K153270 · Bionix Development Corp. · IYE · Mar 8, 2016 · Radiology
Device Facts
Record ID
K153270
Device Name
TruGuard Custom Tongue and Jaw Positioner
Applicant
Bionix Development Corp.
Product Code
IYE · Radiology
Decision Date
Mar 8, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Indications for Use
The TruGuard Custom Tongue and Jaw Positioner from Bionix Development Corporation is intended to be used for repeat positioning and immobilization of patients undergoing or receiving a course of external beam radiation therapy for the treatment of cancer and other diseases. It is intended to be used by or under the direction of a licensed physician.
Device Story
TruGuard is a custom-fit oral positioning device for patients undergoing external beam radiation therapy. It consists of two rigid plastic dental bite trays connected in a bivalve fashion, lined with low-temperature EVA thermoplastic inserts. The device is used in a clinical setting under physician direction. To operate, the patient bites into softened EVA inserts to create a custom dental impression. A plastic indexing tab is then locked into the trays at pre-set angles to maintain jaw opening. An optional tongue depressor restricts tongue movement. The device is secured to the patient's head using a standard low-melt thermoplastic face mask, which is 'sandwiched' between the device and external retaining nuts. The device provides reproducible patient positioning and shields healthy oral tissue from secondary low-energy backscatter radiation caused by metal dental fillings or bridgework during high-energy photon beam treatment. This improves treatment precision and reduces unnecessary radiation exposure to non-target structures.
Clinical Evidence
Bench testing only. Shielding efficacy against backscatter radiation was evaluated using a 6 MeV photon beam and Gafchromic EBT3 film. The study measured the 'Dose Enhancement Ratio' (backscatter dose with device vs. baseline). Results demonstrated that the TruGuard device achieved a Dose Enhancement Ratio of approximately 1.0, indicating effective shielding of backscatter radiation compared to unprotected dental phantoms.
Technological Characteristics
Device consists of rigid plastic bite trays with low-temperature EVA (ethylene vinyl acetate) thermoplastic inserts. Features include a plastic indexing tab for jaw angle adjustment and an optional tongue depressor. The device is MRI safe. No specific ASTM standards cited. Non-powered, mechanical device.
Indications for Use
Indicated for patients undergoing external beam radiation therapy for cancer or other diseases requiring repeat positioning and immobilization of the jaw and tongue.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
Predicate Devices
Dental Tray for use with the Gill-Thomas-Cosman Relocatable Head Holder (K962155)
K232293 — kallisio stentra oral stent · Kallisio, Inc. · Dec 22, 2023
K200652 — Lollipop BiteBlock · Radtec Medical Devices, Inc. · Jul 6, 2020
K962155 — GILL-THOMAS-COSMAN(GTC) DENTAL TRAY · Radionics Software Applications, Inc. · Aug 22, 1996
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling a stylized eagle or bird, composed of three curved lines that suggest the profile of a human face.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 8, 2016
Bionix Development Corporation % James Huttner, M.D., Ph.D. Vice President, Product Development 5154 Enterprise Blvd. TOLEDO OH 43612
Re: K153270
Trade/Device Name: TruGuard Custom Tongue and Jaw Positioner Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: February 25, 2016 Received: March 1, 2016
Dear Dr. Huttner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael O'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# B. Revised Indications for Use :
| DEPARTMENT OF HEALTH AND HUMAN SERVICES | Form Approved: OMB No. 0910-0120 |
|-----------------------------------------|-----------------------------------|
| Food and Drug Administration | Expiration Date: January 31, 2017 |
| | See PRA Statement below. |
**Indications for Use**
| 510(k) Number (if known) | K153270 |
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | TruGuard Custom Tongue and Jaw Positioner |
| Indications for Use (Describe) | The TruGuard Custom Tongue and Jaw Positioner from Bionix Development Corporation is intended to be used for repeat positioning and immobilization of patients undergoing or receiving a course of external beam radiation therapy for the treatment of cancer and other diseases.<br>It is intended to be used by or under the direction of a licensed physician. |
| Type of Use (Select one or both, as applicable) | <label><input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Appended Documents:
- A. Revised 510(k) Summary:
- 510(k) Summary:
Image /page/3/Picture/3 description: The image shows the logo for Bionix. The logo consists of a blue circle with a white "B" inside, followed by the word "BIONIX" in blue, sans-serif font. A small, circled "R" is located to the right of the word "BIONIX", indicating that the name is a registered trademark.
Corporate Office 5154 Enterprise Blvd., Toledo, Ohio 43612
| Date of Application: | 11-4-2015 |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer: | Bionix Development Corporation<br>5154 Enterprise Blvd.<br>Toledo, Ohio 43612 |
| Contact Information: | James Huttner M.D., Ph.D.<br>Vice President, New Product Development<br>Phone: (419) 727-8421<br>Fax: (419) 727-4430<br>Email: jhuttner@bionix.com |
| Trade Name: | TruGuard Custom Tongue and Jaw Positioner |
| Common Name: | Bite Block (positionable) |
| Classification Name: | Accessory to Accelerator, Linear, Medical (Per CFR section 892.5050) |
## Intended Use:
The TruGuard Custom Tongue and Jaw Positioner from Bionix Development Corporation is intended to be used for repeat positioning and immobilization of patients undergoing or receiving a course of external beam radiation therapy for the treatment of cancer and other diseases.
## Predicate Devices:
- 1. Dental Tray for use with the Gill-Thomas- Cosman Relocatable Head Holder, a stereotactic head holder (for LINAC-based stereotactic radiotherapy), manufactured and legally sold by Integra, Plainsboro, NJ. This product is listed under regulation number 892.5050, Accessory to Accelerator, Linear, Medical
{4}------------------------------------------------
and is classified as a Class II device. This product has been approved for sale under K962155.
## Device Description:
## 1. Overview:
This pre-market notification is being submitted in good faith in an effort to satisfy the requirements of the FDAMA guidelines. Additionally, the FDA does not have any voluntary standards for this device and no standards have been applied. The TruGuard Custom Tongue and Jaw Positioner from Bionix Development Corporation has been designed using the design control procedures of the Bionix Quality System in compliance with the Good Manufacturing Practice (GMP) Quality System Regulations of the FDA/CDRH. This pre-market notification will draw information from the Design History File for this device and format it for efficient review.
## 2. Device Design and Principles of Operation.
Patients undergoing external beam radiation therapy for the treatment of oral cancer have unique requirements for positioning. Avoiding irradiation of noninvolved structures such as tongue and buccal mucosa are critical, as is shielding tissue from secondary low-energy backscatter emissions resulting from high energy photons striking metal fillings and dentures.
The TruGuard Custom Tongue and Jaw Positioner from Bionix is designed to provide reproducible opening of the jaw to multiple pre-set angles, with an optional tongue depressor to move the tongue out of the treatment field. It also provides shielding against secondary low-energy backscatter radiation emitted from metal fillings and bridgework.
The TruGuard Custom Tongue and Jaw Positioner consists of two dental bite trays composed of rigid plastic and connected at the base of their arcs in a bivalve fashion so as to allow the upper and lower bite trays to open and close in similar fashion as the upper and lower jaws of the patient. The two bite trays are each lined with an insert of low-temperature EVA (ethylene vinyl acetate) thermoplastic material similar to that used in commercial mouth guards. Projections on the lower and upper bite trays mate securely to a plastic indexing tab component that has holes at pre-set locations. When the plastic indexing tab is joined to the upper and lower bite trays the jaw is held open at pre-set angles, depending on the hole selected. This allows the mouth to be positioned and repositioned as needed, and ensures unrestricted air flow to the patient. An optional tongue depressor interlocks with the lower bite tray to restrict the tongue position to the lower portion of the oral cavity.
To use the TruGuard Custom Tongue and Jaw Positioner to position a patient, the following steps are taken:
{5}------------------------------------------------
- 1. The TruGuard Custom Tongue and Jaw Positioner is placed in hot water to soften the EVA Inserts. The patient then bites down to create dental impressions of their teeth. This allows accurate repositioning of the dental trays between treatments.
- 2. The tongue depressor is locked into place on the lower bite tray, if indicated.
- 3. The plastic indexing tab is aligned over the projections on the bite trays and locked into place, holding the mouth at the desired open angle.
- 4. A standard low-melt thermoplastic face mask is formed over the patient's head with the TruGuard Custom Tongue and Jaw Positioner in place. The upper bite tray of the TruGuard has two threaded posts that can be pushed through the low-melt thermoplastic while it is still soft and pliable. After the low-melt thermoplastic mask cools and hardens, a soft spacer is fitted over the posts and retaining nuts are then screwed onto the posts so as to "sandwich" the low-melt thermoplastic mask between the TruGuard Custom Tongue and Jaw Positioner (on the inside of the mask) and the retaining nuts (on the outside of the mask), holding the TruGuard Custom Tongue and Jaw Positioner in position.
- 5. After the treatment session, the TruGuard Custom Tongue and Jaw Positioner may be detached from the face mask and cleaned and stored until the next therapy session. The individualized nature of the dental impressions ensure that a particular TruGuard Custom Tongue and Jaw Positioner cannot be used for any other patient.
## Comparison to Predicate Device(s):
The TruGuard Custom Tongue and Jaw Positioner from Bionix Development Corporation is similar in design and function to existing bite block systems currently in use as accessories to radiation therapy systems.
One of those systems is K962155, the Dental Tray for use with the Gill-Thomas-Cosman Relocatable Head Holder manufactured and legally sold by Integra, Plainsboro, NJ. The Dental Tray holds an impression of the upper teeth, and is part of the GTC Relocatable Head Holder in stereotactic collimated beam, computer planned, LINAC (linear accelerator) based treatment. The Dental Tray holds the dental impression and allows repeat positioning of the stereotactic frame at the same location.
The Dental Tray for the GTC Relocatable Head Holder uses standard dental impression material to create the impression of the upper teeth. One example of commercially available dental impression material is Vinyl Polysiloxane,
{6}------------------------------------------------
tradename Reprosil®. Reprosil® is a hydrophilic vinyl polysiloxane impression material that provides excellent dimensional accuracy and stability for a variety of crown, bridge and edentulous impressions. It forms a putty-like material that is pressed into the GTC Dental Tray; the patient then bites down to form an impression of the upper teeth, then the material is allowed to harden. Once hardened, the GTC Dental tray allows for accurate patient repositioning for each external radiotherapy treatment.
The following table has been prepared comparing the similarities and differences of the Dental Tray for the GTC Relocatable Head Holder and the TruGuard Custom Tongue and Jaw Positioner.
{7}------------------------------------------------
## Table 1 Comparison of Substantial Equivalence Dental Tray for Use with the Gill-Thomas-Cosman (GTC) Relocatable Head Holder System (Integra) and TruGuard Custom Tongue and Jaw Positioner (Bionix Development Corporation)
| Attribute | Dental Tray for use with the<br>Gill-Thomas-Cosman<br>Relocatable Head Holder | Bionix TruGuard Custom Tongue and<br>Jaw Positioner |
|--------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|
| Intended Use | Holds a dental impression to allow<br>repeat positioning of a patient<br>receiving external beam radiation<br>therapy | Repeat positioning and<br>immobilization of patients<br>receiving external beam radiation<br>therapy |
| Reproducible Patient<br>Positioning | Yes | Yes |
| FDA Classification | Class II | Class II |
| Composition | Molded Plastic Dental Tray<br>(Attaches to Metal GTC Frame) | Molded Plastic |
| Dental Impression<br>Material | Dental Impression Material<br>(Vinyl Polysiloxane ) | Moldable EVA Thermoplastic |
| Dual Bite Trays | No<br>(Upper Bite Tray Only) | Yes |
| Allows for Open Jaw<br>Positioning | No | Yes |
| Positions the Tongue | No | Yes |
| Attaches to Face Mask | No | Yes |
| Provides Shielding from<br>Backscatter Radiation | No Data | Yes |
| MRI Safe | Yes<br>(Dental Tray Only) | Yes |
{8}------------------------------------------------
## Bench Testing:
Bench testing was performed on the TruGuard Custom Tongue and Jaw Positioner to determine shielding efficacy against backscatter radiation produced by high-energy photons striking metal fillings and bridgework.
Test subjects were exposed to a 6 MeV high energy photon beam, and the backscatter radiation dose was evaluated by measuring the exposure density of industry standard Gafchromic EBT3 film placed adjacent to the test subject. Results were plotted as "Dose Enhancement Ratio" vs. "Position Across Teeth". The "Dose Enhancement Ratio" was defined as the backscatter radiation dose produced by the test subject divided by a "Baseline" dose (backscatter radiation produced with no test subject in the photon beam).
Shielding efficacy was indicated by the Dose Enhancement Ratio. Values greater than 1 indicate an increased dose of backscatter radiation to the patient and a poor shielding effect, whereas a ratio of 1 indicates total shielding with no backscatter radiation dose delivered to the patient.
Figure 13 shows the results of this bench testing using various combinations of the phantom (teeth with amalgam fillings) and TruGuard components. The plot showing the Dose Enhancement Ratio of only the phantom (teeth + filling) in the photon beam shows a spike over all teeth, with the greatest Dose Enhancement over the center tooth having the metal amalgam filling, as expected. The plots showing the Dose Enhancement Ratios for the phantom when covered by the EVA insert material only, the Dental Tray material only, and the assembled TruGuard are effectively 1.0, showing near total shielding efficacy of the TruGuard device in shielding against backscatter radiation.
{9}------------------------------------------------
Image /page/9/Figure/0 description: The image is a graph comparing the dose enhancement ratio of teeth with different types of dental protection. The x-axis represents the position across the teeth in millimeters, ranging from 20 to 80 mm. The y-axis represents the dose enhancement ratio, ranging from 0.9 to 1.5. The graph shows that teeth with no protection have the highest dose enhancement ratio, while teeth with EVA insert, dental tray, and TruGuard have lower dose enhancement ratios.
Figure 13 Dose Enhancement Ratio Plot Demonstrating TruGuard Shielding
#### Conclusion:
The similarity of design, features, composition, and function indicate that TruGuard Custom Tongue and Jaw Positioner from Bionix Development Corporation will perform as well as the legally marketed Dental Tray for use with the Gill-Thomas- Cosman Relocatable Head Holder, manufactured and legally sold by Integra, Plainsboro, NJ for the intended use for repeat positioning and immobilization of patients undergoing or receiving a course of external beam radiation therapy for the treatment of cancer and other diseases.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.