K211034 · Best Nomos · MUJ · May 6, 2021 · Radiology
Device Facts
Record ID
K211034
Device Name
Corvus Radiation Therapy Planning System
Applicant
Best Nomos
Product Code
MUJ · Radiology
Decision Date
May 6, 2021
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
CORVUS is intended for use as a planning tool for conformal radiation therapy. For IMRT, using operator-supplied input and patient scans, it creates a plan for treatment delivery systems and generates a set of beam weights that, when applied to a compatible system, facilitates delivery of an intensity-modulated 3D conformal radiation therapy treatment. For Forward Planning, CORVUS helps the user manually create a plan for treatment delivery based on user supplied inputs and patient scans including dose calculation. Forward Planning is supported only for Cobalt-60 based treatment planning. CORVUS is intended only to suggest a delivery plan. It is the physician's responsibility to verify that the dose distributions which would result from plan implementation are appropriate for a particular patient. The CORVUS system is intended to be used as an integrated system with a delivery device for planning and delivery of conformal radiation therapy. CORVUS produces radiation fields to conform to the projected tumor volume plus margins as desired during planning. The system provides statistical calculations to help achieve target goals while sparing sensitive structures.
Device Story
Radiation therapy treatment planning system (TPS) for conformal external beam photon therapy; inputs include patient scans and treatment machine-specific data; outputs include radiation treatment plans with beam weights and dose distributions. Used in clinical settings by medical physicists, dosimetrists, and radiation oncologists. Operates via semi-automatic planning; supports IMRT (system-suggested plans) and Forward Planning (user-designed plans). Optimization methods include simulated annealing and gradient descent. Clinicians review and approve plans to ensure appropriate dose distribution for target volumes while sparing sensitive structures. Facilitates delivery of intensity-modulated 3D conformal radiation therapy.
Clinical Evidence
Bench testing only. Validation included clinical workflow assessment, treatment planning software usability, and dosimetric accuracy. Accuracy was evaluated by comparing system-generated treatment plans against medical physics measurements using 2D arrays and ion chambers. No clinical patient data was required.
Technological Characteristics
Software-based radiation treatment planning system; runs on Mac OS X. Uses finite size pencil beam (FSPB) algorithm for dose calculation. Supports MLC-IMRT, serial tomotherapy, and Cobalt-60 based forward planning. Conforms to IEC 62304 (software life cycle), IEC 62366-1 (usability), and IEC 62083 (safety of radiotherapy planning systems).
Indications for Use
Indicated for medical physicists, dosimetrists, and radiation oncologists to create conformal treatment plans using photon (x-ray, gamma ray) external beam radiation therapy. Valid for use only with external beam photon therapy; calculations for electrons and intracavity sources (Brachytherapy) are not supported.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
May 6, 2021
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Best Nomos % Vineet Gupta, Ph.D. Director R&D One Best Drive PITTSBURGH PA 15202
Re: K211034
Trade/Device Name: CORVUS Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: Class II Product Code: MUJ Dated: April 5, 2021 Received: April 7, 2021
Dear Dr. Gupta:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K211034
Device Name CORVUS
#### Indications for Use (Describe)
The CORVUS system is a radiation treatment planning package designed to allow medical physicists, and radiation oncologists to create conformal treatment plans using photon (x-ray, gamma ray) external beam radiation therapy. The treatment plans generated by CORVUS are based upon treatment machine-specific data and are intended to provide a guide to delivering external beam radiation therapy which conforms to the target volume defined by the radiation oncologist.
The CORVUS system is valid for use only with external beam photon therapy; calculations for electrons and intracavity sources (Brachytherapy) are NOT supported.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Best nomos. The word "Best" is written in a cursive, blue font. To the right of "Best", the word "nomos" is written in a green, sans-serif font.
K211034
# 510(k) Summary
## A. SUBMITTERS NAME
Best NOMOS
#### B. ADDRESS
One Best Drive Pittsburgh, PA 15202
### C. CONTACT
| Name: | Vineet Gupta, Ph.D. |
|--------|------------------------|
| Phone: | (412) 312-6819 |
| Fax: | (412) 312-6701 |
| Email: | vineet.gupta@nomos.com |
#### D. DATE PREPARED
April 5, 2021
#### E. DEVICE NAME
| Device Trade Name: | CORVUS |
|----------------------|-----------------------------------------------------------------------|
| Common Name: | Radiation Therapy Treatment Planning System |
| Classification Name: | Medical Charged-Particle Radiation Therapy System<br>(21CFR 892.5050) |
### F. DEVICE CLASS
Class II Panel: Radiology Product Code: 90-MUJ Regulation Number: 21CFR 892.5050
Image /page/3/Picture/15 description: The image contains the logo for Best healthcare. The logo features the word "Best" in a stylized, blue font. Above the word "Best" is an image of people holding hands in a circle. Below the word "Best" is the phrase "healthcare for everyone" in a smaller, blue font.
One Best Drive (formerly One Conair Drive), Pittsburgh, PA 15202 USA phone 412 312 6700 800 70 NOMOS fax 412 312 6701 www.nomos.com
{4}------------------------------------------------
## G. PREDICATE DEVICES
CORVUS Radiation Therapy Treatment Planning System 14 (K201350)
### H. STATEMENT OF INDICATIONS FOR USE
The intended use and indications for use for the CORVUS Radiation Therapy Treatment Planning System have not changed due to this modification to the product. This modification does not change the fundamental principle of using photons for creating conformal treatment plans which is the same as that of its predicate device.
### Intended Use
CORVUS is intended for use as a planning tool for conformal radiation therapy. For IMRT, using operator-supplied input and patient scans, it creates a plan for treatment delivery systems and generates a set of beam weights that, when applied to a compatible system, facilitates delivery of an intensity-modulated 3D conformal radiation therapy treatment. For Forward Planning, CORVUS helps the user manually create a plan for treatment delivery based on user supplied inputs and patient scans including dose calculation. Forward Planning is supported only for Cobalt-60 based treatment planning.
CORVUS is intended only to suggest a delivery plan. It is the physician's responsibility to verify that the dose distributions which would result from plan implementation are appropriate for a particular patient.
The CORVUS system is intended to be used as an integrated system with a delivery device for planning and delivery of conformal radiation therapy. CORVUS produces radiation fields to conform to the projected tumor volume plus margins as desired during planning. The system provides statistical calculations to help achieve target goals while sparing sensitive structures.
## Indications for Use
The CORVUS system is a radiation treatment planning package designed to allow medical physicists, dosimetrists, and radiation oncologists to create conformal treatment plans using photon (x-ray, gamma ray) external beam radiation therapy. The treatment plans generated by CORVUS are based upon treatment machine-specific data and are intended to provide a guide to delivering external beam radiation therapy which conforms to the target volume defined by the radiation oncologist.
The CORVUS system is valid for use only with external beam photon therapy; calculations for electrons and intracavity sources (Brachytherapy) are NOT supported.
{5}------------------------------------------------
#### I. DEVICE DESCRIPTION
CORVUS is a semi-automatic planning system: For forward planning, the system allows the user to design a treatment plan. For IMRT, rather than simply verifying a user-designed plan, the system itself suggests a plan. A clinician then reviews and approves the plan.
CORVUS is designed to generate plans for treatment delivery systems that can create multiple radiation patterns composed of pencil beams on which the intensity can be individually controlled. The treatment beams are weighted so that when they are projected into the treatment space they superimpose to give the desired dose distribution.
Each radiation field is generated using one of several optimization methods provided with the system, including simulated annealing and gradient descent.
The treatment beams are set not only to deliver the prescribed dose to the identified target volume, but also to keep the dose to other sensitive volumes below user-defined limits. Planning is done volumetrically: the beam weights for treating the entire target volume are generated simultaneously. The dose matrix is volumetric. The dose to each point is calculated to be that received from all beams and from all gantry angles. Dosage is calculated using a finite size pencil beam (FSPB) algorithm based on the beam characterization of clinically measured data. The degree to which a treatment plan is optimized is determined in part by constraints placed on the planning algorithm. The user has direct control over these constraints, which include dose goals to the target structures, dose limits to the sensitive structures, and the specification of arcs or fixed gantry positions in the treatment plan.
CORVUS treatment plans need not have the isocenter located within the target volume. An unlimited number of targets falling within the treatment volume can be planned for at the same time. Dose may be prescribed for up to 32 structures, 29 of them user-selectable, any number of which may be separate targets or radiation-sensitive structures. Each structure can have a separate dose prescription.
## J. PREDICATE DEVICE INFORMATION
The CORVUS 15 system is substantially equivalent to its primary predicate device CORVUS 14 (K201350; Decision Date: 20-September-2020). The CORVUS 14 system was determined to be substantially equivalent to its predicate device as of September 2020.
The fundamental scientific technology of the CORVUS 15 with respect to its predicate device (CORVUS 14) system has not changed. The intended use and indications for use of the device have not changed. Based upon the performance testing results for CORVUS 15 (as detailed in the submission), the system raises no new issues of safety or effectiveness.
{6}------------------------------------------------
## K. COMPARISON TO THE PREDICATE DEVICE
This section provides the summary of comparison of CORVUS 14 to the predicate device.
| | Proposed Device | Predicate Device |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | CORVUS 15 | CORVUS 14 (K201350) |
| | CORVUS is intended for use as a<br>planning tool for conformal radiation<br>therapy. For IMRT, using operator-<br>supplied input and patient scans, it<br>creates a plan for treatment delivery<br>systems and generates a set of beam<br>weights that, when applied to a<br>compatible system, facilitates delivery<br>of an intensity-modulated 3D<br>conformal radiation therapy treatment.<br>For Forward Planning, CORVUS helps<br>the user manually create a plan for<br>treatment delivery based on user<br>supplied inputs and patient scans<br>including dose calculation. Forward<br>Planning is supported only for Cobalt-<br>60 based treatment planning | CORVUS is intended for use as a<br>planning tool for conformal radiation<br>therapy. For IMRT, using operator-<br>supplied input and patient scans, it<br>creates a plan for treatment delivery<br>systems and generates a set of beam<br>weights that, when applied to a<br>compatible system, facilitates delivery<br>of an intensity-modulated 3D<br>conformal radiation therapy treatment.<br>For Forward Planning, CORVUS helps<br>the user manually create a plan for<br>treatment delivery based on user<br>supplied inputs and patient scans<br>including dose calculation. Forward<br>Planning is supported only for Cobalt-<br>60 based treatment planning |
| Intended Use | CORVUS is intended only to suggest a<br>delivery plan. It is the physician's<br>responsibility to verify that the dose<br>distributions which would result from<br>plan implementation are appropriate<br>for a particular patient. | CORVUS is intended only to suggest a<br>delivery plan. It is the physician's<br>responsibility to verify that the dose<br>distributions which would result from<br>plan implementation are appropriate<br>for a particular patient. |
| | The CORVUS system is intended to be<br>used as an integrated system with a<br>delivery device for planning and<br>delivery of conformal radiation<br>therapy. CORVUS produces radiation<br>fields to conform to the projected<br>tumor volume plus margins as desired<br>during planning. The system provides<br>statistical calculations to help achieve<br>target goals while sparing sensitive<br>structures. | The CORVUS system is intended to be<br>used as an integrated system with a<br>delivery device for planning and<br>delivery of conformal radiation<br>therapy. CORVUS produces radiation<br>fields to conform to the projected<br>tumor volume plus margins as desired<br>during planning. The system provides<br>statistical calculations to help achieve<br>target goals while sparing sensitive<br>structures. |
| Indications for<br>Use | The CORVUS system is a radiation<br>treatment planning package designed to | The CORVUS system is a radiation<br>treatment planning package designed to |
| allow medical physicists, dosimetrists,<br>and radiation oncologists to create<br>conformal treatment plans using<br>photon (x-ray, gamma ray) external<br>beam radiation therapy. The treatment<br>plans generated by CORVUS are based<br>upon treatment machine-specific data<br>and are intended to provide a guide to<br>delivering external beam radiation<br>therapy which conforms to the target<br>volume defined by the radiation<br>oncologist.<br><br>The CORVUS system is valid for use<br>only with external beam photon<br>therapy; calculations for electrons and<br>intracavity sources (Brachytherapy) are<br>NOT supported. | allow medical physicists, dosimetrists,<br>and radiation oncologists to create<br>conformal treatment plans using<br>photon (x-ray, gamma ray) external<br>beam radiation therapy. The treatment<br>plans generated by CORVUS are based<br>upon treatment machine-specific data<br>and are intended to provide a guide to<br>delivering external beam radiation<br>therapy which conforms to the target<br>volume defined by the radiation<br>oncologist.<br><br>The CORVUS system is valid for use<br>only with external beam photon<br>therapy; calculations for electrons and<br>intracavity sources (Brachytherapy) are<br>NOT supported. | |
# Table 1 Indications for Use & Intended Use Comparison
{7}------------------------------------------------
| Comparison Item | Proposed Device<br>CORVUS 15 | Predicate Device<br>CORVUS 14 (K201350) |
|----------------------------------------------------------|----------------------------------------------------------------|----------------------------------------------------------------|
| Product Code | 90-MUJ | 90-MUJ |
| Class | II | II |
| Regulation Number | 892.5050 | 892.5050 |
| Operating System | Mac OS X | Mac OS X |
| Treatment Planning Support<br>for X-Rays | MLC-IMRT, Serial Tomotherapy<br>(MIMiC-IMRT) | MLC-IMRT, Serial Tomotherapy<br>(MIMiC-IMRT) |
| Treatment Planning Support<br>for Cobalt-60 (Gamma Rays) | Serial Tomotherapy (MIMiC-IMRT),<br>MLC-IMRT, Forward Planning | Serial Tomotherapy (MIMiC-IMRT),<br>MLC-IMRT, Forward Planning |
## Table 2 General Comparison
CORVUS 15 contains all of the features of CORVUS 14 (K201350) and adds the feature of supporting Cobalt-60 based external beam radiation treatment planning based Forward Planning using jaw defined fields and wedges. The similarities and differences are discussed in detail as part of this submission.
In both products, CORVUS 15 and CORVUS 14, the pencil-beam algorithm is used and dose calculation can be performed using either a homogeneous option or an Effective Path Length (EPL) option. The dose calculation based on the Lateral Disequilibrium Inclusive option is only supported for X-Ray based treatment planning in CORVUS 15 as it was in the predicate
{8}------------------------------------------------
device CORVUS 14 (K201350). The optimization algorithms used in both products are the same.
The target population, intended use, and indications for use are similar to that of the predicate devices. In addition, the fundamental technical characteristics are the same as those of the predicate devices and any minor differences in features do not raise any concerns for safety, performance, or effectiveness of the device. The characteristics/features of CORVUS 15 with respect to the predicate device is described in the comparison chart in the submission.
## L. SUMMARY OF TESTING
Design verification and validation testing was performed to ensure that the device functionality works as per its intended use, all risks are mitigated, is substantially equivalent, and the product conforms to the required standards. The software for this device is considered as a "Major" level of concern. The details of the design verification and validation activities are explained in the submission as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."
The verification activities included system tests, module tests, anomaly verification, code reviews, and run-though integration tests.
The validation activities included clinical workflow, treatment planning software usability, dosimetric accuracy, and import-export. The accuracy of treatment plans was evaluated through comparison with medical physics measurements 2D array and ion chamber measurements.
#### Conclusion:
Detailed results of these tests are included as part of this submission. The verification and validation results demonstrate that the CORVUS 15 system met its design requirements and specifications, is substantially equivalent to its predicate device, and conforms to the applicable sections of standards that includes:
- IEC 62304 Edition 1.1 2015-06 Medical device software Software life cycle processes .
- IEC 62366-1 Edition 1.0 2020-06 Application of usability engineering to medical devices .
- IEC 62083 Edition 2.0 2009-09 Requirements for the safety of radiotherapy treatment . planning systems
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.