CADPLAN BT
Device Facts
| Record ID | K954467 |
|---|---|
| Device Name | CADPLAN BT |
| Applicant | Varian Assoc., Inc. |
| Product Code | JAQ · Radiology |
| Decision Date | Mar 15, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 892.5700 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
CADPLAN BT is designed to allow the dose distribution of an applicator to be individually shaped for each patient.
Device Story
CADPLAN BT is a Windows NT software application for brachytherapy treatment planning; operates in conjunction with VariSource High Dose Rate Remote Afterloader. Inputs include radiograph imaging (stereoshifted and orthogonal films) for applicator placement (up to 20 channels). Device calculates geometrically-optimal dwell times for catheters and shapes dose distribution per patient. Used by clinicians in oncology settings to facilitate radiation therapy planning. Output consists of optimized treatment plans and dwell time calculations, enabling precise radiation delivery to target tissues while sparing healthy anatomy.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Windows NT software application. Supports stereoshifted and orthogonal film imaging. Utilizes linear source model for IR-192. Features automated dwell time calculation algorithm versus manual calculation in predicate. Calculates dose volume histograms.
Indications for Use
Indicated for patients undergoing brachytherapy requiring individualized dose distribution shaping and dwell time calculation for the VariSource High Dose Rate Remote Afterloader.
Regulatory Classification
Identification
A remote controlled radionuclide applicator system is an electromechanical or pneumatic device intended to enable an operator to apply, by remote control, a radionuclide source into the body or to the surface of the body for radiation therapy. This generic type of device may include patient and equipment supports, component parts, treatment planning computer programs, and accessories.
Predicate Devices
- Varian VariSource Treatment Planning System (K945383)
Related Devices
- K081112 — FLEXIPLAN BRACHYTHERAPY TREATMENT PLANNING SYSTEM · Isodose Control B.V. · Jul 29, 2008
- K091145 — FLEXIPLAN 3D AND 3D+ · Isodose Control B.V. · May 27, 2009
- K123263 — HDRPLUS, ADD IN PROSTATE MODULE FOR HDRPLUS · Eckert & Ziegler Bebig GmbH · Apr 4, 2013
- K992762 — BRACHYVISION 6.0 · Varian Assoc., Inc. · Mar 9, 2000
- K141900 — SAGIPLAN, SAGIPLAN PROSTATE MODULE ADD ON, SAGIPLAN INCLUDING PROSTATE MODULE · Eckert & Ziegler Bebig GmbH · May 20, 2015