WaterProof PROFILER

K142431 · Sun Nuclear Corporation · IYE · Dec 5, 2014 · Radiology

Device Facts

Record IDK142431
Device NameWaterProof PROFILER
ApplicantSun Nuclear Corporation
Product CodeIYE · Radiology
Decision DateDec 5, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5050
Device ClassClass 2

Indications for Use

The Model 1203 WaterProof PROFILER is intended for radiotherapy dosimetry measurements for commissioning a treatment planning system (TPS) computer. It is also intended for periodic beam quality assurance (QA) tests as defined by the medical physicist responsible for the QA program.

Device Story

Model 1203 WaterProof PROFILER is a linear array of 127 radiation detectors housed in a waterproof polycarbonate enclosure. It mounts to a 3D SCANNER movement mechanism to measure ionizing radiation dose in a water bath. The device captures charge produced in detectors; control electronics and embedded code manage data transmission to the 3D SCANNER. The 3D SCANNER correlates dose with motor position and transfers data to a computer running SNC Dosimetry software. The software provides a graphical user interface for viewing dose distributions, performing parameter analysis (field edge, field size, beam center, penumbra, wedge angles, symmetry, flatness), and exporting data to treatment planning systems. Used by medical physicists for commissioning treatment planning systems and periodic quality assurance of radiation delivery devices. Benefits include efficient, accurate relative dose measurements for radiotherapy beam characterization.

Clinical Evidence

Bench testing only. No clinical data. Testing compared static and dynamic field measurements against predicate devices during treatment planning system commissioning. Results demonstrated correlation between the WaterProof PROFILER and predicates. Compliance with electrical safety and EMC standards was verified.

Technological Characteristics

Linear array of 127 diodes; 504mm array length; 4mm detector spacing; 0.8x0.8mm detector area; 32nC/Gy sensitivity. Housed in waterproof polycarbonate phantom at 0.8 g/cm2 depth. Connects to 3D SCANNER for positioning and data acquisition. 100ms sampling rate. Embedded firmware manages data transmission.

Indications for Use

Indicated for radiotherapy dosimetry measurements and periodic beam quality assurance tests in clinical radiotherapy settings, performed by medical physicists.

Regulatory Classification

Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 December 5. 2014 Sun Nuclear Corporation % Ms. Noel Downey Project Manager 3275 Suntree Boulevard MELBOURNE FL 32940 Re: K142431 Trade/Device Name: Model 1203 - WaterProof Profiler Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: November 5, 2014 Received: November 10, 2014 Dear Ms. Downey: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Michael D. O'Hara for Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K142431 Device Name Model 1203 - Waterproof Profiler ## Indications for Use (Describe) The Model 1203 WaterProof PROFILER is intended for radiotherapy dosimetry measurements for commissioning a treatment planning system (TPS) computer. It is also intended for periodic beam quality assurance (OA) tests as defined by the medical physicist responsible for the QA program. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) _ Over-The-Counter Use (21 CFR 801 Subpart C) ## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. ## FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for Sun Nuclear Corporation. The logo consists of a blue square with a yellow sun inside on the left, and the words "SUN NUCLEAR" in blue on the right. Below the words "SUN NUCLEAR" is the word "corporation" in a smaller blue font. The logo is simple and modern, and it conveys the company's focus on nuclear energy. # Section 5 - 510(k) Summary Provided in accordance with 21 CFR 807.92 (c) #### 1 General Provisions Date Prepared: August 27, 2014 Submitted by: Sun Nuclear Corporation 3275 Suntree Blvd. Melbourne, FL 32940 Ph: 321-259-6862 Fax: 321-259-7979 Web: www.sunnuclear.com Contact Person: Noel Downey noeld@sunnuclear.com PH: 321-259-6862 ext. 242 Classification Name: Accelerator, Linear, Medical Common Name: Detector Array Proprietary Names: Model 1203 WaterProof PROFILER Establishment Registration Number: 1038814 Classification: Regulation Number: 21 CFR 892.5050 Name: Medical charged-particle radiation therapy system Product code: IYE Primary Predicate Device: | Model Name: | Model 1174 PROFILER 2 | |---------------|--------------------------| | Common Name: | Radiation Field Analyzer | | 510(k) #: | K063021 | | Manufacturer: | Sun Nuclear Corporation | | Submitted: | 10/2/2006 | {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for Sun Nuclear Corporation. The logo consists of a blue square with a yellow sun inside on the left. To the right of the image is the text "SUN NUCLEAR" in blue, with the word "corporation" underneath in a smaller font. Secondary Predicate Device: Model Name: Common Name: 510(k) # Manufacturer: Submitted: Model 1230 3D SCANNER Water Phantom Ionizing Radiation Dosimetry Scanner K101992 Sun Nuclear Corporation July 13, 2010 #### Description and Use: 2 The WaterProof PROFILER, model 1203, is a linear array of radiation detectors that are housed in a water proof enclosure that mounts to the 3D SCANNER's movement mechanisms. The electronics in the array measures charge produced in the detectors due to ionizing radiation. The WaterProof PROFILER provides measurement updates to the 3D SCANNER. The 3D SCANNER correlates the dose information with motor position information and transfers the data to a computer running SNC Dosimetry. SNC Dosimetry then provides a graphical user interface for viewing the measured dose distributions and its parameter analysis; it also provides a tool for exporting the data to treatment planning systems. The WaterProof PROFILER includes five components: - 1. a linear array of radiation detectors; - 2. a water proof enclosure; - 3. a means for connecting to the 3D SCANNER's movement mechanisms; - 4. control electronics and embedded code to manage and transmit the data recorded from the radiation detectors; - 5. a cable that connects the water proof enclosure to the 3D SCANNER The WaterProof PROFILER has applications in: - 1. delivery device data collection for commissioning a treatment planning system, - 2. quality assurance measurements of a treatment delivery device, and - 3. testing of a treatment delivery device at the manufactures facility. To satisfy the applications above, the WaterProof PROFILER is capable of performing relative dose measurements, which may then be used to calculate QA parameters such as: field edge, field size, beam center, penumbra, wedge angles (dynamic and physical), symmetry and flatness. #### Intended Use Statements: 3 The Model 1203 WaterProof PROFILER is intended for radiotherapy dosimetry measurements for commissioning a treatment planning system (TPS) computer. It is also intended for periodic beam quality assurance (QA) tests as defined by the medical physicist responsible for the QA program. {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the logo for Sun Nuclear Corporation. The logo consists of a blue square with a yellow sun-like design inside on the left. To the right of the square, the words "SUN NUCLEAR" are written in large, bold, blue letters, with the word "corporation" written in smaller, blue letters below. #### Technological Characteristics ব The primary technological characteristic of the WaterProof PROFILER is the array geometry, which facilitates measurement of static or dynamic dose deliveries in most beam configurations. The array length is 504mm and detector spacing is 4 mm. The 1D array of diodes is embedded in a waterproof polycarbonate phantom at a depth of 0.8 g/cm2. Connection to the 3D SCANNER (K101992) provides remote control of positioning in a water bath as well as data acquisition in 100 ms update periods. #### Performance Data and Comparison with Predicate 5 The WaterProof PROFILER has been bench tested and shown that these devices perform within their design specifications. Tests that compare known static and dynamic fields required during treatment planning system commissioning have been performed; the results were found to have correlation between the WaterProof PROFILER and its predicate devices. Performance testing also indicated compliance with relevant electrical safety and EMC standards. | | WaterProof<br>PROFILER | PROFILER 2 | 3D SCANNER | |----------------------|------------------------|------------------------|--------------------------------------| | Number of Arrays | 1 | 2 | 0 | | Array Length | 50.4cm | X: 22.4cm<br>Y: 32.8cm | 50cm | | | | | | | Detectors per Array | 127 | X: 57<br>Y: 83 | 1 | | | | | | | Detector Density | 4mm/detector | 4mm/detector | 1 detector translated<br>across 50cm | | Detector Area | 0.8x0.8mm | 0.8x0.8mm | User definable | | Detector Sensitivity | 32nC/Gy | 32nC/Gy | User definable | | Buildup | 1.0 g/cm2 | 0.8 g/cm2 | User definable | | Waterproof | Yes | No | User definable | | Sampling rate | 100ms | 100ms | 50ms | Table of specification differences between WaterProof PROFILER, PROFILER 2 and the 3D SCANNER #### Summary 6 The Model 1203 WaterProof PROFILER is deemed substantially equivalent to the predicate devices. The intended use, performance testing, safety and effectiveness reviews demonstrate that these devices are as safe, as effective, and perform as well or better than the predicate device. The minor technological differences between the WaterProof PROFILER and the predicates do not raise new types of safety or effectiveness questions.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...