The 2-D AUTIMO system is intended for the fabrication and positioning in radiotherapy of individualized (irregular) shielding blocks.
Device Story
Computer-controlled cutting tool for radiotherapy shielding blocks; accepts digital files from treatment-planning systems via LAN; interprets planning data to drive hot-wire-cutting mechanism; produces styrofoam molds for lung blocks and irregular sections; used in clinical radiotherapy settings; operated by trained personnel; output facilitates creation of patient-specific radiation shielding to improve treatment accuracy.
Clinical Evidence
Bench testing only; no clinical data provided. Testing addressed electrical safety and electromagnetic compatibility.
Technological Characteristics
Computer-controlled hot-wire cutting system for styrofoam molds. Connectivity via serial port to LAN-based treatment-planning systems. Hardware-based modification of predicate device.
Indications for Use
Indicated for the fabrication and positioning of individualized (irregular) shielding blocks for use in radiotherapy.
Regulatory Classification
Identification
A radiation therapy beam-shaping block is a device made of a highly attenuating material (such as lead) intended for medical purposes to modify the shape of a beam from a radiation therapy source.
K011845 — MILLCOMP · Math Resolutions, LLC · Dec 4, 2001
Submission Summary (Full Text)
{0}
JUN 13 1997
K971081
# 510(k) Summary
# HEK
Common/Classification Name: Radiation Therapy Beam-Shaping Device, 21 CFR 892.5710
Sponsor: HEK Medizintechnik GmgH
Kaninchenborn 24-28
Postfach 1832
D-23560 Lübeck
GERMANY
Tel: (+49) 451 5300444
FAX: (+49) 451 5300450
Contact: Reinhard Schröder
Prepared: March 13, 1997
# A. LEGALLY MARKETED PREDICATE DEVICES
The AUTIMO 2-D System is substantially equivalent to its predecessor device currently marketed by HEK Medizintechnik, the MCP-70-SE (K844180).
# B. DEVICE DESCRIPTION
The AUTIMO 2-D model is a computer-controlled cutting tool to produce customized shielding blocks for use in radiotherapy. The computer accepts files from existing treatment-planning system operating on a LAN. The GE-Target, Siemens Helax, and Theratronics Theraplan systems are examples of such treatment-planning software. The computer interprets the files from the planning system and communicates with the AUTIMO device through a serial port.
The 2-D model allows automatic hot-wire-cutting of styrofoam molds such as lung blocks and irregular sections.
# C. INTENDED USE
The 2.0-D AUTIMO System is intended for the fabrication and positioning in radiotherapy of irregular shielding blocks.
000021
{1}
D. TECHNOLOGICAL CHARACTERISTICS
The 2.0-D AUTIMO System has the same technological characteristics as the predicate device. The AUTIMO is simply a modification with updated hardware and cosmetic changes from the predicate device.
E. TESTING
HEK Medizintechnik carried out testing to address the following issues:
(1) electrical safety
(2) electromagnetic compatibility
The results from these tests supported the safety and effectiveness of the 2.0-D AUTIMO System and demonstrates that it is substantially equivalent to the predicate device.
F. CONCLUSIONS
The HEK AUTIMO device has the same intended use and target population as the predicate device. HEK has demonstrated through its performance tests on the 2.0-D AUTIMO systems and its comparison of AUTIMO characteristics with those of the predicate device that the 2.0-D AUTIMO system is substantially equivalent to the predicate device.
000022
{2}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
HEK Medizintechnik, GmbH
c/o T. Whit Athey, Ph.D.
Senior Consultant
C. L. McIntosh & Associates, Inc.
12300 Twinbrook Parkway, Suite 625
Rockville, MD 20852
Re: K971001
JUN 13
GmbH AUTIMO 2-D Dose Modification System
Dated: March 19, 1997
Received: March 19, 1997
Regulatory Class: II
21 CFR 892.5710/Procode: 90 IXI
Dear Dr. Athey:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591 for Radiology devices, or 594-4613 for Ear, Nose and Throat devices. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,

Lillian Yin, Ph.D.
Director, Division of Reproductive,
Abdominal, Ear, Nose and Throat,
and Radiological Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{3}
# STATEMENT OF INDICATIONS FOR USE
510(k) Number (if known): _________________________
Device Name: AUTIMO 2-D
Indications For Use:
* The 2-D AUTIMO system is intended for the fabrication and positioning in radiotherapy of individualized (irregular) shielding blocks.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David G. Segura
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K971001
Prescription Use ☑ (Per 21 CFR 801.109)
OR
Over-The-Counter Use
000006
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.