INTRABEAM NEEDLE APPLICATOR

K110590 · Carl Zeiss Surgical GmbH · JAD · Apr 13, 2012 · Radiology

Device Facts

Record IDK110590
Device NameINTRABEAM NEEDLE APPLICATOR
ApplicantCarl Zeiss Surgical GmbH
Product CodeJAD · Radiology
Decision DateApr 13, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5900
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Needle Applicator set (comprising the Needle Applicator and guide shaft) is indicated for use in combination with the INTRABEAM® System to intraoperatively administer radiation to tissue including irradiation of intracranial tumors.

Device Story

The INTRABEAM® System with Needle Applicator is an accessory to the existing INTRABEAM® X-ray radiation therapy system. It consists of a polycarbonate case, stainless steel guide shafts, and a polyetherimide tip. The device is used intraoperatively by physicians to create a tract and position an XRS 4 X-ray source directly at a target site, such as an intracranial tumor. The system delivers a controlled X-ray dose to the target tissue. The device is operated via the PRS500 Control Console, which manages radiation parameters. It benefits patients by enabling precise, localized intraoperative radiation delivery. No modifications were made to the core INTRABEAM® hardware or software.

Clinical Evidence

No clinical testing was conducted. Substantial equivalence is supported by bench testing, including sterilization validation per AAMI/ANSI/ISO 11137, biocompatibility testing per EN ISO 10993, and dosimetric performance testing (isotropy and dose depth curves) compared to the predicate.

Technological Characteristics

Materials: Polycarbonate case, stainless steel shafts, polyetherimide tip. Energy source: XRS 4 X-ray source. Sterilization: Radiation (AAMI/ANSI/ISO 11137-1:2006). Connectivity: Standalone system with PRS500 Control Console. Design: Accessory set for intraoperative positioning of X-ray source.

Indications for Use

Indicated for intraoperative radiation therapy of tissue, including intracranial tumors.

Regulatory Classification

Identification

An x-ray radiation therapy system is a device intended to produce and control x-rays used for radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # KI10580 # 510(k) Summary for APR 1 3 2012 # Carl Zeiss Meditec AG INTRABEAM® System Needle Applicator per 21 CFR 807.92 ## 1. SUBMITTER/510(K) HOLDER Manufacturer: Carl Zeiss Meditec AG Carl-Zeiss-Strasse 22 D-73447 Oberkochen Germany Contact Person: Judith A. Brimacombe, MA Director, Regulatory/Clinical Affairs Carl Zeiss Meditec Inc. 5160 Hacienda Drive Dublin, CA 94568 Phone: (925) 557-4616 FAX: (925) 557-4259 ### 2. DEVICE NAME Proprietary Name: Carl Zeiss INTRABEAM® System with Needle Applicator Common/Usual Name: X-ray radiation therapy system Classification Name: System, Therapeutic, X-ray, (21 CFR 892.5900) Product Code: IAD #### 3. PREDICATE DEVICES - . Photon Radiosurgery System PRS 400 System, K980526 - INTRABEAM® System with Balloon Applicator, most recently cleared in K090584 . # 4. DEVICE DESCRIPTION #### Physical Description The proposed device is a modification of the INTRABEAM® System with Balloon Applicator, K090584, incorporating a Needle Applicator and guide shafts that are used to open a tract for the positioning of the applicator for irradiation of turnors. The INTRABEAM® System that will be used with the new Needle Applicator accessory was most recently described in K090584. There are no modifications to the INTRABEAM® System hardware or software. The INTRABEAM® Core System {1}------------------------------------------------ consists of the following components: - . PRS500 Control Console - . XRS4 X-ray source - . User terminal Components of the Needle Applicator set are as follows: - 1) Inner Sterile Packaging - 2) 5 cm guide shaft - 3) 6 cm guide shaft - 4) Needle Applicator # How the Device Functions The XRS 4 X-ray source generates X-rays according to the parameters set for the required treatment on the PRS500 Control Console. The Needle Applicator allows radiation treatment to be delivered directly to the target (e.g. an intracranial turnor). # Scientific Concepts that form the Basis for the Device X-ray irradiation of tumors is a widely used treatment option in a variety of applications. The INTRABEAM® System delivers a controlled X-ray dose. # Significant Physical and Performance Characteristics of the Device, such as Device Design, Material Used, and Physical Properties The physical design of the Needle Applicator consists of a polycarbonate case which fits over the probe of the XRS 4 X-ray source, a stainless steel shaft, and a polyetherimide tip for the Applicator. The guide shafts are stainless steel. # 5. INTENDED USE The Needle Applicator set (comprising the Needle Applicator and guide shafts) is intended for use in combination with the INTRABEAM® System to intraoperatively administer radiation to tissue including irradiation of intracranial tumors. # 6. Summary of Technological Characteristics Compared to the\Predicate DEVICE Carl Zeiss Meditec AG claims substantial equivalence of the INTRABEAM® System with Needle Applicator to the cited predicate device based on the intended use, design, fundamental technology, and operation characteristics. {2}------------------------------------------------ # 7. SUMMARY OF NON-CLINICAL PERFORMANCE TESTING AS BASIS FOR SUBSTANTIAL EQUIVALENCE The testing summarized below supports the claim of substantial equivalence to the cited predicate by indicating that the Needle Applicator is a sterile device which performs as designed after sterilization and is safe for its intended use as is the predicate device. ## Sterilization Validation Sterilization validation of the Needle Applicator was performed in compliance with AAMI/ANSI/ISO 11137-1:2006, Sterilization of health care products - Radiation - Part 1: Requirements for development, validation, and routine control of a sterilization process for medical devices. Dose setting is performed in accordance with AAMI / ANSI/ISO 11137-2:2006, Sterilization of health care products - Radiation - Part 2: Establishing the sterilization dose, VDmax 25 method. Dosimetric release complies with AAMI/ANSI/ISO 11137-3:2006. Sterilization of health care products - Radiation - Part 3: Guidance on Dosimetric Aspects. The sterilization assurance level is 10 . # Performance Testing The testing of the dosimetry of sterilized and aged INTRABEAM® Needle Applicator was performed to demonstrate that sterilization and accelerated aging exert no adverse effect on the INTRABEAM® Needle Applicator. Five samples were inspected visually before and after sterilization and accelerated aging. The functionality of Needle Applicators subjected to sterilization and accelerated aging equivalent to five years was tested by Carl Zeiss Meditec AG. All applicators passed the tests, were fully functional, and met the requirements of the engineering specifications. Dosimetry testing of the PRS400 with the XRS probe (with sheath) compared to the INTRABEAM® System with Needle Applicator was conducted to demonstrate equivalence. The testing consists primarily of isotropy (radiation characteristics) and transfer functions (using a water phantom) using the same protocol used for past aging studies. The dose depth curves of the Needle Applicators tested matched those of the bare probe with sheath. The isotropy of the tested units met the criteria for passing the test. {3}------------------------------------------------ Biocompatibility Testing Table 5-2. Biocompatibility Testing Performed on the | Material | Test | Comments | Result | |----------------------|--------------------------------------|--------------------------------------------------------------------------|-----------------------------------------------------------| | Needle<br>Applicator | Cytotoxicity EN<br>ISO 10993-5 | L 929 cell cultures, quantitative<br>determination of cell proliferation | No relevant effects observed<br>in comparison to controls | | Guide shaft | Cytotoxicity EN<br>ISO 10993-5 | L 929 cell cultures, quantitative<br>determination of cell proliferation | No relevant effects observed<br>in comparison to controls | | Needle<br>Applicator | Chemical Analysis<br>EN ISO 10993-18 | GC-FID, quantification of organic<br>and inorganic leachables | No relevant effects observed<br>in comparison to controls | ## Final Device Components Based on the information and supporting documentation provided in the premarket notification, the Carl Zeiss INTRABEAM® Needle Applicator is substantially equivalent to the cited predicate devices. Testing demonstrates that the Carl Zeiss INTRABEAM® Needle Applicator fulfills prospectively defined design and performance specifications. # 8. SUMMARY OF CLINICAL TESTING AS BASIS FOR SUBSTANTIAL EQUIVALENCE No clinical testing was conducted to support this submission. ### 9. SUMMARY OF OTHER INFORMATION No other information is provided. #### 10. CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL TESTS Based on the information and supporting documentation provided in the premarket notification, the Carl Zeiss INTRABEAM® System with Needle Applicator is substantially equivalent to the cited predicate device. Testing demonstrates that the Carl Zeiss INTRABEAM® with Needle Applicator fulfills prospectively defined design and performance specifications. {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the department's name around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or bird-like figure. The symbol is composed of three curved lines that form the body and wings of the bird. # DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002 Carl Zeiss Meditec AG Ms. Judith A. Brimacombe. MA Director, Clinical/Regulatory Affairs Carl Zeiss Meditec, Inc. 5160 Hacienda Drive DUBLIN CA 94568 APR 1 3 2012 Re: K110590 Trade/Device Name: Carl Zeiss INTRABEAM® System with Needle Applicator Regulation Number: 21 CFR 892.5900 Regulation Name: X-ray radiation therapy system Regulatory Class: II Product Code: JAD Dated: March 2, 2012 Received: March 5, 2012 ## Dear Ms. Brimacombe: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of {5}------------------------------------------------ #### Page 2 medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely Yours, Janine M. Morris Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ # Indications for Use # 510(k) Number (if known): ヒル0590 Carl Zeiss INTRABEAM® System with Needle Applicator Device Name: Indications for Use: The Needle Applicator set (comprising the Needle Applicator and guide shaft) is indicated for use in combination with the INTRABEAM® System to intraoperatively administer radiation to tissue including irradiation of intracranial tumors. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) M (Division Sign-Off) Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Carl Zeiss Surgical GmbH, Traditional 510(k) INTRABEAM® System with Needle Applicator 510K CONFIDENTIAL Page 4-1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...