RPS-50 BEAM SCANNER
Device Facts
| Record ID | K962080 |
|---|---|
| Device Name | RPS-50 BEAM SCANNER |
| Applicant | Sofield Science Services, Inc. |
| Product Code | IYE · Radiology |
| Decision Date | Aug 23, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 892.5050 |
| Device Class | Class 2 |
Indications for Use
The RPS-50 is intended to be used as an accessory to a linear accelerator under the direct control of a qualified individual (usually a therapeutic radiation physicist.) The scanner may be used for acceptance testing of new accelerators, quality control of existing accelerators, or for the collection of data necessary for the operation of a computerized treatment planning system.
Device Story
RPS-50 Beam Scanner; accessory to clinical linear accelerator. System components: water tank, linear scanning guides, electronic movement control, chamber bias, system gain control. Operation: software-controlled movement of radiation detector through high radiation field; data collection; output as computer screen plots, data tables, or floppy disk files. Usage: clinical setting; operated by therapeutic radiation physicist. Output utility: data used for accelerator quality control and treatment planning system input; assists physicist in verifying radiation beam characteristics.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Water tank, linear scanning guides, electronic movement control, chamber bias, system gain control. Software-controlled signal acquisition. Output formats: screen plot, data table, floppy disk.
Indications for Use
Indicated for use by therapeutic radiation physicists for acceptance testing and quality control of clinical linear accelerators, and for collecting radiation beam data for computerized treatment planning systems.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
Predicate Devices
- A.R.M. beam scanner (K933558)
Related Devices
- K161807 — BEAMSCAN · Ptw-Freiburg Physikalisch Technische-Werksraetten · May 19, 2017
- K200907 — Thales 3D Mr Scanner · Lap GmbH Laser Applikationen · Jun 3, 2020
- K961400 — RFA-300, LDA UTILITY · Scanditronix Medical AB · Jan 3, 1997
- K191646 — Beamscan MR · Ptw-Freiburg Physikalisch-Technische-Werkstaetten · Feb 21, 2020
- K101992 — 3D SCANNER, MODEL 1230 · Sun Nuclear Corp. · Oct 1, 2010