K081066 · Medi-Physics, Inc., Dba GE Healthcare · IWG · May 5, 2008 · Radiology
Device Facts
Record ID
K081066
Device Name
BRACHYTHERAPY SOURCE DEVICE, MODEL 9011
Applicant
Medi-Physics, Inc., Dba GE Healthcare
Product Code
IWG · Radiology
Decision Date
May 5, 2008
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.5730
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Brachytherapy Source Device, with apparent activities between 0.19mCi and 1.01mCi, is indicated for permanent interstitial implantation of selected localized turnors which are of low to moderate radiosensitivity. They may be used either as primary treatment (such as for prostate cancer or unresectable tumors) or for treatment of residual disease after excision of the primary tumor. Seeds in this apparent activity range may be used to treat superficial, intra-abdominal, and intra thoracic tumors. Tumors of the head, neck, lung, pancreas, and prostate (early stages) are commonly treated. The Brachytherapy Source Device, with total apparent activities greater than 1.01mCi, is indicated for interstitial treatment of tumors which have the following characteristics; unresectable, localized, and moderate radiosensitivity. These seeds may be used for selected radiation applications as temporary implants. The Brachytherapy Source Device is indicated to treat residual turnors concurrent with or at the completion of other treatment modalities, such as external beam radiation therapy or chemotherapy. In addition, recurrent tumors may be implanted with seeds.
Device Story
Brachytherapy Source Device (Model 9011) is a radioactive seed for interstitial tumor treatment. Device consists of a welded titanium capsule containing Iodine-125 adsorbed onto a silver rod. Seeds are implanted into localized tumors (e.g., prostate, lung, head/neck) by clinicians. Device provides localized radiation therapy to treat primary, residual, or recurrent tumors. Operates via radioactive decay of I-125. Benefits include targeted radiation delivery to tumor sites while sparing surrounding healthy tissue. Device is provided non-sterile and requires user sterilization prior to implantation.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including ISO 2919 testing for sealed sources and NIST air-kerma rate measurements, alongside design verification and validation against predetermined criteria.
Indicated for permanent interstitial implantation or temporary interstitial treatment of localized, unresectable tumors of low to moderate radiosensitivity, including head, neck, lung, pancreas, and early-stage prostate cancer; used as primary treatment, for residual disease post-excision, or concurrent with external beam radiation/chemotherapy; also for recurrent tumors.
Regulatory Classification
Identification
A radionuclide brachytherapy source is a device that consists of a radionuclide which may be enclosed in a sealed container made of gold, titanium, stainless steel, or platinum and intended for medical purposes to be placed onto a body surface or into a body cavity or tissue as a source of nuclear radiation for therapy.
Special Controls
*Classification.* Class II (special controls). A prostate seeding kit intended for use with a radionuclide brachytherapy source only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
K994317 — I-PLANT, MODEL 3500 · Implant Sciences Corp. · Mar 21, 2000
K992963 — I3RAD 125 I BRACHYTHERAPY SEED, MODEL I125-0401 · International Isotopes, Inc. · Sep 21, 1999
K991526 — ISOSTAR I-125 INTERSTITIAL SEED, MODEL IS-12501 · Imagyn Medical Technologies, Inc. · Jun 17, 1999
K984235 — INTERSOURCE 125, MODEL 1251L · International Brachytherapy, Inc. · Jun 9, 1999
K023242 — I-PLANT MODEL 3500 (125IODINE BRACHYTHERAPY SEEDS)-STERILE · Implant Sciences Corp. · Dec 4, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
K081066
MAY 15 2008
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
The following 510(k) Summary of Safety and Effectiveness is prepared and provided in accordance with the requirements of 21 CFR 807.92 as amended under the Safe Medical Devices Act of 1990 (SMDA).
## Submitter's Information
| Company Name: | Medi-Physics, Inc DBA GE Healthcare |
|---------------|-----------------------------------------------------------------------------------|
| Address: | 101 Carnegie Center<br>Princeton, NJ 08540-6231 |
| Contact Name: | David Risley<br>Director, US Regulatory Co-ordination Group<br>Regulatory Affairs |
| Phone Number: | 609-514-6489 |
April 3, 2008 Summary Prepared Date:
## Subject Device Information
| Trade Name: | Brachytherapy Source Device | |
|---------------------|-----------------------------------------------------------|----------------------|
| Model: | 9011 | |
| Common Name: | Radionuclide Brachytherapy Source | |
| Class: | II | |
| Classification: | 21 CFR 892.5730 | Product Code: 90-IWI |
| Predicate Devices | Legally marketed devices to which equivalence is claimed. | |
| 1. Trade Name: | OncoSeed™ (Iodine-125 seeds) | |
| Common Name: | Radionuclide Brachytherapy Source | |
| Class: | II | |
| Classification: | 21 CFR 892.5730 | Product Code: 90-KXK |
| Cleared 510(k) No.: | K914281, FDA Cleared on November 22, 1991 | |
| Submitted by: | Medi-Physics Inc, dba GE Healthcare | |
. . . .
{1}------------------------------------------------
## Description of Device
The Brachytherapy Source Device, Model 9011, is a welded titanium capsule containing Iodine-125 adsorbed onto a silver rod.
## Intended Use
The Brachytherapy Source Device, with apparent activities between 0.19mCi and 1.01mCi, is indicated for permanent interstitial implantation of selected localized turnors which are of low to moderate radiosensitivity. They may be used either as primary treatment (such as for prostate cancer or unresectable tumors) or for treatment of residual disease after excision of the primary tumor. Seeds in this apparent activity range may be used to treat superficial, intra-abdominal, and intra thoracic tumors. Tumors of the head, neck, lung, pancreas, and prostate (early stages) are commonly treated.
The Brachytherapy Source Device, with total apparent activities greater than 1.01mCi, is indicated for interstitial treatment of tumors which have the following characteristics; unresectable, localized, and moderate radiosensitivity. These seeds may be used for selected radiation applications as temporary implants.
The Brachytherapy Source Device is indicated to treat residual turnors concurrent with or at the completion of other treatment modalities, such as external beam radiation therapy or chemotherapy. In addition, recurrent tumors may be implanted with seeds.
When comparing the Intended Use and Indications for Use of the predicate OncoSeed, Model 6711, device, the Intended Use and Indications for Use of the subject device, Brachytherapy Source Device, Model 9011, has not changed.
## Technological Characteristics
The following modifications have been incorporated in the subject device -- Brachytherapy Source Device, Model 9011, when compared to predicate device, OncoSeed, Model 6711:
| Modified<br>Characteristics | Brachytherapy Source Device<br>Model 9011<br>(Subject Device) | OncoSeed, Model 6711<br>(Predicate Device) |
|-----------------------------|---------------------------------------------------------------|--------------------------------------------|
| Seeds | | |
| Regulatory Status | Subject Application | K914281 |
| Capsule Material | Titanium, type A-40 | Titanium, type A-40 |
| Capsule Length | 4.20 - 4.90mm (0.165 - 0.193in) | 4.20 - 4.90mm (0.165 - 0.193in |
| Capsule Wall<br>Thickness | 0.051 - 0.0635mm (0.002 -<br>0.0025in) | 0.051 - 0.0635mm (0.002 -<br>0.0025in) |
| Capsule Diameter | 0.48 - 0.67mm (0.019 - 0.026in) | 0.77-0.96mm (0.031 - 0.38in) |
| Carrier Substrate | 99.9% silver | 99.9% silver |
| Carrier Length | 2.74 - 2.85mm (0.108 - 0.112in) | 2.74-2.85mm (0.108 - 0.112in) |
| Carrier Diameter | 0.27 - 0.33mm (0.011 - 0.013in) | 0.48 - 0.53mm (0.019 - 0.021in) |
| Isotope | Iodine I-125 | Iodine I-125 |
| Sterilization | Not sterile (User Sterilized) | Not sterile (User sterilized) |
{2}------------------------------------------------
Changes have only been made to the diameter of the capsule and carrier substrate diameter. All other technological characteristics of the subject seed remain unchanged compared to the predicate OncoSeed, Model 6711, 510(k) #K914281.
## Nonclinical Test Data
The Brachytherapy Source Device, Model 9011, has been tested according to ISO 2919 for sealed sources, (test report section E) and activity measurements have been obtained from NIST Report (Report of Air-Kerma rate measurements) (see section F). In addition, Prototype sources have been manufactured and assessed against predetermined criteria for Design Verification and Design Validation (Declaration of Conformity Attachment 1).
## Conclusion:
Upon reviewing the safety and effectiveness information provided in this submission and comparing the intended use, indications for use, method of use, the non-clinical test data and other technological characteristics, it can be concluded that the subject seed is substantially equivalent to the predicate OncoSeed, Model 6711, cleared under 510(k) # K914281.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# MAY 1 5 2008
Mr. David Risley Director, US Regulatory Co-ordination Group Medi-Physics, Inc. dba GE Healthcare 101 Carnegie Center PRINCETON NJ 08540
Re: K081066
Trade/Device Name: Brachytherapy Source Device, Model 9011 Regulation Number: 21 CFR 892.5730 Regulation Name: Radionuclide brachytherapy source Regulatory Class: II Product Code: IWG and KXK Dated: April 14, 2008 Received: April 17, 2008
### Dear Mr. Risley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrb/industry/support/index.html.
Sincerely yours,
Nancy Brogdon
Nancy C. Brogdon Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## EXHIBIT-2
## Indications for Use Form
Page 1 of 1
长081066 510(k) Number (if known): _
Device Name:
Brachytherapy Source Device, Model 9011
## Indications For Use:
The Brachytherapy Source Device, with apparent activities between 0.19mCi and 1.01mCi, is indicated for permanent interstitial implantation of selected localized tumors which are of low to moderate radiosensitivity. They may be used either as primary treatment (such as for prostate cancer or unresectable tumors) or for treatment of residual disease after excision of the primary tumor. Seeds in this apparent activity range may be used to treat superficial, intra-abdominal, and intra-thoracic tumors. Tumors of the head, neck, lung, pancreas and prostate (early stages) are commonly treated.
The Brachytherapy Source Device, with total apparent activities greater than 1.01mCi, is indicated for interstitial treatment of tumors which have the following characteristics; unresectable, localized, and moderate radiosensitivity. These seeds may be used for selected radiation applications as temporary implants.
The Brachytherapy Source Device is indicated to treat residual tumors concurrent with or at the completion of other treatment modalities, such as external beam radiation therapy or chemotherapy. In addition, recurrent tumors may be implanted with seeds.
# PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division S Division of Reproductive, Abdominal and Radiological Devices 510(k) Number _
Prescription Use
(Per 21 CFR 801.109) √
OR
Over-The-Counter Use _
ﻝ
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.