RAPTOR TREATMENT PLANNING SYSTEM

K993359 · Nomos Corp. · IYE · Dec 21, 1999 · Radiology

Device Facts

Record IDK993359
Device NameRAPTOR TREATMENT PLANNING SYSTEM
ApplicantNomos Corp.
Product CodeIYE · Radiology
Decision DateDec 21, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5050
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

The RAPTOR® Radiation Treatment Planning System is intended to be used by radiation oncology professionals for clinical review and judgement of radiation treatment plans and dose estimates based on computation, output and displays. The goal of the system is to produce consistent results using well-documented algorithms. It will provide outputs for radiation delivery equipment including documented and released multileaf collimators [MLCs] for planning, optimization, delivery and reporting of treatments. The device provides output data in the form of displays, hardcopy prints and / or plots, set-up instructions and other control information to guide the physician and / or other competent health care professional in selecting the optimum patient treatment plan.

Device Story

RAPTOR Treatment Planning System is a software suite for radiation oncology; generates radiation dose estimates and treatment plans. Inputs include user-controlled data; processes via documented algorithms; outputs include graphical displays, hardcopy reports, plots, and multileaf collimator (MLC) control instructions. Used in clinical settings by radiation oncologists, physicists, and dosimetrists to guide treatment planning and delivery. System runs on PC platform with Windows NT; supports modular upgrades. Benefits include consistent dose estimation and optimized treatment planning for radiation delivery equipment.

Clinical Evidence

No clinical data. Evidence based on bench testing, verification, and validation performed under NOMOS Corporation Quality System design controls. Testing included hazard analysis, design reviews, and verification against product specifications to ensure safety and effectiveness.

Technological Characteristics

Software-based treatment planning system; runs on PC platform with Microsoft Windows NT OS. Features modular architecture for dose estimation and MLC control. Year 2000 compliant. Utilizes table look-up formations for calculations. Connectivity via standard PC output (displays, printers, plotters).

Indications for Use

Indicated for use by radiation oncology professionals (radiation oncologists, medical physicists, radiation therapists, or dosimetrists) to generate, display, evaluate, and output radiation dose estimates and treatment instructions for clinical review prior to patient treatment.

Regulatory Classification

Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K993359 ## Section 2 510(k) Summary of Safety and Effectiveness RAPTOR® Treatment Planning System Pursuant to Section 513(I) of the Federal food. Drug, and Cosmetic Act and 21 CFR 807.92: | (a) (1) Submitter's name: | NOMOS Corporation | |---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------| | (a) (1) Submitter's address: | 2591 Wexford Bayne Road<br>Sewickley, PA 15143 | | (a) (1) Submitter's telephone number: | (724) 934-8200 | | (a) (1) Contact person: | William O. Chishko<br>Director, Quality and Regulatory Affairs | | (a) (1) Date summary was prepared: | October 4, 1999 | | (a) (2) Trade or proprietary name: | RAPTOR® Treatment Planning System | | (a) (2) Common or usual name: | Radiation Therapy Treatment Planning System | | (a) (2) Classification name: | Accelerator, Linear, Medical, Accessory<br>90 IYE [21 CFR 892.5050] | | (a) (3) Predicate device: | Radiation Oncology Computer Systems, Treatment<br>Planning System K982791<br>Theratronics, TheraPlan Plus K970236<br>NOMOS Corporation, PEACOCK® System K963258 | (a) (4) Device description: The RAPTOR Treatment Planning System is a collection of software modules that execute known and documented algorithms to produce radiation dose estimations. All data is user controlled and is a table look-up formation is presented graphically on CRT screens and hardcopy reports. Various models are available based on the specific features desired by the customer (e.q., asymmetric jaws, electron pencil beam calculations, etc.) to best meet their clinical needs. The software is designed to run on a PC platform utilizing the Microsoft Windows NT® operating system. The RAPTOR Treatment Planning System includes modules for the development of instructions and controls for multileaf collimators. It is designed to upgradable in both hardware and software features. All dates are four digit numbers so the system is year 2000 compliant. (a) (5) Intended use: The RAPTOR® Radiation Treatment Planning System is intended to be used by radiation oncology professionals for clinical review and judgement of radiation treatment plans and dose {1}------------------------------------------------ estimates based on computation, output and displays. The goal of the system is to produce consistent results using well-documented algorithms. It will provide outputs for radiation delivery equipment including documented and released multileaf collimators [MLCs] for planning, optimization, delivery and reporting of treatments. The device provides output data in the form of displays, hardcopy prints and / or plots, set-up instructions and other control information to guide the physician and / or other competent health care professional in selecting the optimum patient treatment plan. ## (a) (6) Technological characteristics: The RAPTOR® Treatment Planning System is designed to run on the Microsoft Windows NT operating system the same as the ROCS Treatment Planning System [K982791]. The RAPTOR® Treatment Planning System uses the same software architecture and hardware platform as the ROCS Treatment Planning System. It was developed and tested using Quality system design control and product release procedures. It differs in that a module for calculating the effects of multileaf collimators and providing instructions for their control like those found in PEACOCK® System is added. (b) (1) Non-clinical tests submitted: The RAPTOR® Treatment Planning System was developed and tested using the design controls set forth in the NOMOS Corporation Quality System. It has a product specification, which was used as the basis for the development of verification and validation plans, tests and acceptance criteria. A rigorous hazard analysis also was performed, which includes mitigation for each identified hazard. Design reviews were conducted at appropriate stages of development. (b) (3) Test Conclusions: Validation and verification testing of the RAPTOR® Treatment Planning System demonstrate that the software is safe and effective. Test completion shows that the device performs and is equivalent to the predicate systems for those features for which this submission is being made as well as overall performance. Product release also is made in accordance with established NOMOS Corporation Quality System Procedures after all tests are completed, reviewed and meet acceptance criteria and regulatory requirements. ainder of page is left blank on purpos {2}------------------------------------------------ Image /page/2/Picture/1 description: The image is a black and white circular logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling an emblem. The emblem is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans. Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 DEC 2 1 1999 William O. Chishko Director, Quality Affairs and Regulatory Assurance NOMOS Corporation 2591 Wexford Bayne Road Sewickley, PA 15143 Re: K993359 RAPTOR Treatment Planning System Dated: October 4, 1999 Received: October 6, 1999 Regulatory Class: II 21 CFR 892.5050/Procode: 90 IYE Dear Mr. Chishko: We have reviewed your premarket notification submission and have found this device to be exempt from the premarket notification requirements of the Federal Food, Drug, and Cosmetic Act (Act). Therefore, you may immediately begin marketing this device as described in your premarket notification. The Federal Register notice exempting your device type was published on January 21, 1998, Vol. 63, No. 13, page 3142, and was effective immediately. Therefore, manufacturers of devices falling within the above classification regulation are now exempt from the premarket notification requirements of the Act if they comply with the classification criteria. Your device's product code, classification regulation and regulatory class are shown above. When listing your device with the Food and Drug Administration, please use this product code. We suggest that you review this above referenced regulation since it may grant other exemptions from certain general controls of the Act. In the future, new but substantially equivalent devices which fall within the above classification regulation name and meet the classification criteria may be marketed without sending a premarket notification submission to the Food and Drug Administration. We suggest, however, that you review the section entitled "Limitations on Exemptions" in the above referenced Federal Register notice to determine whether or not your new device(s) meets the exemption criteria. This Federal Register notice may be accessed on the World Wide Web at "www.fda.gov/cdrh/modact/frclass2.html" or obtained by facsimile from the Division of Small Manufacturers Assistance's Facts On Demand at (800) 899-0381 or (301) 827-0111. The order number for this notice is #394. If you have any questions reqarding this letter, please contact Notification Staff at (301) 594-1190 or the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597. Sincerely yours, CAPT Daniel L.C. Schultz, MD CAPT Daniel G. Schultz, M Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health {3}------------------------------------------------ Page 1 of 1 Fage for 1 510(k) Number [if known]: 《993359 Device Name: RAPTOR® Treatment Planning System Indication for Use: The RAPTOR® Treatment Planning System is used to generate computation, display, evaluation and output radiation dose estimates for clinical review and judgement prior to treating the patient. It uses well-documented algorithms to provide output data in the form of displays, hardcopy prints, plots and treatment instructions to guide the radiation oncology professional and / or physician to select the optimum treatment plan. The MLC module is used to generate plans and instructions for documented and released multileaf collimators. The RAPTOR® Treatment Planning System is intended to be used by a competent health professional such as a radiation oncologist, medical physicist, radiation therapist, or dosimetrist. ## [PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE AS NEEDED] Concurrence of the CDRH Office of Device Evaluation [ODE] Prescription Use V (Per 21 CFR 801.109) OR Over the Counter Use (Optional Format , 2 January 1996 Urint h. Segers (Division Sign-Off) Division of Reproductive, Abdominal, El and Radiological D 510(k) Number
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...