EQUIDOSE SOLID STATE DIODE DETECTORS

K980826 · Cii Mastek, Inc. · IYE · Jun 1, 1998 · Radiology

Device Facts

Record IDK980826
Device NameEQUIDOSE SOLID STATE DIODE DETECTORS
ApplicantCii Mastek, Inc.
Product CodeIYE · Radiology
Decision DateJun 1, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5050
Device ClassClass 2

Indications for Use

EquiDose™ Solid State Diode Detectors are intended to be used for the collection of radiation beam data from radiation therapy treatment machines in air, water, or in other suitable material. The data acquired with this type of detector can be used to compile radiation beam data over time as part of a quality assurance program, to verify treatment machine energy and output, and to measure therapeutic doses delivered to patients during treatment.

Device Story

EquiDose™ Solid State Diode Detectors are passive radiation sensors used in radiation therapy. The device detects ionizing radiation and generates a current signal proportional to the dose. The detector is placed in the radiation field (in air, water, or phantom material) and connected via a 2-meter cable to an external electrometer (e.g., PTW-UNIDOS). The electrometer processes the current signal to quantify radiation dose. Used by medical physicists or radiation therapists in clinical settings to verify treatment machine output and ensure accurate therapeutic dose delivery to patients. The device is waterproof and features higher input impedance compared to the predicate.

Clinical Evidence

No clinical data. Bench testing only.

Technological Characteristics

Solid state diode detector; waterproof; 2-meter cable; high input impedance. Operates as a passive sensor requiring an external electrometer for signal processing. No internal power source or software.

Indications for Use

Indicated for the collection of radiation beam data from radiation-producing machines to verify treatment machine energy, output, and therapeutic doses delivered to patients during treatment.

Regulatory Classification

Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ JUN 1 1838 MasTek D.E.M., Inc. 1465 Cypress Avenue Melbourne, FL 32935 Phone 1-407-253-9800 Fax 1-407-253-0423 K980826 ## EquiDose™ Solid State Diode Detectors 510(k) Submission ### Manufacturer's 510(k) Summary, 21 CFR 807.92: - 1. Company: MasTek D.E.M., Inc. 1465 Cypress Avenue Melbourne, FL 32935 Contact: Jorge A. Fernández Senior Vice President MasTek D.E.M., Inc. (P) 1-407-253-9800 (F) 1-407-253-0423 Date of Submission: February 12, 1998 - Trade/Proprietary Name: 2. EquiDose™ Solid State Diode Detectors Common/Usual Name: Diode Detectors #### 3. Predicate Device(s): ISORAD™ Solid State Diode Detectors, which were cleared to market under 510(k) number K912250. #### 4. Description of Device(s): MasTek's EquiDose™ Solid State Diode Detectors are classical solid state diode detectors used for verification measurements in radiation therapy. All of the diodes share the same design, manufacturing process, are made of the same materials and they are all waterproof. With the exception of their higher input impedance and standard 2 meter cable length, the EquiDose™ Solid State Diode Detectors are exactly the same as the predicate ISORAD™ Solid State Diode Detectors. EquiDose™ Solid State Diode Detectors, when connected to an appropriate electrometer like the PTW-UNIDOS, K951764, are used to collect and verify beam data from radiation therapy treatment machines and to verify therapeutic amounts of radiation delivered during treatments. {1}------------------------------------------------ ## MasTek D.E.M., Inc. 1465 Cypress Avenue Melbourne, FL 32935 Phone 1-407-253-9800 Fax 1-407-253-0423 # EquiDose™ Solid State Diode Detectors 510(k) Submission #### 5. Statement of Intended Use: EquiDose™ Solid State Diode Detectors are intended to be used for the collection of radiation beam data from radiation therapy treatment machines in air, water, or in other suitable material. The data acquired with this type of detector can be used to compile radiation beam data over time as part of a quality assurance program, to verify treatment machine energy and output, and to measure therapeutic doses delivered to patients during treatment. ### 6. Comparison of Technological Characteristics to the Predicate Devices: The indications for use are exactly the same as the predicate devices, ISORAD™ Solid State Diode Detectors which were cleared to market by the FDA under K912250. The designs are exactly the same. The manufacturing and testing, process and procedures are exactly the same. The materials used are the same as in the predicate devices. The specifications are the same as the predicate devices. The indications for use, design, materials, manufacturing, and specifications of the EquiDose™ Solid State Diode Detectors do not raise any issues with regard to safety and effectiveness. MasTek D.E.M., Inc. considers all of these solid state diode detectors equivalent to the predicate devices for radiation therapy beam data measurements. - Note: Any statement made in conjunction with this Summary regarding substantial equivalence to another product was made in relation to the 510(k) premarket approval process and should not be interpreted as an admission or used as evidence in patient infringement litigation. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image is a black and white circular logo for the Department of Health & Human Services - USA. The logo features the department's seal, which includes a stylized representation of a human figure with three faces in profile, suggesting a sense of community and care. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the perimeter of the circle, framing the central image. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 ### JUN 1 1998 Jorge A. Fernandez Senior Vice President MasTek D.E.M., Inc. 1465 Cypress Avenue Melbourne, FL 32935 Re: K980826 EquiDose™ Solid State Diode Detectors Dated: February 12, 1998 Received: March 3, 1998 Regulatory class: II 21 CFR 892.5050/Procode: 90 IYE Dear Mr. Fernandez: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmaldsmamain.html". Sincerely yours, Kiliani Yin Lillian Yin, Ph.D. Director, Division of Reproductive Abdominal, Ear, Nose and Throat and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ 04/29/98 11:01 FAX 407 200 0420 477 01:28 297 201 180 4854 117: 58: 88 H.4.5. L. DEM LN CDRH DRAERD [✓] 02 [✓] 002 . Attn: Dr Chen Page al | <strong>Labels</strong> | <strong>Values</strong> | |---------------------------|--------------------------------------| | 510(k) Number (if known): | K980826 | | Device Name: | Equidose Solid State Diode Detectors | Indications For Use: The EquilDose™ Solid State Diode Detector is intended to be used for the collection of radiation beam data from radiation-producing machines. For use, simply place the EquiDose™ Detector inside the area where radiation shall be administered. The connector end of the detector should be attached to an electrometer that will receive the current signal produced by the detector. Through processing, this signal will then indicate the radiation quantity. PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Kimil A. Wagner (Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices 510(k) Number K480826 Pres ription Use (Per 21 CFR 801.109) OR Over-The-Counter Use (Optional Formal 1-2-96)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...