tumors) or as treatment for residual
{1}------------------------------------------------
disease after excision of primary or recurrent tumors. The 125 lodine Seeds may be used concurrently with or following treatment with other interventions, such as external beam therapy, hyperthermia or chemotherapy. Tumors of the head, neck, lung, pancreas, prostate and other accessible tumors are commonly treated.
# Technological Characteristics Versus Predicate Devices
The Implant Sciences 125 lodine Brachytherapy Seeds use the same titanium capsule material and are the same physical dimensions (length and diameter) as the predicate devices. The range of apparent activity levels is similar to that of the predicate devices. The results of safety testing in accordance with the ANSI Standard for Sealed Radioactive Sources (ANSI/HPS N43.6 - 1997) are consistent with the test results for the predicate devices.
# Substantial Equivalence
This 510(k) Notice supports the position that the Implant Sciences 145 lodine Brachytherapy Seeds are substantially equivalent in design and function to the North American Scientific, Inc. (NASI) loGold™ |-125 Brachytherapv Source (Model MED3631-A) [510(k) K972271], the Medi-Physics, Inc. (Amersham) I-125 Seeds (No. 6711) [510(k) K914281], the International Isotopes, Inc. Imagyn 125 Iodine Brachytherapy Seed [510(k) K982421] and the Theragenics Corp. TheraSeed® Palladium 103 Implants (Model 100) [510/k) K852542 which are brachytherapy radionuclides that have previously been cleared for marketing under the Premarket Notification regulations. Each of these devices has been placed in the same classification category (21 CFR 892.5730) as the ISC Brachytherapy Seeds; and each is appropriate for the same clinical application.
The 510(k) Notice contains summaries of both in vitro studies which were conducted to evaluate the safety, and appropriateness of the ISC Brachytherapy Seeds. Data are presented which demonstrate that the ISC Brachytherapy Seeds satisfy performance requirements for temperature resistance, external pressure resistance, impact resistance, vibration integrity, and puncture resistance as specified by ANSI Standard N43.6-1997. These test results confirm that the ISC 125 lodine Brachytherapy Seeds meet the requirements for radioactive sources in its class and are equivalent to other currently marketed radionuclide brachytherapy sources. The 510(k) Notice also describes further standardized studies which describe the radiation profile of the sources. These profiles are consistent with the data generated from other lodine-125 brachytherapy seeds with similar physical geometries. In addition, capsule leak testing was carried out in accordance with ANSI standards. The results of these studies again support the equivalence of the Implant Sciences device to the predicate devices.
Both the ISC device and the predicate devices are provided non-sterile and must be sterilized by autoclave or ethylene oxide at the hospital prior to use. The manufacturing and cleaning processes for the seeds introduce no new chemical or biological entities to the surface of the device.
{2}------------------------------------------------
Implant Sciences believes that the information provided establishes that similar legally marketed and pre-enactment devices have been used historically for the same types of clinical applications as the Implant Sciences 185 lodine Brachytherapy Seeds. The materials from which the Implant Sciences device is fabricated have an established history of use in medical applications, and the specific materials used by Implant Sciences have been thoroughly tested in accordance with applicable guidelines.
{3}------------------------------------------------
Rockville MD 20850
Food and Drug Administration 9200 Corporate Boulevard
MAY 20 1999
Howard Schrayer Implant Sciences Corporation 107 Audobon Road, No. 5 Wakefield, Massachusetts 01880-1246 Re:
K990193 125 Iodine Brachytherapy Seeds (Model 3000) Dated: April 13, 1999 Received: April 14, 1999 Regulatory Class: II 21 CFR 892.5730/Procode: 90 KXK
Dear Mr. Schrayer:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Dan Setit
APT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
Image /page/3/Picture/14 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized symbol that resembles a bird or a series of flowing lines, possibly representing the organization's mission related to health and well-being.
{4}------------------------------------------------
510(k) Number (if known):
K990193
Device Name: Implant Sciences 125 Iodine Brachytherapy Seeds - (Model 3000)
Indications For Use:
The Implant Sciences 125 Jodine Brachytherapy Seeds with activities from 0.1 to 5.0 mCi are indicated for temporary or permanent interstitial or intracavity implantation or surface application to treat selected localized tumors. They can be used either as primary treatment (such as for prostate cancer or unresectable tumors) or as treatment for residual disease after excision of primary or recurrent tumors. The 125 lodine Seeds may be used concurrently with or following treatment with other interventions, such as external beam therapy, hyperthermia or chemotherapy. Tumors of the head, neck, lung, pancreas, prostate and other accessible tumors are commonly treated.
#### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel A. Seymour
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices Kyaniya 510(k) Number
**Prescription Use**
(Per 21 CFR 801.109) √
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.