MILLENNIUM MULTI-LEAF COLLIMATOR

K050442 · Varian Medical Systems · IYE · Mar 23, 2005 · Radiology

Device Facts

Record IDK050442
Device NameMILLENNIUM MULTI-LEAF COLLIMATOR
ApplicantVarian Medical Systems
Product CodeIYE · Radiology
Decision DateMar 23, 2005
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.5050
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Varian Millennium Multileaf Collimator (MLC) is an accessory X-ray collimator designed to be mounted on a Varian Clinac radiotherapy linear accelerator and is intended to shape the X-ray field perimeter. Field shape can be either static (fixed) or dynamic. Dynamic field shapes are controlled as function of Clinac dose fraction or gantry angle. The intended use is to assist the radiation oncologist in the delivery of radiation to defined target volumes while sparing surrounding normal tissue and critical organs from excess radiation.

Device Story

The Millennium Multi-leaf Collimator (MLC) is an accessory mounted on Varian Clinac radiation therapy linear accelerators. It shapes the X-ray field perimeter to match target volumes. The device operates in static mode (replacing customized shadow blocks) or dynamic mode (leaf positions indexed to Clinac dose fraction or gantry angle). This modification (v7.0) enables large field IMRT treatments, smoother motion control, and faster plan downloads via ethernet. Operated by radiation oncology staff in a clinical setting, the device allows for precise 3D dose distribution, sparing healthy tissue and critical organs from excess radiation.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

X-ray collimator accessory for Varian Clinac linear accelerators. Features multi-leaf design for static or dynamic beam shaping. Connectivity upgraded to ethernet for faster plan downloads. Motion control system optimized for smoothness. Software-controlled leaf positioning based on dose fraction or gantry angle.

Indications for Use

Indicated for use by radiation oncologists to assist in the delivery of radiation to defined target volumes while sparing surrounding normal tissue and critical organs. Used for patients requiring radiation therapy where static or dynamic beam shaping is necessary to create three-dimensional dose distributions.

Regulatory Classification

Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K050442 Page 1 of 2 MAR 2 3 2005 ## 510(k) Summary The following information is provided following the format of 21 CFR 807.92 for the Millennium Multi-leaf Collimator Varian Medical Systems Submitter: 3100 Hansen Way M/S E-110 Palo Alto, CA 94304-1129 Contact Name: Vy Tran Phone: (650) 424-5731 (650) 842-5040 Fax: Email: vy.tran@varian.com Date summary was prepared: February 17, 2005 ## 1. Name of the Device: | Trade/Proprietary Name: | Millennium Multi-Leaf Collimator | |-------------------------|--------------------------------------------------------------------------------------------------------------| | Common or Usual Name: | Accessory to Medical Linear Accelerator | | Classification Name: | Medical Charged Particle Radiation Therapy<br>System<br>21 CFR §892.5050<br>Class II<br>Product Code: 90 IYE | ## 2. Predicate Devices to claim substantial equivalence: a. Varian Medical Systems' Millennium Multi-leaf Collimator (K990085) ## 3. Device Description: Millennium Multi-Leaf Collimator (MLC) Release v7.0 is a modification of the present MLC device (K990085) to include the following: - 1. Large Field IMRT treatments - Smooth motion control 2. - 3. Faster patient plan downloads via ethernet The Varian Millennium Multileaf Collimator (MLC) is an x-ray collimator designed to be mounted on a Varian Clinac® radiation therapy linear accelerator and is intended to shape the x-ray field perimeter. Field shape can either be static (fixed) or dynamic. Dynamic field shapes are controlled as a function of either Clinac® dose fraction or gantry angle. - 5. Intended Use Statement: The Varian Millennium Multileaf Collimator (MLC) is an accessory X-ray collimator designed to be mounted on a Varian Clinac radiotherapy linear accelerator and is intended to shape the X-ray field perimeter. Field shape can be either static (fixed) or dynamic. Dynamic field shapes are {1}------------------------------------------------ K050442 Page 2 of 2 controlled as function of Clinac dose fraction or gantry angle. The intended use is to assist the radiation oncologist in the delivery of radiation to defined target volumes while sparing surrounding normal tissue and critical organs from excess radiation. - 6. Summary of the Technological Characteristics: The Substantial Equivalence Comparison Chart provides a comparison of the technological characteristics to those of the predicate devices. This chart is located in Tab 8 of the submission. {2}------------------------------------------------ Image /page/2/Picture/12 description: The image shows a seal for the Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES" around the perimeter. In the center of the seal is an abstract image of an eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAR 2 3 2005 Ms. Vy Tran Corporate Director, Regulatory Affairs Varian Medical Systems, Inc. 3100 Hansen Way PALO ALTO CA 94304-1038 Re: K050442 Trade/Device Name: Millenium 120 MLC Regulation Number: 21 CFR §892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: February 17, 2005 Received: February 22, 2005 Dear Ms. Tran: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050. {3}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) r ms letter natification. The FDA finding of substantial equivalence of your device to a legally promated predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you does office of Compliance at one of the following numbers, based on the regulation number at the top of this letter: | 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 | |-----------------|----------------------------------|--------------| | 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 | | 21 CFR 892.xxxx | (Radiology) | 240-276-0120 | | Other | | 240-276-0100 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmadsmam.html Sincerely yours, Nancy C. brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ 510(k) Number (if known): K050442 Device Name: Millennium120 MLC Indications For Use: The Varian Millennium 120 Multi-leaf Collimator (MLC) is provided to assist the radiation oncologist in the delivery of radiation to defined target volumes while sparing surrounding normal tissue and critical organs from excess radiation. In static mode, the MLC performs the same function as the customized shadow blocks. In a dynamic mode, a series of MLC leaf positions can be indexed to either Clinac dose fraction or gantry angle to create a changing beam shape while the radiation beam is on to create a three dimensional dose distribution. (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Optional Format 3-10-98) David in Lignon 1 Prescription Use
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...